Leak compensating abilities of six different positive pressure ventilators commonly used to deliver noninvasive positive pressure ventilation, including the bilevel positive airway pressure (BiPAP) S/T-D and Quantum (Respironics Inc, Murrysville, PA, USA), 335 and O'NYX (Mallinckrodt Inc, St Louis, MO, USA), PLV 102 (Respironics), and Siemens Servo 900C (Siemens Inc, Danvers, MA, USA).
Using a test lung model, compensatory capabilities of the ventilators were tested for smaller and larger leaks using the assist/control or timed modes. Back-up rate was 20·min−1, inspiratory pressure was 18 cmH2O, and expiratory pressure was 5 cmH2O.
It was found that even in the absence of air leaking, delivered tidal volume differed substantially between the ventilators during use of pressure-targeted modes, depending on inspiratory flows, inaccuracies in set versus delivered pressures, and inspiratory duration. Also during pressure-targeted ventilation, increasing the tI/ttot up to, but not beyond, 0.5 improved compensation by lengthening inspiratory duration, whereas use of a sensitive flow trigger setting tended to cause autocycling during leaking, interfering with compensation. Leaking interfered with cycling of the BiPAP S/T, inverting the I:E ratio, shortening expiratory time, and reducing delivered tidal volume. Volume-targeted modes achieved limited compensation for small air leaks, but compensated poorly for large leaks.
To conclude, leak-compensating capabilities differ markedly between ventilators but pressure-targeted ventilators are preferred for noninvasive positive pressure ventilation in patients with substantial air leaking. Adequate inspiratory flows and durations should be used, triggering sensitivity should be adjusted to prevent autocycling, and a mechanism should be available to limit inspiratory time and avoid I:E ratio inversion.
Noninvasive positive pressure ventilation (NPPV) is widely used to assist breathing in both acute and chronic forms of respiratory failure 1, 2. In contrast to invasive ventilation, NPPV uses an open circuit design that is inherently leaky. Although ventilation can often be assisted even in the presence of sizeable leaks 3, different ventilators and ventilator modes may be more or less capable of compensating for air leaks. This compensatory capability may be important in optimizing the success of NPPV, but few studies have examined the effectiveness of different ventilators in delivering ventilation in the face of air leaks.
Air leaks during NPPV consist of either mask leaks between the skin and mask, mouth leaks with nasal ventilation, or nose leaks with mouthpiece ventilation. Some air may also escape via the oesophagus, but because of the relatively high impedance posed by the lower oesophageal sphincter 4, this route is likely to be minor in comparison to the others. Studies on patients using nocturnal volume- or pressure-limited ventilation have found that air leaks occur during most of sleep 3, 5. These studies show that ventilation and oxygenation are adequately supported in most patients despite the presence of leaks, but that large leaks may interfere with ventilator cycling, compromise minute ventilation, and cause sleep fragmentation due to leak-associated arousals 3, 5.
Simple bedside interventions may alleviate air leaking. Mask leaks can be reduced by ensuring proper mask fit and by using optimal headstrap tension. The temptation to merely tighten the straps to reduce air leaking must be resisted, because this may reduce patient comfort and tolerance of NPPV, and promote the development of nasal bridge ulcers 6. Air leaking through the mouth that occurs during nasal ventilation can be reduced by encouraging patients to keep their mouths closed, using chin straps, or switching to a mouthpiece or an oronasal mask 6. However, air leaks often persist despite these interventions.
For these reasons, ventilators that are designed to compensate for air leaks are desirable for noninvasive ventilation. Using a lung model testing system, the leak compensating abilities of six positive pressure ventilators commonly used to administer NPPV, including pressure-targeted and volume-targeted devices were evaluated and compared. It was hypothesized that ventilator characteristics such as mode of ventilation, peak inspiratory flow capabilities, triggering sensitivities, and criteria for termination of inspiration as well as lung characteristics such as resistance and compliance, would determine the effectiveness of leak compensation.
Methods
Description of the ventilators
The leak compensation characteristics of commercially available positive-pressure ventilators, five portable and one “critical care”, selected to represent the range of ventilators commonly used to deliver NPPV were evaluated. These included three pressure-limited “bilevel positive airway pressure” (BiPAP) type ventilators, the BiPAP S/T-D and Quantum Pressure Support Ventilator (Respironics, Murrysville, PA, USA), and the Nellcor Puritan Bennett 335 (Mallinckrodt, St. Louis, MO, USA). These are blower-based flow generators capable of delivering continuous positive airway pressure (CPAP) or cycling between inspiratory and expiratory positive airway pressure (IPAP and EPAP, respectively). Other ventilators selected because they are often used for NPPV included the PLV 102 (Respironics), a piston-driven portable volume-targeted ventilator, the blower-driven O′NYX (Mallinckrodt), a ventilator commonly used in Europe but not yet available in North America, and the Siemens Servo 900C (Siemens Inc, Danvers, MA, USA), a critical care ventilator used widely throughout the world. The latter two ventilators deliver either pressure- or volume-targeted breaths.
All ventilators were tested in controlled modes using a back-up rate. The BiPAP ventilator was evaluated in both the spontaneous/timed (S/T) and timed (T) modes, and the Quantum, 335, and PLV 102 were evaluated in the assist/control mode. For the O′NYX and Siemens ventilators, both pressure-targeted (assist controlled pressure ventilation (ACPV) and pressure control ventilation (PCV), respectively) and volume-targeted (assist control ventilation, ACV) modes were evaluated. Important differences between the ventilators are listed in table 1⇓. These include differences in inspiratory termination (cycling) criteria between the ventilators. The Quantum 335 and Siemens (for both modes) use a timer set by the I:E ratio when inspirations are triggered by the timed back-up rate, whereas BiPAP in the S/T mode cycles in response to a decrease in inspiratory flow 7. For this reason, the timed (T) mode on the BiPAP that cycles into expiration based on a preset timer was also tested. Ventilator tubing used was that recommended by the manufacturer. The Whisper SwivelTM, a fixed exhalation valve, was used with the BiPAP ventilators. For the O′NYX, Siemens and PLV 102, separate inspiratory and expiratory circuits and exhalation valves were used as supplied by the manufacturer.
Leak testing system
Each ventilator was tested using a double compartment lung model (Dual Adult Training Test Lung Model 1600, Michigan Instruments, Grand Rapids, MI, USA) connected to a 7 mm internal diameter (ID) endotracheal tube. The lung model was placed inside a body plethysmograph that measured volume displacement of the lung model by means of an externally connected wedge spirometer (fig. 1⇓). Lung model compliances of 0.1, 0.06, and 0.03 L·cm−1 were used. Resistance could be altered by the insertion of one of two nonlinear resistors with a fixed orifice (Rp 5 or Rp 20; Michigan Instruments) between the endotracheal tube and the ventilator circuit. At a flow rate of 60 L·min−1, the resistance of the Rp 5 and Rp 20 was 2.7 cmH2O·L·s−1 and 17.6 cmH2O·L·s−1, respectively.
Ventilators were tested in the absence of leak and in the presence of a smaller or larger leak. A sidearm extended from the ventilator circuit permitted control of the amount of leak (fig. 1⇑). Controlled leaks were created by inserting resistors into the sidearm. The resistors (1.0 and 0.2 cmH2O·L·s−1, respectively, at flow rates of 60 L·s−1) were selected to allow leakage of 1/4 (small leak) and 1/2 (large leak) of the delivered tidal volume (VT) when the BiPAP in the T mode was set at IPAP/EPAP settings that delivered a VT of 1 L in the absence of leak. The leak flow varied widely depending on the ventilator settings, but ranged from 15–30 L·min−1 for the small leak and 30–120 L·min−1 for the large leak. These were selected to span above and below the range previously reported in sleeping patients using the BiPAP S/T (24–38 L·min−1) 3. As the amount of leak was varied, airway pressure, inspiratory and leak flows and VT were measured using inline manometers and pneumotachographs (Fleisch No. 2) and recorded using a Gould 3800 strip chart recorder (Gould, Cleveland, OH, USA).
Experimental settings
For baseline comparisons, respiratory frequency was set at 20 breaths·min−1 in all of the experiments to reflect commonly recommended breathing rates for noninvasive ventilation 8, 9. IPAP/EPAP settings were 18 and 5 cmH2O respectively for the BiPAP (in both T and S/T modes), 335, and Quantum ventilators. Inspiratory and expiratory sensitivity settings on the 335 were 2 and 3, respectively, with 1 being the most and 5 the least sensitive settings 7. In order to match the IPAP/EPAP settings, the O′NYX in the ACPV mode was at a pressure support of 13 cmH2O and a PEEP of 5 cmH2O and the inspiratory trigger was set at 1.5 cmH2O (0 being the most and 3 the least sensitive settings). Inspiratory time/total breath time (tI/ttot) was set at 0.33, and test lung compliance at 0.1 L·cmH2O.
To test the effect of differences in inspiratory duration on delivered VT, tI/ttot was varied (0.25, 0.33 or 0.50) for the BiPAP and PLV. The BiPAP T was additionally tested at a tI/ttot of 0.66 to determine the effect of inverted I‡‡:‡‡E ratios on delivered VT. The influence of inspiratory trigger sensitivity was tested by using the O′NYX at inspiratory sensitivities of 0.5, 1.5 and 3.0 cmH2O. The effect of changes in respiratory system compliance and resistance on tidal volumes delivered in the presence of leak were evaluated using the 335 at test lung compliances of 0.1, 0.06 and 0.03 L·cmH2O. With test lung compliance set at 0.1 L·cmH2O, each of 3 resistance conditions (no resistor, Rp 5 or Rp 20) was also tested. The effect of varying leak on VT delivered by volume-targeted ventilation was assessed using the PLV 102, O'NYX and Siemens at four different tidal volume settings, with test lung compliance set at 0.1 L·cmH2O and tI/ttot at 0.33.
Statistical analysis
Six breaths were analysed and averaged for each experimental setting. All values are reported as mean±sd. Repeated measures analysis of variance was used to compare mean values for each ventilator at different settings, and two way analysis of variance was used to compare mean values between different ventilators. When significant differences were detected, post hoc analysis was performed using Tukey's exact test. Very little variability occurred between breaths, so sd bars are not shown in the figures to improve clarity of presentation.
Results
Leak compensation in pressure-targeted ventilators
During pressure-targeted ventilation, delivered VT differed substantially between ventilators despite equal inspiratory and expiratory pressure settings (18 and 5 cmH20, respectively), even in the absence of leak (fig. 2a⇓). These differences were related to variances in inspiratory flow rates, actual pressures delivered, and inspiratory duration (as determined by tI/ttot) (table 2⇓). For example, the 400-mL greater VT delivered by the BiPAP in the S/T mode as compared to the T mode was related to a greater tI/ttot in the S/T mode, allowing more time for lung inflation. In contrast, the three ventilators using volume-targeted modes consistently delivered the selected volume of 1 L, as would be anticipated.
With all of the ventilators, when a leak was introduced into the system, airway pressure decreased (table 2⇑) and delivered VT was significantly reduced (fig. 2b⇑). When expressed as % of VT delivered in the absence of leak, VT during a small leak was best maintained by the pressure-targeted modes, particularly the O′NYX and Siemens ventilators. The BiPAP in the S/T mode compensated less effectively because of excessive prolongation of the inspiratory duration and, as a group, the volume-targeted modes compensated far less effectively than the pressure-targeted modes (fig. 2b⇑). In the presence of a large leak, the Quantum best maintained delivered VT, which was approximately 65% of the baseline. Once again, the pressure-targeted modes compensated much better than the volume-targeted modes, with the exception of the Siemens PCV (fig. 2b⇑), which failed to sustain high inspiratory flow rates throughout the inspiratory duration (fig. 3⇓).
Surprisingly, leak-compensation between the volume-targeted modes differed considerably, with the O'NYX in the ACV mode performing best by virtue of its ability to increase inspiratory flow rate (table 2⇑). As would be expected with most volume-targeted ventilators, inspiratory flow rates did not change with the PLV, accounting for its poor leak compensating abilities. With the Siemens in the ACV mode, inspiratory flow rate increased slightly in the face of a large leak, but then decreased abruptly, similar to the response with the PCV mode, resulting in no measurable delivered VT (table 2⇑, fig. 3⇑).
Effect of set inspiratory time/total breath time on delivered tidal volume
Among the “bilevel” devices that permit setting of tI/ttot, delivered VT increased as tI/ttot was raised from 0.25, to 0.33 and 0.5, both with small and large leaks (fig. 3⇑). At each setting, the Quantum preserved leak VT better than the 335 or BiPAP T, related to a longer inspiratory duration (table 2⇑). During air leaking, the BiPAP in the S/T mode delivered a lower VT (fig. 4⇓). This was related to the fact that in the S/T mode, the % IPAP control on the BiPAP is inoperative and does not permit adjustments of tI/ttot. Inspiration is terminated by a decrease in inspiratory flow up to a maximum duration of 3 s; therefore, tI/ttot increases during leak because inspiratory flow fails to drop sufficiently to cycle the ventilator. tI/ttot averaged 0.47 in the absence of leak, and increased to 0.8 with a small or large leak. To illustrate this effect of excessive prolongation of tI/ttot, delivered VT using the BiPAP in the T mode fell from 1.23 L to 0.83 L when tI/ttot was lengthened from 0.5 to 0.66, while other settings remained unchanged.
Effect of varying inspiratory trigger sensitivity
The O'NYX, Siemens and 335 have adjustable inspiratory triggers, and experiments were performed to test the effect of inspiratory trigger sensitivity on leak compensation. Trigger sensitivity had no effect on delivered VT with the 335 (data not shown). However, the Siemens and O'NYX both increased delivered rates over the set back-up rates at higher inspiratory trigger sensitivity settings (autocycling) (figs. 5 and 6⇓⇓). With the Siemens, rate increased to 30·min−1 and delivered VT fell when inspiratory sensitivity was set below 0.6 cmH2O for a small leak, and 0.8 cmH2O for a large leak. With the O'NYX in the ACPV mode, autocycling did not occur in the presence of a small leak. However, with a large leak and more sensitive trigger sensitivity settings (0.5 and 1.5 cmH2O), VT dropped dramatically while respiratory rates climbed to 36·min−1, and the ventilator failed to maintain PEEP in the inspiratory tubing (fig. 6⇓). Autocycling did not occur when the sensitivity was set at 3 cmH2O, even with a large leak, and leak compensation was much improved.
Leak compensation with volume-targeted modes
As anticipated, delivered VT with the volume-targeted modes was much lower during leaks than with pressure-targeted ventilators, because of their inability (PLV) or limited ability (O'NYX and Siemens) to compensate by sustaining or increasing inspiratory flow rates (fig. 2⇑, table 2⇑). Some increase in inspiratory flow rate occurred with some of the volume-targeted modes, particularly with the O'NYX ACV mode, accounting for its greater leak compensating abilities compared to the PLV (fig. 2⇑). In contrast to the pressure-targeted ventilators, as set tI/ttot on the PLV 102 was increased from 0.25 to 0.33 to 0.5, delivered VT fell slightly related to proportionate decreases in inspiratory flow rate, regardless of the presence of leak in the system (fig. 7⇓). In addition, introducing a leak into the system reduced delivered VT at all tI/ttot settings more than with the pressure-targeted ventilators (fig. 2b⇑), particularly with the PLV 102. On the other hand, for small leaks, increasing set VT from 500 mL to 2 L increased delivered VT substantially during use of the O'NYX and Siemens and slightly during use of the PLV (fig. 8⇓). However, none of the volume-targeted modes was able to compensate for large leaks, even at the higher set VTs (fig. 8⇓).
Effect of changes in respiratory system impedance on leak compensation
As would be anticipated during pressure-limited ventilation, increases in respiratory system impedance resulting from either reduced lung compliance or increased airway resistance reduced delivered VT, as exemplified by the 335 (fig. 9⇓). In response to air leaking, delivered VT were further reduced, proportionate to the decreases in VT observed in the absence of leak (fig. 9⇓). Responses of the other pressure-limited ventilators was similar, and tI/ttot of the BiPAP S/T was not altered by changes in resistance or compliance. In contrast, the volume-limited PLV 102 maintained VT delivery despite changes in test lung compliance, as would be anticipated (data not shown).
Discussion
The presented study demonstrates that the ability to compensate for air leaks varies substantially between ventilators depending on ventilator mode, leak size, inspiratory airflow capabilities, and inspiratory duration as determined by set tI/ttot or ventilator cycling mechanisms. Although prior investigations have compared the performance characteristics of pressure-targeted portable ventilators 10, 11, few prior published studies have attempted to evaluate the impact of leaks on the efficacy of NPPV. In a recent laboratory study comparing the performance of four positive pressure ventilators, two pressure-targeted ventilators maintained delivered VT during leak, whereas the VT delivered by two volume-targeted ventilators fell by >50% 12. However, since evaluating the effect of leak was not the primary purpose of the study, the amount of leak was not quantified or varied, and information about peak inspiratory flows during leak was not provided.
Even in the absence of leaks at comparable settings, the five pressure-targeted ventilators in the present study delivered quite different VTs. The factors accounting for these differences included inaccuracies in delivered versus set inspiratory and expiratory pressures and a longer inspiratory duration when the BiPAP was set in the S/T mode, allowing more time for lung inflation. Because of this property, the BiPAP in the S/T mode also exhibited the greatest proportional reduction in delivered VT during leaks. This occurred because the sustained high inspiratory flow during leaks prevented termination criteria for inspiration from being reached for up to 3 s, the default setting. At the set respiratory frequency of 20 breaths·min−1, this led to an inversion of the I:E ratio to as high as 4:1. VT fell because of incomplete emptying of the lung due to shortening of the expiratory time.
In the presence of leaks, pressure-targeted modes manifested even larger differences in delivered VT than without leaks. This was related to the ability to increase inspiratory flow and sustain the target inspiratory pressure as well as the duration (inspiratory time). The ventilators that best compensated for leaks (especially large leaks such as the Quantum) were able to increase inspiratory flow rate up to three-fold and lengthened inspiratory duration slightly (table 2⇑). Excessive prolongation of inspiratory time was counterproductive, however, as discussed above. In addition, by demonstrating the reduction in inspiratory flow that occurs with the Siemens ventilator in the face of large leaks (fig. 3⇑), the results illustrate the importance of not only attaining a high inspiratory flow rate during leaks, but also sustaining it for the duration of the inspiratory phase.
In contrast to the pressure-targeted modes, ventilators delivering volume-targeted modes compensated less well for leaks because inspiratory flow and duration are either fixed (PLV 102) or increase slightly (O'NYX and Siemens). Accordingly, significant drops in delivered VT (at least 50%) during small leaks and more than 80% during large leaks were observed. In addition, prolonging tI/ttot from 0.25 to 0.5 did not increase delivered VT as it did with pressure-targeted modes. In fact, delivered VT fell slightly, presumably because of the greater inspiratory duration and resulting longer leak time. In the presence of a small leak, compensation was achieved by increasing set VT to 2 L as delivered VT was raised to 500, 900 and 1200 mL for the PLV 102, Siemens and O'NYX ventilators, respectively. Nevertheless, this strategy for leak compensation is less effective than using pressure-targeted ventilators. Thus, volume-targeted ventilators would not be the first choice for NPPV in patients with substantial air leaking.
Some of the pressure-targeted ventilators permit setting of the I:E ratio (the BiPAP T mode, 335, and Quantum). Prolonging the inspiratory duration by increasing the set tI/ttot from 0.25 to 0.33 or 0.5 improved leak compensation. However, this compensatory effect depends on the rate of lung filling and emptying and the absolute inspiratory duration. For instance, increasing the tI/ttot beyond 0.25 at a rate of 10 would probably have less effect than at a rate of 20 because the inspiratory duration at the lower rate would allow complete lung filling 13. On the other hand, prolonging the inspiratory time to the point of inverting the I:E ratio is counter-productive at a rate of 20, as exemplified by the drop in delivered VT when the tI/ttot was increased from 0.5 to 0.66 on the BiPAP in the T mode. Other potential clinical consequences of inversion of the I:E ratio include the need for patients to activate their expiratory muscles of respiration in order to cycle into EPAP, air trapping due to inadequate expiratory time, and patient-ventilator asynchrony that contributes to patient discomfort and NPPV intolerance. Thus, the capability of limiting tI/ttot to no more than 50% of the respiratory cycle time can help to prevent excessive shortening of expiratory time. In patients with severe COPD, a shorter tI/ttot may be desirable to enhance patient-ventilator synchrony. Previous studies have shown that relatively high inspiratory flow rates (and hence short inspiratory times) reduce work of breathing during pressure support breathing in these patients 14.
The present results also illustrate the potential adverse consequences of sensitive flow triggers during leaking. At the more sensitive inspiratory trigger settings on the O'NYX and Siemens ventilators, air leaking caused autocycling and marked reductions in delivered VT due to the shortening of inspiratory and expiratory times. Flow-triggered ventilators are susceptible to autocycling in the presence of a leak because the leak flow may be interpreted by the ventilator as the onset of inspiration. In a study evaluating three paediatric flow-triggered intensive care ventilators 15, the relative rate of autocycling by the three ventilators was determined by the sensitivity setting, and all of the ventilators autocycled less frequently at decreased sensitivity settings. The presented findings are consistent with these, with rapid respiratory rates occurring during both small and large leaks. These results indicate that monitoring should be performed during use of ventilators with adjustable inspiratory triggers, and trigger sensitivity should be set to avoid autocycling. It should also be noted that certain ventilators are resistant to autocycling by design. For instance, the BiPAP S/T that did not autocycle in our study, has a higher threshold for inspiratory triggering early compared to later during expiration 16.
As anticipated, increases in respiratory impedance caused by increases in test lung elastance or resistance resulted in reductions in VT delivered by the pressure-limited modes, but had little effect on VT delivered by the volume-targeted PLV 102. Thus, in the setting of increased airway resistance or respiratory system elastance, there are several choices available to the clinician. If leaks are minimal, volume-targeted ventilation can be used to assure delivered VT despite the increased impedance. However, because leaks are so common during NPPV, pressure-targeted ventilators may be preferred to compensate for the leaks as long as ventilator pressures or inspiratory times are adjusted to optimize delivered VT. However, the effectiveness of these latter compensatory actions is limited by the patient's tolerance of increased pressures during NPPV, and the increased pressures could exacerbate the leak 17.
A number of limitations should be borne in mind when interpreting the present results. First, a test lung was used in order to compare the ventilators under identical mechanical conditions, but the observations have not been validated in patients. Second, responses were not tested at a range of back-up rates and pressures and only two leak sizes were used. These choices were made in order to simplify data presentation, and to represent the range of settings commonly encountered clinically 10. The two leaks were selected to range below and above those measured in a previous clinical study 3, and ideal settings were used for NPPV based on recommendations in the literature 8, 9. Third, the effects of leak on ventilator triggering or cycling during spontaneous breathing were not tested, so the observations are most relevant to situations where controlled breathing predominates, such as in neuromuscular patients during sleep 3. Fourth, with advances in technology, newer versions of some of the ventilators tested might perform differently. Finally, one representative ventilator of each type was tested and that individual ventilators may differ was allowed for. The study was designed mainly to illustrate certain response patterns of ventilator modes in the presence of air leaking, and should not be construed as an efficacy comparison between specific ventilators in compensating for leaks. The present results should alert practitioners to the need for calibration and close monitoring of ventilators used for NPPV, and for careful selection of ventilator settings.
In conclusion, pressure-targeted modes maintain delivered VT in the presence of leaks better than volume-targeted modes. Hence, pressure-targeted ventilators are preferred over volume-targeted ventilators to provide more effective NPPV in patients with substantial air leaking. To best compensate for air leaks, pressure-targeted ventilators should have high and sustained maximal inspiratory flow capabilities (>3 L·s−1), adjustable I:E ratios or other mechanisms to limit inspiratory duration so that inversion of the I:E ratio is avoided, and adjustable trigger sensitivities or algorithms to prevent autocycling.
Acknowledgments
The authors thank Respironics and Mallinckrodt Nellcor Puritan Bennett for providing the ventilators. The authors would also like to thank MNPB instruments for providing the test lung.