RT Journal Article SR Electronic T1 Introducing the Homburg lung: Efficacy and safety of a minimal-invasive system for extracorporeal CO2 removal JF European Respiratory Journal JO Eur Respir J FD European Respiratory Society SP PA3571 DO 10.1183/13993003.congress-2016.PA3571 VO 48 IS suppl 60 A1 Seiler, Frederik A1 Trudzinski, Franziska A1 Kamp, Annegret A1 Tsitouras, Konstantinos A1 Linn, Oliver A1 Metz, Carlos A1 Böhmer, Philip A1 Kaiser, Ralf A1 Fähndrich, Sebastian A1 Wilkens, Heinrike A1 Haake, Hendrik A1 Bals, Robert A1 Lepper, Philipp YR 2016 UL http://erj.ersjournals.com/content/48/suppl_60/PA3571.abstract AB Introduction: Extracorporeal CO2 removal (ECCO2R) is increasingly used as a therapeutic strategy in hypercapnic lung failure in order to avoid intubation, to allow lung protective ventilator settings, or to facilitate weaning from invasive ventilation. Here, we report on the Homburg Lung, a novel minimal-invasive mid-flow ECCO2R system which was established and successfully used at our ICU. The Homburg Lung is a pump-driven (Maquet Rotaflow) system for veno-venous ECCO2R with ¼” tubing and a 0.8 m2 surface oxygenator (Maquet Quadrox-I pediatric). Vascular access is usually established via a 19F/21cm bilumen cannula in the right internal jugular vein.Methods: To analyse efficacy and safety of the Homburg Lung, we screened our patient registry for ECCO2R procedures since 2013. Patients who underwent ECMO prior to ECCO2R were excluded. Patients who underwent ECCO2R more than one time since 2013 were only included once.Results: A total of 19 patients (age 52.3 ± 12.86 years, 42.1% female) was analysed. Ventilatory failure occurred due to COPD (47.4%), cystic fibrosis (21 %), and other origins (31.6%). The system generated a blood flow of 1.23 ± 0.29 lpm. Gas flow was 5.37 ± 2.98 lpm. Within 4 h, pCO2 could be reduced from 74.03 ± 17.51 mmHg to 55.53 ± 10.85 mmHg. Cannulation-associated complications were transient arrhythmia (1/19 patients), pneumothorax (1/19), and air embolism (1/19). Fatal complications were not registered.Conclusion: The Homburg Lung provides effective CO2 removal in hypercapnic lung failure. The cannulation is a safe procedure with complication rates comparable to those in central venous catheter implantation.