PT - JOURNAL ARTICLE AU - IM Stell AU - S Tompkins AU - AT Lovell AU - JC Goldstone AU - J Moxham TI - An in vivo comparison of a catheter mounted pressure transducer system with conventional balloon catheters AID - 10.1034/j.1399-3003.1999.13e36.x DP - 1999 May 01 TA - European Respiratory Journal PG - 1158--1163 VI - 13 IP - 5 4099 - http://erj.ersjournals.com/content/13/5/1158.short 4100 - http://erj.ersjournals.com/content/13/5/1158.full SO - Eur Respir J1999 May 01; 13 AB - In the assessment of respiratory muscle function balloon catheters have been widely used for pressure measurements. However, this type of investigation is poorly tolerated by acutely ill patients. This study assessed the performance of a possible alternative, a catheter-mounted miniature pressure transducer (CMT). The assessment consisted of a laboratory study of the linearity, frequency response, and stability of gain and baseline of the CMT system, and an in vivo study directly comparing the CMT and balloon catheter systems in seven normal subjects for a range of respiratory manoeuvres. These were: 1) maximal inspiratory and expiratory pressures against a closed airway, 2) twitch transdiaphragmatic pressure elicited by cervical magnetic phrenic nerve stimulation, and 3) tidal breathing, sniffs and coughs in five body positions. The agreement of the two systems was analysed for measurements of 1) absolute pressures, 2) magnitude of changes in pressure, and 3) rate of change of pressure (maximum relaxation rate after sniff manoeuvres). The CMT system was linear, with a high frequency response and stable gain, but showed baseline drift. The two systems agreed well for measurements of change and rate of change of pressure, but less well for measurements of absolute pressure. The CMT system tested is potentially useful for studies of acute changes in respiratory pressures, or studies of respiratory muscle strength, but would be less useful where accurate measurements of absolute pressures are required.