Table 11– Specification of helium meters used for measuring functional residual capacity
ItemSpecification
Linearity0–10% helium
Accuracy±l% of initial deflection
Resolution<0.05% He
Gas flowconstant >200 mL·min−1, gas to be free of CO2 and H2O
Response time95% response time <15 s (meter + spirometer)