Abstract
The interrupter technique is a convenient and sensitive technique for studying airway function in subjects who cannot actively participate in (forced) ventilatory function tests.
Reference values for preschool children exist but are lacking for children >7 yrs. Reference values were obtained for expiratory interrupter resistance (Rint,e) in 208 healthy Dutch Caucasian children 3–13 yrs of age.
A curvilinear relationship between Rint,e and height was observed, similar to published airways resistance data measured by plethysmography. No significant differences in cross-sectional trend or level of Rint,e were observed according to sex. It was found that Z-scores could be used to express individual Rint,e values and to describe intra- and interindividual differences based on the reference equation: 10logRint,e=0.645–0.00668×standing height (cm) kPa·L−1·s−1 and residual SD (0.093 kPa·L−1·s−1).
Expiratory interrupter resistance provides a tool for clinical and epidemiological assessment of airway function in a large age range.
The interrupter technique is one of the few lung function tests that can be used for assessment of airway calibre in young children 1–3. With this technique, measurements of the resistance of the respiratory system (Rint) can be carried out quickly, with minimal cooperation of the child. Rint measurements have been shown to be reproducible 1, 4–6, sufficiently sensitive to detect (sub)clinical airway obstruction 6, 7, and to correlate satisfactorily with measurements of airway resistance 4, 8, 9. The technique can not only be used as a tool to screen for airway obstruction, but also to assess the responses to bronchodilating and bronchoconstricting agents 1, 3, 10. It is especially suitable for preschool children because it only requires passive cooperation. However, passive measurements of airway function may also be required for clinical research in older children or in older children who are unable to perform forced expiratory manoeuvres because of developmental disorders or neuromuscular disease. Reference data are available for young children 6, 7, 11, 12 and adults 13, but not for children of >7 yrs of age. Therefore, the aim of the present study was to expand the previous data set to better describe relationships between expiratory resistance and body size. Normal Rint values were obtained during expiration (Rint,e) in 208 healthy Caucasian children aged 3–13 yrs, from a general population. Measurements were preferably made during expiration rather than during inspiration because Rint,e appears to be more sensitive to detecting changes in resistance within children due to respiratory infections, and to discriminating better between children with and without respiratory symptoms or disease as compared to Rint values obtained during inspiration 6.
Methods
Data set
Rint,e measurements were carried out using identical equipment and the same measurement protocol in two sets of healthy Dutch Caucasian children aged 1–13 yrs, recruited from two daycare centres, two kindergartens and two elementary schools. Information on respiratory symptoms, eczema, allergy, parental smoking, doctor's diagnosis of asthma, and asthma medication was obtained using modified International Study of Asthma and Allergies in Childhood (ISAAC) questionnaires 14. Children were included in the reference population when they had no respiratory symptoms in the month prior to, or during, the measurements. Exclusion criteria were: history of asthma, recurrent rhinitis, eczema, cardiorespiratory or other chronic disease, known anatomical abnormalities of the upper or lower airways, and vocal cord disorders. The authors wanted to obtain reference values from a normal population rather than an ideal population 15. Therefore, mild respiratory symptoms not requiring medical care in the past and involuntary exposure to parental smoking without a history of respiratory symptoms or disease were not part of the exclusion criteria. The study and its protocol were approved by the medical ethics committees of the medical centres and by the principals and boards of the institutes involved. Informed consent was given by the parents of all participating children. When children refused to cooperate, no Rint,e measurements were attempted.
Equipment
Rint,e was assessed using the MicroRint (Micro Medical Ltd, Rochester, UK), as described previously 6. Rint,e was calculated using the back extrapolation technique to t=0 ms after shutter closure during 100 ms 1. Daily calibrations of pressure and flow (volume) were carried out using a manometer and a 2 L precision pump. All measurements were carried out with a filter (Micro Medical Ltd) in place to prevent contamination and dysfunction 16.
Measurement protocol
The protocol has been described previously 6. After the supervisor of the children explained the purpose of the measurements, a measurement was demonstrated on the supervisor and subsequent measurements were carried out in groups of 2–4 children at a time, in a familiar and quiet room. Children were seated and no physical exercise was allowed during 10 min prior to the measurements. During measurements, children were instructed to breathe quietly, sitting upright while the cheeks and chin were supported from behind by the investigator. The head was positioned in slight extension and a nose clip was used. The position of the MicroRint was adjusted on a support arm to facilitate unobstructed breathing. A minimal number of five correct tracings (maximal 10) was obtained at the peak of expiratory tidal flow, because expiratory interruptions appear more sensitive in detecting airways obstruction than those during inspiration 6. Tracings were rejected in the cases of tachypnoea, usage of the vocal cords, extreme neck flexion or extension, or leakage of the mouth piece. Tracings not showing the timing of interruption on the flow tracing or tracings with a horizontal or declining pressure signal suggesting leakage at the mouth or altered ventilation pattern, were also discarded 2 (fig. 1⇓).
Data analysis
The individual Rint,e data were expressed as median values because individual data were not normally distributed 6. Reference values for Rint,e were described based on a model assuming a linear or curvilinear relationship with the following standard independent variables: standing height, weight, and age. Because of physiological similarities between Rint and airway resistance obtained by plethysmography, the current authors hypothesised that an exponential model with standing height would create the best fit for the Rint,e data, as seen in published reference equations for airway resistance (Raw) 17, 18. Trends of residuals with height or age were assessed from linear regression analyses. The threshold for statistical significance was set at p=0.05.
Results
Subjects
The first data set consisted of 135 healthy Dutch children (60 males) studied in Rotterdam, the Netherlands, who were selected from a survey in which the parents of 698 children were asked to participate. Permission was given for 341 (49%) children, 36 refused participation, and 12 failed to complete the measurements. Of the remaining 293 children, 135 (39%) met the inclusion criteria and completed the measurements. This included 54 healthy children described previously 6. The second data set consisted of 79 Dutch children (41 males) studied in Utrecht, the Netherlands. These children were selected from a study in which parents of 445 children were asked to participate. Permission was obtained for 212 (48%) of these children. Reliable Rint,e measurements were carried out in 200 children, of whom 79 (40%) met the inclusion criteria listed above. Both studies were carried out in suburban parts of the cities, inhabited by middle class income Dutch families. Only two of the 24 children who failed to complete the measurements were >4 yrs of age. Anthropometric data of all 214 children are summarised in table 1⇓. The children from data set 1 were slightly younger than those from data set 2 (table 2⇓). The coefficients of variation from data set 1 were larger than those from data set 2 (table 2⇓). This seemed to be explained by the differences in age between centres; at <6 yrs there was a negative correlation between coefficient of variation and age (r=−0.21, p=0.004). The reference equation was based on the 208 children of 3–13 yrs of age, as reliable measurements could only be obtained in six children <3 yrs.
Reference equation for Rint
An inverse, curvilinear relationship was found between Rint,e and the independent variables of standing height, age and weight. When standing height was used in an exponential model instead of a linear model with Rint,e, the explained variance increased from 59 to 63% and the residual standard deviation (RSD) decreased from 0.150 to 0.093 kPa·L·s−1. Residuals of the exponential model were homoscedastically distributed, demonstrating no trend with standing height (fig. 2⇓). When age was added to standing height in an exponential model, the explained variance increased by <3%, and the improvement of RSD was <0.0003 kPa·L·s−1. When age was used as the only independent variable in an exponential model with Rint,e, the explained variance was 64% (RSD=0.091 kPa·L·s−1), but the distribution of the residuals became heteroscedastic for subjects >10 yrs of age. In the children aged >10 yrs, the variability of standing height for age was larger than in the younger subjects. When using weight as the only independent variable, explained variance was 49% (RSD=0.108 kPa·L·s−1).
Reference equations for Rint,e are: for the linear model. r=−0.77, RSD=0.150 kPa·L·s−1 (p<0.001).for the exponential model. r=−0.79, RSD=0.093 kPa·L·s−1 (p<0.001).
Based on the exponential model, the means (sems) of standardised residuals for 97 males and 111 females were −0.079 (0.096) kPa·L·s−1 and 0.071 (0.097) kPa·L·s−1, respectively, with a mean (95% confidence interval (CI)) difference between males and females of 0.15 (−0.12–0.42) kPa·L·s−1. No trends were observed between standing height and standardised residuals for males or females. The separate regression equations from the two data sets did not differ significantly in a multiple regression model that included centre (p=0.33) and interaction between standing height and centre (p=0.59).
Discussion
A few studies have reported reference equations for the interrupter technique in young school children 6, 7, 11, 12, but for older children these are lacking. Because of the possible applications of this technique in a larger age range (epidemiological and clinical research, children unable to participate in active lung function measurements), normal values for Rint,e were obtained in 208 healthy Caucasian children aged between 3–13 yrs. In a previous study by the current authors, a linear model was considered satisfactory to describe the relationship between height and Rint,e 6, but in the present study, due to the larger range in height, an exponential model appeared more appropriate because of a curvilinear relationship. This pattern is consistent with reports of Raw in healthy children 19.
This is the largest study of Rint,e that has been performed in healthy preschool and school children so far. Despite the large number of observations, a significant sex-related difference in airway patency was not detected. This suggests that a possible small difference in airways resistance between sexes is not clinically relevant, or that it cannot be detected when measuring resistance of the respiratory system with this technique.
Height, but not age, was used as an independent variable because of physiological arguments, not because the relationship between height and Rint,e was statistically superior. It is conceivable that body size can function as a proxy for airway calibre, whereas age may also indirectly reflect airway size in children <13 yrs but not in adolescents or adults. Age may be equally valid as a more convenient independent variable in reference equations for Rint,e 7, but this is probably limited to young children only. Indeed, in the present study, the variability of the residuals in the children >10 yrs of age was considerably increased compared to that in younger children, which is explained by a larger variation in height for age.
Because the residuals of the exponential model were normally distributed (with RSD=0.093 kPa·L·s−1), individual measurements can be expressed as Z-scores:This facilitates comparisons within and between individuals. Until there is international standardisation for Rint,e measurements 20, reference equations are likely to differ according to the equipment and protocol of shutter timing and back extrapolation. In the protocol of the present study, interruptions were programmed at peak tidal expiratory flow, which appears to standardise inflation level 6, 9. The linear model of the present study fits remarkably well with the reference equation of van Altena and Gimeno 13: van Altena and Gimeno 13 studied Rint in 172 adults and teenagers, although the exact measurement procedure was not described and the population and equipment differed markedly. The current results are not comparable with those of Klug et al. 11 who programmed inspiratory interruptions at 50 mL above functional residual capacity. The effect of this procedure might have been that with increasing body size, interruptions occurred at progressively lower inflation levels. This could explain the lesser slope with height and lower explained variance of their reference equations.
It is difficult to compare Rint,e values with measurements of Raw or lung resistance (Rl) in healthy populations because of differences in technique and population characteristics, but the reference equation present in this study compares favourably with those of Dab and Alexander 17 and Helliesen et al. 18, respectively: The present authors were able to measure Rint,e reliably in only six children <3 yrs of age and do not recommend routine assessment of Rint,e in children of that age because of low feasibility 6. The use of face masks in children <3 yrs may enhance the feasibility of the test, but Rint measurements obtained using face masks can differ from those obtained with mouthpieces 21 due to differences in the compliance and resistance of the mask and the degree of airways obstruction. Rint measurements using face masks may require specific reference equations.
The interrupter technique remains one of the most convenient and sensitive tests of airway function in young children >3 yrs of age, and has the potential for use over a wide age range.
Acknowledgments
The authors would like to thank all the children, parents, and personnel of the participating institutions for their enthusiastic contribution to this study.
- Received August 8, 2001.
- Accepted March 19, 2002.