Abstract
The American Thoracic Society/European Respiratory Society Task Force underlined that the use of in-line filters during respiratory function tests “is an area of controversy”. The aim of the present study was to measure the contamination occurring during forced expiration downstream from a screen pneumotachograph (SP) with and without an in-line filter (Pall PF30S). A total of 40 healthy subjects performed eight consecutive maximal expiratory manoeuvres into four sterile apparatuses (A1: no filter, no SP; A2: filter-only; A3: SP-only; A4: filter and SP) in random order. A blood agar plate was fixed downstream from the apparatus. Colony-forming units (CFUs) were counted after 24 h incubation at 37°C. Of the 40 plates obtained with each apparatus, 13 were sterile with A1 (range 0–679 CFUs), 25 with A2 (0–49 CFUs), 30 with A3 (0–35 CFUs) and 39 with A4 (one CFU in the only positive plate). A1 versus A2 and also A3 versus A4 gave different values for the CFU number, but A2 and A3 showed similar contamination levels.
The authors conclude that: 1) the in-line filter does not perform better than a screen pneumotachograph; 2) it does not eliminate the need to decontaminate the pneumotachograph; and 3) equipment placed downstream from an in-line filter and a screen pneumotachograph is almost protected from contamination.
Public and private healthcare providers are increasingly concerned over control of hospital infections and the limitation of cross-contamination. This preoccupation is particularly acute in cystic fibrosis patients who are colonised with respiratory pathogens 1, 2 and in whom pulmonary function tests are routinely performed 3. Although various circumstantial evidence points to contaminated spirometers being possibly involved in cross-contamination 4, 5, there is no direct or even indirect evidence that infection transmission occurs with the use of flow-type systems, which seem less susceptible to bacterial contamination than water-sealed spirometers 6.
Recommendations regarding hygiene in respiratory function testing were restated recently by the American Thoracic Society (ATS)/European Respiratory Society (ERS) Task Force 7. Prevention of transmission by direct contact can be achieved by disinfecting after each use (or discarding if disposable) mouthpieces, nose clips and any other equipment coming into direct contact with mucosal surfaces. This recommendation appears easy to follow. Transmission of infection can also occur by indirect contact through aerosol droplets, therefore, the ATS/ERS committee recommends that any equipment surface showing visible condensation from expired air should be disinfected or sterilised before re-use. Proximal tubing is easily interchangeable on volume-type spirometers, where contamination occurs most frequently 8, 9, and on certain flow-type pneumotachometers. However, as pointed out by the ATS/ERS Task Force, spirometric equipment also sometimes employs complex valve manifolds whose internal surfaces are exposed to aerosol deposition from expired air. Such equipment is not easy to decontaminate and in some cases it is unrealistic to recommend disinfection after each use. At present, the optimal frequency for disinfection or sterilisation of tubing, valves and manifolds has still not been established.
The appropriate placement of filters with ideally 100% effectiveness for trapping pathogenic organisms from exhaled air during pulmonary function testing would thus appear to be the most practical method for eliminating transmission of pathogenic organisms between patients 10, and, inasmuch as in-line filters have been shown to be effective in removing pathogens from the expiratory flow 11 without affecting the clinical utility of respiratory function tests 3, 12–14, their use might be encouraged in the clinical laboratory. Patient advocate groups, with the help of their medical committees, have published recommendations including the mandatory use of in-line filters during spirometry, and many laboratory managers have generalised the use of such filters. For instance, a survey of the 19 respiratory function laboratories of the public hospitals of Paris (France) showed that 16 out of the 17 laboratories answering the questionnaire were already using in-line filters for spirometry in 1996, just four years after they came onto the market 15. However, as explicitly mentioned by the ATS/ERS Task Force 7 and confirmed by a recent exchange of letters 16–18, “the use of such filters is an area of controversy”. The ability of these filters to remove bacteria from the high flow produced by true 19 or simulated 20 forced expiration has been seriously questioned and the use of in-line filters does not eliminate the need for regular cleaning and decontamination of lung function equipment. Furthermore, combined with a classical screen pneumotachograph, the least efficient (in terms of bacterial retention) and least resistive of four commercially available specially designed filters 20 already reaches the upper limit of resistance to flow set by the ATS/ERS Task Force guidelines for spirometry measurement 12, 21. Finally, while the reduction in filter-related respiratory parameters is clinically unimportant, it nevertheless depends on the resistance to air flow 22 and is presumably correlated to filtration efficiency. Filters have been shown to reduce forced expiratory volume in one second 12, 14, forced vital capacity 14 and peak expiratory flow 12, 14, indicating that in research and in multicentre studies especially, results obtained with and without a filter should be analysed separately 14.
The most recent lung function equipment is built around a flow-measuring device (usually connected to a pressure transducer) held close to the mouth and so theoretically, although not necessarily 6, it is more exposed to contamination than the inner surface of the bellow of a volume-type spirometer. Manufacturers have developed sensors that are now easier to decontaminate, if not disposable. As standard spirometry can be carried out with just this one sensor (connected to the electronic equipment), it could conceivably be replaced for each patient, or at least for those most at risk from cross-infection, provided that the calibration procedures are not too time-consuming. As pointed out by the ATS/ERS Task Force, difficulties arise when the flow-type spirometer is connected to a more complex system such as those used for plethysmography or gas transfer measurements, where decontamination is a problem. Screen pneumotachographs (Lilly type) are widely used as flow-measuring devices. Made of a fine wire mesh screen with low resistance to air flow, the screen pneumotachograph might also be useful in protecting complex equipment from contamination by pathogens.
The aim of the present study was to measure the natural contamination occurring during forced expiration downstream from a screen pneumotachograph with and without an in-line filter.
MATERIAL AND METHODS
Material
A special coupling device (fig. 1⇓) was made comprising of the following three parts. 1) A screw lid in which a hole has been drilled and a connector has been glued. Two identical lids with different connectors were made because the diameters of the pneumotachograph and the filter were slightly different. 2) A corresponding PVC tubing in which 16 holes (0.8 cm in diameter) had been drilled to allow forced expiration. 3) A plastic ring used to maintain the agar plate. The plane of the agar plate was at a distance of 1.5 cm from the centres of the holes. This device, designed to hold an agar plate, was built in order to test natural contamination occurring during forced expiration in four different situations: in open air with (A2) and without (A1) a filter and downstream from a screen pneumotachograph (fig. 2⇓) with (A3) and without (A4) a filter.
Four apparatuses (A1 to A4) were used to represent the four situations tested (fig. 3⇓), these consisted of the following. A1: Mouthpiece (MP), elbow connector (EC); L = 16.1 cm; A2: MP, in-line filter (Pall PF30SJE; Pall Corporation, East Hills, NY, USA), EC; L = 23 cm; A3: MP, EC, screen pneumotachograph (Jaëger PT36; Jaëger, Wursburg, Germany); L = 23.4 cm; and A4: MP, in-line filter, EC, screen pneumotachograph; L = 30.3 cm; with L equalling distance between the mouthpiece opening and the surface of the agar plate.
Note that the distances between the MP opening and the surface of the agar plate are similar in A2 and A3. The EC is a standard connector used with Jaëger Systems (Ref: 852352; Jaëger). The MP is a disposable plastic one. The Pall PF30SJE filter is the same as the widely used Pall PF30S with connector dimensions adapted to the Jaëger pneumotachograph. This filter was chosen because, whilst one of the commercially available filters offering least resistance, when combined with the screen pneumotachograph used in the current study, it already reaches the limit of resistance recommended by the ATS/ERS Committee 12. The screen pneumotachograph was used nonheated. A trypticase soy agar plate with 5% horse blood was fixed downstream from the apparatus in the coupling device (fig. 1⇑).
The expired gas was flushed along the agar plate, and driven through the 16 holes whose total surface (8 cm2) was higher than the minimum surface of the connection tubing (5 cm2) ensuring that the expiration was not limited by the device.
Subjects and protocol
A total of 40 healthy subjects, students of the Faculty of Medicine, University of Caen (Caen, France), or members of staff, with no history of respiratory disease, gave their informed consent and performed eight consecutive maximal expiratory manoeuvres from total pulmonary capacity into each apparatus, with a 30-s pause between expirations. The study was approved by the regional ethical committee (Consultation Committee for the Protection of People in Biomedical Research, Basse-Normandy). All subjects were trained for pulmonary function testing. They were respiratory technicians, hygienists or medical doctors of the hospital or medical students in pneumology working in the laboratory. The number of expirations was chosen in line with the ATS/ERS Task Force recommendation, as eight manoeuvres is generally a practical upper limit for most subjects 21. Immediately before the series of expirations, all the components of the apparatus were reassembled together with a fresh agar plate. All parts (except the pneumotachograph screen) were wiped with alcohol and allowed to dry immediately before reassembly. Handling was done on a table disinfected with Surfanios® (Laboratoire Anios, Hellemmes, France). A dry sterile metallic screen (120°C sterilisation) was inserted in the pneumotachograph (A3 and A4) and a new filter was used (A2 and A4). The order of use of the apparatus was randomised according to a Latin square design with 10 identical repetitions. Colony-forming units (CFUs) were counted after 24 h of incubation at 37°C under inspection with a magnifying glass.
Statistical analysis
The plates were defined as negative (no contamination) or positive (≥1 CFU) and a Cochran Q test was performed between the four situations. Individual comparisons between apparatuses were carried out with a McNemar's test. The number of CFUs in the four apparatuses were also compared with a nonparametric test (Friedman test), because the data was not normally distributed; individual comparisons were performed with a Wilcoxon paired-sample test. All the CFUs were counted. However the highest count (679 CFU in one A1 plate) is approximate as some of the CFU were confluent.
The difference between A1 and A2 represents the filter's ability to prevent contamination of the pneumotachograph. The difference between A3 and A4 represents the filter's ability to prevent contamination of the equipment placed downstream to the screen pneumotachograph. The difference between A2 and A3 represents the screen pneumotachograph's efficacy as a respiratory filter as compared with the Pall PF30SJE. A p-value <0.05 was considered significant.
RESULTS
Visual inspection of the blood agar plates in situation A1 showed that vapour condensation occurred on the entire surface of the plate. A summary of the results is given in figure 4⇓.
Of the 40 plates obtained with each apparatus, 13 were sterile with A1 (range 0–679 CFUs), 25 with A2 (range 0–49 CFUs), 30 with A3 (range 0–35 CFUs) and 39 with A4 (p<0.001; Cochran Q test). A1 produced significantly fewer sterile plates than A2 (p<0.01; McNemar's test) as did A3 versus A4 (p<0.01), A2 versus A4 (p<0.001) and A1 versusA3 (p<0.001) but there was no significant difference between A2 and A3 (p = 0.07).
The five plates with >150 CFUs (679, 342, 279, 277 and 240) were obtained with A1. The number of CFUs was affected by the apparatus (p<0.001; Friedman test). A1 had significantly more CFUs than A2 (p<0.001; Wilcoxon paired-sample test), as did A3 versus A4 (p<0.01), A1 versus A3 (p<0.001) and A2 versus A4 (p<0.001) but the contamination in A2 and A3 was not significantly different (p = 0.97). There was only one CFU in the one positive plate of A4. For this subject, there were 679 CFUs in A1, 21 CFUs in A2 and 31 CFUs in A3.
DISCUSSION
The present results indicate that with healthy subjects performing maximal expirations according to the ATS/ERS standards for studying the expiratory flow–volume curve, contamination occurs downstream from an in-line filter supposed to prevent cross-contamination during pulmonary function testing. These results are in accordance with the study of Leeming et al. 19 who showed that the same filter allowed the transmittance of approximately one-third of expired CFUs in 37 patients with respiratory diseases, the highest CFU count being obtained in cystic fibrosis patients. Accordingly, as pointed out by the ATS/ERS Task Force: “The use of in-line filters does not eliminate the need for regular cleaning and decontamination of lung function equipment” 7.
Despite the fact that all subjects were free from current or past respiratory disease, the current authors observed a large inter-subject variability in the number of CFUs, with five subjects producing >150 CFUs (one producing >500) in the control condition while 13 subjects produced no contamination of the agar plate. Such variability has already been mentioned 11, 23 and may be linked to high inter-subject variability in the number of small-size particles produced during mouth breathing 24, possibly relating to the physical properties of the lung lining fluid 25. Several mechanisms have been described to explain the retention properties of filters in terms of particle size. Direct interception filters large particles (>1 µm), inertial impaction removes smaller particles (0.5–1.0 µm in diameter) by collision within the filter material and very small particles, because of their Brownian motion, can be trapped by diffusional interception. Other properties, hydrophobic or electrostatic, are held to improve a filter's efficiency.
Contamination after the filter is similar to that observed downstream from a screen pneumotachograph whose resistance to flow is in the same order of magnitude. Therefore, the Pall filter is no better than a screen pneumotachograph. The resistance of the Pall PF30S was measured as 0.56 cm·H2O·s·L−1 at 12 L·s−1 12, close to the value given by the manufacturer (0.6 cm·H2O·s·L−1). The resistance of the same screen pneumotachograph used in the same setting (non-heated) can be estimated from the graphical representation at ∼0.7 to 0.8 cm·H2O·s·L−1 at 12 L·s−1 22. The pneumotachograph used was nonheated in order to keep a similar manipulation time for the four apparatuses. However, a high temperature around the proximal tubing could inhibit bacterial growth 6; therefore, heating the pneumotachograph, while not changing its efficiency as a bacterial filter, could reduce the bacterial growth and the risk of inter-subject contamination.
In one experimental design, the SensorMedics MicroGard® Filter (SensorMedics Corp., Yorba Linda, CA, USA) and the KOKO MOE® filter (KK; Pulmonary Data Service Instrumentation, Inc., Louisville, CO, USA) have demonstrated greater bacterial removal efficiency than the Pall PF30S 20. However, the measured resistance of such filters is found to be slightly higher 26 and the experimental design is far from the conditions of natural contamination occurring in pulmonary function testing. In particular, the droplets produced by this experimental condition (3–5 µm) are much larger than the droplets (<1 µm) produced by normal breathing 25 and even coughing 24. During standard spirometry, the SensorMedics filter was considered no different from the Pall filter 27.
The controversy over using in-line filters for lung function testing resembles the one over the use of filters in mechanical ventilation 28: it is a balance between the advantages and disadvantages. Several issues must be considered; some, but not all of which are related to hygiene. 1) As there is substantial inter-individual variability in the number of CFUs expired during forced expiration, the potential for any one person to contaminate the equipment is unpredictable. 2) In-line filters reduce but do not eliminate contamination of the downstream equipment. 3) The ability of in-line filters to prevent cross-contamination has not been demonstrated; however, neither has it been demonstrated that cross-infection has occurred due to the use of pulmonary function testing equipment. 4) In-line filters introduce an undesirable resistance to flow. 5) Use of an in-line filter has a statistical effect on the results of certain lung function parameters, at least in those nonheated pneumotachographs. 6) The widely used pneumotachograph (screen or Fleisch type) may be subject to mucus deposition or vapour condensation that could considerably alter the flow and volume measurements, giving rise to a gross overestimation of respiratory parameters 29. 7) In-line filters can behave as heat and moisture exchange filters 30. Thus, the body, temperature, ambient pressure, saturated with water vapour (BTPS) corrections factor for the expiratory gas should be adapted to the actual gas condition in the pneumotachograph resulting from the use of the filter and its heat and moisture exchange characteristics. The error might be as large as 10% 21 and will not be taken in account when calibrating with the filter in place. At the present time, this aspect would seem not to have been taken into account by the lung equipment manufacturers, as for most equipment the presence of a filter is at best mentioned to correct the resistance measurements or the dead space of the equipment. Different heat and moisture exchange characteristics and resulting BTPS correction or different methods of BTPS correction may explain why, independently of the resistance, spirometry results are 12, 13, or are not 3, affected by the presence of a filter. One advantage of this potential characteristic is that the need to heat the pneumotachograph to prevent vapour condensation can be reconsidered 31.
In conclusion, the present study shows that, combined with the screen pneumotachograph, a low resistance filter virtually protects complex downstream equipment from contamination. This leads us to believe that in clinical situations where such contamination of complex equipment could occur, the use of an in-line filter might be justified, provided that measurements remain accurate, that the question of body, temperature, ambient pressure, saturated with water vapour correction is solved, while keeping in mind that the pneumotachograph must be decontaminated. Conversely, such a filter seems to have limited efficiency in terms of bacterial retention, no better in fact than the pneumotachograph itself. In the absence of complex equipment, the current authors therefore believe that, where necessary, decontamination of the pneumotachograph is a preferable solution, if possible, unless other advantages of the filter in terms of mucus retention or heat exchange are demonstrated.
Acknowledgments
The authors would like to thank A. Lemarié and Y. Liégard (University Hospital of Caen, Caen, France) for technical assistance throughout this project and to P. Thibon (University Hospital of Caen, Caen, France) for helpful advice in the statistical analysis.