Oxygenation index, an indicator of optimal distending pressure during high-frequency oscillatory ventilation?

Intensive Care Med. 2002 Aug;28(8):1151-6. doi: 10.1007/s00134-002-1368-5. Epub 2002 Jun 20.

Abstract

Objective: To test the hypothesis that, during high-frequency oscillatory ventilation (HFOV) of pigs with acute lung injury, the oxygenation index (OI = Paw*FIO(2)*100/PaO(2)) is minimal at the lowest continuous distending pressure (Paw), where the physiological shunt fraction is below 0.1 (Paw(optimal)).

Design and setting: Prospective, observational study in a university research laboratory.

Subjects: Eight Yorkshire pigs weighing 12.0+/-0.5 kg, with lung injury induced by lung lavage.

Interventions: After initiation of HFOV, the pigs were subjected to a stepwise increase of Paw to obtain under-inflation, optimal inflation and over-distension of the lungs (inflation) in series, followed by a similar decrease of Paw (deflation).

Measurements and results: At each Paw level, the OI and physiological shunt fraction were determined. The OI reached a minimum of 6.2+/-1.4 at Paw 30+/-4 cmH(2)O during inflation and a minimum of 2.4+/-0.3 at Paw 13+/-2 cmH(2)O during deflation. Paw(optimal) was 32+/-6 cmH(2)O on the inflation limb and 14+/-2 cmH(2)O on the deflation limb. The difference between the Paw at minimal OI and Paw(optimal) was -1.9+/-4.2 cmH(2)O (NS) during inflation and -1.5+/-1.6 cmH(2)O (p<0.05) during deflation. In 15 out of the 16 comparisons, the difference in Paw was within one step (+/-3 cmH(2)O).

Conclusion: The minimal OI is indicative for the Paw where oxygenation is optimal during HFOV in surfactant-depleted pigs.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • High-Frequency Ventilation*
  • Oxygen / supply & distribution*
  • Prospective Studies
  • Respiratory Distress Syndrome
  • Swine