Authors | Type of camera | Trademark and model | Estimated clinical spatial resolution of PET camera (FWHM) | Fasting | Injected activity | Acquisition time per bed position (emission/transmission) | Quantification value used | SUV normalisation | Incubation time between injection and image acquisition | SUV obtained from clinical report versus specific review | Glycaemia levels accepted |
Agarwal et al. [34] | PET and PET/CT | GE Discovery DST and GE Advance | 5 mm (NA for PET alone) | 4 h | 370 MBq average | 3 min emission (NA for PET alone) | SUVmax | Body weight | 60 min | Report (a minority review) | NA |
Borst et al. [16] | PET | Siemens/CTI | 8 mm | >6 h | 350–550 MBq | 10–15 min emission/transmission | SUVmax | Body weight and glucose corrected | 60 min | Review | No limit |
de Jong et al. [23] | PET | Siemens ECAT (two centres) | 5 mm | >6 h | 220–370 MBq | 5 min emission/
3 min transmission | SUVmax | Body weight | 60–90 min | Both | NA |
Eschmann et al. [20] | PET | GE Advance | 5 mm | >12 h | 350–450 MBq | NA | SUVmax and SUVmean | Body weight | 60 min | Both | 150 mg·dL-1 |
Goodgame et al. [27] | PET and PET/CT | Simens ECAT Exact, Phillips/ADAC CPET Plus and Siemens Biograph LSO Duo Hybrid | NA | NA | NA | NA | SUVmax | NA | NA | NA | NA |
Higashi et al. [13] | PET | Shimazu Headtome IV | 5 mm | >4 h | 185 MBq | 5 min emission/5 min transmission | SUVmean | Body weight | 40 min | Clinical report | 120 mg·dL-1 |
Hoang et al. [29] | PET | GE 4096 plus and GE Advance Unit | 5 mm | 4 h | 370–555 MBq | NA | SUVmax | Body weight | 60±15 min | Review | <200 mg·dL-1 |
Unpublished data (Meert) | PET/CT | GE Discovery LS | 8 mm | >6 h | 3.7–7.4 MBq·kg-1 | 4 min emission | SUVmax | Body weight | 60–75 min | Review | 150 ml·dL-1l |
Port et al. [18] | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
Sasaki et al. [19] | PET | CTI Siemens HR | 8 mm | >6 h | 370–555 MBq | 5 min emission/
3 min transmission | SUVmax | Body weight | 60 min | Review | 150 mg·dL-1 |
van Baardwijk et al. [25] | PET | Siemens ECAT EXACT 922 and CTI Siemens 931/08/12 | 7 mm | >6 h | 3.5 MBq·kg-1 | 8 min emission/
4 min transmission | SUVmax | Body weight | 60 min (range 45–120 min) | Both | ∼150 mg·dL-1 |
Vansteenkiste et al. [11] | PET | CTI Siemens 931/08/12 | 8 mm | >6 h | 6.5 MBq·kg-1 | NA | SUVmax | Body weight | 50–70 min | Review | NA |