RT Journal Article SR Electronic T1 Comparison of a cardiorespiratory device versus polysomnography for diagnosis of sleep apnoea JF European Respiratory Journal JO Eur Respir J FD European Respiratory Society SP 1505 OP 1510 DO 10.1183/09031936.02.00297402 VO 20 IS 6 A1 J.M. Calleja A1 S. Esnaola A1 R. Rubio A1 J. Durán YR 2002 UL http://erj.ersjournals.com/content/20/6/1505.abstract AB This study assessed the accuracy of a cardiorespiratory monitoring device versus polysomnography for the diagnosis of suspected sleep apnoea/hypoponea syndrome (SAS). A total of 86 patients (89% male, mean age 52 yrs) that had been referred to a sleep laboratory with a clinical diagnosis of SAS underwent cardiorespiratory polygraphy in an unattended mode using an ambulatory device (MERLIN). Analysis was carried out both automatically and manually. Conventional overnight full-channel polysomnography was performed simultaneously. Valid polygraphical recordings were obtained from 79 patients. The mean±sd apnoea/hypopnoea index (AHI) was 34.4±29.2. The results obtained with manual scoring were superior to automatic scoring for all AHI thresholds. For an AHI of ≥5, which is diagnostic SAS, the optimum cut-off value for the manual respiratory event index was 6.7 and the cardiorespiratory monitoring device had 97.1% sensitivity and 90.9% specificity. Correct classification according to the different cut-off points obtained via polysomnography and the corresponding cut-off points in the MERLIN manual index were confirmed in 90–96% of patients. The MERLIN device is a useful diagnostic approach for the initial assessment of adult patients with clinical suspicion of sleep apnoea/hypopnoea syndrome. Manual scoring is clearly better than automatic scoring in terms of agreement with the apnoea/hypopnoea index and to discern patients with sleep apnoea/hypopnoea syndrome. This study was supported by a grant from the Dept of Health, Basque Government.