Abstract
OBJECTIVE:
To determine whether ventilated, low birth weight infants treated with closed versus open tracheal suction in a neonatal intensive care unit (NICU) differ as to airway bacterial colonization, nosocomial pneumonia, bloodstream infection (BSI), incidence and severity of bronchopulmonary dysplasia (BPD), neonatal mortality, frequency of suction, reintubation, and nurse preference.
STUDY DESIGN:
A total of 175 low birth weight infants (≤1250 gm) consecutively born (1997 to 1999), intubated, and ventilated in the delivery room were randomized on admission to the NICU to a closed (Trach Care MAC) or open suction group. Closed multi-use catheters were changed daily; open catheters were changed after every use. Two-pass endotracheal suctioning (both groups) was performed every 8 hours or as needed. Side-port connectors were not used; thus open suction required disconnection from ventilators. Tracheal aspirate cultures were obtained on admission and weekly thereafter. Nosocomial BSI (occurring after 48 hours of life) was documented by positive blood cultures. Radiographs taken before, during, and after tracheal aspirate cultures or BSIs were graded using a semiquantitative system for pneumonia and a modified score for BPD. Nurse preference regarding suction method was recorded.
RESULTS:
Of the original 175 patients, 10 (5 from each group) died and 32 others (16 from each group) were extubated at or before 7 days of life. The study population comprised 67 patients in the closed group and 66 in the open group who were ventilated longer than 1 week. Groups were not statistically different in terms of demographic and clinical characteristics, such as birth weight (837 vs 876 gm), ventilation (27 vs 26 days), and length of stay (49 vs 40 days). Airway colonization with Gram-positive cocci occurred in the majority of patients by 2 weeks of life, regardless of group. A total of 39% of infants in the closed group and 44% of infants in the open group became airway colonized with Gram-negative bacilli; differences were statistically significant. No Gram-negative bacilli species was more likely to be associated with either suction. Nosocomial pneumonia was diagnosed in five patients from each group. Nosocomial BSIs occurred in six closed suction infants and five open suction infants. A comparable number of infants in each group developed severe BPD and were discharged from the hospital on oxygen. A total of 28% of closed suction patients and 27% of open suction patients died. Infants in the closed versus open group were suctioned on average 4.4 and 4.1 times per day and were reintubated 9.7 and 8.6 times per 100 ventilator days, respectively. A total of 40 of 44 NICU nurses considered closed suction to be easier to use, less time-consuming, and better tolerated by the patient.
CONCLUSIONS:
Closed suction obviates the physiological disadvantage of ventilator disconnection without increasing the rate of bacterial airway colonization, frequency of endotracheal suction and reintubation, duration of mechanical ventilation, length of hospitalization, incidence of nosocomial pneumonia, nosocomial BSI, severity of BPD, and neonatal mortality. Although slightly more expensive, closed suction is perceived by nursing staff to be easier, less time-consuming, and better tolerated by small premature infants requiring mechanical ventilation for ≥1 week.
This is a preview of subscription content, access via your institution
Access options
Subscribe to this journal
Receive 12 print issues and online access
$259.00 per year
only $21.58 per issue
Buy this article
- Purchase on Springer Link
- Instant access to full article PDF
Prices may be subject to local taxes which are calculated during checkout
Similar content being viewed by others
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Cordero, L., Sananes, M. & Ayers, L. Comparison of a Closed (Trach Care MAC) With an Open Endotracheal Suction System in Small Premature Infants. J Perinatol 20, 151–156 (2000). https://doi.org/10.1038/sj.jp.7200330
Published:
Issue Date:
DOI: https://doi.org/10.1038/sj.jp.7200330
This article is cited by
-
Closed versus open endotracheal tube suction in mechanically ventilated neonates: a randomized controlled trial
European Journal of Pediatrics (2022)
-
The effects of closed tracheal suctioning plus volume guarantee on cerebral hemodynamics
Journal of Perinatology (2011)
-
Beatmungsassoziierte Pneumonie bei Kindern
Monatsschrift Kinderheilkunde (2011)
-
Experimental Contamination of a Closed Endotracheal Suction System: 24 h vs 72 h
Infection (2009)
-
Empfehlung zur Prävention nosokomialer Infektionen bei neonatologischen Intensivpflegepatienten mit einem Geburtsgewicht unter 1500 g
Bundesgesundheitsblatt - Gesundheitsforschung - Gesundheitsschutz (2007)