PT - JOURNAL ARTICLE AU - B Hannhart AU - JP Haberer AU - C Saunier AU - MC Laxenaire TI - Accuracy and precision of fourteen pulse oximeters AID - 10.1183/09031936.93.04010115 DP - 1991 Jan 01 TA - European Respiratory Journal PG - 115--119 VI - 4 IP - 1 4099 - http://erj.ersjournals.com/content/4/1/115.short 4100 - http://erj.ersjournals.com/content/4/1/115.full SO - Eur Respir J1991 Jan 01; 4 AB - Two sets of seven pulse oximeters (Criticare CSI-502; Nellcor-N200; Datex-Satlite; Physio-Control-Lifestat 1600; Critikon-Oxyshuttle; Ohmeda-Biox 3700; Ohmeda-Biox 3740; Radiometer-Oxi; Spectramed-Pulsat; Kontron-7840; Biochem-Ox2000; Invivo-4500; Engstrom-EOS; Novametrix-505) were studied in two groups of eight healthy subjects, aged 26-50 yrs. The transcutaneous oxygen saturation (SpO2) was compared with arterial oxygen saturation (SaO2) measured in simultaneously with drawn blood samples (OSM2 Radiometer) at four 20 min steady-state levels of inspired oxygen fraction (FIO2) (0.21, 0.10, 0.08 and 0.07; SaO2 99-55%) in a conditioned chamber. Both the error in accuracy (mean SpO2-SaO2 difference), and the error in precision (SD of the differences) remained less than 3% for the two highest FIO2 levels (SaO2 greater than 83%) but, during deeper hypoxia, they were increased to 8% and 5%, respectively. An instrumental systematic bias affected accuracy in particular. We concluded that a good agreement between SpO2 and SaO2, as reflected by the Bartko's intraclass coefficient, was observed in nine instruments.