Accuracy of autotitrating CPAP to estimate the residual Apnea–Hypopnea Index in patients with obstructive sleep apnea on treatment with autotitrating CPAP

  • Original Article
  • Published:
Sleep and Breathing Aims and scope Submit manuscript

Abstract

Objective

Autotitrating continuous positive airway pressure (auto-CPAP) devices now have a smart card (a pocket-sized card with embedded integrated circuits which records data from the CPAP machine such as CPAP usage, CPAP pressure, large leak, etc.) which can estimate the Apnea–Hypopnea Index (AHI) on therapy. The aim of this study was to determine the accuracy of auto-CPAP in estimating the residual AHI in patients with obstructive sleep apnea (OSA) who were treated with auto-CPAP without a CPAP titration study.

Patients and Methods

We studied 99 patients with OSA from April 2005 to May 2007 who underwent a repeat sleep study using auto-CPAP. The estimated AHI from auto-CPAP was compared with the AHI from an overnight polysomnogram (PSG) on auto-CPAP using Bland–Altman plot and likelihood ratio analyses. A PSG AHI cutoff of five events per hour was used to differentiate patients optimally treated with auto-CPAP from those with residual OSA on therapy.

Results

Bland and Altman analysis showed good agreement between auto-CPAP AHI and PSG AHI. There was no significant bias when smart card estimates of AHI at home were compared to smart card estimates obtained in the sleep laboratory. An auto-CPAP cutoff for the AHI of six events per hour was shown to be optimal for differentiating patients with and without residual OSA with a sensitivity of 0.92 (95% confidence interval (CI) 0.76 to 0.98) and specificity of 0.90 (95% CI 0.82 to 0.95) with a positive likelihood ratio (LR) of 9.6 (95% CI 5.1 to 21.5) and a negative likelihood ratio of 0.085 (95% CI 0.02 to 0.25). Auto-CPAP AHI of eight events per hour yielded the optimal sensitivity (0.94, 95% CI 0.73 to 0.99) and specificity (0.90, 95% CI 0.82 to 0.95) with a positive LR of 9.6 (95% CI 5.23 to 20.31) and a negative LR of 0.065 (95% CI 0.004 to 0.279) to identify patients with a PSG AHI of  ≥ 10 events per hour.

Conclusion

Auto-CPAP estimate of AHI may be used to estimate residual AHI in patients with OSA of varying severity treated with auto-CPAP.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Young T et al (2008) Sleep disordered breathing and mortality: eighteen-year follow-up of the Wisconsin sleep cohort. Sleep 31(8):1071–1078

    PubMed  Google Scholar 

  2. Young T et al (1993) The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med 328(17):1230–1235. doi:10.1056/NEJM199304293281704

    Article  CAS  PubMed  Google Scholar 

  3. Kushida CA et al (2006) Practice parameters for the use of continuous and bilevel positive airway pressure devices to treat adult patients with sleep-related breathing disorders. Sleep 29(3):375–380

    PubMed  Google Scholar 

  4. Littner MR et al (2005) Practice parameters for clinical use of the multiple sleep latency test and the maintenance of wakefulness test. Sleep 28(1):113–121

    PubMed  Google Scholar 

  5. Morgenthaler TI et al (2008) Practice parameters for the use of autotitrating continuous positive airway pressure devices for titrating pressures and treating adult patients with obstructive sleep apnea syndrome: an update for 2007. An American Academy of Sleep Medicine report. Sleep 31(1):141–147

    PubMed  Google Scholar 

  6. Cross MD et al (2006) Comparison of CPAP titration at home or the sleep laboratory in the sleep apnea hypopnea syndrome. Sleep 29(11):1451–1455

    PubMed  Google Scholar 

  7. Masa JF et al (2004) Alternative methods of titrating continuous positive airway pressure: a large multicenter study. Am J Respir Crit Care Med 170(11):1218–1224. doi:10.1164/rccm.200312-1787OC

    Article  PubMed  Google Scholar 

  8. Woodson BT et al (2003) Nonattended home automated continuous positive airway pressure titration: comparison with polysomnography. Otolaryngol Head Neck Surg 128(3):353–357. doi:10.1067/mhn.2003.35

    Article  PubMed  Google Scholar 

  9. Bonnet M, Carley D, Carskadon M et al (1992) EEG arousals: scoring rules and examples: a preliminary report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders Association. Sleep 15(2):174–184

    Google Scholar 

  10. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310

    CAS  PubMed  Google Scholar 

  11. Hanley JA, McNeil BJ (1982) The meaning and use of the area under a receiver operating characteristic (ROC) curve. Radiology 143(1):29–36

    CAS  PubMed  Google Scholar 

  12. El Solh AA, Grant BJB (1996) A comparison of severity of illness scoring systems for critically ill obstetric patients. Chest 110:1299–1304. doi:10.1378/chest.110.5.1299

    Article  CAS  PubMed  Google Scholar 

  13. Ruxton GD (2006) The unequal variance t-test is an underused alternative to Student’s t-test and the Mann–Whitney U test. Behav Ecol 17:688–690. doi:10.1093/beheco/ark016

    Article  Google Scholar 

  14. Katz D (2001) Clinical epidemiology and evidence-based medicine: a primer for the clinician. Sage, Thousand Oaks, pp 57–62

    Google Scholar 

  15. Kushida CA et al (2008) Clinical guidelines for the manual titration of positive airway pressure in patients with obstructive sleep apnea. J Clin Sleep Med 4(2):157–171

    PubMed  Google Scholar 

  16. Torre-Bouscoulet L et al (2008) Autoadjusting positive pressure trial in adults with sleep apnea assessed by a simplified diagnostic approach. J Clin Sleep Med 4(4):341–347

    PubMed  Google Scholar 

  17. Stammnitz A et al (2004) Automatic CPAP titration with different self-setting devices in patients with obstructive sleep apnoea. Eur Respir J 24(2):273–278. doi:10.1183/09031936.04.00074304

    Article  CAS  PubMed  Google Scholar 

  18. Baltzan MA et al (2006) Prevalence of persistent sleep apnea in patients treated with continuous positive airway pressure. Sleep 29(4):557–563

    PubMed  Google Scholar 

  19. Pittman SD et al (2006) Follow-up assessment of CPAP efficacy in patients with obstructive sleep apnea using an ambulatory device based on peripheral arterial tonometry. Sleep Breath 10(3):123–131. doi:10.1007/s11325-006-0058-x

    Article  PubMed  Google Scholar 

  20. Becker HF et al (2003) Effect of nasal continuous positive airway pressure treatment on blood pressure in patients with obstructive sleep apnea. Circulation 107(1):68–73. doi:10.1161/01.CIR.0000042706.47107.7A

    Article  PubMed  Google Scholar 

  21. Peker Y et al (2002) Increased incidence of cardiovascular disease in middle-aged men with obstructive sleep apnea: a 7-year follow-up. Am J Respir Crit Care Med 166(2):159–165. doi:10.1164/rccm.2105124

    Article  PubMed  Google Scholar 

  22. Rigau J et al (2006) Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices. Chest 130:350–361. doi:10.1378/chest.130.2.350

    Article  PubMed  Google Scholar 

  23. Series F et al (2008) Reliability of home CPAP titration with different automatic CPAP devices. Respir Res 9:56

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Work was performed at: Western New York Veteran Affairs Healthcare System, Buffalo, New York.

Himanshu Desai, M.D., has no financial or personal conflict of interest in presenting this manuscript.

Anil Patel, M.D., has no financial or personal conflict of interest in presenting this manuscript.

Pinal Patel, M.B.B.S., has no financial or personal conflict of interest in presenting this manuscript.

Brydon J. B. Grant, M.D., has no financial or personal conflict of interest in presenting this manuscript.

M Jeffrey Mador, M.D., has no financial or personal conflict of interest in presenting this manuscript.

Funding

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. Jeffery Mador.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Desai, H., Patel, A., Patel, P. et al. Accuracy of autotitrating CPAP to estimate the residual Apnea–Hypopnea Index in patients with obstructive sleep apnea on treatment with autotitrating CPAP. Sleep Breath 13, 383–390 (2009). https://doi.org/10.1007/s11325-009-0258-2

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11325-009-0258-2

Keywords

Navigation