Skip to main content

Main menu

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • ERS Guidelines
  • Authors/reviewers
    • Instructions for authors
    • Submit a manuscript
    • Open access
    • Peer reviewer login
  • Alerts
  • Subscriptions
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

User menu

  • Log in
  • Subscribe
  • Contact Us
  • My Cart

Search

  • Advanced search
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

Login

European Respiratory Society

Advanced Search

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • ERS Guidelines
  • Authors/reviewers
    • Instructions for authors
    • Submit a manuscript
    • Open access
    • Peer reviewer login
  • Alerts
  • Subscriptions

Detecting active pulmonary tuberculosis with a breath test using nanomaterial-based sensors

Morad K. Nakhleh, Raneen Jeries, A'laa Gharra, Anke Binder, Yoav Y. Broza, Mellissa Pascoe, Keertan Dheda, Hossam Haick
European Respiratory Journal 2014 43: 1522-1525; DOI: 10.1183/09031936.00019114
Morad K. Nakhleh
1Dept of Chemical Engineering and Russell Berrie Nanotechnology Institute, Technion - Israel Institute of Technology, Haifa, Israel
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Raneen Jeries
1Dept of Chemical Engineering and Russell Berrie Nanotechnology Institute, Technion - Israel Institute of Technology, Haifa, Israel
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
A'laa Gharra
1Dept of Chemical Engineering and Russell Berrie Nanotechnology Institute, Technion - Israel Institute of Technology, Haifa, Israel
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Anke Binder
2Division of Pulmonology and UCT Lung Institute, Dept of Medicine, Groote Schuur Hospital Observatory, Cape Town, South Africa
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Yoav Y. Broza
1Dept of Chemical Engineering and Russell Berrie Nanotechnology Institute, Technion - Israel Institute of Technology, Haifa, Israel
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Mellissa Pascoe
2Division of Pulmonology and UCT Lung Institute, Dept of Medicine, Groote Schuur Hospital Observatory, Cape Town, South Africa
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Keertan Dheda
2Division of Pulmonology and UCT Lung Institute, Dept of Medicine, Groote Schuur Hospital Observatory, Cape Town, South Africa
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Hossam Haick
1Dept of Chemical Engineering and Russell Berrie Nanotechnology Institute, Technion - Israel Institute of Technology, Haifa, Israel
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • For correspondence: hhossam@tx.technion.ac.il
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

To the Editor:

Detecting active tuberculosis (TB) remains a major global public health challenge [1]. The tuberculin skin test does not distinguish latent from active TB [2]. The interferon-γ release assays have similar limitations [3]. Acid-fast bacilli staining of sputum has a high false-negative rate (up to 50%) [4]. Nucleic acid amplification tests (NAATs), such as GeneXpert MTB/RIF (Cepheid, Sunnyvale, CA, USA), are accurate but require a good infrastructure and the necessity to obtain a good quality sputum sample, which is often unobtainable in more than a third of HIV-infected persons [5, 6].

Given these unmet needs, we explored the use of a novel, rapid, simple and inexpensive point-of-care test for the diagnosis of TB [7]. The approach is based on the detection of volatile organic compounds (VOCs) that are emitted from infected cells and released in exhaled breath [8, 9]. Using gas chromatography linked with mass spectrometry, researchers have previously reported identification of TB-related VOCs in the exhaled breath, though there has been low accuracy in detection (80–85%) [10]. In this study, we explore the possibility of active TB detection via the analysis of exhaled breath using a novel technology of organically modified nanomaterial-based sensors. Such cross-reactive sensors are highly sensitive to the collective changes in the VOCs spectrum [8].

Therefore, we designed a case–control study, in which breath and sputum sampling was performed in 210 adult participants, after informed consent, at three sites in Cape Town, South Africa, between November 2011 and March 2012. The study population consisted of two main subgroups. The first subgroup included those in whom Mycobacterium tuberculosis culture was proven (n=64). The control subgroup consisted of two main subcategories: 1) healthy volunteers (n=67); and 2) a group of TB-negative subjects (n=67). The latter comprised those in whom TB was suspected yet presented a negative smear microscopy, culture and GeneXpert MTB/RIF. The two groups were age and sex matched, yet differed in their smoking habits, HIV infection rates and treatment for TB.

Exhaled alveolar breath was collected in a controlled manner as reported elsewhere [11]. Briefly, inhaled air was cleared of ambient contaminants by repeatedly inhaling via a mouthpiece that contained a charcoal filter cartridge (Eco Medics, Dürnten, Switzerland), then, 750 mL of end tidal expired air was collected and trapped in two-bed ORBO 420 Tenax TA sorption tubes for gas sampling (Sigma-Aldrich, Shanghai, China). Sealed tubes were stored under refrigeration at 4°C, until the day of analysis.

Thermal desorption was used to release the VOCs trapped in the tubes and delivered into a stainless-steel test chamber containing 12 nanomaterial-based sensors (S1–S12), which are based on either: 1) molecularly-modified gold nanoparticles; or 2) molecularly-modified single-walled carbon nanotubes. More details on the fabrication and modification of these sensors can be found in [12, 13]. The interaction between the films of the modified particles and the VOCs results in a time-dependent and reversible change in resistance, which was recorded and then normalised according to a daily calibration measurement of fixed standard.

Receiver operating characteristic (ROC) curves were applied to all sensors. The sensitivity, specificity, accuracy, positive predictive value (PPV) and negative predictive value (NPV) were calculated for both the training and validation sets. Statistical tests were performed using SAS JMP, version 10.0 (SAS Institute Inc., Marlow, UK).

Of the eligible 198 samples, 138 (44 TB positive, 47 TB negative and 47 healthy) were of known identity and were used as training set. The remaining independent 60 samples, of unknown identity (20 of each subgroup), were used for blind validation.

We used the ROC-derived Youden's index as a cut-point (best sensitivity + specificity − 1) to evaluate the examined sensors as a binary classifier. Three of the 12 sensors showed >80% accuracy in classification. However, the best classifier was obtained from S1, a chemiresistor based on dodecanthiol-capped gold nanoparticles. S1 correctly classified 121 of the 138 training samples, an accuracy of 88%; the sensitivity was 85% and the specificity 89% (fig. 1a). The ROC's area under curve (AUC) was 94.8% (fig. 1b). In the validation set, sensitivity was 90%, specificity was 93% and accuracy was 92%. Moreover, when applying a cut-off to rule out TB (a threshold scoring 95% sensitivity in the training phase), the specificity, PPV and NPV were 83%, 78%, and 96%, respectively, in the training set. In the blinded validation set, the relevant values were 88%, 76% and 94%, respectively. In the threshold to rule-in TB (a specificity of 96% in the training phase), the sensitivity, PPV and NPV were 68%, 93% and 87%, respectively, in the training set. In the blind test, the relevant values were 78%, 76% and 94%, respectively (fig. 1a).

Figure 1–
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 1–

a) Dot-plots of the response of sensor S1. Each circle/star represents one sample. The central dashed line represents Youden's cut-point, and the outer dashed lines represents the cut-points to rule-in and rule-out tuberculosis. Samples from the validation set with responses lower than the threshold were classified as tuberculosis positive (open stars) or non-tuberculosis positive (closed stars) according to Youden's cut-point. b) Receiver operating characteristic (ROC) curve of sensor S1. C) ROC curves of sensor S1 when comparing smoking, HIV and treatment status. AUC: area under curve.

The effects of potential confounding factors were examined. S1 showed a random classification ability between smokers versus non-smokers, HIV-positive versus HIV-negative subjects and treated versus untreated TB (ROC AUC of 56%, 54% and 55%, respectively) (fig. 1c). This proved that the results were unaffected by confounding factors.

In the present study, we used NAAT as a complementary diagnostic test to verify a true-negative result in subjects who were found to be negative with a sputum test [6], and were known to have relatively low sensitivity, yet a very high (∼100%) specificity [4].

One obvious advantage of breath analysis over sputum tests is the simplicity of obtaining the sample; this is in contrast to the difficulties in obtaining sputum both in children and HIV patients [5, 6]. The lack of confounding by HIV is another clear advantage over the existing TB diagnostic tests, which exhibit up to 43% misclassification in HIV-TB co-infected persons [14]. These results could be explained by the fact that nanomaterial-based sensors are broadly cross-reactive, so that each responds to a substantial number of VOCs of interest (e.g. 2,4-dimethylheptane, acetylaldhedye and 2-butanone [10]). Thus, the response is the result of the whole mixture of breath VOCs of interest, rather than a single VOC that may be linked with HIV [8, 15].

Our study has several limitations. The case–control design may over-estimate sensitivity and specificity; thus, our results must be interpreted with caution. We also didn't exclude latent TB cases. The presented results only apply to a high HIV-TB setting in Africa; further studies are therefore required in different settings. Due to safety regulations, healthy subjects were sampled in a separate location from subjects with suspected TB, which could raise concerns regarding possible geographical bias affecting the results. We compared the healthy samples collected in one location with TB-negative samples collected in other locations. The results did not show any difference between the two control subgroups (AUC ROC 58%), thus dismissing the fear of a potential geographical effect. Another limitation of the study is that many TB patients were treated prior to breath sampling (median of 5 days), which may have influenced our results. However, this was unavoidable for ethical reasons, and the analysis of the subgroups did not show any effect of TB treatment.

In summary, as TB diagnostics are usually either inaccurate or too expensive and complicated for use as a point-of-care detection tool, we examined a noninvasive, relatively simple and cheap (US$1 per sensor) methodology to fill the gap. The preliminary proof-of-concept results indicate that a nanomaterial-based sensor may be a useful rule-in or rule-out tool for the diagnosis of TB (92% accuracy). A larger cohort study is required to validate the results.

Acknowledgments

The authors acknowledge Dr Falk Fish from Orgenics Ltd (Yavne, Israel). Orgenics Ltd have expressed interest in commercialising this technology and provided funding this study. The company played no role in the study design, analysis of the data, or drafting of the manuscript.

Footnotes

  • Conflict of interest: Disclosures can be found alongside the online version of this article at www.erj.ersjournals.com

  • Received January 29, 2014.
  • Accepted February 14, 2014.
  • ©ERS 2014

References

  1. ↵
    World Health Organization. Global Tuberculosis Control 2010. Geneva, World Health Organization, 2010.
  2. ↵
    American Thoracic Society/Centers for Disease Control and Prevention. Targeted tuberculin testing and treatment of latent tuberculosis infection. ATS/CDC Statement Committee on Latent Tuberculosis Infection membership list, June 2000. http://www.cdc.gov/MMWR/PREVIEW/MMWRHTML/rr4906a1.htm.
  3. ↵
    1. Dheda K,
    2. van Zyl Smit R,
    3. Badri M,
    4. et al
    . T-cell interferon-γ release assays for the rapid immunodiagnosis of tuberculosis: clinical utility in high-burden vs. low-burden settings. Curr Opin Pulm Med 2009; 15: 188–200.
    OpenUrlCrossRefPubMedWeb of Science
  4. ↵
    1. Lawn SD,
    2. Zumla AI
    . Tuberculosis. Lancet 2011; 378: 57–72.
    OpenUrlCrossRefPubMedWeb of Science
  5. ↵
    1. Whiley DM,
    2. Tapsall JW,
    3. Sloots TP
    . Nucleic acid amplification testing for Neisseria gonorrhoeae: an ongoing challenge. J Mol Diagn 2006; 8: 3–15.
    OpenUrlCrossRefPubMedWeb of Science
  6. ↵
    1. Theron G,
    2. Peter J,
    3. Meldau R,
    4. et al
    . Accuracy and impact of Xpert MTB/RIF for the diagnosis of smear-negative or sputum-scarce tuberculosis using bronchoalveolar lavage fluid. Thorax 2013; 68: 1043–1051.
    OpenUrlAbstract/FREE Full Text
  7. ↵
    1. Wallis RS,
    2. Pai M,
    3. Menzies D,
    4. et al
    . Biomarkers and diagnostics for tuberculosis: progress, needs, and translation into practice. Lancet 2010; 375: 1920–1937.
    OpenUrlCrossRefPubMedWeb of Science
  8. ↵
    1. Broza YY,
    2. Kremer R,
    3. Tisch U,
    4. et al
    . A nanomaterial-based breath test for short-term follow-up after lung tumor resection. Nanomedicine 2013; 9: 15–21.
    OpenUrlCrossRefPubMed
  9. ↵
    1. Wheelock CE,
    2. Goss VM,
    3. Balgoma D,
    4. et al
    . Application of 'omics technologies to biomarker discovery in inflammatory lung diseases. Eur Respir J 2013; 42: 802–825.
    OpenUrlAbstract/FREE Full Text
  10. ↵
    1. Phillips M,
    2. Cataneo RN,
    3. Condos R,
    4. et al
    . Volatile biomarkers of pulmonary tuberculosis in the breath. Tuberculosis 2007; 87: 44–52.
    OpenUrlCrossRefPubMedWeb of Science
  11. ↵
    1. Peled N,
    2. Hakim M,
    3. Bunn PA Jr.,
    4. et al
    . Non-invasive breath analysis of pulmonary nodules. J Thorac Oncol 2012; 7: 1528–1533.
    OpenUrlCrossRefPubMed
  12. ↵
    1. Dovgolevsky E,
    2. Tisch U,
    3. Haick H
    . Chemically sensitive resistors based on monolayer-capped cubic nanoparticles: towards configurable nanoporous sensors. Small 2009; 5: 1158–1161.
    OpenUrlPubMedWeb of Science
  13. ↵
    1. Peng G,
    2. Tisch U,
    3. Adams O,
    4. et al
    . Diagnosing lung cancer in exhaled breath using gold nanoparticles. Nat Nanotechnol 2009; 4: 669–673.
    OpenUrlCrossRefPubMed
  14. ↵
    1. Elliott A,
    2. Namaambo K,
    3. Allen B,
    4. et al
    . Negative sputum smear results in HIV-positive patients with pulmonary tuberculosis in Lusaka, Zambia. Tubercle and Lung Disease 1993; 73: 191–194.
    OpenUrl
  15. ↵
    1. Greulich T,
    2. Hattesohl A,
    3. Grabisch A,
    4. et al
    . Detection of obstructive sleep apnoea by an electronic nose. Eur Respir J 2013; 42: 145–155.
    OpenUrlAbstract/FREE Full Text
View Abstract
PreviousNext
Back to top
View this article with LENS
Vol 43 Issue 5 Table of Contents
European Respiratory Journal: 43 (5)
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
Email

Thank you for your interest in spreading the word on European Respiratory Society .

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Detecting active pulmonary tuberculosis with a breath test using nanomaterial-based sensors
(Your Name) has sent you a message from European Respiratory Society
(Your Name) thought you would like to see the European Respiratory Society web site.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Print
Citation Tools
Detecting active pulmonary tuberculosis with a breath test using nanomaterial-based sensors
Morad K. Nakhleh, Raneen Jeries, A'laa Gharra, Anke Binder, Yoav Y. Broza, Mellissa Pascoe, Keertan Dheda, Hossam Haick
European Respiratory Journal May 2014, 43 (5) 1522-1525; DOI: 10.1183/09031936.00019114

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero

Share
Detecting active pulmonary tuberculosis with a breath test using nanomaterial-based sensors
Morad K. Nakhleh, Raneen Jeries, A'laa Gharra, Anke Binder, Yoav Y. Broza, Mellissa Pascoe, Keertan Dheda, Hossam Haick
European Respiratory Journal May 2014, 43 (5) 1522-1525; DOI: 10.1183/09031936.00019114
del.icio.us logo Digg logo Reddit logo Technorati logo Twitter logo CiteULike logo Connotea logo Facebook logo Google logo Mendeley logo
Full Text (PDF)

Jump To

  • Article
    • Acknowledgments
    • Footnotes
    • References
  • Figures & Data
  • Info & Metrics
  • PDF

Subjects

  • Respiratory infections and tuberculosis
  • Tweet Widget
  • Facebook Like
  • Google Plus One

More in this TOC Section

Agora

  • Airway immune responses to COVID-19 vaccination in COPD patients
  • Wider access to rifapentine-based regimens is needed for TB care globally
  • Association between immunosuppressants and outcomes of COVID-19
Show more Agora

Research letters

  • Lessons from 2 years of use of the Post-COVID-19 Functional Status scale
  • Beta-blockade improves right ventricular diastolic function in exercising PAH
  • Trends in COVID-19-associated mortality in PH
Show more Research letters

Related Articles

Navigate

  • Home
  • Current issue
  • Archive

About the ERJ

  • Journal information
  • Editorial board
  • Press
  • Permissions and reprints
  • Advertising

The European Respiratory Society

  • Society home
  • myERS
  • Privacy policy
  • Accessibility

ERS publications

  • European Respiratory Journal
  • ERJ Open Research
  • European Respiratory Review
  • Breathe
  • ERS books online
  • ERS Bookshop

Help

  • Feedback

For authors

  • Instructions for authors
  • Publication ethics and malpractice
  • Submit a manuscript

For readers

  • Alerts
  • Subjects
  • Podcasts
  • RSS

Subscriptions

  • Accessing the ERS publications

Contact us

European Respiratory Society
442 Glossop Road
Sheffield S10 2PX
United Kingdom
Tel: +44 114 2672860
Email: journals@ersnet.org

ISSN

Print ISSN:  0903-1936
Online ISSN: 1399-3003

Copyright © 2023 by the European Respiratory Society