Abstract
Airway inflammation in asthma is not measured routinely in clinical practice. Fractional exhaled nitric oxide (FENO), a marker of airway inflammation, is increasingly used as an outcome measure in asthma intervention studies and yet the reproducibility of FENO measurements is unknown.
The reproducibility, daytoday, diurnal variation and perception of standardised FENO measurements were examined in 59 subjects (40 children aged 7–13 yrs and 19 adults aged 18–60 yrs), both healthy (n=30) and with mild (n=29) asthma. FENO was measured on five consecutive days (four measurements on the same day) for adults and twice on the same day for children.
The coefficient of reproducibility expressed as the mean pooled standard deviation (n=59, 675 estimations) was 2.11 parts per billion (ppb) and intraclass correlation coefficient was 0.99 in both children and adults. FENO was significantly higher in asthma subjects (32.3 ppb) than in healthy subjects (16.3 ppb). There was no diurnal or daytoday variation, or a learning effect, as the result of FENO measurements were identical at results of the beginning and at the end of the study.
It was concluded that fractional exhaled nitric oxide measurements are simple, reproducible, free from diurnal and daytoday variation, and acceptable by both healthy and asthmatic adults and children, as a part of their routine visit to a physician.
This study was supported by Aerocrine AB, Stockholm, Sweden.
Asthma is an inflammatory disease, yet airway inflammation is not measured directly and routinely in clinical practice 1. This makes management of asthma difficult, because it is based on indirect measurements of airway inflammation, such as symptoms and lung function. Symptoms may not accurately reflect the extent of underlying inflammation due to differences in perception, and lung function tests may have little room for improvement in mild asthma. None of these parameters is able to distinguish the effect of different doses of inhaled corticosteroids and both may be affected by bronchodilators. The latter is particularly important because of a recent trend towards use of lower doses of inhaled corticosteroids in combination with longacting β2‐agonists.
Current invasive (bronchoscopy), or semiinvasive (sputum induction) direct methods to measure airway inflammation are difficult to use repeatedly in clinical practice. The use of sputum induction is limited by its proinflammatory effect 2, and a considerable bronchospasm has been reported during sputum induction in moderate (14%) and severe (25%) asthma, as a result of the procedure 3.
There has been an explosion of research into exhaled nitric oxide (NO) since levels were found to be increased in asthma 1, 4. Standardised 5, 6 measurements of fractional exhaled NO (FENO) provide a completely noninvasive means of monitoring airway inflammation and antiinflammatory treatment in asthma 1, including a dosedependent onset and duration of action of inhaled corticosteroids 7. It may be useful in patients using fixed combination inhalers (corticosteroids and longacting β2‐agonist) to ensure that inflammation is controlled, as this may be difficult to assess from symptoms when a longacting bronchodilator is taken.
The changes in serial FENO, as a lossofcontrolmarker 8, have higher predictive values for diagnosing the deterioration of asthma 9 than does single measurements 10, 11.
It can be argued, however, that these changes in FENO may be due to measurement error and/or the natural variability of airway inflammation over time. Therefore, design and interpretation of clinical studies in asthma and the use of FENO in routine clinical practice depends greatly on the reproducibility and the safety of FENO measurements. Reproducibility of FENO measurements was studied by several groups, but either the statistical analysis was inappropriate (correlation coefficient) 12, or exhalation flow rate was either not registered 13, or was different from the American Thoracic Society (ATS) recommendations 14.
Because of uncertainty regarding the reproducibility and the safety of standardised FENO measurements made in accordance with the ATS recommendations 5, these issues were examined in asthmatic and healthy subjects, including both adults and children.
Methods
Subjects and protocol
Sixty nonsmoking subjects (26 males and 34 females, with an adult mean±sd age of 35.6±9.39 yrs and children 10.7±1.79 yrs) were recruited for this prospective, open, singlecentre study, performed in accordance with the guidelines of Good Clinical Practice. The healthy subjects (10 adults and 20 children) had neither atopy nor history of asthma or other diseases, and normal lung function. All patients were diagnosed with atopic mild steroidnaive asthma (10 adults and 20 children). The asthmatic subjects had no history of asthma exacerbation or upper respiratory infection in the preceding 4 weeks, and their lung function was normal (forced expiratory volume in one second (FEV1 101±3% pred and 98±4% pred, for adults and children, respectively). Subjects were not allowed to eat any food within 1 h prior to the exhalation session, but could drink water (water intake was recorded). There was no change in asthma medication during the study.
Procedures on the first visit included information about the study and obtaining informed consent from adults and the parents of children, medical and asthma history, physical examination, allergen skin testing, baseline spirometry and practice of exhalation procedure with the NO analyser for measuring exhaled NO.
Children underwent two FENO measurements consisting of sessions of three measurements in each session, the first between 09:00–10:00 h and the second between 14:00–15:00 h with a standard lunch between 12:00–13:00 h. Adults had their FENO measured once on five consecutive days between 09:00–10:00 h during visits 1, 2, 3, and 4, and during the last visit (fifth) FENO was measured four times, between 09:00–10:00 h, 11:00–12:00 h, 14:00–15:00 h, and 16:00–17:00 h with a standard lunch between 12:00–13:00 h. Children and/or their parents and adults were asked to report any unpleasant events associated with the measurement sessions, and questions regarding their subjective experience of the FENO measurements were asked at the end of the study. The study was approved by the Ethics Committee of the Royal Brompton and Harefield NHS Trust.
Measurements of fractional exhaled nitric oxide
FENO was measured according to the ATS recommendations 5 using the NIOX® system (Aerocrine AB, Stockholm, Sweden). Any exhalation, which did not meet the ATS requirements, was not accepted by the NIOX® system and the subjects were asked to perform a new exhalation manoeuvre. At each session three correctly executed exhalations were recorded.
Statistical methods
Reproducibility of FENO measurements was assessed in three different ways: 1) by the Bland and Altman analysis, which is used when the measurements are on the same scale; 2) by means of the intraclass correlation coefficient (ICC), a dimensionless statistic that describes the reproducibility of repeated measurements in the same population; 3) by the pooled sd, as the final measure of repeatability for all studied healthy and asthmatic subjects. BlandAltman analysis was also applied to study diurnal variation in adults using the first and last measurement during day 5. In children the morning and the afternoon sessions were compared. The “learning effect” was studied in adults by comparing the morning sessions on day 1 and day 5. Daytoday variability, influence of asthma, sex, height, weight, age and food intake on the FENO value, and the safety of FENO measurements performed under conditions, which were close to a routine clinical setting, were examined.
The value of ICC is bounded by 0 and 1, and in a stable population ICC values in excess of 0.6 are thought to be clinically significant, and those <0.6 are probably not 15. The authors decided to choose an ICC value of >0.75 3 to represent a clinically useful method, as this cutoff level was used recently in a reproducibility study of another approach to monitor asthma, i.e. sputum induction 3.
Comparisons of FENO in healthy and asthmatic children and adults were performed using mean values of FENO measurements from all sessions. Descriptive parametric statistical methods were used to describe and summarise the results from the study. Mean values were presented as mean±sd. Betweengroup significance tests were performed using twosample t‐tests with due correction according to Satterthwaite 16, for degrees of freedom when variances were inhomogeneous according to the Levene 17 test, and within group changes were analysed using paired samples t‐test. Mean individual FENO values were correlated to sex, age, height, weight, and food intake using Pearson correlation analyses. The KolmogorovSmirnov test of normality of data was applied. Sample size calculations were made by using the following formulae.
Onesample (one treatment) trials:Twosample (treatment versus placebo) trials:
where δ2 is effect size expressed as relative change (%), σ2 is sd of relative change (%), and t is the t‐test.
Statistics were presented using both complete data or sometimes with outliers excluded. An outlier was defined as a value outside ±sd from the mean. FENO values of one of the adult asthmatic patients (1.13) were not included in the final analysis because the levels were well above 2sd (110–160 parts per billion (ppb)) of the studied population. This patient with mild stable atopic asthma and characteristically high FENO levels, which may become even higher during asthma exacerbations, has been seen in the authors' clinic for several years. Other research groups 18 have also reported patients with very high FENO values between 150–300 ppb, which suggests that there is a small number of patients with difficult to explain high FENO values.
Results
Patient enrolment and effect of demographic and other factors on fractional exhaled nitric oxide
The majority (89%) of adults and 78% children completed the study between September–December 2000, hence outside the grass pollen season. In all, 675 exhalations (of 696 theoretically accessible exhalations) made during 229 measurements (of 232 theoretically accessible sessions) were performed and analysed. At three sessions (21 exhalations) the NIOX® device was out of order.
Patients with asthma (adults and children) had a higher FENO (32.3±25.9 ppb) than healthy subjects (16.3±8.4 ppb; p<0.005). When analysed separately, adults with asthma had higher FENO (48.8±27.2 ppb) than healthy adults (17.8±6.8 ppb; p<0.010). Children with asthma had a borderline significantly higher FENO (24.9±22.3 ppb) than healthy children (15.6±9.2 ppb; p<0.096).
Healthy adult females (n=7) had significantly lower FENO levels (14.85±3.34 ppb) than healthy males (n=3, 24.6±8.59 ppb; p<0.03). No correlations were found between FENO and duration of asthma, season, food intake, body weight or height. There was no significant difference in FENO between the subjects who consumed water before the measurements (41.24±57.40 ppb) and subjects who did not consume water (37.83±49.05 ppb, p>0.05).
Repeatability and reproducibility
Data was approximately normally distributed (KolmogorovSmirnov test, p>0.20). The mean pooled sd of all measurements was 2.1±1.25 ppb.
FENO measurements made at different visits and time points were highly reproducible throughout the study (table 1⇓), and there was no significant diurnal or daytoday variation (fig. 1⇓) of FENO in adults and no diurnal variation in children. BlandAltman analysis of distance from session means using all 675 FENO measurements yielded BlandAltman “limits of agreement” of ±4.01 ppb (i.e. 2×sd). After removal of outliers (i.e. 38 measurements >2sd from session mean, 2×2.01=4.01 ppb) the repeatability of the FENO measurements as measured by the BlandAltman limits of agreement was ±2.92 ppb (fig. 2⇓).
Means±sem for fractional exhaled nitric oxide (FENO) in healthy (○) and asthmatic (•) adult subjects during successive sessions. ppb: parts per billion.
BlandAltman analysis for the repeatability of fractional exhaled nitric oxide (FENO) values (n=637 measurements). ppb: parts per billion.
Reproducibility of fractional exhaled nitric oxide measurements
In order to assess the possibility of reducing the number of measurements from three to two per session, the precision of measurements using one and two measurements per session was analysed. The FENO value from each one of the three measurements in a session was compared with the mean FENO value from all the three measurements. The difference in the worst case was 0.28±5.8 ppb (95% confidence interval (CI)). When the mean FENO from two measurements was compared with the mean FENO value from all three measurements, the worst case difference was −0.26±2.92 ppb (95% CI) (table 2⇓). All ICC values found were above the chosen accepted limit of 0.75. In fact all ICC values were ≥0.90 (table 3⇓).
Mean distance from session mean of fractional exhaled nitric oxide (FENO) measured during one or two exhalations in studied population groups
Statistical comparison of the reproducibility of fractional exhaled nitric oxide (FENO) measurements between first and second versus first and fourth and morning versus afternoon visits
There was no “learning effect”, as there was no statistically significant difference between the FENO measurements made at the beginning and at the end of the study. Morning measurements during the first and last day did not differ significantly (p=0.138). The mean difference was 2.58 ppb. The upper limit of agreement was 16.64 and the lower limit −11.48 ppb in a BlandAltman analysis. Sample size calculations based on the data from adult asthmatic patients for one and two clinical samples are presented in figure 3⇓.
The sample size estimates are based on the variability in the outcomes over the two samples (treatment versus placebo (•)) and one sample (one treatment (▪)) study.
Patients perception of fractional exhaled nitric oxide measurements with NIOX®
The use of the NIOX® system was safe throughout the study, and there were no adverse incidents recorded. The majority of the subjects found FENO measurements easy to perform, simple and potentially acceptable for use in routine clinical practice (table 4⇓).
Subjective experience of fractional exhaled nitric oxide measurements with NIOX®
Discussion
It has been demonstrated that standardised FENO measurements performed with a NIOX® system are simple, highly reproducible (pooled sd=2.11 ppb and ICC=0.99), free from diurnal and daytoday variation and acceptable by healthy and asthmatic adults and children as a part of their routine visit to a physician.
Airway inflammation in asthma has been shown to exhibit considerable biological variability, yet it is not measured directly and routinely in clinical practice. This makes asthma monitoring and its antiinflammatory treatment difficult. Several approaches are currently used to measure airway inflammation in asthma. They are, however, either invasive or semiinvasive (bronchoscopy and sputum induction), or less direct (provocative concentration causing a 20% fall in the FEV1 (PC20) and lung function), or may be affected by the patient's perception or bronchodilators (asthma symptoms). The reproducibility of these approaches is also variable.
Bronchial hyperreactivity test (PC20) is noninvasive and reasonably reproducible (ICC for PC20 methacholine 0.74 4 and 0.94 19).
Bronchial biopsy, sputum induction and bronchial hyperreactivity tests may only be performed in specialised centres, are costly and time consuming, and cannot be used in patients with severe asthma or in small children. It is unlikely that any of these methods of airway inflammation assessment will be used in routine clinical practice.
FEV1 is the most reproducible indicator of airway inflammation in asthma. ICC values between 0.87 20 and 0.93 4 have been shown for the measurements made in specialised lung function (LF) laboratories, and coefficient of variation of 3% for FEV1 and forced vital capacity can be achieved by trained research workers 21. Conversely, repeated FEV1 measurements made in patients at the primary care, by either trained or untrained physicians, had significantly lower sensitivity (41% of the LF laboratory values as the standard) and a higher falsenegative rate (59%) 22. The reproducibility of peak expiratory flow (PEF) may also be very low, and there are reports that some patients may manipulate the PEF meter in a way that could repeatedly produce clinically significant changes in readings 23. Both, home recording of PEF and symptom diaries are easily faked and are generally unreliable in both adults 24 and children 25.
In addition, symptoms may not accurately reflect the extent of underlying inflammation due to differences in perception. The fact that lung function and symptoms may be affected by bronchodilators is particularly important because of a recent trend towards use of the lower doses of inhaled corticosteroids in combination with longacting β2‐agonists (LABA).
Reproducibility of the standardised FENO measurements according to the ATS standards within a single day in both adults (ICC 0.94) and children (ICC 0.94) is superior to any conventional methods of airway inflammation monitoring in asthma. This adds significantly to other major advantages of FENO measurements, such as their strong association with airway inflammation 2, even in nonsymptomatic asthma patients 26, their high sensitivity to steroid treatment 27, insensitivity to β2‐agonists 2 and noninvasiveness.
Repeated FENO measurements, therefore, can be used much more frequently and will not disturb the body system, in contrast to the invasive or semiinvasive procedures currently used in clinical research to monitor inflammation status 2. The advantage of repetitive FENO measurements has been studied and an increase in FENO and asthma symptoms was seen before any significant deterioration in airway hyperresponsiveness, sputum eosinophils, or lung function during asthma exacerbation induced by steroid reduction 10, 11. These data suggest that FENO may be used as a lossofcontrolmarker in asthma 8. Recently, it was confirmed that changes in FENO measured over time have higher predictive values, sensitivities, and specificities both for predicting and diagnosing loss of control than did single measurements, which clearly indicates the need for repeated tests 9. When measured longitudinally the changes in FENO correlated significantly not only with changes in sputum eosinophils and hyperresponsiveness, but also with lung function and asthma symptoms 9.
There are several important practical implications for the FENO measurements regarding the data comparison of spirometry versus FENO examination. Firstly, it has been shown that high reproducibility of FENO measurements in both children and adults may allow the medical practitioner to perform two instead of three exhalations to obtain reliable results (table 2⇑). This may be of great advantage, as it will shorten the time needed for the measurement procedure.
Secondly, FENO measurements with recently developed NO analysers are fully automated and therefore incorrect exhalation manoeuvres by a patient (<10 s or outside the limits of the exhaled flow) will not be accepted by the analyser. In addition to the automation, staff training will be minimal. Finally, the advantage of FENO measurements is that patients do not require extra encouragement, as may be the case with PEF measurements. Indeed, a significant difference may be seen between unobserved data and encouraged PEF readings 28.
BlandAltman analysis of FENO measurements between the sessions separated by either hours or days allowed for the investigation of whether there was a “learning effect” (implying that the subsequent, but not the first FENO measurements are the most reliable). The authors did not find any such “learning effect” or systematic error of serial FENO measurements. The simplicity and high reproducibility of FENO measurements in the present study are probably the major reasons for this. In this study, a single nurse made all the FENO measurements after a short tutorial was given. The simplicity of the standardised FENO measurements was further enhanced by the design of the NO analyser used, which controlled the exhalation parameters and ensured that the measurements were not accepted unless they were performed according to the guidelines 5.
Recently, much greater variability of FENO measurements was shown in 107 healthy children using a different NO analyser when some of the ATS requirements were not fulfilled, for example exhalation flow rate of 184 mL·s−1 was used instead of 50 mL·s−1 (ATS standard), nasal clips used (shall not be used according to the ATS standard) and NO free air was not considered 29. This interesting study demonstrated that although there was no diurnal variation of exhaled NO levels in the studied population, the intraindividual coefficient of variation was rather high (25.9%) and the intrameasurement coefficient of variation in a selected group of children was 6.5 (4.2–8.6)%.
The mean pooled sd of all measurements was 2.1±1.25 ppb. These results suggest that if a patient's exhaled FENO level changes >4 ppb between sessions, it is more likely due to the inflammatory process rather than inaccuracy of the NO analyser. This finding is valuable for the potential use of FENO in routine clinical practice. Shortterm monitoring, when the measurements of airway inflammation are made frequently, for example every day or twice a day as in the case of PEF, is particularly important. This is because of a recent trend towards use of lower doses of inhaled corticosteroids in combination with LABA when antiinflammatory and clinical effect of combination treatment may be seen within hours and days.
Dose adjustment in both clinical practice and clinical research is an important issue, in which high reproducibility of FENO measurements and sensitivity of FENO to corticosteroids may substantially reduce the cost of medical care and research. Recently, the authors demonstrated a dosedependent onset and cessation of antiinflammatory action of inhaled corticosteroids on FENO and asthma symptoms in a small number (n=28) of mild asthma patients who were treated with 100 or 400 µg budesonide, or placebo once daily for 3 weeks in a doubleblind, placebocontrolled, parallel group study 7.
Samplesize determination is often an important step in planning such studies. According to the present data (fig. 3⇑), a small number (between 7–20) of asthmatic subjects, either adults or children, will be required to demonstrate 25–80% effect of a studied drug in a clinical trial. Based on the knowledge of the individual variability of FENO measurements, like individual PEFs, individual FENO values should be established and monitored, and when the levels are above or below a certain reference level, steroid treatment should be either reduced or increased 30.
The data suggest that standardised exhaled nitric oxide measurements may provide a useful clinical tool to monitor airway inflammation in patients with asthma, and acceptable by both healthy and asthmatic adults and children, as a part of their routine visit to a physician. Longterm prospective clinical studies are now necessary to prove that exhaled nitric oxide can be used to optimise dosing of antiinflammatory treatment.