Abstract
Bronchodilator drugs produce variable improvements in forced expiratory volume in 1 s (FEV1), but larger changes in end-expiratory lung volume (EELV) in chronic obstructive pulmonary disease (COPD), which were suggested to be related to the presence of expiratory flow limitation (EFL) at rest.
We tested this concept in 42 COPD patients (FEV1 42.3±13.8% predicted) during spontaneous breathing before and after 5 mg nebulised salbutamol. EFL was detected by within-breath changes in respiratory system reactance measured by a multifrequency forced oscillation method, while changes in EELV were assessed by inspiratory capacity (IC). Bronchodilation (BD) increased IC (from 1.8±0.5 to 2.1±0.6 L, p<0.001) and reduced inspiration resistance (R̄insp) at 5 Hz (from 5.1±1.6 to 4.2±1.5 cmH2O·s·L−1, p<0.001). R̄insp identified BD responders with a discriminative power of 80.1%.
In total, 20 patients were flow-limited before BD. They showed worse spirometry and higher residual volume, but significant improvements in IC were seen in all patients irrespective of flow limitation. Changes in R̄insp were confined to flow-limited patients, as were reactance changes. BD reduced the degree of heterogeneity in the respiratory system, a change best seen with inspiratory values.
BD has complex effects on lung mechanics in COPD, and EFL affects both this and the response of some respiratory variables to treatment. However, changes in EELV are consistently seen, irrespective of the presence of flow limitation at rest.
- Chronic obstructive pulmonary disease
- forced oscillation technique
- respiratory system reactance
- within-breath impedance
Chronic obstructive pulmonary disease (COPD) is defined by the presence of incompletely reversible expiratory airflow limitation (EFL) 1, which occurs at much lower flows for any given lung volume when compared with healthy subjects. Initially, flow limitation is only present during maximal or near maximal respiratory efforts, but as lung disease progresses, EFL develops at rest in many, but not all, individuals 2. The presence of resting EFL may identify COPD patients who behave differently and who develop dynamic hyperinflation 3, at least during exercise 4.
Bronchodilator drugs improve lung emptying, and this leads to variable increases in forced expiratory volume in 1 s (FEV1), mainly by reducing lung volume rather than changing the FEV1/forced vital capacity (FVC) ratio 5. However, the reproducibility and predictive value of testing for FEV1 reversibility is relatively poor 6, 7, while the change in resting inspiratory capacity has been shown to be a better predictor of improvement in exercise performance 8, 9. Again, patients with EFL have been reported to show improvements in inspiratory capacity after bronchodilators 10, which may relate to an improvement in exercise performance 11.
Previous workers have used the negative expiratory pressure method to detect EFL 12, but that study samples a relatively small number of breaths and not all tests are suitable for analysis 2. We have developed an effort-independent method to determine flow limitation during tidal breathing using the forced oscillatory technique to identify within-breath differences in respiratory system reactance 13, 14. This method allows the assessment of more breaths, is equivalent to the negative expiratory pressure approach when both can be recorded 2, and adds a potential “quantitative” assessment of how close the patent is to the threshold of EFL 14. Modelling simulations based on these data suggest that EFL will influence other measurements of oscillatory mechanics during expiration, and this will reduce the sensitivity of expiratory impedance data to change after interventions, such as bronchodilators. Although this effect can be identified when within-breath analysis is performed 13, most published reports of oscillatory mechanics in COPD only report total respiratory cycle data 15–17.
In the current study, we tested the hypothesis that the changes in lung volume (specifically inspiratory capacity) and oscillatory lung mechanics of COPD patients given an inhaled bronchodilator drug would differ when EFL was present, and whether this would be unrelated to the presence of reversibility defined spirometrically. Additionally, we extended our observations of within-breath impedance using a forced oscillation method from single to multiple forcing frequencies. This approach allowed us to test whether bronchodilator drugs improve resistance and the intrapulmonary homogeneity of lung mechanics, avoiding the confounding effects of EFL on impedance data that would be corrupted when adopting the conventional multifrequency approach. Finally, we examined the changes in resting lung and respiratory system mechanics in those individuals who were no longer flow-limited after the bronchodilator drug.
METHODS
Subjects
We recruited clinically stable outpatients who met the diagnostic criteria for COPD 18 and were either current or ex-smokers. All patients were using short- and long-acting inhaled bronchodilators, which were omitted before study for 3–24 h, as appropriate. The study was approved by the institutional research ethical review committee (South Sefton Research Ethics Committee, Liverpool, UK), and written informed consent was given by each subject.
Measurements
Forced expiratory flow, lung volume and subdivisions were measured by a constant-volume body plethysmograph (Medgraphic Autolink 1085D; Medical Graphics, St Paul, MN, USA). All measurements met current standards for acceptable data quality 19. We report FEV1, FVC, FEV1/FVC, inspiratory capacity (IC), residual volume (RV), thoracic gas volume (TGV) and total lung capacity (TLC) both as absolute values and % predicted (% pred). Predicted values were those recommended by the European Respiratory Society (ERS) 20.
We measured breathing pattern and oscillatory mechanics using previously described methods 13. Briefly, we recorded pressure and flow at the airway opening by a transducer connected to the mouthpiece (PXLA0025DN; Sensym, Milpitas, CA, USA) and by a screen-type pneumotachograph (3700A, Hans Rudolph, Kansas City, MO, USA) connected to another pressure transducer (PXLA02X5DN, 0–2.5 cm; Sensym). All the signals were sampled at 200 Hz and recorded onto a PC. The flow signal was integrated to give lung volume, and volume drift was removed by selecting 2–3 min of stable quiet breathing and estimating the linear trend on the integrated signal. This trend was then removed from the traces.
From these signals we measured tidal volume (VT), respiratory frequency, total cycle duration, inspiratory time, expiratory time and inspiratory duty cycle (fig. 1⇓). We derived minute ventilation (V′E), mean inspiratory flow rate and mean expiratory flow rate from these data.
Forced oscillations
Patients were studied while being oscillated by the following two different waveforms: 1) a 5 Hz sinusoidal signal, and 2) a pseudo-random noise (PRN) with three components at 5, 11 and 19 Hz chosen to be non-sum non-difference of order 3 21. For both the waveforms, the peak-to-peak pressure amplitude measured at the mouth was ∼1–2 cmH2O. In order to have comparable total energy at 5 Hz in both the sinusoidal and the PRN signals, and to keep the total energy of the PRN signal low, the relative amplitude of the 5 Hz component of the PRN has been slightly increased.
The experimental set-up for forced oscillation technique (FOT) measurement was similar to that described previously 13. The pressure signal generated by a loudspeaker was transferred from the box to the subject's mouthpiece through a connecting tube (22 cm long, 19 mm internal diameter). A low-resistance, highly inert tube (1.5 m long, 22 mm internal diameter) in parallel with the loudspeaker allowed the subjects to breathe room air without significant loss of forcing pressure. A bias flow of ∼15 L·min−1 reduced the equipment dead space to the volume of the pneumotachograph and the mouthpiece 22. The frequency response of the whole measuring system was assessed up to 30 Hz, as described previously 2, and was flat.
Experimental protocol
Patients attended on one occasion, when all measurements were made in the same order. Plethysmographic lung volume measurements were followed by recording oscillatory mechanics at 5 Hz and breathing pattern with the patients seated, wearing a nose-clip and with an operator firmly supporting the cheeks to reduce upper airways shunt 23. The patients breathed spontaneously through the FOT system for 1 min, then performed an IC manoeuvre and resumed spontaneous breathing.
After 10-min rest with the patient disconnected from the measuring circuit, the FOT measurements were repeated by following the same sequence of manoeuvres, but with the multifrequency PRN signal applied.
Next, patients received 5 mg of nebulised salbutamol from an oxygen-driven Acorn® nebuliser (MedicAid, Pagham, UK) and after 45 min of spirometry, plethysmography and the two FOT measurements were repeated as described previously.
Data analysis
Within-breath respiratory system input impedance (Zrs) was determined by using a least squares algorithm taking advantage of the a priori knowledge of the frequency spectrum components of the forcing signals 24, 25.
From the complete FOT recording and impedance tracing, we selected ∼10 breaths starting from 45 s after the first IC, to avoid possible alterations of the breathing pattern after the manoeuvre. Breaths in which Zrs tracings showed spikes or oscillations due to swallowing or glottis closure were discarded. For each breath, the values of several breathing pattern parameters and Zrs indices were computed and averaged for all the breaths in each subject and condition. Within-breath respiratory system resistance (Rrs) was characterised by the mean values it assumed during inspiration, expiration and during the whole breath (R̄insp, R̄exp and R̄tot, respectively). As the frequency dependence of Rrs is related to the heterogeneity of airway obstruction 22, 26, we also computed the difference between Rrs measured at our lowest and highest frequency (5 and 19 Hz, R̄5–R̄19).
Respiratory system reactance (Xrs) was characterised by its average value during a breath and its within-breath fluctuations were quantified by computing its average value during inspiration (X̄insp) and expiration (X̄exp). Their difference (ΔX̄rs = X̄insp-X̄exp) was used to detect EFL. A breath was considered flow-limited if ΔX̄rs was greater than a threshold of 2.8 cmH2O·s·L−1, a value that in our previous studies 2, 13, 14, enabled identification of flow-limited breaths with very high sensitivity and specificity. A subject was classified as flow-limited if the majority of his selected breaths were flow-limited.
Data are expressed as mean±sd, unless otherwise stated. All data comparisons were made relative to that individual’s baseline value, although we did conduct an exploratory analysis of the spirometry data based on the reversibility criteria recommended by the American Thoracic Society (ATS)/ERS to identify bronchodilator responsiveness (FEV1 change >12% from the baseline and >200 mL) 27. Significant differences in the physical characteristics, spirometric data, and Rrs and Xrs indices values of the different groups were evaluated using paired or unpaired t-tests, as appropriate. To allow for the multiple comparisons to be made between groups, only p<0.01 was considered to be statistically significant. Our primary outcome was the change in IC after administration of the bronchodilator. We calculated that a study with 15 patients would have an 80% chance of showing a difference of 200 mL at the 5% significance level between the groups. As we anticipated identifying flow-limited and nonflow-limited patients, we aimed to recruit 40 individuals to increase our ability to detect differences between the subgroups.
RESULTS
The baseline characteristics of the 42 COPD patients recruited in this study are reported in table 1⇓. All patients performed the measurements correctly, with no reports of discomfort. From these patients, a total of 788 breaths were selected and analysed (408 before and 380 after bronchodilator). In figure 1⇑, an experimental tracing of volume and multifrequency impedance data are shown for a representative flow-limited patient. The presence of flow limitation is clearly shown by the large decrease of Xrs during expiration compared with inspiration. Figure 1⇑ also shows that the presence of EFL affects within-breath variations of Zrs at all frequencies but, as predicted by the model simulation 13, the intra-breath Xrs swings decrease in amplitude with increasing frequencies.
The 5 Hz component of the multifrequency forcing gave similar results to those of the single 5 Hz frequency and, therefore, in the rest of this study only data recorded during multifrequency forcing are reported. A full account of the comparison of the single and multifrequency testing is presented in the online supplementary material.
Group mean data post-bronchodilator without accounting for tidal expiratory flow limitation
The lung function, breathing pattern and impedance indices for the whole patient group, measured after the bronchodilator are presented in tables 1⇑ and 2⇓. Bronchodilation (BD) produced statistically significant improvements in all the measured plethysmographic variables except for FEV1/FVC and TLC (table 1⇑). There was a significant increase in V′E and a fall in mean inspiratory and expiratory flow (table 2⇓).
In total, 18 patients met the ATS/ERS criteria for reversibility of airway obstruction. There were no differences between responder and nonresponder groups in their baseline plethysmographic or oscillatory variables. The changes in plethysmographic variables post-bronchodilator were similar for the two spirometrically defined groups, while the oscillometric indices differed between the responders and nonresponders. Specifically R̄tot and R̄insp at 5 Hz and mean difference in resistances at 5 and 19 Hz (R̄5–R̄19) fell significantly more (p = 0.002, p = 0.001 and p = 0.002, respectively), while no differences were seen in Xrs between the groups. The discriminative power tested by the receiver operated characteristic curves was greater when R̄insp was used compared with R̄tot (80.1% and 73.5%, respectively) 28. More details on lung volumes, breathing pattern and Zrs data for responders and nonresponders groups are shown in table E4 in the online supplementary material.
In general, changes in Xrs indices were statistically significant at all forcing frequencies, while the fall in Rrs only occurred consistently when measured during inspiration. As a result R̄tot only decreased significantly relative to baseline at 5 Hz while R̄exp did not change significantly at any frequency. Considering all the patients, R̄insp at 5 Hz was statistically greater than at 19 Hz both before and after BD. However, it was possible to identify a subgroup of seven patients in which this difference was statistically different before BD (p = 0.008), but not after (p = 0.204) BD. These patients were, on average, less obstructed (FEV1 was 56.6±16.02 % predicted (% pred) pre- and 66.9±18.6 % pred post-BD) than the others, enforcing the concept that R̄5–R̄19 can be used as a sensitive index of heterogeneity in airway obstruction. Indeed, R̄5–R̄19 showed a statistically significant decrease, suggesting that the pattern of airway obstruction was on average more homogeneous after BD (table 2⇑, fig. 2⇓).
Effects of expiratory flow limitation on pre- and post-bronchodilator lung function and impedance measurements
Of the 42 patients, 20 were flow-limited at rest pre-bronchodilator and EFL was present in the majority of breaths studied both before and after BD (fig. 3⇓).
At baseline, FEV1 was clearly lower and RV higher in the flow-limited patients, but the differences in other plethysmographic variables did not reach our adjusted significance level.
Of the oscillometric measurements R̄insp, R̄5–R̄19 and all the reactance indices were greater in the flow-limited patients, with nonsignificant differences being seen in the other resistance measurements.
Post-bronchodilator, both groups improved in all spirometric and lung volume variables except for FEV1/FVC and TLC (table 3⇓, fig. 4a⇓). There were significant decreases in R̄insp and R̄tot in patients with EFL, but not in those without EFL where the pre-bronchodilator values for these variables were significantly lower (table 3⇓, fig. 4b⇓). Xrs was significantly less negative after BD at all frequencies in flow-limited patients, while in nonflow-limited subjects, the change in Xrs was significant only at high frequencies.
The effect of flow limitation on indices of lung homogeneity pre- and post-BD
To further investigate the effect of the bronchodilator without the confounding effect of expiratory flow limitation, we considered both R̄tot and R̄insp, and dynamic elastance (i.e. Xrs multiplied by -2πf, where f is the forcing frequency) over the forcing frequencies used in this study (fig. 2⇑). Unlike patients without flow limitation, EFL patients showed a clear pattern of frequency dependence that became more evident when the expiratory phase was excluded. After the bronchodilator, Rrs decreased at all frequencies in nonflow-limited patients, and the changes were similar whether R̄tot or R̄insp data were selected. By contrast, the change in Rrs was much more evident (and statistically significant) in EFL patients when R̄insp was used (fig. 2⇑).
Effects of a bronchodilator on the presence of expiratory flow limitation
Of the 20 flow-limited patients at baseline, eight patients became nonflow-limited after BD, while no patient initially without EFL developed it. The patients where flow limitation was abolished had nonsignificantly different ΔX̄rs values at baseline and changes in the ΔX̄rs after salbutamol compared with those patients where flow-limitation persisted. There was no relationship between inspiratory capacity and ΔX̄rs changes overall in the flow-limited patients (r = 0.107).
DISCUSSION
The development of expiratory flow limitation during tidal breathing identifies a group of COPD patients whose ability to increase their VT to maintain gas exchange is significantly limited 29. Inhaling a bronchodilator drug can potentially have multiple effects in COPD, which may be influenced by the presence of tidal EFL. These include a reduction in airways’ resistance, an abolition of EFL at that operating lung volume or a shift in the distribution of choke points within the lung, all of which can lead to a fall in end-expiratory lung volume that in turn may lead to the persistence of EFL at rest. We used the forced oscillation method to identify the presence of expiratory flow limitation on a breath-by-breath basis, to measure respiratory system mechanics during tidal breathing and to quantify the heterogeneity of lung obstruction in COPD. To do this we used a within-breath multifrequency method that allowed us to assess the heterogeneity of the obstruction [26, 30], and produced comparable data to that measured using the single frequency approach. Our data in a more homogeneous patient group suggest that the response to bronchodilators is more complex than initially proposed 10.
The effect of the high dose β-agonist on resting lung mechanics and breathing pattern we observed in the group as a whole was similar to that reported in other studies of hyperinflated COPD patients 5, 9, 31, with significant increases in FEV1 and inspiratory capacity and falls in RV and TGV. V′E increased, as did mean inspiratory and expiratory flow rates, compatible with the decrease in total and inspiratory resistance. Bronchodilator reversibility defined spirometrically is common in COPD 32. Although nearly half of our patients met the current criteria for a response 27, there was no difference in the magnitude of the IC change in spirometric responders and nonresponders, which helps to explain why these tests are only poorly predictive of the patient's subsequent clinical course 6, 7, 33. However, oscillatory mechanics changed differently in responders and nonresponders, with the Rrs values tracking the changes in FEV1, unlike the reactance values, which followed the inspiratory capacity data. A similar discrepancy between resistance and reactance measurements pattern has been reported during recovery from COPD exacerbations 34, 35, where changes in breathlessness follow those in inspiratory capacity.
Resting expiratory flow limitation was present in just over half the patients. All the breaths studied were consistently classified, except for eight cases where the degree of EFL varied from breath to breath and the classification was based on a majority decision. As expected, flow-limited patients had worse spirometry and a higher RV with a general tendency for higher lung volumes, although these differences were less consistent between groups. Despite this, the response to bronchodilators was almost identical with similar changes in flow and volume indices irrespective of the presence of flow limitation. This does not preclude a different behaviour during exercise in the patients who were flow-limited at rest, but the improvement in exercise performance post-bronchodilator has been consistently related to changes in resting inspiratory capacity, without reference to whether these occurred in flow-limited patients 33, 36.
The changes in oscillatory mechanics noted above were largely driven by changes in flow-limited patients, presumably because the fall in lung volume in the nonflow-limited patients compensated for any improvement in resting resistance or reactance. The fall in group mean Rrs in the flow-limited subjects was due mainly to a reduction in R̄insp of the respiratory system.
Multifrequency testing generated large amounts of data, which have been retained for completeness along with the breathing pattern data as online supplementary material. In general, these showed qualitatively similar changes in response to the bronchodilator to the data measured at 5 Hz. We characterised the heterogeneity of the lung with an index of the frequency dependence of resistance, R̄5–R̄19. It is possible to identify at least four different sources of heterogeneity: serial distribution of airway geometry 37, heterogeneous parallel airway constriction pattern 38, airway wall shunting 30, and heterogeneity of alveolar tissue in heterogeneous parenchymal diseases, such as emphysema 39. Serial distribution of airway geometry affects impedance data mainly for frequencies >100 Hz, providing only negligible contribution at the forcing frequencies used in this study 40. Tissue heterogeneity should not be affected by the administration of BD. As the frequency dependence of Rrs changed statistically significantly after bronchodilator application in most of the patients, with some of them showing no frequency dependence at all after BD, the tissue heterogeneity should not be the dominant mechanism affecting R̄5–R̄19. Parallel airway constriction and airway wall shunting are not easy to differentiate, and it is likely that they are all together contributing to the definition of R̄5–R̄19.
Between-frequency comparisons showed that nonflow-limited patients had a relatively homogeneous distribution of resistance, and their response to the bronchodilator was similar whether R̄tot or R̄insp was plotted. By contrast, flow-limited patients showed a much greater frequency dependence at baseline, suggesting a highly heterogeneous pattern of obstruction. This pattern was apparently little affected by the bronchodilator when total Rrs was considered. However, a clear fall in frequency dependence of Rrs was evident when inspiratory data, unaffected by the artefacts due to EFL, were used. This suggests that BD has a great effect in homogenising time constants throughout the airway tree in flow-limited patients, as also supported by the changes in dynamic elastance with frequency, which are also in agreement with the model prediction of Lutchen et al. 26.
In some patients, bronchodilators abolished expiratory flow limitation in all or in the majority of breaths. This has been seen in other reports 31, although the potential for breath-to-breath variation in the presence of flow limitation complicates the interpretation of data when only a few breaths are sampled, as with the negative expiratory pressure method for identifying EFL. This subset of patients did not differ either in their baseline physiological characteristics before testing or in their degree of within-breath reactance change either before or after treatment.
Our data have some limitations. All studies were conducted at rest and seated, and changes in lung mechanics may not translate to data during exercise, although as already noted there is a good relationship between resting operating lung volumes and exercise performance. Oscillatory signals can be influenced by the shunt compliance provided by the upper airway in COPD. However, each subject is their own control in our data before and after the bronchodilator. Changes related to technical factors, such as the presence of expiratory flow limitation, provide a plausible explanation for the limited bronchodilator response previously reported in COPD using total Zrs data and attributed to upper airway factors 15. Our data have been reported using the multifrequency pseudo-random noise signal, which might have yielded different results to previous single frequency oscillation studies. However, as indicated in the online supplementary material, any differences seen with the systems are likely to relate to physiological differences between breath variation in the degree of flow limitation rather than systematic methodological error. This issue is considered in more detail in the online supplementary material.
In summary, expiratory flow limitation during tidal breathing has an important influence on the changes in resting lung mechanics after bronchodilator drugs in COPD, but it does not predict the magnitude of the subsequent improvement in operating lung volume, at least not at rest. Noninvasive measurements of tidal lung mechanics using the forced oscillation method are an attractive alternative to more usual effort dependent tests of pulmonary function and others have shown that such tests are a sensitive way of detecting bronchodilator effects in these patients 41. However, the change in the total Zrs after a bronchodilator may underestimate the true effects of therapy if expiratory impedance data are not excluded from the analysis in flow-limited patients. Despite these limitations, forced-oscillation data add considerable insight into the way treatment works in COPD and, as a noninvasive, effort-independent methodology, is well suited for monitoring patients in clinical settings where reliable clinical measurement has until now been difficult.
Support statement
The present work was partially supported by the British Lung Foundation. R.L. Dellacà has recieved a European Respiratory Society Fellowship (no. 43).
Statement of interest
Statements of interest for R.L. Dellacà, P.P. Pompilio and A. Pedotti can be found at www.erj.ersjournals.com/misc/statements.dtl
Footnotes
This article has supplementary material accessible from www.erj.ersjournals.com
- Received September 10, 2008.
- Accepted December 29, 2008.
- © ERS Journals Ltd