Skip to main content

Main menu

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • ERS Guidelines
  • Authors/reviewers
    • Instructions for authors
    • Submit a manuscript
    • Open access
    • Peer reviewer login
    • WoS Reviewer Recognition Service
  • Alerts
  • Subscriptions
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

User menu

  • Log in
  • Subscribe
  • Contact Us
  • My Cart
  • Log out

Search

  • Advanced search
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

Login

European Respiratory Society

Advanced Search

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • ERS Guidelines
  • Authors/reviewers
    • Instructions for authors
    • Submit a manuscript
    • Open access
    • Peer reviewer login
    • WoS Reviewer Recognition Service
  • Alerts
  • Subscriptions

Noninvasive cardiac output measurements in patients with pulmonary hypertension

Jonathan D. Rich, Stephen L. Archer, Stuart Rich
European Respiratory Journal 2013 42: 125-133; DOI: 10.1183/09031936.00102212
Jonathan D. Rich
Dept of Medicine, Section of Cardiology, University of Chicago Medical Center, Chicago, IL, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • For correspondence: jonathan.rich@uchospitals.edu
Stephen L. Archer
Dept of Medicine, Section of Cardiology, University of Chicago Medical Center, Chicago, IL, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Stuart Rich
Dept of Medicine, Section of Cardiology, University of Chicago Medical Center, Chicago, IL, USA
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

Pulmonary hypertension (PH) is characterised by a progressive decline in cardiac output (CO) and right heart failure. NICOM® (noninvasive cardiac output monitor) is a bioreactance-based technology that has been broadly validated, but its specific application in right heart failure and PH is unknown.

Cardiac catheterisation was performed in 50 consecutive patients with PH. CO measurements were performed using three different methods (thermodilution, Fick and NICOM) at baseline and after vasodilator challenge. We compared the precision (coefficient of variation) and accuracy of NICOM compared to thermodilution and Fick.

The mean CO (L·min−1) at baseline as measured by the three methods was 4.73±1.15 (NICOM), 5.69±1.74 (thermodilution) and 4.84±1.39 (Fick). CO measured by NICOM was more precise than by thermodilution (3.5±0.3% versus 9.6±6.1%, p<0.001). Bland–Altman analyses comparing NICOM to thermodilution and Fick revealed bias and 95% limits of agreement that were comparable to those comparing Fick to thermodilution. All three CO methods detected an increase in CO in response to vasodilator challenge.

CO measured via NICOM is precise and reliably measures CO at rest and changes in CO with vasodilator challenge in patients with PH. NICOM may allow for the noninvasive haemodynamic assessment of patients with PH and their response to therapy.

Introduction

Pulmonary arterial hypertension (PAH) is a disease characterised by an elevated pulmonary vascular resistance (PVR) that ultimately results in a progressive decline in cardiac output (CO) due to right ventricular failure [1]. Whereas the severity of pulmonary artery pressure elevation does not reliably predict mortality, the status of the right ventricle as measured by CO has consistently been shown to be among the strongest predictors of outcomes in this fatal disease [2, 3]. From a clinical perspective, the ability to measure and serially follow the CO response to treatment is of significant value, perhaps even more so than serial measurements of pulmonary artery pressure. However, the measurement of CO has traditionally required cardiac catheterisation, which has inherent risks and expense [4]. Thus, the ability to serially measure CO noninvasively in patients with PAH is clinically attractive.

Transthoracic bioreactance is a recently introduced technology that allows for the noninvasive measurement of CO (NICOM®; Cheetah Medical Inc., Portland, OR, USA) [5, 6]. The bioreactance signal is determined by measuring blood flow-dependent changes in the phase shifts between an oscillating electrical current applied across the thorax and the resulting voltage signal. This signal has been shown to be directly proportional to aortic blood flow [5]. Importantly, this technology differs from bioimpedance, which relies on the detection of changes in voltage signal amplitude, which has been shown to be limited by factors such as body habitus, pleural effusions and body motion [7–9]. Because bioreactance is generally unaffected by these factors, it yields a favourable signal-to-noise ratio [5]. The accuracy and precision of the NICOM device has been validated against invasive measurements of CO in several different clinical settings [6, 10]. However, the ability of NICOM to reliably measure CO in patients with pulmonary hypertension (PH) and right heart failure has never been evaluated. Therefore, since it is unknown how changes in pulmonary blood flow may affect the measurement of CO with NICOM, we tested the accuracy of NICOM to measure CO as compared to the reference standards of thermodilution (TD) and the indirect Fick method in a consecutive cohort of patients with PH.

Methods

The current study was approved by the University of Chicago Institutional Review Board (IRB # 10-179-B). Written informed consent was obtained from all patients.

We enrolled 50 consecutive patients referred to the cardiac catheterisation laboratory for a haemodynamic assessment of presumed or previously confirmed PH. We included patients from all five World Health Organization (WHO) PH groups, but excluded patients who carried a diagnosis of WHO group 1 PAH secondary to congenital heart disease, to eliminate technological issues with all three methods that may be affected by the presence of an intracardiac shunt. With each patient lying quietly and supine on the catheterisation table, venous access was achieved with an 8-French sheath placed in either the internal jugular or femoral vein. A 7.5-French Swan–Ganz catheter (Edwards Lifesciences, Irvine, CA, USA) was inserted, followed by measurements of right atrial, right ventricular, pulmonary artery and pulmonary capillary wedge pressures. After the pressure measurements were performed, we measured CO as close to simultaneously as possible according to the NICOM, TD and indirect Fick methods (see below). The same clinical investigator, expertly trained and experienced in the performance of cardiac catheterisation, performed every measurement in this study.

NICOM

Noninvasive bioreactance CO monitoring was performed using the NICOM system, the methodology of which has been described previously [5]. The NICOM calculation of stroke volume (SV) is given by:Embedded Image

where C is a constant of proportionality and VET is ventricular ejection time, which is determined from the NICOM and ECG signals. The value of C accounts for patient age, sex and body size [6]. Finally, CO is calculated as the product of SV and heart rate.

At the start of the preparation of each case, the NICOM device was connected to each patient by placing the four double electrodes on the patient's thorax, which were connected to the NICOM device by four wires. After the NICOM electrodes were applied, the device underwent a 1-min autocalibration and the patient was subsequently prepared and draped in sterile fashion for cardiac catheterisation. The NICOM measured the CO continuously with average values displayed every minute throughout the procedure. The CO data were electronically stored within the monitor.

Thermodilution

CO via TD was performed via the indicator-dilution technique [11]. For each patient, we confirmed the proper positioning of the Swan–Ganz catheter by documenting the presence of a right atrial pressure waveform from the proximal port of the Swan–Ganz catheter and fluoroscopic visualisation of the distal tip of the catheter in the mid-pulmonary artery before every measurement. CO was measured by the injection of 10 mL of sterile, isotonic (0.9%) saline, injected through the proximal lumen of the catheter, and the time-course of change of temperature was recorded at the distal thermistor. Three consecutive bolus saline injections were performed for individual estimations of CO, and the mean value of the three measurements used as the final value for CO.

Indirect Fick method

Total body oxygen consumption (V′O2/BSA), in mL·min−1·m−2, was estimated via the formula of LaFarge and Miettinen [12]:Embedded Image

where V′O2 is oxygen consumption in mL·min−1, age is presented in years, heart rate (HR) is in units of beats per minute, and BSA is body surface area in m2.

Simultaneous blood samples were taken from the aorta and pulmonary artery for determination of the oxygen saturation and haemoglobin concentrations. CO was calculated according to the Fick equation:Embedded Image

where CO is defined as CO in L·min−1, V′O2 is oxygen consumption in L·min−1·m−2, CaO2 is arterial oxygen content in mg·L−1 (1.36×Hbg×SaO2)+(PaO2×0.003) and CvO2 is defined as mixed venous oxygen content in mg·L−1 (1.36×Hbg×SvO2)+(PvO2×0.003). Hbg is the haemoglobin level in g·L−1, SaO2 is the arterial oxygen saturation, PaO2 is the arterial oxygen tension in mmHg, SvO2 is the mixed venous oxygen saturation and PvO2 is the mixed venous oxygen tension in mmHg.

Simultaneous CO measurements

In order to perform the CO measurements as nearly simultaneously as possible, blood samples for the arterial saturations were drawn immediately prior to the initial TD bolus injection. Because the NICOM device is continuously sampling and providing a measured CO every minute, all NICOM measurements were determined based on the value displayed on the monitor as close as possible to each bolus TD injection. Thus, in each patient, we were able to consistently make the CO measurements according to each modality within ∼1–5 min of one another. This was critical in order to best ensure that all measurements were obtained as simultaneously as possible, thus avoiding any significant changes in the clinical status of the patient during the course of the measurements.

Acute vasodilator challenge

Vasodilator testing using i.v. adenosine was performed as clinically indicated (n=36). All vasodilator studies were initiated at a dose of 50 μg·kg−1·min−1 and up-titrated to a maximum dose of 150 μg·kg−1·min−1 or to the highest dose tolerated by the patient (whichever occurred first). After achieving a steady state at the final adenosine dose, haemodynamic measurements were repeated followed by a repeat of the near-simultaneous measurements of CO by NICOM, TD and Fick methods, as described above.

Echocardiography

The majority of patients in the study underwent cardiac imaging using 2D transthoracic echocardiography within 30 days of the cardiac catheterisation (n=44). We tabulated the results of assessments made of overall right ventricle size and function, presence and severity of tricuspid regurgitation (TR) and pulmonic insufficiency, respectively, to see if they affected the CO measurements by any of the techniques.

Statistical analyses

Baseline demographic and clinical variables were collected. Continuous variables are described as mean±sd and categorical variables are described as percentages. Correlations between parameters of interest were determined according to the Pearson correlation method. A t-test and/or a one-way repeated-measure ANOVA with Bonferroni correction was performed to determine statistical significance of differences between continuous variables. A p-value <0.05 was considered statistically significant. Coefficient of variance was calculated to determine the precision of the NICOM and TD methods, respectively. Bland–Altman analyses were performed to estimate the general degree of agreement among the three methods. The mean bias and 95% limits of agreements were calculated. Sensitivity and specificity for directional change in response to adenosine vasodilator challenge for each CO method was calculated. A false-positive directional change was considered to have occurred when an increase in CO of ≥10% was not detected by the remaining two modalities and a false-negative change was considered to have occurred when the CO method did not detect the increase in CO of ≥10% while the other two methods did. All measured CO values were included in these analyses, even when obvious outliers occurred.

Results

Baseline demographics of the patient cohort are shown in table 1 and key haemodynamic and echocardiographic data are shown in tables 2 and 3. The mean±sd age of the patients was 54±15 years and the majority of patients had WHO group 1 PAH. The majority of patients had moderate-to-severe PH as demonstrated by the elevated mean pulmonary artery pressure (40.7±13.7 mmHg) and PVR (6.0±4.3 Wood Units) and most patients had at least moderate-to-severely reduced right ventricular function by echocardiography. Although almost all patients had some degree of TR, severe TR was relatively uncommon (n=5, 11.6%).

View this table:
  • View inline
  • View popup
Table 1– Baseline characteristics of the pulmonary hypertension cohort
View this table:
  • View inline
  • View popup
Table 2– Invasive haemodynamic measurements in the pulmonary hypertension cohort
View this table:
  • View inline
  • View popup
Table 3– Echocardiographic findings in the pulmonary hypertension cohort

The mean±sd CO at baseline as measured by the three methods was 4.73±1.15 L·min−1 by NICOM, 5.69±1.74 L·min−1 by TD and 4.84±1.39 L·min−1 by Fick. There was no difference in CO between NICOM and Fick (p=0.58), but CO according to TD was higher than CO measured by both NICOM and Fick (p<0.01 for both comparisons) (fig. 1). NICOM measurements were significantly more precise than TD (coefficient of variation 3.5±0.3% versus 9.6±6.1%, respectively; p<0.001) (fig. 2). The above findings were similar when we restricted our analyses to the subgroup of patients with WHO group 1 PAH only.

Figure 1–
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 1–

Comparison of simultaneous measurements of cardiac output at rest between noninvasive cardiac output measurement (NICOM®), thermodilution (TD) and Fick methods in patients with pulmonary hypertension. The mean resting cardiac output measured by TD tended to be slightly higher than both NICOM and Fick (p<0.01 for both comparisons). Conversely, there was no difference in resting cardiac output between NICOM and Fick. Data are presented as median (line), interquartile range (box), range (whiskers) and outliers (>1.5× interquartile range above upper quartile).

Figure 2–
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 2–

Precision of cardiac output measurements at rest using noninvasive cardiac output measurement (NICOM®) compared with thermodilution (TD) in patients with pulmonary hypertension. Cardiac output measurements are significantly more precise with NICOM as compared with TD. p<0.001. The figures for coefficient of variation quoted in the text are mean values, whereas the horizontal lines in these plots represent median values. Data are presented as median (line), interquartile range (box), range (whiskers) and outliers (>1.5× interquartile range above upper quartile).

A significant correlation was seen among all three CO methods: NICOM and TD (r=0.60, p<0.001), NICOM and Fick (r=0.54, p<0.001) and TD and Fick (r=0.83, p<0.001). Bland–Altman analyses (fig. 3) revealed the following: NICOM compared to TD showed a mean bias of -0.81 with 95% limits of agreement of -3.54–1.92. NICOM compared to Fick showed a mean bias of 0.02 with 95% limits of agreements of -2.41–2.44. TD compared to Fick showed a mean bias of 0.83 with 95% limits of agreement of -0.98–2.63.

Figure 3–
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 3–

Bland–Altman plots of a) noninvasive cardiac output measurement (NICOM®) versus thermodilution (TD); b) NICOM versus Fick; and c) TD versus Fick in patients with pulmonary hypertension. TD measurements of cardiac output tended to be higher than those measured by NICOM or Fick (mean bias of -0.81 L·min−1 and 0.83 L·min−1, respectively) whereas a negligible difference was observed between NICOM and Fick (mean bias of 0.02 L·min−1). All three modalities demonstrated sufficiently wide 95% limits of agreement, thus cautioning against the interchangeability of one method with another in an individual patient.

Similar findings were seen in those patients undergoing acute vasodilator challenge (n=36). Following vasodilator challenge, the mean CO was 5.53±1.46 L·min−1 (NICOM), 7.02±1.84 L·min−1 (TD) and 5.83±1.75 L·min−1 (Fick). There was no difference in CO between NICOM and Fick (p=0.22) but CO measured by TD was higher than CO measured by both NICOM and Fick (p<0.01 for both comparisons). NICOM measurements were significantly more precise than TD (3.4±2.1% versus 8.0±6.4%, p<0.001). All three CO methods detected a mean increase in CO: 18.8±16.8% (NICOM), 26.8±22.2% (TD) and 21.0±19.2% (Fick). Taken together, all three methods detected an increase in CO ≥10% of its baseline value in 24 out of 36 cases. In three out of 36 cases, all three methods determined that the CO did not increase by ≥10%. In the remaining cases (n=9), there was disagreement in the directional change of similar magnitude (i.e. ≥10%). In four cases, NICOM did not detect the increase in CO but TD and Fick both did. In four cases, TD did not detect the increase in CO but NICOM and Fick both did. In one case, Fick did not detect the increase in CO but NICOM and TD both did. Thus, the sensitivity and specificity, respectively, of detecting a directional change in CO according to each method was: 88.9% and 100% for NICOM, 88.9% and 100% for TD, and 97.2% and 100% for Fick. There was not a single case where only one of the methods detected an increase in CO while the other methods did not.

Discussion

We demonstrate that a noninvasive measurement of CO in patients with PH is feasible and produces results comparable to the existing invasive “reference standard” methods commonly used in clinical practice. Moreover, NICOM demonstrated superior precision to TD and reliably detected dynamic, directional changes in CO following vasodilator challenge. Among the many variables obtained during an invasive haemodynamic assessment, the CO has arguably the most prognostic relevance and affects clinical decision strategies more than any other value in patients with PH [2, 13–15]. Accordingly, having the ability to easily measure and detect changes in CO noninvasively may have importance in the clinical care and assessment of treatments in patients with PH.

Numerous studies over the years have compared different technologies to measure CO against one another in a variety of clinical settings, although few studies have specifically addressed this question in patients with PH [16]. Despite the multitude of studies, equipoise persists regarding the superiority (or lack thereof) of one particular method over another. While the gold standard of CO determination is the direct Fick method, this method is rarely used in the clinical setting because of the technical and logistical demands involved in directly measuring V′O2, which makes this method impractical. As a result, the current reference standards most commonly employed in clinical practice in the evaluation of patients with PH, and the cardiac output methods employed in the largest PH registries, are the indirect Fick and TD methods, both of which are invasive and possess certain inherent limitations [2, 14, 15, 17]. For example, TD suffers from imprecision because of the potential influences from catheter migration, differences in injector technique (with intra- and inter-investigator variability), the influence of the different phases of the respiratory cycle, and other factors [18–21]. The lack of precision seen with TD also affects the usefulness of using TD as a reference technique when comparing with other methods of measurement of CO [22]. Some investigators have found TD to be less reliable in the setting of severe TR and/or low CO due to loss of indicator [23, 24], although others have not found this to be true [25–27]. These conditions are particularly common in patients with PH.

The indirect Fick method also has limitations that may affect measurement accuracy, most notably the need to estimate V′O2, which when estimated erroneously will lead to an error of the same magnitude in the estimation of CO. Other factors that may affect the Fick method include errors in the measurement of oxygen saturations and haemoglobin levels and the variable influence of bronchial and thebesian venous drainage on the mixed venous saturation [26, 28, 29]. The indirect Fick method, however, is likely to be highly accurate and reliable in assessing percentage changes in CO in response to an acute intervention, since the only value that often changes in response to the intervention is the mixed venous saturation, and thus the other factors that go into the estimation will cancel out when calculating percentage changes in CO. Hoeper et al. [26] compared CO measured by TD and by Fick in a cohort of patients with PAH and found that both methods correlated well, and that the presence of TR did not appear to influence the CO as determined by either method. However, the wide limits of agreement as seen with Bland–Altman analyses led the authors of that study to conclude that TD and Fick should not be considered interchangeable. The findings in our study are similar to those by Hoeper et al. [26], in that all three methods of CO correlated with one another and all methods reliably detected a change in CO from drug challenge. However, while all three methods showed acceptable overall agreement, the 95% limits of agreement by Bland–Altman analyses were also sufficiently wide that we too caution against assumption of interchangeability between the methods [30]. In this study, although the vast majority of patients carried a diagnosis of WHO group 1 PAH, we specifically included patients with PH from all five WHO groups, including those with lung disease, left ventricular diastolic dysfunction and obesity, to ensure the ability to apply NICOM across the entire PH spectrum. Also, similar to the findings of Hoeper et al. [26], the presence of severe TR did not seem to influence the CO measurements, although this finding should be interpreted with caution in our study as this subgroup of patients was small (11.6%).

Although no single method of CO measurement employed in current clinical practice can directly and precisely measure CO, the current reference standards (i.e. TD and indirect Fick) are still heavily relied upon in the evaluation of PH patients because they provide reasonably precise CO measurements and are able to consistently detect directional changes in CO. These two factors (i.e. precision and the ability to detect clinically relevant changes) are arguably the most important features of a CO methodology [31]. Because CO, as measured by either TD or indirect Fick, is an important predictor of outcomes in PH [2, 14, 15], clinical PH specialists often track changes in CO over time to assess response to therapy or the need to escalate care. For example, McLaughlin et al. [13] showed that an improvement in cardiac index after initiation of epoprostenol therapy in patients with PAH predicted improved survival.

The purpose of the present study was not to validate the NICOM technique itself, as this technology has been previously validated against a variety of different CO methods in animals and in humans in the intensive care unit and other clinical settings [6, 10, 32, 33]. Rather, this is the first study to assess its ability to measure CO in patients with PH and right ventricular dysfunction. Similar to those studies, we show that NICOM in PH patients is more precise than TD and that all three methods show comparable sensitivity and specificity in detecting even small changes in CO in the majority of patients, thus potentially allowing the clinician to track response to therapy and disease progression in the ambulatory setting.

The success of the bioreactance technology used by NICOM to measure CO in comparison to the noninvasive bioimpedance technology is largely attributable to the more favourable signal-to-noise ratio in bioreactance. Whereas bioimpedance measures changes in signal amplitude, bioreactance measures changes in signal frequency (analogous to the difference in signal strength in AM versus FM radio) [5, 6]. As such, the precision of bioimpedance has been shown to be negatively affected by any variable that affects the signal amplitude, including relative distance of electrode placement, the presence of pulmonary oedema and/or pleural effusions, and body habitus [34]. In contrast, the bioreactance technology was developed specifically to overcome the limitations that hindered the use of bioimpedance, while retaining the simple and noninvasive nature of the CO measurement methodology, thus improving its clinical utility across many clinical settings [5]. While the bioreactance technology has helped overcome certain obstacles encountered by previous noninvasive technologies, it too has potential limitations. It should be noted that the system detects and mutes itself during periods of electrocautery when used continuously (i.e. in the operating room), and external pacemakers can potentially interfere with signal quality. Also, the device has not been tested in the setting of severe aortic insufficiency, which could theoretically result in an overestimation of the net forward CO (although this is rarely present in patients with PAH).

The results of this study suggest many possible clinical applications of NICOM in patients with PH. Because CO measured by NICOM can be performed quickly and noninvasively, it could be performed in the ambulatory setting, thus allowing for serial measurements of CO to track response to therapy and disease progression. Given the uncertainty over the use of multiple pulmonary vasodilators to treat PH patients [14] and the continued high mortality in patients with current treatments [15], NICOM has the potential to favourably influence the clinical management of symptomatic patients, although this would need to be rigorously studied. It could also be applied to patients admitted to the intensive care unit with acute decompensated right ventricular failure because of its ability to provide rapid, continuous CO measurements, thus potentially avoiding invasive procedures in these very sick patients. Finally, another potentially useful application of NICOM would be during i.v. prostacyclin initiation, when close monitoring of CO is strongly advised [1].

Limitations

This study has several important limitations. First, this was a single-centre study, which may limit the generalisability of the findings. Secondly, in addition to TD, we also used the indirect Fick CO method and thus did not directly measure V′O2. However, this was not unintentional. The direct Fick method is cumbersome and generally impractical for daily use in busy clinical practices. Thus, because the technique is not widely used clinically in PH patients, we felt that a comparison of NICOM to the currently accepted reference standards (i.e. TD and indirect Fick) was most appropriate. Also, we performed the TD CO measurements in triplicates. While this conforms with the practice of many clinicians and investigators, others suggest that taking the average of five measurements would yield more precise results. Finally, although we attempted to perform exactly simultaneous CO measurements to compare the three methods, the measurements were made within a few minutes of each other. Thus, although the patients remained stable and in the supine position, slight changes in CO may have occurred and produced disagreements in some of the measurements.

Conclusions

The noninvasive measurement of CO with NICOM in PH patients is feasible, precise and reliably detects clinically relevant changes. Although practice guidelines stress the importance of an initial cardiac catheterisation to confirm the diagnosis of PH and assess response to vasodilators, this technology could minimise the need for serial invasive measurements to determine disease progression and response to therapy. While the use of the NICOM in the clinical management of patients with PH has not been studied, the noninvasive bioreactance technology should be added to the array of CO monitoring tools used in PH patients and further studies evaluating this promising technology in patients with PH are warranted.

Footnotes

  • Conflict of interest: None declared.

  • Received July 3, 2012.
  • Accepted August 27, 2012.
  • ©ERS 2013

References

  1. ↵
    1. McLaughlin VV,
    2. Archer SL,
    3. Badesch DB,
    4. et al
    . ACCF/AHA 2009 expert consensus document on pulmonary hypertension a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents and the American Heart Association developed in collaboration with the American College of Chest Physicians; American Thoracic Society, Inc.; and the Pulmonary Hypertension Association. J Am Coll Cardiol 2009; 53: 1573–1619.
    OpenUrlCrossRefPubMedWeb of Science
  2. ↵
    1. D'Alonzo GE,
    2. Barst RJ,
    3. Ayres SM,
    4. et al
    . Survival in patients with primary pulmonary hypertension. Results from a national prospective registry. Ann Intern Med 1991; 115: 343–349.
    OpenUrlCrossRefPubMedWeb of Science
  3. ↵
    1. Sandoval J,
    2. Bauerle O,
    3. Palomar A,
    4. et al
    . Survival in primary pulmonary hypertension. Validation of a prognostic equation. Circulation 1994; 89: 1733–1744.
    OpenUrlAbstract/FREE Full Text
  4. ↵
    1. Matthay MA,
    2. Chatterjee K
    . Bedside catheterization of the pulmonary artery: risks compared with benefits. Ann Intern Med 1988; 109: 826–834.
    OpenUrlPubMedWeb of Science
  5. ↵
    1. Keren H,
    2. Burkhoff D,
    3. Squara P
    . Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. Am J Physiol Heart Circ Physiol 2007; 293: H583–H589.
    OpenUrlAbstract/FREE Full Text
  6. ↵
    1. Squara P,
    2. Denjean D,
    3. Estagnasie P,
    4. et al
    . Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med 2007; 33: 1191–1194.
    OpenUrlCrossRefPubMedWeb of Science
  7. ↵
    1. Engoren M,
    2. Barbee D
    . Comparison of cardiac output determined by bioimpedance, thermodilution, and the Fick method. Am J Crit Care 2005; 14: 40–45.
    OpenUrlAbstract/FREE Full Text
    1. Imhoff M,
    2. Lehner JH,
    3. Lohlein D
    . Noninvasive whole-body electrical bioimpedance cardiac output and invasive thermodilution cardiac output in high-risk surgical patients. Crit Care Med 2000; 28: 2812–2818.
    OpenUrlCrossRefPubMedWeb of Science
  8. ↵
    1. Leslie SJ,
    2. McKee S,
    3. Newby DE,
    4. et al
    . Non-invasive measurement of cardiac output in patients with chronic heart failure. Blood Press Monit 2004; 9: 277–280.
    OpenUrlCrossRefPubMedWeb of Science
  9. ↵
    1. Raval NY,
    2. Squara P,
    3. Cleman M,
    4. et al
    . Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput 2008; 22: 113–119.
    OpenUrlCrossRefPubMed
  10. ↵
    1. Pavek K,
    2. Boska D,
    3. Selecky FV
    . Measurement of cardiac output by thermodilution with constant rate injection of indicator. Circ Res 1964; 15: 311–319.
    OpenUrlAbstract/FREE Full Text
  11. ↵
    1. LaFarge CG,
    2. Miettinen OS
    . The estimation of oxygen consumption. Cardiovasc Res 1970; 4: 23–30.
    OpenUrlAbstract/FREE Full Text
  12. ↵
    1. McLaughlin VV,
    2. Shillington A,
    3. Rich S
    . Survival in primary pulmonary hypertension: the impact of epoprostenol therapy. Circulation 2002; 106: 1477–1482.
    OpenUrlAbstract/FREE Full Text
  13. ↵
    1. Benza RL,
    2. Miller DP,
    3. Gomberg-Maitland M,
    4. et al
    . Predicting survival in pulmonary arterial hypertension: insights from the Registry to Evaluate Early and Long-Term Pulmonary Arterial Hypertension Disease Management (REVEAL). Circulation 2010; 122: 164–172.
    OpenUrlAbstract/FREE Full Text
  14. ↵
    1. Humbert M,
    2. Sitbon O,
    3. Chaouat A,
    4. et al
    . Survival in patients with idiopathic, familial, and anorexigen-associated pulmonary arterial hypertension in the modern management era. Circulation 2010; 122: 156–163.
    OpenUrlAbstract/FREE Full Text
  15. ↵
    1. Reuter DA,
    2. Huang C,
    3. Edrich T,
    4. et al
    . Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 2010; 110: 799–811.
    OpenUrlCrossRefPubMedWeb of Science
  16. ↵
    1. Bonderman D,
    2. Wexberg P,
    3. Martischnig AM,
    4. et al
    . A noninvasive algorithm to exclude pre-capillary pulmonary hypertension. Eur Respir J 2011; 37: 1096–1103.
    OpenUrlAbstract/FREE Full Text
  17. ↵
    1. Elkayam U,
    2. Berkley R,
    3. Azen S,
    4. et al
    . Cardiac output by thermodilution technique. Effect of injectate's volume and temperature on accuracy and reproducibility in the critically ill patient. Chest 1983; 84: 418–422.
    OpenUrlCrossRefPubMedWeb of Science
    1. Nishikawa T,
    2. Dohi S
    . Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 1993; 40: 142–153.
    OpenUrlPubMedWeb of Science
    1. Renner LE,
    2. Morton MJ,
    3. Sakuma GY
    . Indicator amount, temperature, and intrinsic cardiac output affect thermodilution cardiac output accuracy and reproducibility. Crit Care Med 1993; 21: 586–597.
    OpenUrlPubMedWeb of Science
  18. ↵
    1. Stevens JH,
    2. Raffin TA,
    3. Mihm FG,
    4. et al
    . Thermodilution cardiac output measurement. Effects of the respiratory cycle on its reproducibility. JAMA 1985; 253: 2240–2242.
    OpenUrlCrossRefPubMedWeb of Science
  19. ↵
    1. Cecconi M,
    2. Rhodes A,
    3. Poloniecki J,
    4. et al
    . Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output. Crit Care 2009; 13: 201.
    OpenUrlCrossRefPubMed
  20. ↵
    1. Cigarroa RG,
    2. Lange RA,
    3. Williams RH,
    4. et al
    . Underestimation of cardiac output by thermodilution in patients with tricuspid regurgitation. Am J Med 1989; 86: 417–420.
    OpenUrlCrossRefPubMedWeb of Science
  21. ↵
    1. Hillis LD,
    2. Firth BG,
    3. Winniford MD
    . Analysis of factors affecting the variability of Fick versus indicator dilution measurements of cardiac output. Am J Cardiol 1985; 56: 764–768.
    OpenUrlCrossRefPubMedWeb of Science
  22. ↵
    1. Buffington CW,
    2. Nystrom EU
    . Neither the accuracy nor the precision of thermal dilution cardiac output measurements is altered by acute tricuspid regurgitation in pigs. Anesth Analg 2004; 98: 884–890.
    OpenUrlCrossRefPubMedWeb of Science
  23. ↵
    1. Hoeper MM,
    2. Maier R,
    3. Tongers J,
    4. et al
    . Determination of cardiac output by the Fick method, thermodilution, and acetylene rebreathing in pulmonary hypertension. Am J Respir Crit Care Med 1999; 160: 535–541.
    OpenUrlPubMedWeb of Science
  24. ↵
    1. Konishi T,
    2. Nakamura Y,
    3. Morii I,
    4. et al
    . Comparison of thermodilution and Fick methods for measurement of cardiac output in tricuspid regurgitation. Am J Cardiol 1992; 70: 538–539.
    OpenUrlCrossRefPubMed
  25. ↵
    1. Fakler U,
    2. Pauli C,
    3. Hennig M,
    4. et al
    . Assumed oxygen consumption frequently results in large errors in the determination of cardiac output. J Thorac Cardiovasc Surg 2005; 130: 272–276.
    OpenUrlCrossRefPubMedWeb of Science
  26. ↵
    1. Kendrick AH,
    2. West J,
    3. Papouchado M,
    4. et al
    . Direct Fick cardiac output: are assumed values of oxygen consumption acceptable? Eur Heart J 1988; 9: 337–342.
    OpenUrlAbstract/FREE Full Text
  27. ↵
    1. Critchley LA,
    2. Critchley JA
    . A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999; 15: 85–91.
    OpenUrlCrossRefPubMedWeb of Science
  28. ↵
    1. Squara P,
    2. Cecconi M,
    3. Rhodes A,
    4. et al
    . Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med 2009; 35: 1801–1808.
    OpenUrlCrossRefPubMedWeb of Science
  29. ↵
    1. Marque S,
    2. Cariou A,
    3. Chiche JD,
    4. et al
    . Comparison between Flotrac-Vigileo and bioreactance, a totally noninvasive method for cardiac output monitoring. Crit Care 2009; 13: R73.
    OpenUrlCrossRefPubMed
  30. ↵
    1. Maurer MM,
    2. Burkhoff D,
    3. Maybaum S,
    4. et al
    . A multicenter study of noninvasive cardiac output by bioreactance during symptom-limited exercise. J Card Fail 2009; 15: 689–699.
    OpenUrlCrossRefPubMedWeb of Science
  31. ↵
    1. Moshkovitz Y,
    2. Kaluski E,
    3. Milo O,
    4. et al
    . Recent developments in cardiac output determination by bioimpedance: comparison with invasive cardiac output and potential cardiovascular applications. Curr Opin Cardiol 2004; 19: 229–237.
    OpenUrlCrossRefPubMedWeb of Science
PreviousNext
Back to top
View this article with LENS
Vol 42 Issue 1 Table of Contents
European Respiratory Journal: 42 (1)
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
Email

Thank you for your interest in spreading the word on European Respiratory Society .

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Noninvasive cardiac output measurements in patients with pulmonary hypertension
(Your Name) has sent you a message from European Respiratory Society
(Your Name) thought you would like to see the European Respiratory Society web site.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Print
Citation Tools
Noninvasive cardiac output measurements in patients with pulmonary hypertension
Jonathan D. Rich, Stephen L. Archer, Stuart Rich
European Respiratory Journal Jul 2013, 42 (1) 125-133; DOI: 10.1183/09031936.00102212

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero

Share
Noninvasive cardiac output measurements in patients with pulmonary hypertension
Jonathan D. Rich, Stephen L. Archer, Stuart Rich
European Respiratory Journal Jul 2013, 42 (1) 125-133; DOI: 10.1183/09031936.00102212
del.icio.us logo Digg logo Reddit logo Technorati logo Twitter logo CiteULike logo Connotea logo Facebook logo Google logo Mendeley logo
Full Text (PDF)

Jump To

  • Article
    • Abstract
    • Introduction
    • Methods
    • Results
    • Discussion
    • Footnotes
    • References
  • Figures & Data
  • Info & Metrics
  • PDF

Subjects

  • Pulmonary vascular disease
  • Tweet Widget
  • Facebook Like
  • Google Plus One

More in this TOC Section

Original Article

  • Lung volumes and survival in chronic lung allograft dysfunction
  • Prior CV disease and mortality COPD pneumonia patients
  • Factors associated with FEV1 decline in CF
Show more Original Article

Pulmonary Vascular Disease

  • Effects of sotatercept on haemodynamics and right heart function
  • Minimally important differences for cardiac MRI end-points in PAH
  • Heterogeneity of treatment effects by risk in PAH
Show more Pulmonary Vascular Disease

Related Articles

Navigate

  • Home
  • Current issue
  • Archive

About the ERJ

  • Journal information
  • Editorial board
  • Press
  • Permissions and reprints
  • Advertising

The European Respiratory Society

  • Society home
  • myERS
  • Privacy policy
  • Accessibility

ERS publications

  • European Respiratory Journal
  • ERJ Open Research
  • European Respiratory Review
  • Breathe
  • ERS books online
  • ERS Bookshop

Help

  • Feedback

For authors

  • Instructions for authors
  • Publication ethics and malpractice
  • Submit a manuscript

For readers

  • Alerts
  • Subjects
  • Podcasts
  • RSS

Subscriptions

  • Accessing the ERS publications

Contact us

European Respiratory Society
442 Glossop Road
Sheffield S10 2PX
United Kingdom
Tel: +44 114 2672860
Email: journals@ersnet.org

ISSN

Print ISSN:  0903-1936
Online ISSN: 1399-3003

Copyright © 2023 by the European Respiratory Society