Abstract
Airway dimensions are difficult to quantify bronchoscopically because of optical distortion and a limited ability to gauge depth. Anatomical optical coherence tomography (aOCT), a novel imaging technique, may overcome these limitations. This study evaluated the accuracy of aOCT against existing techniques in phantom, excised pig and in vivo human airways.
Three comparative studies were performed: 1) micrometer-derived area measurements in 10 plastic tubes were compared with aOCT-derived area; 2) aOCT-derived airway compliance curves from excised pig airways were compared with curves derived using an endoscopic technique; and 3) airway dimensions from the trachea to subsegmental bronchi were measured using aOCT in four anaesthetised patients during bronchoscopy and compared with computed tomography (CT) measurements.
Measurements in plastic tubes revealed aOCT to be accurate and reliable. In pig airways, aOCT-derived compliance measurements compared closely with endoscopic data. In human airways, dimensions measured with aOCT and CT correlated closely. Bland–Altman plots showed that aOCT diameter and area measurements were higher than CT measurements by 7.6% and 15.1%, respectively.
Airway measurements using aOCT are accurate, reliable and compare favourably with existing imaging techniques. Using aOCT with conventional bronchoscopy allows real-time measurement of airway dimensions and could be useful clinically in settings where knowledge of airway calibre is required.
- Airway dimensions
- computed tomography
- interventional bronchoscopy
- optical coherence tomography
- quantitative bronchoscopy
Bronchoscopy is a widely performed procedure with many indications. Several of these, such as airway stenosis, tracheomalacia and the deployment of valves for endoscopic lung volume reduction, require knowledge of airway dimensions in order to optimise the interventions and/or to assess disease progression longitudinally. To date, bronchoscopists have relied on pre-procedure imaging, usually with computed tomography (CT), to quantify airway dimensions such as calibre or length of stenosis. CT has a proven capacity to provide such measurements, but it cannot be utilised during a procedure. This presents a major limitation when, for example, during a stenting procedure the bronchoscopist wishes to confirm the stenosis dimensions to ensure selection of an appropriate stent, as complications can arise from poorly sized stents 1, 2.
While it is possible to derive quantitative information directly from bronchoscopic images, two major impediments make doing so difficult. First, bronchoscopes display three-dimensional anatomy as a two-dimensional image, thereby limiting the ability of an observer to gauge depth. Secondly, the wide-angle lens at the tip of the bronchoscope, while employed to facilitate the widest possible view, results in image distortion, particularly at the peripheries. Several techniques have been developed to permit quantification from bronchoscopic images 3–9. However, these techniques entail complex correction algorithms, and for accurate calibration often require targets of known size (e.g. biopsy forceps or measuring devices) to be placed in the field of view.
This paper describes a novel imaging technique that addresses many of these problems. Optical coherence tomography (OCT) is a light-based imaging technique based on the principle of low-coherence interferometry. In its conventional form, OCT allows subsurface airway analysis to a depth of 2–3 mm, providing an “optical biopsy” approaching histological resolution. Respiratory applications focus on the detection of malignant and dysplastic epithelial changes 10, 11. Anatomical OCT (aOCT) is a modification of conventional OCT designed to range over larger distances to permit real-time macroscopic imaging of hollow organs 12. Applications in the human upper airway demonstrate the in vivo utility of this technique, but it has yet to be validated in the tracheobronchial tree 13–16.
Our group has previously detailed the technical considerations of applying aOCT to the central airways 17. The purpose of this paper is to validate aOCT for use in the human tracheobronchial tree.
METHODS
The accuracy of aOCT was systematically validated in plastic tubes, pig bronchi and human airways by comparing measurements made with aOCT and other established imaging techniques. The Human Research Ethics Committee at our hospital approved this study and patients provided informed consent. The animal study was approved by our institutional Animal Ethics Committee. Further details of the following methods can be found in an accompanying online repository.
Anatomical optical coherence tomography
The aOCT unit consists of a 1.3-mm wide fibre-optic probe housed within a plastic catheter (outer diameter 2.2 mm) as previously described 12–14, 17, 18. The catheter is passed through the working channel of a standard bronchoscope and into an airway of interest. The probe rotates at 2.5 Hz, directing a broadband light beam (central wavelength 1310 nm, bandwidth 32 nm) perpendicular to the airway, thus building up a rotational scan of the internal airway lumen that is displayed on a monitor. The probe is capable of scanning over a radius of 36 mm with an optimal lateral resolution of 156 μm at a distance of 2.35 mm, although this varies with distance. A custom-built computer program enables the user to trace manually the internal lumen and diameter immediately post-acquisition. While rotating, the probe may also be mechanically retracted along an airway segment at 0.4 mm·s−1, to generate a volumetric dataset. We refer to this as a “pullback” scan.
Plastic tubes
A plastic airway phantom (online repository, E1) consisting of 10 round tubes (diameters 2.6–50.0 mm) was constructed and the tubes were measured using internal micrometers (Bowers Metrology, Bradford, UK) accurate to 0.01 mm. Three separate sites along each tube were imaged by one investigator using aOCT. The 30 images were then randomised prior to measurement of diameter and internal area (Ai). To evaluate intra-observer repeatability, measurements were repeated after 3 weeks. For inter-observer repeatability, a second investigator repeated the measurements. To assess the effect on measurement accuracy of the bends in the fibre-optic probe during bronchoscopy, in vivo conditions were simulated by repeating the measurements through a bronchoscope inserted into a resuscitation mannequin (online repository, E2).
Excised pig airways
A trachea and two bronchial segments were dissected from the lungs of an 8-week-old pig. Side branches were ligated with surgical silk and the segments (length ∼60 mm) were placed horizontally in a perspex chamber of Krebs solution at 37°C as described elsewhere 19. The ends were cannulated, one being connected to a reservoir containing Krebs solution, the height of which could be adjusted to vary the transmural pressure, thereby manipulating airway size (fig. 1⇓).
The airway lumina were visualised using a videoendoscopy technique, whereby a rigid fibre-optic endoscope (SES1711D; Olympus, Tokyo, Japan) was inserted into the airway through an outlet in the organ chamber (fig. 1⇑) 19. At the site selected for analysis, the mucosal surface was circumferentially stained with dye. Transmural pressures were increased in increments of 5 cmH2O from -15 cmH2O to +30 cmH2O, then decreased back to -15 cmH2O. Photographs of the dye–mucosa interface, depicting the inner airway perimeter, were captured and transferred to a computer workstation for analysis of lumen Ai using ImageJ software, version 1.38t (National Institutes of Health, Bethesda, MD, USA; http://rsb.info.nih.gov/ij). The endoscope was then removed and the aOCT catheter was passed into the airway lumen and advanced to the dye ring, where the measurements were repeated. The order of imaging (aOCT or endoscope) was randomised. We compared the two techniques by constructing airway compliance curves from the plot of transmural pressure against Ai. Compliance was described first from the slope of the curve between 0 and 10 cmH2O (the region of maximal change) and secondly by fitting the positive pressure data with an exponential expression (also used to describe the nonlinear elastic behaviour of the lungs), of the form V = A - B × exp(-KP), where V is airway lumen area, P is transmural pressure and A, B and K are constants. The Colebatch shape factor K defines the nonlinearity of the area–pressure curve, and is an index of distensibility independent of airway size 20.
To investigate the potential use of aOCT in the setting of tracheobronchial stenosis, immediately following measurement of airway compliance, a stenosis was simulated in the excised pig trachea by tying suture material around its centre. The aOCT probe was advanced through the tracheal lumen beyond the “stenosis” then mechanically retracted across the stenosis while recording continuously. This pullback scan was analysed using a three-dimensional software viewer (VolView; Kitware, Clifton Park, NY, USA) to measure stenosis dimensions.
In vivo human airways
Subjects were recruited from patients scheduled for bronchoscopy who also required chest CT as part of their medical care. The 64-slice scanner (Philips Brilliance CT 95089; Philips Medical Systems, Best, The Netherlands) scanned from the glottis to lung bases (220 mA, 120 Kvp, voxel size 0.625×0.625×0.625 mm and using a lung-enhanced reconstruction algorithm). On the same day, a bronchoscopy was performed under propofol anaesthesia titrated to a depth sufficient to permit spontaneous breathing.
To ensure that scans were obtained at a similar lung volume during both the CT and bronchoscopy, impedance plethysmography bands (Respitrace 10.9230; Ambulatory Monitoring, Ardsley, NY, USA) were positioned over the ribcage and abdomen and calibrated using an isovolume manoeuvre. The summed abdominal and ribcage signal is a measure of total chest wall displacement (lung volume change) and was used to monitor lung volume changes during CT and bronchoscopy 21. During CT scanning, respiration was suspended at functional residual capacity (FRC) using the summed Respitrace signal to guide a voluntary breath-hold. FRC decreases during general anaesthesia 22. To match FRC during bronchoscopy to the preceding CT, we noted baseline FRC using a period of quiet tidal breathing prior to anaesthetic delivery and, once stable anaesthesia had been achieved, applied positive end expiratory pressure and pressure support to raise FRC back to baseline (online repository, E3). In each subject, between 7 and 11 normal airway sites (from the trachea to the subsegmental airways) were imaged using aOCT over six respiratory cycles each, taking ∼20 min.
After the bronchoscopy, Ai and short-axis diameter were measured from the aOCT and CT images and compared. CT images were analysed using Pulmonary Workstation 2.0 software (VIDA Diagnostics, Iowa City, IA, USA), which reformats CT data into three-dimensional images and calculates airway area and diameter perpendicular to the airways. At each site, measurements from three adjacent CT slices were averaged. For the aOCT images, measurements were performed at end-expiration (FRC) on three successive breaths and averaged.
Statistical analyses
The coefficient of variation of repeated Ai measurements from the plastic phantom was expressed as a percentage. Intra- and inter-observer repeatability coefficients were determined using the Bland and Altman method and a paired t-test compared compliance measurements derived using aOCT with the endoscopic technique 23. Human airway dimensions measured by CT and aOCT were compared using Bland–Altman plots 23. Statistical analyses were performed using SPSS for Windows 15.0.0 (SPSS Inc, Chicago, IL, USA).
RESULTS
Comparative studies
Plastic tubes
aOCT-derived Ai measurements in 10 plastic tubes (range 5.3–1,963.5 mm2) showed very close agreement to the micrometer Ai measurements (fig. 2⇓), with the mean difference across the tubes being -0.1±5.6 mm2 (-0.3±3.5%) with even scatter about the zero difference line. The coefficient of variation of repeated measurements was 0.8%. Measurements repeated through the resuscitation mannequin showed a small bias towards overestimating the Ai, and a paired t-test showed the difference of the means to be 2.1 mm2 (95% confidence interval 0.7–3.5 mm2, p<0.01). Bland–Altman plots of intra- and inter-observer variability demonstrated good agreement (fig. 3⇓). The calculated intra- and inter-observer repeatability coefficients were 5.7 and 7.1 mm2, respectively, which represent <2% of the mean area of all 10 tubes.
Excised pig airways
Three airways were dissected and four sites (A–D) were marked with dye (one bronchus was marked at two sites). Airway Ai ranged 9.0–170.0 mm2. At each site, ascending and descending limbs of the pressure/Ai curve were plotted using both endoscope and aOCT-derived Ai (fig. 4⇓). Compliance curves for each technique were similar and the mean shape factor, K, was not significantly different (mean K = 0.11 (endoscopy) versus K = 0.12 cmH2O−1 (aOCT); 95% confidence interval for the difference of means -0.03–0.01; p = 0.24). Compliance measured using the slope of the curve between 0 and 10 cmH2O, was not different between techniques, being 2.2 and 2.1 mm2·cmH2O−1 (95% confidence interval for the difference of means -0.3–0.2; p = 0.64).
Two- and three-dimensional reconstruction of the tracheal stenosis is shown in figure 5⇓. At the narrowest point, the transverse dimensions measured 7.9×7.6 mm, with a stenosis length of 23 mm.
In vivo human airways
CT and aOCT measurements were obtained at 36 airway sites from six airway generations in four patients (fig. 6⇓ and online repository, E4). Ai ranged 6.2–282.0 mm2 and short-axis diameters ranged 2.3–16.9 mm. There was close correlation between CT and aOCT-based Ai and diameter measurements (r = 0.99; p<0.001; online repository, E5). Bland–Altman plots showed aOCT measurements to be slightly greater than CT measurements by a mean of 6.2±18.1 mm2 for Ai (fig. 7a⇓ and b) and 0.4±1.3 mm for diameter (fig. 7c⇓ and d).
DISCUSSION
Despite a revolution in the field of thoracic imaging, contemporary bronchoscopists still lack a reliable tool to quantify endobronchial anatomy during a procedure. This paper describes the use of a novel imaging technique, aOCT, in plastic tubes, excised pig airways and in vivo human airways. As aOCT operates independently of conventional bronchoscopic imaging, it overcomes the major limitation of bronchoscopic airway quantification: distorted images from which airway dimensions are difficult to derive, requiring post-procedure analysis. The results of our validation studies indicate that aOCT is a suitable complementary bronchoscopic technique for real-time measurements of airway dimensions.
This study demonstrates the utility and accuracy of aOCT in three experimental settings. In plastic tubes, aOCT provided accurate and repeatable Ai measurements over a wide range of diameters. Relative to micrometer measurements, Ai measured when the aOCT catheter was passed through a mannequin was greater, on average, by 2.1 mm2. This may be due to bends imposed on the catheter/probe during in vivo scanning, causing torsion and uneven probe rotation. Regardless, the magnitude of this difference (representing <1% of the mean area of the 30 measurements) is clinically insignificant. Compliance measurements in excised pig airways using aOCT and an endoscopic technique were also similar, highlighting the potential for aOCT to determine physiological properties of airways, such as regional compliance during bronchoscopy. Lastly, comparison of aOCT with CT measurements in humans also showed close agreement. A strength of our approach was that great care was taken to control subject lung volume, which can affect airway size 24. Indeed, to our knowledge, among quantitative videobronchoscopy studies, this is the first such study to do so. However, while every attempt was made to standardise conditions during CT and aOCT scanning, the measurements were performed several hours apart, making comparison of the exact site within the airway difficult. Simultaneous imaging would have been preferable, but this would be difficult to achieve as simultaneous bronchoscopy and CT scanning is complex and would cause CT image artefact.
While, as discussed above, there was close agreement between aOCT and CT, overall, aOCT yielded measurements of diameter and Ai that were 7.6% and 15.1% greater than those produced CT measurements, respectively. This finding contrasts with the study by Coxson et al. 25 who, using conventional OCT, found measurements of Ai in small airways (mean airway size 12.5±6.1 mm2) to be lower when measured with OCT than with CT, by an average of 31%. The discrepancy is readily attributable to the higher CT lung volumes in the study of Coxson et al. 25 (total lung capacity) relative to their bronchoscopic OCT images (tidal breathing during conscious sedation). With the use of impedance plethysmography bands, our study carefully controlled lung volume, allowing us to estimate FRC during CT scanning and approximate this lung volume during bronchoscopy.
We found the relative difference between CT and aOCT measurements to be greatest in smaller airways (fig. 7c⇑ and d). We have identified two mechanisms for this disparity. Firstly, it is possible that the presence of the aOCT catheter could mechanically increase small airway calibre, leading to an overestimation of Ai, although we measured only airways larger than the catheter with a partial air space evident between the two (fig. 6f⇑). Secondly, CT partial volume averaging, which results in underestimation of Ai, is particularly evident in airways <2–3 mm in diameter and increases with the obliqueness of the imaging plane 26, 27. Also, partial volume averaging may have been more significant in our study as we scanned at FRC rather than total lung capacity. This is less of a concern with aOCT, where the “slice thickness” equivalence of an axial aOCT image in, for example, a 5-mm diameter airway is ∼0.2 mm, compared with 0.625 mm using modern CT scanners. Further, unlike CT, the aOCT probe lies within the plane of the airway and scanning occurs close to the airway axis, avoiding oblique imaging. Even in the largest airways, in which the relative size of the bronchoscope to airway lumen increases the likelihood of oblique scanning, an angle approaching 18° would be required to produce a 5% area error (online repository, E6).
Potential applications
aOCT could prove beneficial in several bronchoscopic settings. The deployment of stents for airway stenoses requires pre-bronchoscopy imaging, usually with CT. This provides important information about the site, number and dimensions of stenoses. However, CT is usually performed at total lung capacity and scans are not always recent at the time of procedure. It is still, therefore, incumbent upon the bronchoscopist to estimate stenosis length and calibre, which can be highly subjective. The ability to scan a stenosis and derive immediate quantitative dimensions would be highly advantageous for stent size selection. By virtue of real-time imaging, aOCT could also prove useful in the management of the malacia disorders, where quantitative assessment of dynamic airway collapse is desirable. This may be particularly useful in the setting of paediatric bronchology, where longitudinal assessment of tracheo-bronchomalacia is required, ideally without ionising radiation. Finally, measurement of airway size is required during endobronchial valve deployment for endoscopic lung volume reduction. This relatively new field might benefit from the capacity of aOCT to size airway diameters optimising valve selection.
Limitations
Although aOCT can image a wide range of airway sizes, the current prototype has limited subsurface detail and, therefore, cannot measure airway wall thickness. The catheter size could also be considered a limitation as the probe–catheter system limits the size of measureable airways to ∼2.5 mm. This gets close to but does not image the small airways of greatest interest in asthma and chronic obstructive pulmonary disease. However, this technical limitation is temporary as smaller probes are in development.
Probe rotation speed was found to be adequate for scanning during normal breathing. Faster acquisition speeds would improve the capacity for three-dimensional reconstructions to account for respiratory and cardiac motion. Despite this, we have recently demonstrated the capacity to “gate” images to specific phases of the respiratory cycle 28. Finally, as with any new technology, cost is a factor. Our device is a custom-built prototype, making a unit cost difficult to estimate. Larger centres specialising in interventional procedures will probably be the first to realise the potential of and to benefit from this technique.
Conclusion
This paper demonstrates the accuracy and precision of aOCT in phantom, pig and human airways and demonstrates its capacity to reconstruct three-dimensional images. An advantage of this technique over existing methods of bronchoscopic airway quantification is the ability to operate independent of the distorting effects of the bronchoscope. Such a tool could be of considerable value to interventional bronchoscopists, by improving assessment of disease severity and its progression, sizing dimensions of airway stenoses in preparation for stenting and for the deployment of endobronchial valves. Future studies examining the effectiveness of aOCT in these and other clinical settings will further define the optimal role of this emerging technique.
Support statement
This study was supported by an Australian National Health and Medical Research Council (NHMRC; Canberra, Australia) Project Grant No. 513854. J.P. Williamson is funded by a NHMRC Postgraduate Research Scholarship No. 463926, a Sir Charles Gairdner Hospital (Perth, Australia) Research Grant and a University of Western Australia (Perth, Australia) top-up grant. P.R. Eastwood is funded by a NHMRC Senior Research Fellowship No. 513704. R.A. McLaughlin is funded by the Raine Medical Research Foundation (Nedlands, Australia). P.B. Noble is funded by an Australian-based Biomedical Fellowship, No. 513921.
Statement of interest
Statements of interest for J.P. Williamson, J.J. Armstrong, R.A. McLaughlin, M.J. Phillips, D.D. Sampson, D.R. Hillman and P.R. Eastwood can be found at www.erj.ersjournals.com/misc/statements.dtl
Acknowledgments
The authors wish to acknowledge the enthusiastic support of N. Hicks, P. Muir and C. Smith from the Radiology Dept of Sir Charles Gairdner Hospital (SCGH; Perth, Australia). They also wish to thank C. Smith and M. Schmidt from the Dept of Anaesthesia, SCGH, for their anaesthetic expertise and J. Graham and W. Lilly (SCGH) for assistance with the mannequin study. The authors are sincerely grateful to R. Kap and T. Deans from the Dept of Medical Technology and Physics, SCGH, for assistance with construction of the airway phantom and technical support and to B. Manners from the Audiovisual Production Unit, SCGH, for illustration support.
Footnotes
-
This article has supplementary material accessible from www.erj.ersjournals.com
- Received March 14, 2009.
- Accepted May 23, 2009.
- © ERS Journals Ltd