To the Editors:
Plasma osteopontin is a very interesting marker for pulmonary diseases, being proposed for evaluating prognosis in patients with nonsmall cell lung cancer 1 or tuberculosis 2. Osteopontin has also been implicated in the pathogenesis of idiopathic pulmonary fibrosis 3 and chronic obstructive pulmonary disease 4. A more debated use of this marker has been its diagnostic utility in malignant pleural mesothelioma 5, 6. Two major commercial kits are available, supplied by R&D systems, Minneapolis, MN, USA (Quantikine Human Osteopontin Immunoassay) and by Immuno-Biological Laboratories Co., Ltd, Gunma, Japan (Human osteopontin ELISA). It has been suggested in the literature that the results of osteopontin dosage are highly dependent on the choice of the ELISA kit. Indeed a previous publication reported important differences between available assays with little correlation between the results of the two commercial kits 7. However, correlation is not the only analysis that should be performed when comparing two tests that measure the same variable 8.
We performed a comparative evaluation of the two commercial ELISA kits in a series of 75 samples of plasma retrieved from 13 patients treated for tuberculosis. The samples were obtained at diagnosis, after 1, 2, 4 and 8 weeks, and 4 and 6 months of treatment. A standard operation procedure was in place for retrieving the samples and all patients gave informed consent. After centrifugation, plasma samples were immediately transferred in a -80°C freezer until assayed. A mean of six samples per patient (interquartile range 5–6) were available.
A correlation analysis between the two measurements yielded a Spearman ρ of only 0.4924 but this was statistically significant (p<0.001; fig. 1a⇓). If we used the Pearson correlation coefficient (and thus assuming a normal distribution of data) this correlation was also low but statistically significant (r = 0.4515; p<0.001). However, when we used a Bland–Altman plot to analyse the results, we obtained an almost straight line (fig. 1b⇓). Thus, the lack of strong correlation between the two methods is, in fact, the result of a systematic bias between the two methods, which is proportional with the measured value. This type of result, called “proportional error” is, in fact, the result of the nonlinearity of the two measures and probably is a consequence of the amplification generated by different enzymes in the ELISA assay. Since we do not have a “gold standard”, we cannot say exactly which method is biased and in which direction. Mainly, both assays measure the same target molecule despite using different antibodies. These data must be taken into account when interpreting and comparing the results between different publications as well as choosing an assay for use in practice.
a) Correlation between the values obtained for osteopontin in plasma assayed with the Quantikine Human Osteopontin Immunoassay (R&D systems, Minneapolis, MN, USA) (R&D) and the Human osteopontin ELISA (Immuno-Biological Laboratories Co., Ltd, Gunma, Japan) (IBL). Spearman ρ = 0.4924. b) Bland–Altman plot of the measurements for osteopontin showing a strong proportional error between the two methods.
Support statement
This research has been funded by the National Council for Scientific Research in Higher Education (CNCSIS) by the grant number 1191/2007.
Statement of interest
None declared.