Test | Range/accuracy (BTPS) | Flow range L·s−1 | Time s | Resistance and back pressure | Test signal |
VC | 0.5–8 L, ±3% of reading or ±0.050 L, whichever is greater | 0–14 | 30 | 3-L Calibration syringe | |
FVC | 0.5–8 L, ±3% of reading or ±0.050 L, whichever is greater | 0–14 | 15 | <1.5 cmH2O·L−1·s−1 (0.15 kPa·L−1·s−1) | 24 ATS waveforms, 3-L Cal Syringe |
FEV1 | 0.5–8 L, ±3% of reading or ±0.050 L, whichever is greater | 0–14 | 1 | <1.5 cmH2O·L−1·s−1 (0.15 kPa·L−1·s−1) | 24 ATS waveforms |
Time zero | The time point from which all FEVt measurements are taken | Back extrapolation | |||
PEF | Accuracy: ±10% of reading or ±0.30 L·s−1 (20 L·min−1), whichever is greater; repeatability: ±5% of reading or ±0.15 L·s−1 (10 L·min−1), whichever is greater | 0–14 | Mean resistance at 200, 400, 600 L·min−1 (3.3, 6.7, 10 L·s−1) must be <2.5 cmH2O·L−1·s−1 (0.25 kPa·L−1·s−1) | 26 ATS flow waveforms | |
Instantaneous flows (except PEF) | Accuracy: ±5% of reading or ±0.200 L·s−1, whichever is greater | 0–14 | <1.5 cmH2O·L−1·s−1 (0.15 kPa·L−1·s−1) | Data from manufacturers | |
FEF25–75% | 7.0 L·s−1, ±5% of reading or ±0.200 L·s−1, whichever is greater | ±14 | 15 | Same as FEV1 | 24 ATS waveforms |
MVV | 250 L·min−1 at VT of 2 L within ±10% of reading or ±15 L·min−1, whichever is greater | ±14 (±3%) | 12–15 | <1.5 cmH2O·L−1·s−1 (0.15 kPa·L−1·s−1) | Sine wave pump |
BTPS: body temperature and ambient pressure saturated with water vapour; VC: vital capacity; FVC: forced vital capacity; ATS: American Thoracic Society; FEV1: forced expiratory volume in one second; FEVt: forced expiratory volume in t seconds; PEF: peak expiratory flow; FEF25–75%: mean forced expiratory flow between 25% and 75% of FVC; MVV: maximum voluntary ventilation; VT: tidal volume.