Abstract
Nasal high flow (NHF) reduces minute ventilation and ventilatory loads during sleep but the mechanisms are not clear. We hypothesised NHF reduces ventilation in proportion to physiological but not anatomical dead space.
11 subjects (five controls and six chronic obstructive pulmonary disease (COPD) patients) underwent polysomnography with transcutaneous carbon dioxide (CO2) monitoring under a metabolic hood. During stable non-rapid eye movement stage 2 sleep, subjects received NHF (20 L·min−1) intermittently for periods of 5–10 min. We measured CO2 production and calculated dead space ventilation.
Controls and COPD patients responded similarly to NHF. NHF reduced minute ventilation (from 5.6±0.4 to 4.8±0.4 L·min−1; p<0.05) and tidal volume (from 0.34±0.03 to 0.3±0.03 L; p<0.05) without a change in energy expenditure, transcutaneous CO2 or alveolar ventilation. There was a significant decrease in dead space ventilation (from 2.5±0.4 to 1.6±0.4 L·min−1; p<0.05), but not in respiratory rate. The reduction in dead space ventilation correlated with baseline physiological dead space fraction (r2=0.36; p<0.05), but not with respiratory rate or anatomical dead space volume.
During sleep, NHF decreases minute ventilation due to an overall reduction in dead space ventilation in proportion to the extent of baseline physiological dead space fraction.
Abstract
Nasal high flow decreases dead space ventilation in both controls and patients with COPD during sleep http://ow.ly/uJHd30jAzwA
Introduction
Nasal high flow (NHF) is a method of ventilatory support increasingly used in several clinical settings. During NHF, warm and humidified air is delivered to the patient's nose at high flow rates (2–60 L·min−1) [1]. NHF can assist ventilation and prevent intubation in both adults and children with respiratory failure [2–5]. Several mechanisms have been proposed to explain the responses observed with the use of this therapy. NHF could lead to improvements in respiratory mechanics [6], better humidification of airways [7], reductions in anatomical dead space [8, 9], increases in end-expiratory lung volume [10–12] and decreases in ventilatory demand due to a reduction in work of breathing [4, 9, 13–15]. The decrease in ventilatory demand has been attributed to washout of the anatomical nasal dead space [8]; however, dynamic measures of dead space volume and carbon dioxide (CO2) production during tidal breathing were not performed, leaving the actual mechanism of action for the effects produced by NHF still unknown.
Sleep is a unique opportunity to study physiological mechanisms of NHF as we can avoid potential conscious confounders. We have previously shown that ventilatory responses to NHF are different during wakefulness and sleep [16]. These differences may explain the heterogeneous responses reported when examining mechanisms of NHF in the clinical setting. In fact, when controlled for sleep, we observed a more homogeneous response to NHF. Both minute ventilation and work of breathing decrease, while CO2 levels remain constant [9]. The reduction in ventilation is possibly due to a decrease in dead space ventilation due to nasal dead space washout [8, 17]. However, by decreasing work of breathing, NHF can also decrease CO2 production and, therefore, reduce the amount of minute ventilation required to maintain constant levels of CO2.
To determine whether the reduction in minute ventilation is driven by the decrease in work of breathing or by a reduction in dead space ventilation, we measured the effects of NHF on CO2 production during sleep in controls and chronic obstructive pulmonary disease (COPD) patients. Specifically, using a metabolic hood and standard polysomnography with CO2 monitoring, we measured CO2 production and calculated alveolar ventilation with and without the use of NHF during stable sleep. We hypothesised that NHF would reduce dead space ventilation and that the amount of the reduction would depend on the extent of dead space. We also hypothesised that, in contrast to controls, COPD patients would show a reduction in CO2 production, thereby also reducing alveolar ventilation.
Methods
Participants were recruited from the Johns Hopkins pulmonary clinics and surrounding community in the USA. Subjects >18 years old and consenting to participate were included. Exclusion criteria were exacerbation of COPD or severe illnesses within 8 weeks prior to the study, severe heart and liver disease, home use of oxygen (O2), and/or chronic use of opioids, benzodiazepines or other sedatives.
In a previous study [9], we detected a 30–40% reduction in minute ventilation, work of breathing and tidal volume with the use of NHF. As we planned to investigate whether changes in metabolic rate contributed to minute ventilation reduction, we estimated that 10 subjects would be required to reproduce the previously observed effect on ventilation and to study the contribution of variation in metabolic rate.
The study was approved by the Johns Hopkins Medical Institution Human Investigation Review Board.
Equipment setup
Subjects enrolled underwent an overnight sleep study under a metabolic hood to measure CO2 production while using NHF. During the study, we collected standard polysomnography signals including transcutaneous CO2 (TCM3; Radiometer Medical, Brønshøj, Denmark). Ventilation was monitored with respiratory inductive plethysmography (Respitrace; AMI, Ardsley, NY, USA). At the beginning of the protocol, subjects wore a face mask attached to a pneumotachograph [18] and end-tidal CO2 monitor (VacuMed, Ventura, CA, USA). Subjects breathed through the mask for ∼30 s to calibrate the respiratory inductive plethysmography with the integrated pneumotachograph signal. Additionally, participants performed a slow exhalation manoeuvre to elicit an accurate end-tidal CO2 reading. We scaled the transcutaneous CO2 measurement to the obtained end-tidal CO2 values. After calibration procedures, the mask was removed and subjects wore only a nasal cannula for the intermittent delivery of NHF.
Once participants were fully monitored and comfortable in bed, we placed a metabolic hood (Quark RMR; Cosmed, Rome, Italy) covering their head and chest to collect the air exhaled by the subject (figure 1). This system allows the measurement of O2 consumption and CO2 production in an individual without wearing a mask. The long sheet attached to the hood sits around the upper part of the subject's body, avoiding air leaks. A pump continuously aspirates air from the hood to the analysis equipment, directing the exhaled air towards the hood exit and away from other potential leak points. The exhaled air is collected and analysed to compute CO2 production, O2 consumption and energy expenditure.
In our setting, we sought to determine CO2 production and energy expenditure during the NHF therapy, which would alter the flow delivered to the interior of the hood. Therefore, a high flow rate at the pump collecting air from the hood (50 L·min−1) was maintained to avoid leaks and to ensure constant experimental conditions. The incoming flow to the hood was the sum of the NHF flow (20 L·min−1) and a room air inlet flow (30 L·min−1). Whenever the NHF to the subject's nose was turned off, we initiated the delivery of an equal flow to the subject's chest via a secondary cannula, maintaining the overall flow to the interior of the hood constant (figure 1). As it was important to maintain a constant flow to the metabolic hood, we performed a series of pilot experiments using a constant source of CO2 to ensure that alternating NHF would not interfere with metabolic measurements.
Study protocol
Subjects were admitted to the sleep centre at approximately 19:00 h to get accustomed to the experimental procedures and to standardise their level of activity prior to the sleep studies. Once participants were fully monitored, we started the delivery of a flow of 20 L·min−1, 85% humidity and 32°C through a nasal cannula during wakefulness for acclimatisation. We used a modified S8 continuous positive airway pressure machine (ResMed, Bella Vista, Australia) that was able to generate constant flows up to 30 L·min−1. The device was also designed with a special hose with active heating that maintained warm and humid delivered air. Between 22:00 and 23:00 h, subjects initiated sleep with NHF on at 20 L·min−1. When subjects were in stable non-rapid eye movement (NREM) stage 2 sleep, we alternated periods of NHF on/off at 5- to 10-min intervals (figure 2). During the periods when NHF was turned off, the flow to the second cannula (to the subject's chest) was turned on at the same rate, humidity and temperature of the NHF. For each subject, pairs of NHF on/off trials were selected when intervals were absent of interruptions in sleep and there was no significant reduction in pulse wave amplitude of >50% from baseline as a marker for cortical activity or skin nerve activity. On average, three pairs of NHF on/off trials for each subject were obtained. Measurements were summarised to produce one mean value of “NHF off” and “NHF on” for each subject, which were used for the statistical comparisons.
Parameters calculation and statistical analysis
Data were recorded using a RemLogic polysomnography recording station (Natus, Pleasanton, CA, USA) and transferred to IgorPro data analysis software (WaveMetrics, Lake Oswego, OR, USA). We derived minute ventilation, respiratory rate and tidal volume from the calibrated respiratory inductive plethysmography signal. Physiological dead space ventilation was calculated using the equation for alveolar ventilation: PaCO2=0.863×V′CO2/(V′E–V′D), where PaCO2 is arterial CO2 tension (for which calibrated transcutaneous CO2 measurements were used), V′CO2 is CO2 production (measured with the metabolic hood), V′E is minute ventilation and V′D is the physiological dead space ventilation per minute [19].
Tidal dead space volume was calculated by dividing dead space ventilation by respiratory rate. Dead space fraction was computed as dead space volume divided by tidal volume. Anatomical dead space volume was estimated from participant's height, as previously described [20–22].
Measurements with and without NHF were compared by the paired t-test. The degree of response in controls and COPD patients was compared by the nonpaired t-test. The correlation of changes in dead space to baseline dead space was analysed using Pearson correlation coefficient. We considered p-values <0.05 as statistically significant.
Results
12 participants were recruited (six controls and six COPD patients), but one subject (with COPD) could not sleep under the metabolic hood and was excluded. Table 1 shows the demographics of the population enrolled.
We detected an awakening in 12 out of the 98 NHF on/off trials, some of them apparently induced by the transition of the state of NHF (on to off or vice versa). Once an arousal was detected, we waited longer periods until subjects reached stable sleep or excluded the analysis on that segment if sleep was not resumed.
Effect of NHF on respiratory pattern during sleep under a metabolic hood
In figure 2, we show the effect of the NHF on and off conditions on ventilation during sleep in one individual (COPD1). It can be seen that our experimental setup of turning NHF on and off during sleep (see “NHF tag” trace) did not disturb sleep continuity (see electro-oculogram, EEG and electromyogram traces) or lead to an activation of autonomic nerve activity, as illustrated by the stability of the heart rate (pulse) and pulse wave amplitude in this patient. The effects of NHF on ventilation are illustrated by comparing the tidal volume signal (see respiratory plethysmography trace) with the O2 saturation, transcutaneous CO2, CO2 production and energy expenditure. While tidal volumes increased during the NHF off condition, there was no change in transcutaneous CO2, O2 saturation, CO2 production or energy expenditure, indicating that alveolar ventilation did not change.
Effect of NHF on ventilation
Pooled data are shown figure 3. At baseline, compared with controls, COPD patients had a slightly but not statistically significant higher minute ventilation (controls 4.9±0.5 L·min−1 versus COPD patients 6.4±0.6 L·min−1; not significant), respiratory rate (controls 15.8±1.0 breaths·min−1 versus COPD patients 18.4±1.3 breaths·min−1; not significant) and dead space ventilation (controls 1.8±0.3 L·min−1 versus COPD patients 3.2±0.6 L·min−1; not significant).
In response to NHF, we noticed a significant reduction in minute ventilation in all subjects, as shown in figure 3a (NHF off 5.6±0.4 L·min−1 versus NHF on 4.8±0.4 L·min−1; p<0.05). The reduction in ventilation was associated with a reduction in tidal volumes (figure 3b) from NHF off (340±30 mL) to NHF on (300±30 mL; p<0.05). There was no significant change in respiratory rates (figure 3c) and a minimal reduction in transcutaneous CO2 (figure 3d), which fell from NHF off (41.7±1.1 mmHg) to NHF on (41.3±1.2 mmHg; p<0.05). Changes in these parameters in response to NHF were similar for controls and COPD patients. Of note, there was one participant who had moderate COPD (female, 55 years, forced expiratory volume in 1 s 38% of predicted) who reduced minute ventilation from NHF off (7.7 L·min−1) to NHF on (6.7 L·min−1) due to a 17% decrease in respiratory rate from 18.6 to 15.5 breaths·min−1 without a significant change in her tidal volume (from 413 to 433 mL).
In figure 4a, we show pooled data for alveolar ventilation which remained unchanged in controls and COPD subjects. However, there was a substantial (∼40%) reduction in dead space ventilation (NHF off 2.5±0.4 L·min−1 versus NHF on 1.6±0.4 L·min−1; p<0.05; figure 4b), dead space fraction (NHF off 0.42±0.03 versus NHF on 0.31±0.05; p<0.05; figure 4c) and dead space volume (NHF off 144±19 mL versus NHF on 98±22 mL; p<0.05; figure 4d) with the use of NHF. The reduction in dead space ventilation was similar for both controls and COPD patients. Of note, the individual with moderate COPD who reduced the respiratory rate from 18.6 to 15.5 breaths·min−1 on NHF, as mentioned previously, did not have a significant change in either alveolar ventilation (NHF off 3.3 L·min−1 versus NHF on 3.4 L·min−1) or transcutaneous CO2 (NHF off 40.3 mmHg versus NHF on 39.5 mmHg). However, she had a substantial reduction in dead space ventilation (1.1 L·min−1) that was due to both the reduction in respiratory rate and a reduction in tidal dead space volume (NHF off 236 mL versus NHF on 214 mL).
Determinants of reductions in dead space ventilation
There was no correlation between baseline respiratory rate, minute ventilation or anatomical dead space volume with either the reduction in dead space ventilation or the reduction in tidal dead space volume. In contrast, the degree of physiological dead space at baseline correlated with the reduction of dead space ventilation with NHF: the greater the dead space fraction of an individual at baseline, the greater the reduction in dead space ventilation with NHF. As can be seen in figure 5, for each 10% increase in dead space fraction there is a 208 mL·min−1 reduction in dead space ventilation with the use of NHF (r2=0.36; p<0.05).
Discussion
In the present work, we evaluated the effects of the use of NHF during sleep on ventilation and directly measured CO2 production using a metabolic hood and polysomnography. First, we observed that responses in ventilation to NHF during sleep were similar in COPD patients and controls. Second, the use of NHF led to a substantial decrease in minute ventilation due to a reduction in dead space ventilation without a significant change in alveolar ventilation, CO2 production, energy expenditure or transcutaneous CO2. Third, the reduction in dead space ventilation depended on the physiological dead space fraction and not anatomical dead space volume, respiratory rate or minute ventilation at baseline NHF off. Although we did not observe a reduction in either alveolar ventilation or CO2 production during our short trials of NHF, it is still possible that longer trials of NHF may also affect these parameters.
Several previous studies have demonstrated a washout of the anatomical dead space with NHF using imaging techniques in upper airway models [8] or the human upper airway in volunteers [17]. These studies, however, could not elucidate the effect of NHF on dead space clearance because of technical restrictions, and did not evaluate CO2 production and metabolism [17]. We were able to show that baseline physiological dead space volume is related to the amount of dead space washout at a given flow rate. Physiological dead space is known to directly correlate with tidal volume [23]. We now show that the lower the tidal dead space fraction, the lower the reduction in dead space ventilation with NHF. Thus, it appears that individuals with a higher tidal volume due to increases in the dead space fraction benefit more from the use of NHF compared with those with a lower dead space fraction.
NHF has been extensively used for treating respiratory failure in different clinical settings. The warming and humidification of the air allow the delivery of higher flows, creating a low level of positive end-expiratory pressure and increasing end-expiratory lung volume [10–12]. Furthermore, NHF reduces minute ventilation during both sleep and wakefulness, without increasing CO2 levels [9, 16, 24, 25]. The respiratory pattern response to NHF apparently differs when individuals are awake or sleeping [16], thus explaining potential differences in respiratory rate responses between individuals [16]. To exclude this confounder, we determined ventilatory responses during sleep, and showed that NHF reduced minute ventilation and dead space ventilation similarly in controls and patients with mild to moderate COPD. Moreover, the reduction in dead space ventilation was associated with no change in alveolar ventilation, CO2 production or transcutaneous CO2. This is in contrast to previous studies, in which we and others have shown a significant reduction in work of breathing following the reduction in minute ventilation [4, 9, 13–15].
There are several explanations for these discrepancies. First, it is possible that our on/off trials of NHF were too short to translate into changes in CO2 production and energy expenditure. Second, during sleep, energy expenditure is low and for individuals with a low amount of ventilation, we expect a low level of energy expenditure related to respiratory muscle activity. Thus, changes in work of breathing could have been too small to be detected by measurements of energy expenditure. Third, ventilatory responses to NHF were similar in controls and COPD patients, which suggests that breathing mechanics were not markedly different between these groups. Additionally, the number of patients enrolled was too small to exclude minor differences between the responses in COPD patients and controls. We also could not compare the differential responses between sexes. The complexity of the experimental design prevented the enrolment of a larger sample. Furthermore, our patients did not have severe obstruction. Patients with more severe COPD disease may have shown a differential response to NHF.
There are also several strengths in our study. First, this study performed on/off experiments in a very controlled setting without influencing the variables of the metabolic hood. Second, polysomnography was used to control for biological variables related to changes in sleep stages, movements and conscious influences. As a consequence, we obtained data from stable NREM stage 2–3 sleep in which breathing pattern and metabolism are very regular, allowing the detection of even small changes in CO2 production. Third, we quantified ventilation with calibrated respiratory inductive plethysmography, which is devoid of any artificial changes in ventilation that may arise with the use of either nasal or face masks. Taken together, our data indicate that the immediate ventilatory responses to NHF are due to a reduction in dead space ventilation. Whether these changes translate into reductions in energy expenditure may depend on NHF being used over prolonged periods of time or in patients with high ventilatory demands.
Our study has immediate clinical implications. First, we demonstrate that alveolar ventilation remains constant during the use of NHF and the driver for the reduction in minute ventilation is a reduction in dead space ventilation. Therefore, NHF could be particularly useful in patients with large dead space ventilation, maximising its potential for unloading the respiratory system. Second, we now provide an explanation for the heterogeneous response to NHF observed in previous studies. While some individuals lowered their arterial CO2 in response to NHF, others did lower their respiratory rate [24]. We hypothesise that individuals with higher physiological dead space ventilation showed a reduction in arterial CO2. Finally, although CO2 production remained constant during our short on/off trials of NHF, it is possible that longer use of NHF would lead to reductions in work of breathing and CO2 production, and could prevent nocturnal hypercapnia. However, studies with longer exposure to NHF are required to further evaluate these hypotheses.
In conclusion, we show that the mechanism for minute ventilation reduction during short-term use of NHF during sleep is not related to a reduction in CO2 production. Rather, it is caused by a reduction in dead space ventilation. The degree of the reduction in minute ventilation is correlated to the amount of baseline physiological dead space, measured as the dead space fraction. Therefore, NHF could be used to unload the respiratory system, particularly in patients with a very high dead space fraction.
Footnotes
Conflict of interest: P. Biselli reports grants from the National Institutes of Health (NIH) (HL105546), and grants from the Brazilian funding agencies FAPESP and CNPq, during the conduct of the study. K. Fricke reports grants from the NIH (HL105546), during the conduct of the study. L. Grote reports grants from ResMed Foundation and Philips Foundation, grants, personal fees and nonfinancial support from ResMed (speaker's bureau and study support), personal fees from Philips (speaker's bureau), personal fees from Itamar (speaker's bureau and study support), and personal fees from Weinmann (for consultancy), outside the submitted work. A.T. Braun reports grants from the NIH (HL105546), during the conduct of the study. J. Kirkness has received grants from the NIH (HL105546), and is director of clinical affairs for Fisher & Paykel Healthcare, during the conduct and publication of the study. P. Smith reports grants from the NIH (HL105546), during the conduct of the study. A. Schwartz reports grants from the NIH (HL105546), during the conduct of the study. H. Schneider reports grants from the NIH (HL105546), personal fees for consulting and nonfinancial support with devices in patients with CF from Fisher & Paykel Healthcare, grants from ResMed (for sponsored research on the effect of NHF in COPD), and personal fees from TNI Medical (for consultancy; high flow in COPD), during the conduct of the study; personal fees for consultancy from Fisher & Paykel Healthcare and TNI Medical, outside the submitted work; in addition, H. Schneider has a US patent: 7,080,645 issued to TNI Medical (anti-snoring device, method for reducing snoring, and a nasal air cannula).
Support statement: This study was funded by the Ministério da Ciência, Tecnologia e Inovação, Conselho Nacional de Desenvolvimento Científico e Tecnológico (grant 200817/2012-4), Fundação de Amparo à Pesquisa do Estado de São Paulo (grant 2012/05190-0) and the US Dept of Health and Human Services, National Institutes of Health, National Heart, Lung, and Blood Institute (grant 105546). Funding information for this article has been deposited with the Crossref Funder Registry.