Oxygen device | Flow rate L·min−1 | Dispersion distance cm | Ref. |
HFNC | 60 | 17.2±3.3 | [6] |
30 | 13.0±1.1 | [6] | |
10 | 6.5±1.5 | [6] | |
Simple mask | 15 | 11.2±0.7 | [7] |
10 | 9.5±0.6 | [7] | |
Non-rebreathing mask | 10 | 24.6±2.2 | [7] |
Venturi mask at FIO2 0.4 | 6 | 39.7±1.6 | [7] |
Venturi mask at FIO2 0.35 | 6 | 27.2±1.1 | [7] |
Summary of studies evaluating oxygen delivery devices using a high-fidelity human simulator with smoke particles of <1 µm (an aerosol of solid particles). The smoke was illuminated by a laser light-sheet and high-definition video was used to measure dispersion distance away from the manikin. Indicated dispersion distances give an idea of proximity of contaminated bio-aerosols, to which healthcare workers may be directly exposed. HFNC: high-flow nasal cannula; FIO2: inspiratory oxygen fraction.