Sleep detection with an accelerometer actigraph: comparisons with polysomnography

Physiol Behav. 2001 Jan;72(1-2):21-8. doi: 10.1016/s0031-9384(00)00355-3.

Abstract

Two validation studies were conducted to optimize the sleep-detection algorithm of the Actillume. The first study used home recordings of postmenopausal women (age range: 51 to 77 years), which were analyzed to derive the optimal algorithm for detecting sleep and wakefulness from wrist activity data, both for nocturnal in-bed recordings and considering the entire 24 h. The second study explored the optimal algorithm to score in-bed recordings of healthy young adults (age range: 19 to 34 years) monitored in the laboratory. In Study I, the algorithm for in-bed recordings (n=39) showed a minute-by-minute agreement of 85% between Actillume and polysomnography (PSG), a correlation of.98, and a mean measurement error (ME) of 21 min for estimates of sleep duration. Using the same algorithm to score 24-h recordings with Webster's rules, an agreement of 89%, a correlation of.90, and 1 min ME were observed. A different algorithm proved optimal to score in-bed recordings (n=31) of young adults, yielding an agreement of 91%, a correlation of.92, and an ME of 5 min. The strong correlations and agreements between sleep estimates from Actillume and PSG in both studies suggest that the Actillume can reliably monitor sleep and wakefulness both in community-residing elderly and healthy young adults in the laboratory. However, different algorithms are optimal for individuals with different characteristics.

Publication types

  • Clinical Trial
  • Comparative Study
  • Evaluation Study
  • Research Support, U.S. Gov't, P.H.S.
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Calibration
  • Evaluation Studies as Topic
  • Female
  • Humans
  • Male
  • Middle Aged
  • Motor Activity / physiology*
  • Polysomnography*
  • Sleep / physiology*