PT - JOURNAL ARTICLE AU - Gabriella Abbate AU - Sebastiano Colombo AU - Keibun Liu AU - Clayton Semenzin AU - Noriko Sato AU - Carmen Ainola AU - Emily Wood AU - Davide Chiumello AU - Jacky Suen AU - Jo Pauls AU - John Fraser AU - Gianluigi Li Bassi TI - Structural characteristics of a novel catheter for esophageal pressure monitoring AID - 10.1183/13993003.congress-2021.PA3229 DP - 2021 Sep 05 TA - European Respiratory Journal PG - PA3229 VI - 58 IP - suppl 65 4099 - http://erj.ersjournals.com/content/58/suppl_65/PA3229.short 4100 - http://erj.ersjournals.com/content/58/suppl_65/PA3229.full SO - Eur Respir J2021 Sep 05; 58 AB - Introduction: Esophageal pressure monitoring allows estimation of transpulmonary pressure and precise application of lung-protective ventilation.Aims and Objectives: We describe structural characteristics of a novel esophageal catheter comprising a polyurethane balloon in comparison with commercially available alternatives.Methods: We tested a novel catheter (AspisafeNG+, Aspisafe Solution, NY, USA), designed to monitor esophageal pressure and prevent gastro-esophageal reflux vs. four commercially available catheters. We used a caliper to measure esophageal balloon outer diameter (OD) and length. Balloon compliance was measured via 0.5-mL increments of air up to an internal pressure of 40 cmH2O. Minimum volume (Vmin) was defined as the inflated volume that caused a linear increase in balloon internal pressure. Each experiment was made in triplicate.Results: Balloon OD varied among catheters (N 15, p=0.012) and ranged from median (IQR) of 6.6 (0.5) mm (Cooper®), up to 25.2 (0.2) mm (AspisafeNG+). Balloon lengths as well differed among catheters (N 15, p=0.013) from median (IQR) of 39.1 (2.5) mm (Aspisafe NG+), up to 100.7 (0.7) mm (Nutrivent®). Vmin (N 15, p=0.010) varied among catheters, as reported in Fig.1, while compliance did not (N 103, p=0.714).Measurements and Main Results: The novel esophageal balloon is characterized by large OD and Vmin. Manufacturers produce highly heterogeneous catheters for esophageal pressure monitoringFootnotesCite this article as: European Respiratory Journal 2021; 58: Suppl. 65, PA3229.This abstract was presented at the 2021 ERS International Congress, in session “Prediction of exacerbations in patients with COPD”.This is an ERS International Congress abstract. No full-text version is available. Further material to accompany this abstract may be available at www.ers-education.org (ERS member access only).