1. Position the tip of the bronchoscope close to the segmental area corresponding to radiographic infiltrates. |
2. Advance the PSB 3 cm out of the distal tip of the bronchoscope. |
3. Push the inner cannula of the PSB to eject the polyethylene glycol plug. |
4. Wedge the brush in the subsegmental area or sample secretions if these are visualized |
5. Retract the brush into the inner cannula and the inner cannula into the outer cannula, and remove the PSB from the bronchoscope |
6. Once the PSB is out of the bronchoscope, the distal portion of the inner cannula is wiped with 70% alcohol solution. |
7. The brush is advanced and cut with sterile scissors into a sterile solution containing 1 mL of Ringer's lactate or saline. |
8. The tube with the PSB and Ringer's lactate or saline solutions is submitted immediately to the microbiology laboratory for processing. |