Skip to main content

Main menu

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • For authors
    • Instructions for authors
    • Submit a manuscript
    • Author FAQs
    • Open access
    • COVID-19 submission information
  • Alerts
  • Podcasts
  • Subscriptions
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

User menu

  • Log in
  • Subscribe
  • Contact Us
  • My Cart
  • Log out

Search

  • Advanced search
  • ERS Publications
    • European Respiratory Journal
    • ERJ Open Research
    • European Respiratory Review
    • Breathe
    • ERS Books
    • ERS publications home

Login

European Respiratory Society

Advanced Search

  • Home
  • Current issue
  • ERJ Early View
  • Past issues
  • For authors
    • Instructions for authors
    • Submit a manuscript
    • Author FAQs
    • Open access
    • COVID-19 submission information
  • Alerts
  • Podcasts
  • Subscriptions

Components of respiratory resistance monitored in mechanically ventilated patients

B. Babik, F. Peták, T. Asztalos, Z.I. Deák, G. Bogáts, Z. Hantos
European Respiratory Journal 2002 20: 1538-1544; DOI: 10.1183/09031936.02.00000802
B. Babik
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
F. Peták
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
T. Asztalos
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Z.I. Deák
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
G. Bogáts
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Z. Hantos
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

The interrupter technique is commonly adopted to monitor respiratory resistance (Rrs,int) during mechanical ventilation; however, Rrs,int is often interpreted as an index of airway resistance (Raw).

This study compared the values of Rrs,int provided by a Siemens 940 Lung Mechanics Monitor with total respiratory impedance (Zrs) parameters in 39 patients with normal spirometric parameters, who were undergoing elective coronary bypass surgery. Zrs was determined at the airway opening with pseudorandom oscillations of 0.2–6 Hz at end inspiration. Raw and tissue resistance (Rti) were derived from the Zrs data by model fitting; Rti and total resistance (Rrs,osc=Raw+Rti) were calculated at the actual respirator frequencies. Lower airway resistance (Rawl) was estimated by measuring tracheal pressure.

Although good agreement was obtained between Rrs,osc and Rrs,int, with a ratio of 1.07±0.19 (mean±sd), they correlated poorly (r2=0.36). Rti and the equipment component of Raw accounted for most of Rrs,osc (39.8±11.9 and 43.0±6.9%, respectively), whereas only a small portion belonged to Rawl (17.2±6.3%).

It is concluded that respiratory resistance may become very insensitive to changes in lower airway resistance and therefore, inappropriate for following alterations in airway tone during mechanical ventilation, especially in patients with relatively normal respiratory mechanics, where the tissue and equipment resistances represent the vast majority of the total resistance.

  • airway resistance
  • forced oscillations
  • interrupter technique
  • respiratory monitoring
  • tissue resistance

This study was supported by Ministry of Health grant No. 591/1996 06.

The accurate bedside monitoring of anaesthetised or critically ill patients involves a careful assessment of respiratory mechanics. Most of the recent commercially available ventilators are able to display the flow, pressure and volume changes throughout the respiratory cycles. From these signals, respiratory compliance (Crs) and resistance (Rrs) are also calculated on-line.

Crs is a combined index characterising the elastic properties of the lung parenchyma and the chest wall. Many diseases, including emphysema and interstitial oedema 1, alter the elasticity of the lung parenchyma directly. Furthermore, Crs is primarily affected by a loss in functional lung volume resulting from an increased closing capacity 2 leading to air trapping, or by abdominal distension 3.

Rrs is often considered to be a marker of airway resistance (Raw) 4–10 and is therefore, used to follow the respiratory strategy 8 and the effects of bronchoactive agents 5–7, 9, 10. Nevertheless, it has been well established that a significant proportion of Rrs can be attributed to energy loss in the respiratory tissues (tissue resistance (Rti)) 11–21. Since the parameter Rrs is complex in nature, alterations in it may reflect pathological changes at different sites in the lungs. An increased Rrs may reflect an elevated Raw resulting from chronic obstructive pulmonary disease (COPD), an acute bronchospasm or mucus deposit in the airways. Alternatively, increases in Rti in diseases affecting the pulmonary parenchyma, such as adult respiratory distress syndrome 4, 5, 22–24, COPD 20, cardiac failure 1 or congenital heart disease 25, 26, also elevate Rrs. Additionally, Rrs is often measured in such a way that it includes the resistance of the endotracheal tube (ETT), the connectors, the filters and the connecting tubing of the respirator 7–9, 22, i.e. the equipment resistance (Req).

Although it is evident that Raw, Rti and Req are all included in Rrs, their relative contributions have not been quantified. Accordingly, this study was performed in order to determine the airway, respiratory tissue and respiratory equipment components of Rrs displayed on a commonly used bedside ventilatory monitor. Since exact monitoring of the Raw is crucial in following the response to bronchodilator therapy, the authors also aimed to verify whether Rrs calculated by the monitor during tidal ventilation adequately indicates changes in bronchial tone.

Methods

Subjects and preparations

Thirty-nine patients (23 males, 16 female, 62±9 (mean±sd) yrs of age (range 40–78 yrs)) undergoing elective coronary bypass surgery were studied in the supine position before the surgical procedure. The protocol was approved by the Human Research Ethics Committee of Szeged University (Szeged, Hungary), and the patients gave their informed consent to the study. None of the patients had a history of COPD, and they did not have any respiratory problems at the time of the study. Indices of forced expiration were established 1 day before the surgery with a portable spirometer (Model PDD-301/s; Piston Ltd, Budapest, Hungary); the mean values of the predicted forced vital capacity and forced expiratory volume in one second of the group were 100±20 and 93±19%, respectively. The patients were premedicated with i.m. morphine (0.1 mg·kg−1), midazolam (0.07 mg·kg−1) and atropine (0.5 mg) 1 h before the operation. Anaesthesia was induced with i.v. fentanyl (5–15 µg·kg−1) and midazolam (0.05–0.07 mg·kg−1). Muscle paralysis was achieved with an i.v. bolus of pipecuronium bromide (0.1 mg·kg−1). The trachea was intubated with a cuffed ETT 7, 8 or 9 mm in internal diameter (ID), and the patients were ventilated with a Siemens 900C Servo Ventilator (Solna, Sweden) in the constant flow mode by adjusting the inspiratory/expiratory ratio to ∼1:3. The ventilator frequency was set to 12–17 min−1, and a tidal volume (VT) of 8–12 mL·kg−1 and a positive end-expiratory pressure (PEEP) of 2 cmH2O was applied. Maintenance doses of the anaesthetics and muscle relaxants were administered as needed.

Oscillatory measurements

The set-up used for the impedance measurements during short intervals of suspended mechanical ventilation is shown schematically in figure 1⇓. Two collapsible latex tube segments (A and B) were clamped alternately to switch the ETT from the respirator to the oscillatory device and back as follows. During mechanical ventilation, A was open and B closed. Before the first oscillatory measurement, the lungs were inflated to a pressure of ∼30 cmH2O to standardise the volume history. Segment B was then opened for a few ventilatory cycles before data acquisition, in order to equilibrate the pressures in the lungs and the loudspeaker chambers. The respirator was stopped during the end-inspiratory plateau and segment A was clamped. In the resulting apnoeic period, small-amplitude (<2 cmH2O peak-to-peak) pseudorandom pressure excitations were introduced into the trachea. The forcing signal contained 30 integer-multiple components of the 0.2-Hz fundamental frequency, in the frequency range 0.2–6 Hz. The component amplitudes were progressively enhanced with decreasing frequency so that a sufficient signal-to-noise ratio was provided over the whole frequency range. In order to decrease the involvement of nonlinear phenomena, the phases of the components were optimised so as to give the minimum pressure excursion at the airway opening 27.

Fig. 1.—
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 1.—

Schematics of the measurement set-up. V′: central airflow; Pao: airway opening pressure; Ptr: tracheal pressure; A and B: collapsible segments; PTG: pneumotachograph; F: antibacterial filter.

Tracheal flow (V′) was measured with a 28-mm ID screen pneumotachograph connected to a differential pressure transducer (ICS model 33NA002D; ICSensors, Milpitas, CA, USA). Airway opening pressure (Pao) was detected with an identical pressure transducer. To estimate the contribution of the impedance of the ETT and the antibacterial filter to the measurements, the tracheal pressure (Ptr) was measured with a third ICS sensor through a 1.5-mm OD polyethylene catheter, the tip of which was furnished with several lateral holes and which was positioned 1.5–2 cm beyond the distal end of the ETT.

The recordings of oscillatory flow and pressures lasted for 15 s. Four to five measurements were recorded, and >2-min intervals of ventilation were interposed between successive measurements. The input respiratory impedances (Zrs) of the total and the lower respiratory system (Zrsl) were computed from the power spectra of Pao and V′ and Ptr and V′, respectively. The Zrs and Zrsl data were then ensemble-averaged separately and the data points at frequencies close to the cardiac frequency or its harmonics, and hence exhibiting large scatter, were discarded in both. The mean Zrs data were fitted by a four-parameter model 13 containing a frequency-independent Raw and inertance (Iaw) and a constant-phase tissue compartment characterised by the coefficients of damping (G) and elastance (H): Embedded Image where f is frequency, j is the imaginary unit and α=(2/B)tangent−1(H/G). The Rti and the total respiratory resistance (Rrs,osc) at the frequency of mechanical ventilation (fB) were respectively calculated as: Embedded Image and Embedded Image The Zrsl data were evaluated analogously to determine lower airway resistance (Rawl), lower airway inertance (Iawl), etc. pertinent to the lower respiratory system. Req was calculated as: Embedded ImageThe effective oscillatory compliance of the respiratory system at fB was also computed, as: Embedded Image

Conventional mechanics

The mechanical parameters were monitored by a Siemens Servo Computer Module (Type 990) attached to the respirator. The Rrs,int and Crs,int values were read before and after the collection of impedance data, and their respective mean values were applied. According to the manual of the module, Rrs,int was calculated as the ratio of the pressure drop (dP) due to the interruption of inspiratory flow to the flow before interruption; dP was determined according to: Embedded Image where Pmax is the pre-interruption pressure and Pplateau was established at the end of the interruption interval (10% of the ventilation period time). Crs,int was calculated as follows: Embedded Image

Statistical methods

Scatters in the parameters were expressed as sd values. Paired t-tests and linear regression analysis were used to study the relationships between oscillatory and ventilatory mechanical parameters. Bland-Altman analysis 28 was performed to determine the agreement between the respiratory resistances measured by the ventilatory monitor (Rrs,int) and by the forced oscillations (Rrs,osc). A p-value <0.05 was considered to be significant.

Results

Figure 2⇓ displays typical Zrs and Zrsl spectra obtained from a representative patient, with the best-fitting model curves, in terms of the real (Rrs) and imaginary part (reactance (Xrs)). The successive measurements yielded reproducible data at all frequencies except those coinciding with the heart frequency and its harmonics. Overall, the model was consistent with the measured data; only the slight increase in the measured Rrs>∼4–5 Hz, presumably a consequence of the frequency-dependent inertial distortion of the velocity profile and/or shunt effects in the central airways 29, led to systematic fitting errors. The decreases in both Rrs and Rrsl with frequency are attributed to the common tissue component, whereas the asymptotic levels of Rrs and Rrsl correspond to the airway components; the difference between them reflects the largely frequency-independent contribution of the equipment resistance. The Xrs and lower airway reactance (Xrsl) curves depart from each other with increasing frequency: the steeper increase in Xrs resulting in a resonant frequency at ∼3 Hz is a consequence of the substantial inertance of the equipment components.

Fig. 2.—
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 2.—

An example of respiratory input impedance data in one subject. Mean values from five successive measurements of a) total respiratory resistance (Rrs,osc: •) and lower respiratory resistance (Rrsl,osc: ▪) and b) total respiratory reactance (Xrs,osc: •), and lower respiratory reactance (Xrsl,osc: ▪) with the best-fitting model curves (continuous and dashed lines for the total and lower impedances, respectively). fB: frequency of ventilation. For clarity, the sd values are indicated for total impedance only. Open symbols indicate values corrupted by cardiac noise, exhibiting higher sd values and omitted from the model fitting. The difference between Rrs,osc(fB) and Rrsl,osc(fB) corresponds to the equipment resistance.

The relationships between the interrupter and oscillatory resistance values are illustrated in figures 3⇓ and 4. The mean values of Rrs,int and Rrs,osc(fB) were very similar (10.1±1.9 and 10.6±1.9 cmH2O·L−1·s, respectively). Although the correlation between Rrs,int and Rrs,osc(fB) was statistically significant (p<0.005), the correlation coefficient was rather low, and Bland-Altman analysis revealed a lack of agreement between the two parameters, with limits of agreement of 3.1 cmH2O·L−1·s and 4.1 cmH2O·L−1·s (fig. 4⇓). The values of Raw (5.97±1.19 cmH2O·L−1·s) and Rawl (1.76±0.81 cmH2O·L−1·s) were much lower than those of Rrs,osc(fB) and Rrs,int. Additionally, the correlations between the forced oscillatory airway resistances and Rrs,int were weak, although statistically significant (p<0.002 for both). Rti and Req accounted for most of Rrs,osc (39.8±11.9% and 43.0±6.9%, respectively), whereas only a much smaller portion of the total resistance could be attributed to the lower airways (17.2±6.9%). There was a stronger correlation (p<0.001) between the compliance estimates (fig. 5⇓), with Crs,int values being significantly higher (p<0.0001) than Crs,osc(fB) (a ratio of 1.4±0.19).

Fig. 3.—
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 3.—

Relationship of interrupter resistance (Rrs,int) to lower airway resistance (Rawl=−0.43+0.22 Rrs,int; r2=0.263), b) total airway resistance (Raw=3.0+0.30 Rrs,int; r2=0.240), and c) total oscillatory resistance at ventilation frequency (Rrs,osc(fB)=4.1+0.6 Rrs,int; r2=0.365). Solid line: linear regression; dashed line: line of identity.

Fig. 4.—
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 4.—

The differences between interrupter resistance (Rrs,int) and total oscillatory resistance at ventilation frequency Rrs,osc(fB) plotted against the mean of Rrs,int and Rrs,osc(fB) (Bland-Altman plot). The mean (solid line) ±2 sd levels (dashed lines: mean+2 sd=3.09; mean-2 sd=−4.11) indicate the limits of agreement.

Fig. 5.—
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig. 5.—

Relationship of dynamic respiratory compliance from tidal ventilation (Crs,int) to forced oscillatory compliance (Crs,osc) calculated at ventilation frequency (fB). Solid line: linear regression; dashed line: line of identity. r2: 0.803; Crs,int=1.14+1.40 Crs,osc.

Discussion

Previous investigations reporting total respiratory or pulmonary resistance often associated these parameters with the airways only 4–10. Accordingly, the present study was designed to quantify the components of total respiratory resistance measured by a common bedside monitor. The results demonstrate that, during tidal ventilation of an intubated patient, a very high proportion of the dissipation in the respiratory system originates from the respiratory tissues and the flow resistance of the equipment. Rawl, the resistance component that can change because of an alteration in airway tone, comprised only a small fraction of the monitored resistance.

The low-frequency forced oscillation technique has been applied to partition total respiratory impedance into airway and tissue components in intubated patients free of cardiorespiratory disease 11, in normal subjects 15, COPD patients 20 and asthmatic patients receiving the optimal ventilatory waveform 21, and in healthy sedated infants 17, 18. The interrupter technique has also been demonstrated to provide separate estimates for both airway resistance and tissue viscoelasticity 23, 24, 30–32. If the flow in the central airway is interrupted rapidly, the tracheal pressure responds with a transient of two distinct phases, a sudden drop and a subsequent quasi-exponential decay. The initial change in pressure results from the cessation of flow in the Newtonian resistances, whereas the slow decay corresponds to the combined process of stress relaxation of the respiratory tissues and the interregional equilibration of inhomogeneous alveolar pressures. The interruption method is therefore a potential tool with which to separate the mechanical properties of the airways and the tissues in a similar manner as offered by the low-frequency oscillation technique 12, 20. Identification of these two phases, however, requires a fast interruption and a good-quality pressure recording for further analysis, both made as close to the airway opening as possible 30–32. Since the long closure time and the proximal placement of the shutter in the respirator available for this study precluded the identification of the two phases, the current authors applied low-frequency oscillations for the separation of the airway and tissue components, and the intratracheal pressure recording to identify the instrumental component.

Although the Rrs,osc and Rrs,int values were obtained at the same end-inspiratory position, there are essential differences between these methods, and the relevance of the oscillatory ratio Raw/Rti in Rrs,int is therefore not straightforward. Rti was overestimated, with the oscillatory signal being smaller than the tidal deformation because of the negative amplitude dependence of tissue resistance 16. However, Raw was underestimated because of the use of oscillatory flows smaller than the inflation flow rate, which may have exceeded the limits of linearity (this is also true for Req, and especially its ETT component). These two opposing factors possibly cancelled each other out in the total resistance, which may explain the good agreement between the mean values of <1?show=[to]>Rrs,int and Rrs,osc(fB), but they must have modified the proportions of the tissue and airway components in Rrs,int from those predicted from the oscillatory data. Nevertheless, it is unlikely that these nonlinear effects could have led to a substantially higher airway/tissue resistance ratio in Rrs,int than in Rrs,osc(fB). It should also be noted that Rrs,int was calculated from a relatively short (0.4–0.5 s) end-inspiratory pause, which most likely resulted in overestimation of the values of Pplateau and hence Rrs,int that would have been obtained by more accurate interruption techniques with longer (3–5 s) end-inspiratory pauses 32.

Although the frequency-independent resistive component of Zrs is fundamentally determined by the pulmonary system 14, 19, the chest wall has also been shown to have a frequency-independent resistive component (Rw) with a relative contribution reported to be 10–37% of the total Newtonian resistance 14, 19. Since Zrs spectra were used for the estimation of the resistive parameters of the airways in the present study, both Raw and Rawl included Rw. Hence, the contribution of the intrathoracic airways to Rrs,osc and thus to Rrs,int is likely to be even smaller than the modest figure of 17.2%.

Whereas the mean values of Rrs,int and Rrs,osc(fB) were very similar (Rrs,osc/Rrs,int=1.07±0.19) and their correlation was significant, Bland-Altman analysis revealed unacceptably wide limits of agreement between these parameters. Since the difference between the total resistance values obtained by the two methods did not display any systematic tendency (fig. 4⇑), a random error was most probably responsible for this substantial discrepancy. Although the reasons for the inconsistency are not completely clear, temporal changes in the respiratory mechanics between the Rrs,int and Rrs,osc recordings, the poor resolution of the monitor displaying Rrs,int without showing decimals, or the noisy pressure signal limiting the accuracy of the estimation of the pressure drop following the occlusion might all have played a role.

The major finding of the present study is the very small contribution of the intrathoracic airways to the total resistance value displayed by the respiratory monitor. To estimate the specificity and sensitivity of Rrs,int in the detection of changes in airway calibre, the maximum possible change in Rawl that may remain hidden in the Rrs,int value monitored with poor resolution, as in the present case, was calculated. For a typical value of Rrs,int of 10 cmH2O·L−1s, this round figure may represent any value between 9.51 and 10.50 cmH2O·L−1s (a range of 0.99 cmH2O·L−1s). By assuming the mean contribution of Rawl to Rrs,int to be the same as that to Rrs,osc (i.e. 17% of 10 cmH2O·L−1s), increases in Rawl not exceeding 0.99 cmH2O·L−1s may reach up to 67% and still remain undetectable in the monitored Rrs,int. It can therefore, be concluded that Rrs,int may become a rather insensitive measure of the airway properties, especially when the studied population, as in the present investigation, is free of airway diseases.

It is noteworthy that estimations of respiratory compliance were not subject to the problems associated with the measurement of resistance. A good correlation was found between the total respiratory compliances measured by forced oscillation and the respiratory monitor, though the latter were systematically greater (fig. 5⇑). This difference can be explained by the principles of the two measurement methods. Crs,osc was estimated from the oscillations applied around the end-inspiratory plateau and characterises the incremental elastic properties at that level. By contrast, Crs,int represents an average compliance for the tidal excursion between end-expiration and end-inspiration at lower mean airway pressures (Paw,mean). Both the lower Paw,mean 18 and the larger excursions 14, 16 result in higher Crs,int than Crs,osc, whereas, as noted above in connection with Rrs,int, the short end-inspiratory pause has an opposite effect on Crs,int. Since the contributions of these factors depend on the subject and the respirator settings, they may explain both the systematic difference and the random scatter in the relationship of the compliance values.

To conclude, the results of the present study demonstrate that the total respiratory resistance displayed by a standard bedside respiratory monitor in intubated patients with normal respiratory mechanics is primarily determined by the viscous resistance of the respiratory tissues and the flow resistance of the devices of the ventilatory circuit, the contribution of the airway resistance being relatively small. The present data imply that the poor accuracy of the bedside monitor may even result in the changes following a severe increase (by more than two-thirds) in the lower airway resistance remaining undetected. In the event of heterogeneous narrowing of peripheral airways, elevations in Raw are accompanied by increases in Rti 13, 16, 19, 21, which would augment the changes in Rrs,int; however, it is also possible that the alterations in Raw and Rti are independent 19. Accordingly, both in routine clinical practice and under experimental conditions, much improved bedside monitoring techniques are advocated in order to follow the changes in airway tone and tissue mechanics during mechanical ventilation.

Besides a knowledge of the instrumental resistance component, a prerequisite for higher measurement sensitivity, it is of primary importance to choose methods with the ability to separate the airway and tissue components of total respiratory resistance, such as low-frequency forced oscillations 11, 17, 20, 21 and appropriately instrumented airway interruptions 12, 23, 24, 32.

Acknowledgments

The authors would like to thank L. Vígh and I. Kopasz for their excellent technical assistance and the staff of the Division of Cardiac Surgery for their cooperation.

  • Received January 8, 2002.
  • Accepted July 19, 2002.
  • © ERS Journals Ltd

References

  1. ↵
    Guntupalli KK. Acute pulmonary edema. Cardiol Clin 1984;2:183–200.
    OpenUrlPubMed
  2. ↵
    Estenne M, Gevenois PA, Kinnear W, Soudon P, Heilporn A, De Troyer A. Lung volume restriction in patients with chronic respiratory muscle weakness: the role of microatelectasis. Thorax 1993;48:698–701.
    OpenUrlAbstract/FREE Full Text
  3. ↵
    Gilroy RJ Jr, Lavietes MH, Loring SH, Mangura BT, Mead J. Respiratory mechanical effects of abdominal distension. J Appl Physiol 1985;58:1997–2003.
    OpenUrlAbstract/FREE Full Text
  4. ↵
    Wright PE, Bernard GR. The role of airflow resistance in patients with the adult respiratory distress syndrome. Am Rev Respir Dis 1989;139:1169–1174.
    OpenUrlPubMedWeb of Science
  5. ↵
    Wright PE, Carmichael LC, Bernard GR. Effect of bronchodilators on lung mechanics in the acute respiratory distress syndrome (ARDS). Chest 1994;106:1517–1523.
    OpenUrlCrossRefPubMedWeb of Science
  6. ↵
    Kil HK, Rooke GA, Ryan-Dykes MA, Bishop MJ. Effect of prophylactic bronchodilator treatment on lung resistance after tracheal intubation. Anesthesiology 1994;81:43–48.
    OpenUrlCrossRefPubMedWeb of Science
  7. ↵
    Eames WO, Rooke GA, Wu RS, Bishop MJ. Comparison of the effects of etomidate, propofol, and thiopental on respiratory resistance after tracheal intubation. Anesthesiology 1996;84:1307–1311.
    OpenUrlCrossRefPubMedWeb of Science
  8. ↵
    Berry A, Brimacombe J, Keller C, Verghese C. Pulmonary airway resistance with the endotracheal tube versus laryngeal mask airway in paralyzed anesthetized adult patients. Anesthesiology 1999;90:395–397.
    OpenUrlCrossRefPubMedWeb of Science
  9. ↵
    Kim ES, Bishop MJ. Endotracheal intubation, but not laryngeal mask airway insertion, produces reversible bronchoconstriction. Anesthesiology 1999;90:391–394.
    OpenUrlCrossRefPubMedWeb of Science
  10. ↵
    Laude EA, Emery CJ, Suvarna SK, Morcos SK. The effect of antihistamine, endothelin antagonist and corticosteroid prophylaxis on contrast media induced bronchospasm. Br J Radiol 1999;72:1058–1063.
    OpenUrlAbstract
  11. ↵
    Navajas D, Farré R, Canet J, Rotger M, Sanchis J. Respiratory input impedance in anesthetized paralyzed patients. J Appl Physiol 1990;69:1372–1379.
    OpenUrlAbstract/FREE Full Text
  12. ↵
    Bates JHT, Daróczy B, Hantos Z. A comparison of interrupter and forced oscillations measurements of respiratory resistance in the dogs. J Appl Physiol 1992;72:46–52.
    OpenUrlAbstract/FREE Full Text
  13. ↵
    Hantos Z, Daróczy B, Suki B, Nagy S, Fredberg JJ. Input impedance and peripheral inhomogeneity of dog lungs. J Appl Physiol 1992;72:168–178.
    OpenUrlAbstract/FREE Full Text
  14. ↵
    Hantos Z, Adamicza Á, Govaerts E, Daróczy B. Mechanical impedances of lungs and chest wall in the cat. J Appl Physiol 1992;73:427–433.
    OpenUrlAbstract/FREE Full Text
  15. ↵
    Lutchen KR, Yang K, Kaczka DW, Suki B. Optimal ventilation waveforms for estimating low-frequency respiratory impedance. J Appl Physiol 1993;75:478–488.
    OpenUrlAbstract/FREE Full Text
  16. ↵
    Barnas GM, Sprung J, Kahn R, Delaney PA, Agarwal M. Lung tissue and airway impedances during pulmonary edema in normal range of breathing. J Appl Physiol 1995;78:1889–1197.
    OpenUrlAbstract/FREE Full Text
  17. ↵
    Sly PD, Hayden MJ, Peták F, Hantos Z. Measurement of low-frequency respiratory impedance in infants. Am J Respir Crit Care Med 1996;154:161–166.
    OpenUrlCrossRefPubMedWeb of Science
  18. ↵
    Peták F, Hayden MJ, Hantos Z, Sly PD. Volume dependence of respiratory impedance in infants. Am J Respir Crit Care Med 1997;156:1172–1177.
    OpenUrlPubMedWeb of Science
  19. ↵
    Suki B, Peták F, Adamicza Á, Daróczy B, Lutchen KR, Hantos Z. Airway and tissue constrictions are greater in closed than in open-chest conditions. Respir Physiol 1997;108:129–141.
    OpenUrlCrossRefPubMedWeb of Science
  20. ↵
    Farré R, Ferrer M, Rotger M, Torres A, Navajas D. Respiratory mechanics in ventilated COPD patients: forced oscillation versus occlusion techniques. Eur Respir J 1998;12:170–176.
    OpenUrlAbstract
  21. ↵
    Kaczka DW, Ingenito EP, Israel E, Lutchen KR. Airway and lung tissue mechanics in asthma. Effects of albuterol. Am J Respir Crit Care Med 1999;159:169–178.
    OpenUrlPubMedWeb of Science
  22. ↵
    Broseghini C, Brandolese R, Poggi R, Bernasconi M, Manzin E, Rossi A. Respiratory resistance and intrinsic positive end-expiratory pressure (PEEPi) in patients with the adult respiratory distress syndrome (ARDS). Eur Respir J 1988;1:726–731.
    OpenUrlAbstract/FREE Full Text
  23. ↵
    Pelosi P, Cereda M, Foti G, Giacomini M, Pesenti A. Alterations of lung and chest wall mechanics in patients with acute lung injury: effects of positive end-expiratory pressure. Am J Respir Crit Care Med 1995;152:531–537.
    OpenUrlPubMedWeb of Science
  24. ↵
    Morina P, Herrera M, Venegas J, Mora D, Rodriguez M, Pino E. Effects of nebulized salbutamol on respiratory mechanics in adult respiratory distress syndrome. Intensive Care Med 1997;23:58–64.
    OpenUrlCrossRefPubMedWeb of Science
  25. ↵
    Freezer NJ, Lanteri CJ, Sly PD. Effect of pulmonary blood flow on measurements of respiratory mechanics using the interrupter technique. J Appl Physiol 1993;74:1083–108.
    OpenUrlAbstract/FREE Full Text
  26. ↵
    Lanteri CJ, Kano S, Duncan AW, Sly PD. Changes in respiratory mechanics in children undergoing cardiopulmonary bypass. Am J Respir Crit Care Med 1995;152:1893–900.
    OpenUrlPubMedWeb of Science
  27. ↵
    Daróczy B, Hantos Z. Generation of optimum pseudorandom signals for respiratory impedance measurements. Int J Biomed Comput 1990;25:21–31.
    OpenUrlCrossRefPubMed
  28. ↵
    Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–310.
    OpenUrlCrossRefPubMedWeb of Science
  29. ↵
    Finucane KE, Dawson SV, Phelan PD, Mead J. Resistance of intrathoracic airways of healthy subjects during periodic flow. J Appl Physiol 1975;38:517–530.
    OpenUrlAbstract/FREE Full Text
  30. ↵
    Bates JHT, Rossi A, Milic-Emili J. Analysis of the behavior of the respiratory system with constant inspiratory flow. J Appl Physiol 1985;58:1840–1848.
    OpenUrlAbstract/FREE Full Text
  31. Sly PD, Bates JH, Milic-Emili J. Measurement of respiratory mechanics using the Siemens Servo Ventilator 900C. Pediatr Pulmonol 1987;3:400–405.
    OpenUrlPubMedWeb of Science
  32. ↵
    D'Angelo E, Robatto FM, Calderini E, et al. Pulmonary and chest wall mechanics in anesthetized paralyzed humans. J Appl Physiol 1991;70:2602–2610.
    OpenUrlAbstract/FREE Full Text
PreviousNext
Back to top
View this article with LENS
Vol 20 Issue 6 Table of Contents
  • Table of Contents
  • Index by author
Email

Thank you for your interest in spreading the word on European Respiratory Society .

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Components of respiratory resistance monitored in mechanically ventilated patients
(Your Name) has sent you a message from European Respiratory Society
(Your Name) thought you would like to see the European Respiratory Society web site.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Print
Alerts
Sign In to Email Alerts with your Email Address
Citation Tools
Components of respiratory resistance monitored in mechanically ventilated patients
B. Babik, F. Peták, T. Asztalos, Z.I. Deák, G. Bogáts, Z. Hantos
European Respiratory Journal Dec 2002, 20 (6) 1538-1544; DOI: 10.1183/09031936.02.00000802

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero

Share
Components of respiratory resistance monitored in mechanically ventilated patients
B. Babik, F. Peták, T. Asztalos, Z.I. Deák, G. Bogáts, Z. Hantos
European Respiratory Journal Dec 2002, 20 (6) 1538-1544; DOI: 10.1183/09031936.02.00000802
del.icio.us logo Digg logo Reddit logo Technorati logo Twitter logo CiteULike logo Connotea logo Facebook logo Google logo Mendeley logo
Full Text (PDF)

Jump To

  • Article
    • Abstract
    • Methods
    • Results
    • Discussion
    • Acknowledgments
    • References
  • Figures & Data
  • Info & Metrics
  • PDF
  • Tweet Widget
  • Facebook Like
  • Google Plus One

More in this TOC Section

  • Automatic regulation of the cuff pressure in endotracheally-intubated patients
  • Efficiency of cold passover and heated humidification under continuous positive airway pressure
Show more Original Articles: Technical Note

Related Articles

Navigate

  • Home
  • Current issue
  • Archive

About the ERJ

  • Journal information
  • Editorial board
  • Reviewers
  • CME
  • Press
  • Permissions and reprints
  • Advertising

The European Respiratory Society

  • Society home
  • myERS
  • Privacy policy
  • Accessibility

ERS publications

  • European Respiratory Journal
  • ERJ Open Research
  • European Respiratory Review
  • Breathe
  • ERS books online
  • ERS Bookshop

Help

  • Feedback

For authors

  • Instructions for authors
  • Submit a manuscript
  • ERS author centre

For readers

  • Alerts
  • Subjects
  • Podcasts
  • RSS

Subscriptions

  • Accessing the ERS publications

Contact us

European Respiratory Society
442 Glossop Road
Sheffield S10 2PX
United Kingdom
Tel: +44 114 2672860
Email: journals@ersnet.org

ISSN

Print ISSN:  0903-1936
Online ISSN: 1399-3003

Copyright © 2021 by the European Respiratory Society