RT Journal Article SR Electronic T1 Inaccuracy of tidal volume delivered by home mechanical ventilators JF European Respiratory Journal JO Eur Respir J FD European Respiratory Society SP 338 OP 341 DO 10.1034/j.1399-3003.2000.15b20.x VO 15 IS 2 A1 F Lofaso A1 R Fodil A1 H Lorino A1 K Leroux A1 A Quintel A1 A Leroy A1 A Harf YR 2000 UL http://erj.ersjournals.com/content/15/2/338.abstract AB Ideally, the inspired (tidal) volume (V(T)) provided by a volume-controlled ventilation device should not change when the pressure imposed on the ventilator varies. A bench study evaluation of V(T) versus pressure was performed on 10 commercially available devices. The difference between the desired V(T) and the observed V(T) reached 100 mL for some devices when inspiratory resistance was at its lowest, rising to 150 mL when inspiratory resistance was increased to obtain peak airway pressure of 60 cmH2O. The present data indicate that some home ventilators are inaccurate in delivering the preset tidal volume when the pressure imposed on the ventilator is increased to simulate high airway resistance.