Skip to main content

Main menu

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • Authors/reviewers
    • Instructions for authors
    • Submit a manuscript
    • Open access
    • COVID-19 submission information
    • Peer reviewer login
  • Alerts
  • Podcasts
  • Subscriptions
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

User menu

  • Log in
  • Subscribe
  • Contact Us
  • My Cart
  • Log out

Search

  • Advanced search
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

Login

European Respiratory Society

Advanced Search

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • Authors/reviewers
    • Instructions for authors
    • Submit a manuscript
    • Open access
    • COVID-19 submission information
    • Peer reviewer login
  • Alerts
  • Podcasts
  • Subscriptions

Biomarker reproducibility in exhaled breath condensate collected with different condensers

P. P. Rosias, C. M. Robroeks, A. Kester, G. J. den Hartog, W. K. Wodzig, G. T. Rijkers, L. J. Zimmermann, C. P. van Schayck, Q. Jöbsis, E. Dompeling
European Respiratory Journal 2008 31: 934-942; DOI: 10.1183/09031936.00073207
P. P. Rosias
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
C. M. Robroeks
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
A. Kester
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
G. J. den Hartog
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
W. K. Wodzig
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
G. T. Rijkers
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
L. J. Zimmermann
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
C. P. van Schayck
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Q. Jöbsis
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
E. Dompeling
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

Optimal collection and analysis of exhaled breath condensate (EBC) are prerequisites for standardisation and reproducibility of assessments. The present study aimed to assess reproducibility of EBC volume, hydrogen peroxide (H2O2), 8-isoprostane and cytokine measurements using different condensers, including a newly developed glass condenser.

At four points in time, 30 healthy subjects performed sequential EBC collections randomly using the following four condensers: glass, silicone, EcoScreen® (Erich Jaeger GmbH, Hoechberg, Germany) and an optimised glass condenser. In small EBC samples, H2O2 was measured by spectrophotometer, 8-isoprostane by enzyme immunoassay, and cytokines by multiplexed xMAP® technology (Luminex Corporation, Austin, TX, USA).

The optimised glass condenser yielded significantly more EBC volume (median 2,025 µL, interquartile range 1,600–2,525). The reproducibility of EBC volume, yielded by the new glass condenser, was comparable with EcoScreen® (19–20 coefficients of variation (CV)%), but was significantly better compared with silicone and glass (29–37 CV%). The new condenser was associated with significantly more detections of H2O2, 8-isoprostane, interleukin-2, -4, -5 and -13, and tumour necrosis factor-α. Isoprostane concentrations were significantly higher using the new condenser, whereas H2O2 and cytokine concentrations were not. Reproducibility of biomarkers was equally variable for all condenser types.

In conclusion, significantly more exhaled breath condensate volume and biomarker detections were found using the optimised glass condenser, including higher 8-isoprostane levels. However, biomarker reproducibility in exhaled breath condensate in healthy adults was not influenced by the type of condenser.

  • Condenser
  • cytokines
  • exhaled breath condensates
  • hydrogen peroxide
  • 8-isoprostane
  • multiplex array

The collection of exhaled breath condensate (EBC) is a noninvasive, safe technique with which to obtain direct samples from the lower respiratory tract, without disturbing an ongoing inflammation 1–3. Analysis of EBC reveals the presence of inflammatory markers, such as eicosanoids, hydrogen peroxide (H2O2) and cytokines 1–3. Although the American Thoracic Society and European Respiratory Society Task Force on EBC published general methodological recommendations on the collection and analysis of EBC, there are still some unresolved methodological pitfalls, as illustrated by the use of various nonstandardised collection systems 1–3.

Optimal condensate collection, and optimal biomarker detection and measurement in EBC are reciprocal prerequisites for any standardisation. However, current condensation systems are suboptimal, with relatively short-measured, open-ended designs and loss of noncondensed exhaled breath, as reflected by variable EBC volumes and biomarker reproducibility 1. Logically, modification by using guided breath flows, enlarged condensation surface, and optimised condensate recovery may improve condensation. Moreover, current designs have different inner coatings, featuring different adhesive interactions with exhaled markers 4. Recently, loss of biomarker within the sampling system was demonstrated, both in vitro and in vivo, for 8-isoprostane and albumin, at the expense of nonglass condenser systems, and in favour of glass and silicone condensers 4. To assess clinical relevance, a study on the reproducibility of these biomarker measurements in vivo, using different condenser coatings including glass, is needed 4. Furthermore, conventional biomarker assays are not always suitable for use in even large sample volumes of condensate, a biofluid highly “diluted” by water vapour 5, whereas new analytical techniques are rapidly emerging and may offer new perspectives 6, 7. Recently, multiplexed cytometric bead array was used in children to simultaneously measure different cytokines in only 50-µL condensate samples; however, the detection level did not reach 50% 6. Liquid bead-based multiplexing xMAP® technology (Luminex Corporation, Austin, TX, USA), based on flow cytometry and (faster) liquid microspheres reaction kinetics, is less laborious, highly sensitive and specific, allows simultaneous measurements in small sample sizes, and improves interarray reproducibility 7. Therefore, it was hypothesised that optimised condensate collection with minimal adhesive properties improves the reproducibility of different measurements in EBC.

The aim of the present study was to assess the effect of four different condensers (glass, silicone, EcoScreen® (Erich Jaeger GmbH, Hoechberg, Germany) and a new, optimised glass condenser) on the reproducibility of breath condensate volume, and the detection, concentration and reproducibility of inflammatory biomarkers, including H2O2, 8-isoprostane, interleukin (IL)-2, -4, -5, -6, -8, -10 and -13, and tumour necrosis factor (TNF)-α, in EBC in healthy nonasthmatic adults.

METHODS

Study subjects

A total of 30 eligible healthy nonasthmatic adult volunteers were recruited (table 1⇓) among medical students and staff at the University Hospital Maastricht (Maastricht, the Netherlands), based on the criterion that each volunteer was able to breath tidally into a mouthpiece for ≥15 min. Exclusion criteria were as follows: history of asthma, upper or lower airway infection, and use of antibiotics, corticosteroids, cromoglycate, nedocromil, theophylline or leukotriene antagonist. Nonasthmatic healthy adults were chosen to eliminate possible confounding factors attributable to heterogeneous disease expression and/or variability of disease control.

View this table:
  • View inline
  • View popup
Table 1—

Subject characteristics of healthy nonasthmatic adult volunteers#

Study design

To assess within-day, between-day and between-week reproducibility, each volunteer was asked to perform the following four tests. Test 1: on the first day in the morning; test 2: on the first day in the afternoon; test 3: the second day at the same time as test 1; and test 4: 1 week after test 1 at the same time. In turn, each test consisted of four sequential EBC collections, using random different types of condenser.

EBC collection

EBC was collected using either of the following condensers: the commercial Teflon-like EcoScreen®, a condenser with exchangeable inner cylinder of silicone or glass, as described previously 4, and a new, optimised glass condenser that was developed in close collaboration with the Dept of Instrument Development Engineering and Evaluation of the Maastricht University (patent number EP 07102586), as described in figure 1⇓. Briefly, the inclined condensation surface is enlarged (using a length of 90 cm), condensate recovery is optimised (using a downwards moveable plunger), and breath flows are turbinately directed towards the condenser wall (by the plunger's multiple breath channels).

Fig. 1—
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 1—

a) A diagrammatic description of the new, optimised glass condenser system. 1: inclined glass condenser with a tube length of 90 cm (to enlarge the condensation surface) and downwards moveable plunger (to optimise condensate recovery from the inner condenser wall), with tangential and axial breath flow channels (to turbinately direct the tidally exhaled breath flows towards the cooled inner condenser wall). 2: swan-neck tubing (serving as gravitational saliva trap) and two-way nonrebreathing valve, connected to a mouthpiece. 3: entrance of inspired room air. 4: cooling unit consisting of a counter-current circulating ice-water pump (▓). 5: removable, cooled glass sample vial (to collect exhaled breath condensate). b) The glass condenser tube with plunger in situ, and c) detail of the glass condenser tube with the downwards moveable plunger in situ. The plunger has three tangential and axial breath flow channels (to turbinately direct the tidally exhaled breath flows towards the cooled inner condenser wall).

To perform one EBC collection, each subject was asked to exhale tidally, while using a nose-clip, through a mouthpiece and two-way nonrebreathing valve connected with the condenser, during a fixed period of 15 min.

Condensate sample processing

Immediately after collection, condensate samples were snap-frozen at -78°C using dry ice and stored at -80°C. Analysis was performed within 3 months from sampling time.

H2O2 was measured in 50-μL EBC, in duplicate by spectrophotometer (UV-VIS Lambda 10 Spectrometer; Perkin Elmer, Shelton/Norwalk, CT, USA) with a lower detection limit of 0.05 µM, as described previously 8.

Isoprostane was measured in 100 μL of EBC, by specific enzyme immunoassay (Cayman Chemical, Ann Arbor, MI, USA), which was modified to reach a lower limit of detection of 1.0 pg·mL−1. The standard curve of this assay ranged 250–1.95 pg·mL−1. It was possible to report 8-isoprostane values as low as 1.0 pg·mL−1, as a logit/log transformation was used. Isoprostane recovery experiments were performed, and coefficients of variation (CV) of the absorption signals were assessed. When spiking for the lower 8-isoprostane values of 3.9 and 7.8 pg·mL−1, an isoprostane recovery of 92% (CV of concentration 16%) and 95% (CV of concentration 15%) were found, respectively. A CV (of absorption signals) <15% was considered highly acceptable. Hence, the corresponding CV of concentrations may be higher. Therefore, the CVs found in the 8-isoprostane recovery experiments using these low values were considered good. Standard curves, patient samples and quality control samples of 2.5 and 10 pg·mL−1 8-isoprostane were assayed in triplicate. Intra-assay variation of standard curves, patient samples and quality control samples had to be <15%, otherwise all samples measured in that assay were excluded and reanalysed. Finally, in all accepted samples, 8-isoprostane concentrations were determined from mean absorption signal intensities.

Cytokines were measured in 50-μL EBC, using the liquid bead-based multiplexing xMAP® technology. Multiplex immunoassay was performed as described previously 7, 9, 10. The corresponding lower limits of cytokine and chemokine detection were as follows (in pg·mL−1): IL-2 (1.0), IL-4 (1.2), IL-5 (1.2), IL-6 (0.4), IL-8 (1.1), IL-10 (1.2), IL-13 (1.0) and TNF-α (1.3). Measurement and data analysis were performed using the Bioplex 100 system and Bioplex Manager software version 3.0 (Bio-Rad Laboratories, Hercules, CA, USA). All multiplex immunoassays were performed in a 96-well format, 1.2-μm filter bottom plates (Millipore, Amsterdam, the Netherlands) and a 12-point standard curve in duplicate was included on every plate. In order to minimise interassay variation, positive and negative control samples were included. As far as possible, EBC samples from one donor series of experiments were run on one plate. Three EBC samples were spiked with either 100 or 10 pg·mL−1 cytokines (IL-2, -4, -5, -6, -8, -10 and -13, and TNF-α), and were measured in quadruplicate. The mean CV of these cytokine measurements was 12.2%. The mean recovery of 100 pg·mL−1 spiked cytokine was 103% (range 71–129%). At 10 pg·mL−1, the mean recovery was somewhat lower at 89% (ranging from 64% in IL-13 to 111% in IL-4). It was concluded that there was a slight matrix effect of EBC, but this did not result in an overestimation of cytokine measurements.

Statistics

Not normally distributed data were expressed as median (interquartile range). Normally distributed data were expressed as mean±se. To estimate variance within a single method of measurement (coating), CV were used.

CV = (SD/mean) × 100% (1)

The CV was calculated as the mean of individual CVs, calculated over the two or four relevant measurements. Within-subject CVs were used as data distribution was normal. When the distribution of individual CVs was not normal, the nonparametric Friedman test was used to see if these CVs were different between coatings. ANOVA was used to test for differences among normally distributed repeated measures. The Chi-squared test was used to statistically evaluate the differences between proportions for four groups in a data set. Condensate samples with a biomarker concentration below the lower detection limit, i.e. in strictu sensu negative detections, were not considered as missing values because they actually informed that marker concentrations were below the lower detection limit. Therefore, these negative detections were given an arbitrary value between zero and the lower detection limit as follows: 0.025 µM for H2O2, 0.1 pg·mL−1 for 8-isoprostane and 0.1 pg·mL−1 for cytokines. Samples were defined as missing if the EBC volume, yielded after 15 min collection time, was either zero or insufficient to analyse.

Power calculation

Ideally, power calculations should be based on the expected changes in biomarker concentrations and on their variability. However, these changes in biomarker concentrations and their variability were unknown. In fact, the objective of the present study was to answer this question. Therefore, the uncertainty of the estimated variability was taken as a basis for power calculations. Data from 30 volunteers were thus available for each of the estimated variances (within-day, between-day and between-week). Using standard results for the variance of a Chi-squared random variable, the present authors inferred that, with the 30 subjects, the relative confidence limits (relative to the observed value of the variance) would be 0.634 and 1.807. For the sd, this implies relative limits of 0.80 and 1.34 11.

Ethics

All parents gave written informed consent. The study was approved by the Medical Ethics Committee of the University Hospital of Maastricht.

RESULTS

EBC volume

In five out of 480 manoeuvres performed, the present authors were unable to collect condensate due to erroneously connected tubing. EBC collection by the optimised glass condenser yielded significantly more median condensate volume compared with the other condensers (p = 0.001, Friedman test; table 2⇓). Within-day, between-day, between-week and overall reproducibility of EBC volume (expressed as CV%) were comparable in the new condenser and EcoScreen® (p = 0.715, Chi-squared test), but was significantly better compared with silicone and glass (p<0.028, Chi-squared test).

View this table:
  • View inline
  • View popup
Table 2—

Volume of exhaled breath condensate(EBC), and reproducibility (expressed as coefficients of variation (CV)) in four different condenser systems

H2O2 measurements in EBC

Overall, 29% of H2O2 measurements were missing (table 3⇓). Significantly more positive H2O2 detections were found using the optimised glass condenser, compared with silicone and glass (p<0.050, Chi-squared test). Median H2O2 concentrations and reproducibility (expressed as CV%) did not significantly differ between the four condensers (p = 0.286 and p>0.080, respectively, Friedman test).

View this table:
  • View inline
  • View popup
Table 3—

Hydrogen peroxide(H2O2) measurements in exhaled breath condensate (EBC) presented as detection, concentration and reproducibility (expressed as coefficients of variation (CV)) in four different condenser systems

8-Isoprostane measurements in EBC

Overall, 13% of 8-isoprostane measurements were missing (table 4⇓). Significantly more positive 8-isoprostane detections were found using the optimised glass condenser, compared with silicone and glass (p<0.023, Chi-squared test). The median concentration of 8-isoprostane was significantly higher using the new condenser compared with the other three condensers (p = 0.001, Friedman test). Statistically, 8-isoprostane reproducibility (expressed as CV%) did not significantly differ between the four condensers (p>0.151, Friedman test).

View this table:
  • View inline
  • View popup
Table 4—

8-Isoprostane (IP) measurements in exhaled breath condensate (EBC) presented as detection, concentration and reproducibility (expressed as coefficients of variation (CV)) in four different condenser systems

Cytokine measurements in EBC

Overall, 20% of cytokine measurements were missing (table 5⇓). The levels of cytokine detection using xMAP® technology were as follows: IL-2 (68%), IL-4 (97%), IL-5 (73%), IL-6 (46%), IL-8 (46%), IL-10 (61%), IL-13 (70%) and TNF-α (64%). The optimised glass condenser had significantly more positive detections of IL-2, -4, -5, -13 and TNF-α, compared with silicone and glass (p<0.050, Chi-squared test), and more IL-5 and -13 detections, compared with EcoScreen® (p<0.021, Chi-squared test). Cytokine concentrations were not significantly different between the four condenser types (p>0.113, Friedman test or ANOVA, respectively; table 6⇓). Reproducibility (expressed as CV%) of cytokine measurements did not significantly differ, either overall (table 7⇓) or within-day, between-day and between-week (data not shown). The best range of overall CV% was found for the measurement of IL-6 (11–14%), IL-8 (2–15%), IL-10 (11–30%) and TNF-α (8–22%). When CVs were evaluated using the positive values only (without negative and/or arbitrary values), again no differences between condensers were found (data not shown).

View this table:
  • View inline
  • View popup
Table 5—

Number of positive to negative detections of cytokines in exhaled breath condensate, collected with four different condenser systems

View this table:
  • View inline
  • View popup
Table 6—

Concentration of cytokine measurements in exhaled breath condensate(EBC), collected with four different condenser systems

View this table:
  • View inline
  • View popup
Table 7—

Overall reproducibility(expressed as coefficients of variation) of cytokine measurements in exhaled breath condensate, collected with four different condenser systems

DISCUSSION

The present authors have demonstrated that EBC volume and the detection of biomarkers were significantly influenced by the condenser system, in favour of the new glass condenser, whereas biomarker reproducibility was not influenced by the type of condenser. The proposed optimised glass condenser yielded significantly more condensate volume compared with the silicone, glass and EcoScreen® condensers. Reproducibility of EBC volume was comparable for the new condenser and EcoScreen®, and was significantly better compared with the other two condenser types.

In EBC collected with this new condenser, significantly more positive H2O2, 8-isoprostane, IL-2, -4, -5 and -13, and TNF-α detections were found, supporting improved sampling of EBC. Moreover, 8-isoprostane concentrations were significantly increased in EBC yielded by the new condenser compared with the other three condensers, which is in accordance with former findings 4. Conversely, reproducibility of H2O2, 8-isoprostane and cytokine measurements in EBC did not significantly differ between the four condensers, suggesting no significant influence of the type of condenser coating on reproducibility. In the literature to date, no formal study addressing this issue has been published, although reproducibility using other analytical techniques has been reported 12–14, and/or could not be calculated due to the small number of subjects 15–17.

Variations in biomarker measurements in EBC may be attributed to variations in the dilution and/or quality of condensate that is influenced by different collection techniques and procedures, sample processing and storage conditions, and/or sensitivity of the analytical techniques used 17. The levels of the highly volatile H2O2 in EBC may be susceptible to different cooling temperatures during collection 18, circadian rhythm 19, flow dependency 20, different methods of measurement with widely varying values (even in healthy subjects) close to lower detection limits 3, 17, 21, 22, and high chemical reactivity with salivary and exhaled compounds by which (some) H2O2 is consumed over time during collection and storage 3, 17, 22. Isoprostanes are relatively stable end-products of in vivo arachidonate peroxidation and are measured in EBC by immunoassay, which might be influenced by cross-reactivity with closely related substances 17. Other confounding factors may be age, diets rich in antioxidants and smoking habits 17.

The use of CV is not always the ideal way to express variability: when mean values are low, CV values can be abnormally high. Therefore, results in the present study were also expressed as sd values and intraclass correlation coefficients. When using both of these alternative expression methods, results were comparable: no difference in biomarker variability between condensers was found (data not shown). Furthermore, biomarker levels were assessed in condensate originating from healthy adults and, thus, may have been more pronounced in a steroid-naïve population with documented chronic respiratory inflammation and comparable levels of disease control. Hence, increased mean concentrations and lowered CVs in diseased subjects could be expected. Conversely, with these very low concentrations of cytokines and 8-isoprostane, an influence of analytical variability cannot be ruled out.

Cytokines were simultaneously measured by xMAP® technology in small, 50-μL EBC samples. The overall level of detection was 46–97%, which was much better compared with cytometric bead array in small samples in children (<50%), and in large 1,000–2,000-µL lyophilisated samples in adults (3–100%) 6, 23–26. When compared with conventional (solid-phase) immunoassays, multiplexed immunoassays detect bioactive and inactive molecules, have a growing analytical range, are rapid (take hours instead of days to perform), have good precision (CV 10–15%), are not interfered with by drugs, and have simple protocols 7.

Currently, different nonstandardised techniques to collect EBC are in use 1–3. All systems are based on the cooling of exhaled breath, whereas their design may vary from immersed plastic tubing over glass distilling columns, to commercial systems, such as the lamellar Teflon-like EcoScreen® condenser, the hand-held disposable polypropylene RTubeTM (Respiratory Inc., Charlottesville, VA, USA), and the thermostatically controlled polyethylene Turbo-Deccs (ItalChill, Parma, Italy) 1–3, 27. These designs implicate relatively short-measured and open-ended systems that tolerate the needless loss of noncondensed exhaled breath, whether initially, or after a prolonged collection time. Moreover, biomarkers may also be lost within these collection systems, as recently demonstrated for 8-isoprostane and albumin measurements, both in vitro and in vivo, in nonglass condensers (including EcoScreen®) 4. This superiority of glass coatings may be mainly related to the behaviour of water as bipolar vehicle 4. Three other studies report the influence of sampling systems on biomarkers in EBC 28–30. Tufvesson and Bjermer 28 proposed to coat EcoScreen® collection surfaces with bovine serum albumin and Tween-20, to measure cytokines and eicosanoids, respectively. However, they also reported possible false-positive (eicosanoid) results, and the need of sample concentration, by vacuum centrifugation, prior to analysis. Soyer et al. 29 found significantly higher cysteinyl leukotrienes and eotaxin using EcoScreen®, compared with RTubeTM, due to susceptibility of pre-cooled RTubeTM sleeves to (increased) ambient temperatures during collection, and due to different materials that could affect sample recovery. Prieto et al. 30 compared RTubeTM and EcoScreen®, and reported that EBC pH values are dependent on the collection device used. Furthermore, the EcoScreen® has been associated with deposition of frozen condensate on its lamellar walls. For these reasons, an optimised glass condenser system was developed by the present authors. The new glass condenser had an improved condensation process and condensate recovery, using an inclined and enlarged condensation surface, with a condensate sweeping plunger, having tangentially and axially guiding breath-flow channels.

In the present study, these improvements resulted in significantly increased EBC volumes, and increased biomarker detections with the new glass condenser compared with silicone, glass and EcoScreen®. This suggests both an improved condensation process and an increased opportunity to perform a broad spectrum of analyses. Moreover, the optimised glass and EcoScreen® condensers were both significantly associated with less variation in the generated EBC volume compared with the other condensers, thereby reducing a possible confounding influence of the variable quantity of EBC collected over a given time, and even within individuals.

Optimisation of EBC collection using the modified new glass condenser with statistically equivalent CVs compared with the commercial EcoScreen®, and optimisation of EBC analysis using rapid, multiplexed measurement of cytokines in small EBC sample sizes may, in combination, open a window of opportunities, even in strained collection procedures, such as in young or dyspnoeic subjects (with less-sustained efforts to cope with sampling procedures), by allowing the search for, and identification of, particular profiles of different exhaled markers involved in the regulation of chronic respiratory inflammation for diagnostic and monitoring purposes.

In conclusion, the optimised glass condenser yielded significantly more exhaled breath condensate volume, and with good reproducibility. Furthermore, significantly more positive detections of hydrogen peroxide, 8-isoprostane, interleukin-2, -4, -5 and -13, and tumour necrosis factor-α were found in exhaled breath condensate collected with the new condenser, thereby offering an increased capacity to analyse for complex biomarker profiles. Moreover, concentrations of 8-isoprostane were significantly increased using the optimised glass condenser compared with the other three condensers. However, reproducibility of biomarker measurements in exhaled breath condensate was not influenced by the type of condenser.

Statement of interest

None declared.

Acknowledgments

The authors wish to thank J. Suykerbuyk (Dept of Paediatrics, University Hospital Maastricht, Maastricht, the Newtherlands) for her technical support in the collection of exhaled breath condensate; M. Meers (Dept of Clinical Chemistry, University Hospital Maastricht) for her technical support in the measurement of 8-isoprostane; N. van Uden (Dept of Immunology, University Medical Centre Wilhelmina Children's Hospital, Utrecht, the Netherlands) for her technical support in the measurement of cytokines; and, last but not least, the volunteers for their participation.

  • Received June 19, 2007.
  • Accepted December 19, 2007.
  • © ERS Journals Ltd

References

  1. ↵
    Rosias PPR, Dompeling E, Hendriks JJE, Heijnens JWCM, Donckerwolcke RAMG, Jöbsis Q. Exhaled breath condensate in children: pearls and pitfalls. Pediatr Allergy Immunol 2004;15:4–19.
    OpenUrlCrossRefPubMedWeb of Science
  2. Jöbsis Q, Rosias PPR. Analysis of exhaled breath condensate in children. In: Montuschi P, ed. New Perspectives in Monitoring Lung Inflammation. London, Taylor & Francis Publishers, 2004; 105–111
  3. ↵
    Horváth I, Hunt J, Barnes PJ. on behalf of the ATS/ERS Task Force. Exhaled breath condensate: methodological recommendations and unresolved questions. Eur Respir J 2005;26:523–548.
    OpenUrlAbstract/FREE Full Text
  4. ↵
    Rosias PP, Robroeks CM, Niemarkt HJ, et al. Breath condenser coatings affect measurement of biomarkers in exhaled breath condensate. Eur Respir J 2006;28:1036–1041.
    OpenUrlAbstract/FREE Full Text
  5. ↵
    Huszar E, Szabo Z, Jakab A, Barta I, Herjavecz I, Horvath I. Comparative measurement of thromboxane A2 metabolites in exhaled breath condensate by different immunoassays. Inflamm Res 2005;54:350–355.
    OpenUrlCrossRefPubMedWeb of Science
  6. ↵
    Robroeks CM, Jöbsis Q, Damoiseaux JGMC, et al. Cytokines in exhaled breath condensate of children with asthma and cystic fibrosis. Ann Allergy Asthma Immunol 2006;96:349–355.
    OpenUrlPubMedWeb of Science
  7. ↵
    de Jager W, Rijkers GT. Solid-phase and bead-based cytokine immunoassay: a comparison. Methods 2006;38:294–303.
    OpenUrlCrossRefPubMedWeb of Science
  8. ↵
    Dekhuijzen PN, Aben KK, Dekker I, et al. Increased exhalation of hydrogen peroxide in patients with stable and unstable chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1996;154:813–816.
    OpenUrlCrossRefPubMedWeb of Science
  9. ↵
    de Jager W, Prakken BJ, Bijlsma JW, Kuis W, Rijkers GT. Improved multiplex immunoassay performance in human plasma and synovial fluid following removal of interfering heterophilic antibodies. J Immunol Methods 2005;300:124–135.
    OpenUrlCrossRefPubMedWeb of Science
  10. ↵
    de Jager W, te Velthuis H, Prakken BJ, Kuis W, Rijkers GT. Simultaneous detection of 15 human cytokines in a single sample of stimulated peripheral blood mononuclear cells. Clin Diagn Lab Immunol 2003;10:133–139.
    OpenUrlCrossRefPubMed
  11. ↵
    Rosner B. Fundamentals of Biostatistics. 6th Edn. Belmont, Thomson Brooks/Cole, 2006; p. 200
  12. ↵
    Brooks WM, Lash H, Kettle AJ, Epton MJ. Optimising hydrogen peroxide measurement in exhaled breath condensate. Redox Rep 2006;11:78–84.
    OpenUrlCrossRefPubMedWeb of Science
  13. Gerritsen WB, Zanen P, Bauwens AA, van den Bosch JM, Haas FJ. Validation of a new method to measure hydrogen peroxide in exhaled breath condensate. Respir Med 2005;99:1132–1137.
    OpenUrlCrossRefPubMedWeb of Science
  14. ↵
    Gonzalez-Reche LM, Musiol AK, Muller-Lux A, Kraus T, Goen T. Method optimization and validation for the simultaneous determination of arachidonic acid metabolites in exhaled breath condensate by liquid chromatography-electrospray ionization tandem mass spectrometry. J Occup Med Toxicol 2006;1:5
    OpenUrlCrossRefPubMed
  15. ↵
    Van Hoydonck PG, Wuyts WA, Vanaudenaerde BM, Schouten EG, Dupont LJ, Temme EH. Quantitative analysis of 8-isoprostane and hydrogen peroxide in exhaled breath condensate. Eur Respir J 2004;23:189–192.
    OpenUrlAbstract/FREE Full Text
  16. Kharitonov SA. Exhaled markers of inflammatory lung diseases: ready for routine monitoring?. Swiss Med Wkly 2004;134:175–192.
    OpenUrlPubMed
  17. ↵
    Rahman I, Biswas SK. Non-invasive biomarkers of oxidative stress: reproducibility and methodological issues. Redox Rep 2004;9:125–143.
    OpenUrlCrossRefPubMedWeb of Science
  18. ↵
    Goldoni M, Caglieri A, Andreoli R, et al. Influence of condensation temperature on selected exhaled breath parameters. BMC Pulmon Med 2005;5:10
    OpenUrlCrossRef
  19. ↵
    Nowak D, Kalucka S, Bialasiewicz P, Krol M. Exhalation of H2O2 and thiobarbituric acid reactive substances (TBARs) by healthy subjects. Free Radic Biol Med 2001;30:178–186.
    OpenUrlCrossRefPubMedWeb of Science
  20. ↵
    Schleiss MB, Holz O, Behnke M, Richter K, Magnussen H, Jorres RA. The concentration of hydrogen peroxide in exhaled air depends on expiratory flow rate. Eur Respir J 2000;16:1115–1118.
    OpenUrlAbstract
  21. ↵
    Rahman I. Reproducibility of oxidative stress biomarkers in breath condensate: are they reliable?. Eur Respir J 2004;23:183–184.
    OpenUrlFREE Full Text
  22. ↵
    Silkoff PE, Erzurum SC, Lundberg JO, et al. ATS workshop proceedings: exhaled nitric oxide and nitric oxide oxidative metabolism in exhaled breath condensate. Proc Am Thorac Soc 2006;3:131–145.
    OpenUrlCrossRefPubMed
  23. ↵
    Sack U, Scheibe R, Wötzel M, et al. Multiplex analysis of cytokines in exhaled breath condensate. Cytometry A 2006;69:169–172.
    OpenUrlPubMed
  24. Gessner C, Scheibe R, Wötzel M, et al. Exhaled breath condensate cytokine patterns in chronic obstructive pulmonary disease. Respir Med 2005;99:1229–1240.
    OpenUrlCrossRefPubMedWeb of Science
  25. Matsunaga K, Yanagisawa S, Ichikawa T, et al. Airway cytokine expression measured by means of protein array in exhaled breath condensate: correlation with physiologic properties in asthmatic patients. J Allergy Clin Immunol 2006;118:84–90.
    OpenUrlCrossRefPubMedWeb of Science
  26. ↵
    Schumann C, Triantafilou K, Krueger S, et al. Detection of erythropoietin in exhaled breath condensate of nonhypoxic subjects using a multiplex bead array. Mediators Inflamm 2006;5:18061
    OpenUrl
  27. ↵
    Caglieri A, Goldoni M, Acampa O, et al. The effect of inhaled chromium on different exhaled breath condensate biomarkers among chrome-plating workers. Environ Health Perspect 2006;114:542–546.
    OpenUrlPubMedWeb of Science
  28. ↵
    Tufvesson E, Bjermer L. Methodological improvements for measuring eicosanoids and cytokines in exhaled breath condensate. Respir Med 2006;100:34–38.
    OpenUrlCrossRefPubMedWeb of Science
  29. ↵
    Soyer OU, Dizdar EA, Keskin O, Lilly C, Kalayci O. Comparison of two methods for exhaled breath condensate collection. Allergy 2006;61:1016–1018.
    OpenUrlCrossRefPubMedWeb of Science
  30. ↵
    Prieto L, Ferrer A, Palop J, Domenech J, Llusar R, Rojas R. Differences in exhaled breath condensate pH measurements between samples obtained with two commercial devices. Respir Med 2007;101:1715–1720.
    OpenUrlCrossRefPubMedWeb of Science
PreviousNext
Back to top
View this article with LENS
Vol 31 Issue 5 Table of Contents
European Respiratory Journal: 31 (5)
  • Table of Contents
  • Index by author
Email

Thank you for your interest in spreading the word on European Respiratory Society .

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Biomarker reproducibility in exhaled breath condensate collected with different condensers
(Your Name) has sent you a message from European Respiratory Society
(Your Name) thought you would like to see the European Respiratory Society web site.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Print
Citation Tools
Biomarker reproducibility in exhaled breath condensate collected with different condensers
P. P. Rosias, C. M. Robroeks, A. Kester, G. J. den Hartog, W. K. Wodzig, G. T. Rijkers, L. J. Zimmermann, C. P. van Schayck, Q. Jöbsis, E. Dompeling
European Respiratory Journal May 2008, 31 (5) 934-942; DOI: 10.1183/09031936.00073207

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero

Share
Biomarker reproducibility in exhaled breath condensate collected with different condensers
P. P. Rosias, C. M. Robroeks, A. Kester, G. J. den Hartog, W. K. Wodzig, G. T. Rijkers, L. J. Zimmermann, C. P. van Schayck, Q. Jöbsis, E. Dompeling
European Respiratory Journal May 2008, 31 (5) 934-942; DOI: 10.1183/09031936.00073207
del.icio.us logo Digg logo Reddit logo Technorati logo Twitter logo CiteULike logo Connotea logo Facebook logo Google logo Mendeley logo
Full Text (PDF)

Jump To

  • Article
    • Abstract
    • METHODS
    • RESULTS
    • DISCUSSION
    • Statement of interest
    • Acknowledgments
    • References
  • Figures & Data
  • Info & Metrics
  • PDF
  • Tweet Widget
  • Facebook Like
  • Google Plus One

More in this TOC Section

  • Validation of assays for inflammatory mediators in exhaled breath condensate
  • Extracellular purines are biomarkers of neutrophilic airway inflammation
Show more Original Articles: Bench to bedside: exhaled biomarkers

Related Articles

Navigate

  • Home
  • Current issue
  • Archive

About the ERJ

  • Journal information
  • Editorial board
  • Reviewers
  • Press
  • Permissions and reprints
  • Advertising

The European Respiratory Society

  • Society home
  • myERS
  • Privacy policy
  • Accessibility

ERS publications

  • European Respiratory Journal
  • ERJ Open Research
  • European Respiratory Review
  • Breathe
  • ERS books online
  • ERS Bookshop

Help

  • Feedback

For authors

  • Instructions for authors
  • Publication ethics and malpractice
  • Submit a manuscript

For readers

  • Alerts
  • Subjects
  • Podcasts
  • RSS

Subscriptions

  • Accessing the ERS publications

Contact us

European Respiratory Society
442 Glossop Road
Sheffield S10 2PX
United Kingdom
Tel: +44 114 2672860
Email: journals@ersnet.org

ISSN

Print ISSN:  0903-1936
Online ISSN: 1399-3003

Copyright © 2022 by the European Respiratory Society