RT Journal Article SR Electronic T1 Prevention of thermal and condensation errors in pneumotachographic recordings of the maximal forced expiratory manoeuvre JF European Respiratory Journal JO Eur Respir J FD European Respiratory Society SP 198 OP 201 DO 10.1183/09031936.94.07010198 VO 7 IS 1 A1 MR Miller A1 T Sigsgaard YR 1994 UL http://erj.ersjournals.com/content/7/1/198.abstract AB When recording respiratory manoeuvres using a pneumotachograph (PT), it is customary to heat the pneumotachograph head to avoid condensation errors. It is technically demanding to heat the PT correctly, and this increases the complexity of the equipment. We have studied whether a simpler method can eradicate condensation errors and preserve thermal stability of the PT. We tested the accuracy of a PT by recording the discharge of 1,000 ml of air through the PT before and after a subject had blown through it, thus causing both condensation within the PT and a rise in its temperature. We then determined whether the thermal and condensation errors could be avoided by placing the PT, between blows, on a fan that was blowing air at a constant 1.65 l.s-1. With ambient temperature at 20 degrees C, a single blow through the PT was found to cause a subsequent 4% over-reading in volume, due to condensation and temperature changes in the PT. After five consecutive blows the error was 7%. Placing the PT on the fan for 15 s after a single blow abolished these errors, whereas a wait of 15 s did not. At an ambient temperature of 14 degrees C there was a 5% over-reading of volume after a single blow, and an over-reading of 9% after five blows. At an ambient temperature of 26 degrees C these errors were 3 and 5%, respectively. All of these errors were abolished by placing the PT on the fan for 15 s and 30 s, respectively.(ABSTRACT TRUNCATED AT 250 WORDS)