Abstract
Exhaled breath condensate collection is not yet standardised and biomarker measurements are often close to lower detection limits. In the current study, it was hypothesised that adhesive properties of different condenser coatings interfere with measurements of eicosanoids and proteins in breath condensate.
In vitro, condensate was derived from a collection model using two test solutions (8-isoprostane and albumin) and five condenser coatings (silicone, glass, aluminium, polypropylene and Teflon). In vivo, condensate was collected using these five coatings and the EcoScreen® condenser to measure 8-isoprostane, and three coatings (silicone, glass, EcoScreen®) to measure albumin.
In vitro, silicone and glass coatings had significantly higher albumin recovery compared with the other coatings. A similar trend was observed for 8-isoprostane recovery. In vivo, median (interquartile range) 8-isoprostane concentrations were significantly higher using silicone (9.2 (18.8) pg·mL-1) or glass (3.0 (4.5) pg·mL-1) coating, compared with aluminium (0.5 (2.4) pg·mL-1), polypropylene (0.5 (0.5) pg·mL-1), Teflon (0.5 (0.0) pg·mL-1), and EcoScreen® (0.5 (2.0) pg·mL-1). Albumin in vivo was mainly detectable using glass coating.
In conclusion, a condenser with silicone or glass coating is more efficient for measurement of 8-isoprostane or albumin in exhaled breath condensate, than EcoScreen®, aluminium, polypropylene or Teflon. Guidelines for exhaled breath condensate standardisation should include the most valid condenser coating to measure a specific biomarker.
The collection of exhaled breath condensate (EBC) has been rediscovered as a simple and noninvasive technique to measure mediators of airway inflammation. EBC consists not only of water vapor, but also contains aerosolised respiratory fluid droplets released from the respiratory epithelial lining fluid (ELF). These fluid droplets contain traces of nonvolatile solutes, which can be recovered in EBC samples 1. EBC is collected by guiding and cooling exhaled air of a tidally breathing subject in a condenser system. EBC does not affect the airway; in contrast to bronchial biopsy, bronchoalveolar lavage and induced sputum. It can be obtained with minimal risk and minimal inconvenience for both adults and children 2. Although the American Thoracic Society and European Respiratory Society Task Force on EBC recently published general methodological recommendations on the collection of EBC, there are still some methodological pitfalls and unresolved questions 3, 4. The most commonly used condensers are the EcoScreen® (Erich Jaeger GmbH, Hochberg, Germany), RTube® (Respiratory Research Inc., Charlottesville, VA, USA) and home-made glass or Teflon devices.
Two important groups of inflammatory biomarkers in EBC of adults and children are eicosanoids and cytokines 3–10. Eicosanoids, such as 8-isoprostane, are formed by lipid peroxidation of arachidonic acid during oxidative stress 11. 8-Isoprostane (molecular weight 354 Dalton) can be measured in EBC of adults and children, although reproducibility remains controversial 12, 13. Cytokines are low molecular weight (<80,000 Dalton) proteins involved in mediating inflammation and tissue repair. Reports on the detection of cytokines in EBC are incidental, with the exception of a report on interleukin (IL)-6 by Carpagnano et al. 14. Albumin, a protein with a similar molecular weight (66,000 Dalton), was demonstrated in induced sputum and EBC in asthmatic subjects 15, 16.
The reported mean biomarker values in EBC are usually situated in the lower range of detection 3, 6. It may be that the biomarker concentration in ELF is intrinsically low or that only minor amounts are released from ELF into exhaled air. An alternative explanation may be that adhesive properties of different inner condenser coatings interfere with the detection of various inflammatory markers in EBC 13, 17.
The aim of the present study was to investigate the influence of different inner condenser coating surfaces (silicone, borosilicate glass, aluminium, polypropylene, Teflon and EcoScreen® with a Teflon-like coating) on the measurement of 8-isoprostane and albumin in EBC using both an in vitro and in vivo approach.
METHODS
In vitro studies
In close collaboration with the Dept of Instrument Development Engineering & Evaluation of the University of Maastricht (Maastrict, the Netherlands) an in vitro EBC collection model was developed consisting of a rechargeable condenser system connected in series with an ultrasonic nebuliser (Ultra-NebTM2000; DeVilbiss, Langen, Germany) and a ventilator (Servo-900C; Siemens, Solna, Sweeden; fig. 1a⇓). The rechargeable condenser system consisted of a fixed 30-cm metal outer cylinder (diameter 4 cm; Felix Philips Maatmetaal BV, Maastricht, the Netherlands) connected to a counter-current circulating ice-water pump at 0°C, and an exchangeable inner cylinder (diameter 2cm), as shown in figure 1b⇓. The exchangeable inner cylinder was available in five different coatings: silicone, borosilicate glass, aluminium, polypropylene and Teflon. The borosilicate glass tubes were manufactured by Louwers Glass and Ceramic Technologies (Hapert, the Netherlands). Identical glass tubes were coated internally with silicone by Imbreglon BV (Beuningen, the Netherlands). The aluminium tubes were manufactured by Felix Philips Maatmetaal BV. Identical aluminium tubes were coated internally with Tempcoat 1011F (Teflon) by Imbreglon BV. The polypropylene tubes were produced by Applikon BV (Schiedam, the Netherlands). The outer surface of the inner cylinder was in direct contact with the counter current circulating ice water, while its inner surface was in direct contact with the exhaled breath or the volatile substances derived from the nebuliser and propelled by the ventilator. The nebuliser was filled with a biomarker test solution (42 pg·mL-1 8-isoprostane or 0.1 mg·mL-1 albumin in a 22 mL saline 0.9% solution), which was nebulised and propelled by a ventilator through the condenser system, simulating tidal breathing for 15 min (ventilator settings: inspiration time 33%, respiratory frequency 20·min-1, tidal volume 200 mL). The construction of the collection device only allowed contact between the propelled air/exhaled breath and one specific type of inner coating that was chosen at that moment (fig. 1b⇓). The condensate was collected at the open end of the condenser device directly into vials. For each coating, 15 experiments were performed (see the power analysis in the Statistics section). After each experiment, the coated tubes were rinsed repeatedly with bi-distilled water and subsequently dried at room air for ≥24 h. Multiple coated tubes of each type were available to allow a smooth continuation of the experiments, without any delay due to the drying of the coated tube that was used in the preceding test. The biomarker condensate recovery percentage was defined as the ratio of biomarker concentration in condensate, to biomarker concentration in the test solution.
In vivo 8-isoprostane study
As 8-isoprostane has been detected in EBC of healthy subjects, the influence of condenser coatings on 8-isoprostane measurements in vivo was studied in 28 healthy volunteers (see the power analysis in the Statistics section). Standardised questionnaires from the International Study of Asthma and Allergies in Childhood (ISAAC) were completed to exclude the presence of respiratory infections, asthma or allergic disease 18. Each subject was asked to breathe tidally, while wearing a nose-clip, into a mouthpiece connected to a two-way non-rebreathing valve that was in turn connected by tubing to the condenser. The two-way valve and tubing also served as a saliva trap. In this way, each subject exhaled for 15 min along each of the five coatings in random order, as well as along the EcoScreen®, a commercial condenser system with a modified Teflon coating.
In vivo albumin study
In contrast to 8-isoprostane, the in vivo albumin study was planned in asthmatic children, as the chance of positive albumin detection is larger in a disease population 16. EBC was collected in 40 asthmatic children. Inclusion criteria were doctor-diagnosed mild or moderate persistent asthma requiring daily inhaled corticosteroids for ≥6 months 19. Exclusion criteria were clinical evidence of respiratory infection 4 weeks prior to study, and comorbidity (mental retardation, respiratory tract anomalies or cardiac defects). In contrast to the in vivo 8-isoprostane study, the asthmatic children in the albumin study were only asked to exhale tidally along one type of coating (silicone, glass or EcoScreen®), to allow parallel group comparison. Children were randomly allocated to one out of three coating groups.
Sample processing
Immediately after collection, the condensate samples were snap-frozen at -78°C using dry ice and subsequently stored at -80°C until analysis. A reaction mixture of the immunoassay was not added to the condensate samples before storage. In order to maintain optimal condensate sample preservation, for subsequent 8-isoprostane and albumin measurement, samples were analysed within 6–8 weeks of storage at -80°C. Samples were only defrosted once, at the time of analysis. All isoprostane and albumin concentrations were determined in duplicate by specific enzyme immunoassay (Cayman Chemical®, Ann Arbor, MI, USA) and sandwich ELISA respectively 16. The lower limit of detection was 2.45 and 32 pg·mL-1, respectively.
Statistics
The condensate recovery percentage of a specific coating was the primary outcome measure of the in vitro experiments. Mean values and standard deviations of recovery percentages were calculated. A comparison between the five different coatings and between the two biomarkers in vitro was made by means of univariate ANOVA. Power analysis demonstrated a required number of 15 experiments in each group, in order to detect a relevant difference between coatings of 25% with a power of 90%, an alpha of 0.05 and an SD of 20% (the SD in the first set of 10 experiments was 20%).
In the in vivo studies, the number of positive detections, median and mean values, SD (based on the first 15 subjects) and interquartile ranges (IQRs) were calculated. Power calculation showed that 23 patients are needed to detect a 10% difference in biomarker concentration with a power of 90%, an alpha of 0.05 and an SD of 14% (paired t-test). Due to a number of negative detections, the Wilcoxon paired signed-rank test was used to compare data within subjects. A sample with negative detection is not considered as a missing value, because it actually informs that the measurement of a marker was below the detection limit. Therefore, the negative detection is given an arbitrary value of 0.5.
Ethics
All subjects gave written informed consent. The study was approved by the Medical Ethics Committee of the University Hospital of Maastricht.
RESULTS
In vitro study of albumin and 8-isoprostane
Albumin was detected in 93% of condensate samples (fig. 2⇓). Overall, the mean albumin recovery in condensate was only 52% using this experimental model. The albumin recovery percentages varied from 22.9% using a Teflon coating, to 69.4 and 94.7% using a glass and silicone coating, respectively. The albumin recovery percentage was significantly higher using the silicone and glass coating compared with the other coatings (p = 0.03, ANOVA). No other statistically significant differences were present.
Isoprostane was detected in 77 out of 78 condensate samples (fig. 3⇓). The isoprostane recovery percentages varied from 93.1% using an aluminum coating to 102.7 and 105.6% using a glass and silicone coating, respectively, which did not reach statistical significance (ANOVA, F-test, p = 0.09). Overall, the mean 8-isoprostane recovery in condensate was close to 100%, which was significantly higher than the overall albumin recovery percentage of 52% (ANOVA, p = 0.0001).
In vivo study of 8-isoprostane
Subject characteristics are shown in table 1⇓. Isoprostane was detected in 41% of condensate samples (fig. 4⇓). For each coating, the number of positive/negative isoprostane detections in EBC was 23/five (silicone), 15/13 (glass), 11/17 (aluminium), seven/21 (polypropylene), five/23 (Teflon) and eight/20 (EcoScreen®). The median (IQR) 8-isoprostane concentrations in EBC in 28 healthy subjects ranged from 0.5 pg·mL-1 for aluminium (IQR 0.5–2.9), polypropylene (0.5–1.0), Teflon (0.5–0.5) and the EcoScreen® (0.5–2.5), to 3.0 (0.5–5.0) pg·mL-1 and 9.2 (5.1–23.9) pg·mL-1 for the glass and silicone coating, respectively. EBC 8-isoprostane concentrations were significantly higher using silicone coating compared with each of the other coatings (p<0.001, Wilcoxon signed-rank test). Similarly, glass coating was better than polypropylene, Teflon and the EcoScreen® (p<0.02, Wilcoxon test). The 8-isoprostane concentrations obtained with aluminium, polypropylene, Teflon and the EcoScreen® did not differ from each other (p = 0.763, Friedman test). As median values and IQRs in smokers and nonsmokers were comparable, smoking was not a confounder in this study.
In vivo study of albumin
Subject characteristics are shown in table 1⇑. Albumin was detected in eight out of 40 condensate samples (0.04–0.24 μg·mL-1). Positive detections were almost exclusively found with a glass coating: seven out of 13 samples using a glass coating, one out of 14 samples using the EcoScreen® and none out of 13 samples using a silicone coating (p<0.008, Chi-squared test). Disease stability, lung function and atopic status were not confounding factors, as these characteristics were equally distributed across the three coating groups (data not shown).
DISCUSSION
Both in vitro and in vivo, the present study demonstrates a significant influence of the different inner condenser coatings on the measurement of eicosanoid and protein biomarkers in EBC. In particular, silicone and glass coatings were superior to aluminum, polypropylene and Teflon with respect to the recovery of both 8-isoprostane and albumin in EBC in vitro and in vivo.
A careful MEDLINE online database search for all available studies on 8-isoprostane in EBC, demonstrates data in line with the results of the present study. When considering an identical assay (enzyme- or radio-immunoassay), higher 8-isoprostane concentrations were reported using a glass coating compared with the EcoScreen®. In vivo using the EcoScreen®, present 8-isoprostane concentrations (median (range) 3.5 (2.5–4.8) pg·mL-1) were comparable to the values reported by Van Hoydonck et al. 20 (4.6 (3.9–7.7) pg·mL-1). Specific data on detectability of isoprostane in EBC are scarce. The current study reports 41% of detectability in volunteers, which is in agreement with the findings (36%) of Van Hoydonck et al. 20 in smokers.
In vitro, the overall mean albumin condensate recovery percentage was 52%, compared with 100% isoprostane recovery. This suggests a more difficult condensate recovery of protein biomarkers, which is supported by the present limited number of positive albumin detections in vivo. Also, the relatively few positive studies on cytokines or proteins in EBC versus several positive reports on 8-isoprostane and leukotrienes in EBC may suggest that the recovery of proteins in EBC is less easy than that of eicosanoids. Recently, Tufvesson and Bjermer 21 also showed that the detection of eicosanoids and cytokines in EBC in asthmatic subjects may be facilitated by coating of all collection surfaces with Tween 20 and bovine serum albumin, respectively, combined with vacuum-centrifugation of condensate samples. Moreover, 30-fold differences in condensate levels of IL-4 in asthmatic children were found using different condenser systems and enzyme immunoassays 22–24. The latter finding indicates that variability of biomarker levels in condensate may also be attributed to intrinsic analytical problems of currently used immunoassays, as illustrated recently for thromboxane-A2 metabolite measurements in EBC 24.
In vitro, the mean 8-isoprostane condensate recovery percentages did not differ significantly between the condenser coatings (p = 0.09). One explanation may be the higher overall mean condensate recovery of isoprostane, in contrast to albumin, which may illustrate a different behaviour of these biomarkers during the expiratory phase of the EBC collection. In turn, this may be a reflection of difficulties experienced by molecules with a higher molecular weight, in their transition from respiratory fluid droplets and exhaled air, to condensate. Conversely, overload in the 8-isoprostane in vitro tests seems unlikely when considering that the absolute amount in the test tubes must have been very close to concentrations found in vivo. Hence, it may be that the nebuliser set-up is producing 8-isoprostane aerosols that are more effectively condensed as compared to the real EBC in vivo. Therefore, further studies on the optimal condenser coating should always include in vivo experiments.
The EcoScreen® is a commonly used and commercially available condenser system with distinctive physical characteristics. In contrast to the current authors' collection system and the RTube®, which collect EBC in the liquid phase, the EcoScreen® collects EBC mainly as ice, which may be associated with more drastic freeze–thaw cycles that can affect the molecules collected 23, 25. The EcoScreen® is a condenser system based on a modified inner Teflon coating. The 8-isoprostane EBC concentrations were similar for the EcoScreen® and the Teflon coating, suggesting no significant influence from the specific physical characteristics of the EcoScreen® condenser device in vivo.
A different design of the in vivo 8-isoprostane and albumin study was chosen, although ideally both study parts should have had a similar design. However, considerations in support of differing designs were decisive: 1) subjects with asthma are more likely to have positive albumin detection in EBC than healthy controls 16; and 2) a 90-min EBC collection period, like in the in vivo 8-isoprostane study, was neither realistic nor feasible in the asthmatic children of the in vivo albumin study. Therefore, a between-subject (parallel groups), rather than an intra-subject, comparison of condenser coatings was chosen for the in vivo albumin study.
Standardisation of the EBC collection method implicates more than applying one identical collection technique. It implicates the development of a condenser system with the highest and least varying recovery percentages. Optimisation of condensate recovery percentages may lead to an improvement of biomarker reproducibility. Therefore, the current authors recommend the use of a condenser system with an inner glass or silicone coating for the measurement of 8-isoprostane and albumin in EBC. The reproducibility of measurements by using different condenser coatings should be evaluated in vivo.
An explanation for this superiority of glass and silicone coatings may be that water acts as a bipolar vehicle. The pH of exhaled breath ranges from 6.50–7.80, whereas the isoelectrical point of human albumin is situated at pH 4.80 3, 26. In other words, at higher pH values (such as reported in EBC) albumin acts as a negatively charged molecule. Therefore, the negatively charged albumin in exhaled breath may be repulsed by glass that is also negatively charged. This may promote the recovery of albumin in EBC. In contrast, the same albumin molecule may be attracted by aluminium metal ions that are positively charged. This may hamper the recovery of albumin in EBC in this case. Furthermore, 8-isoprostane acts in an extremely hydrophobic manner and will stick to neutral surfaces (such as Teflon) in the presence of aqueous solutions, to lipids and to detergents.
Not only adhesive and electrical properties of the condensation equipment may affect whether a compound is collected in EBC, but also issues such as the length of the condensation tube, thermal conduction of the coating, solute solubility, volatility, and its dissociation constant (pKa = −log Ka).
Finally, these findings in 8-isoprostane and albumin do not necessarily hold for all other biomarkers. Every biomarker has its own chemical and physical characteristics. Currently, one may speculate that most eicosanoids act in a similar way. Likewise, small proteins such as cytokines may act differently from eicosanoids, but similarly when comparing one cytokine to another.
The present authors conclude that a condenser system with a silicone or glass coating is more efficient for the measurement of 8-isoprostane or albumin in exhaled breath condensate, compared with the EcoScreen®, or condensers with aluminum, polypropylene or Teflon coating. The increased efficiency of the borosilicate glass and silicone coatings may be due to less adhesive properties resulting in more biomarker recovery in the condensate. To establish methodological standardisation of the collection of exhaled breath condensate, the most valid and inert condenser coating has to be applied for the measurement of a specific inflammatory biomarker.
Acknowledgments
The authors wish to thank: M. Dentener and R. Cloots for their technical support in the analysis of the albumin condensate samples (Dept Respiratory Medicine, University Hospital of Maastricht); M. Hadfoune for his technical support in the analysis of the 8-isoprostane condensate samples (General Surgery, Nutrition and Toxicology Research Institute Maastricht, Maastricht University); P. Laeven for his technical support in the development of the rechargeable condenser system (Dept of Instrument Development Engineering & Evaluation, University of Maastricht); J. Sijbrandij for her statistical support (Dept of Epidemiology, University of Maastricht); and last but not least the volunteers for their participation.