Copyright ©ERS Journals Ltd 2007 Bronchoscopy-guided radiofrequency ablation as a potential novel therapeutic tool1 First Dept of Internal Medicine and 2 Dept of Anesthesiology, Shinshu University School of Medicine, Asahi Matsumoto, Japan. CORRESPONDENCE: K. Tsushima, First Department of Internal Medicine, School of Medicine, Shinshu University, 3-1-1 Asahi Matsumoto Nagano 390-8621, Japan. Fax: 81 263363722. E-mail: tsushima{at}hsp.md.shinshu-u.ac.jp Keywords: Fibreoptic bronchoscopy, internal cooled-radiofrequency ablation, radiofrequency ablation, standard noncooled-radiofrequency ablation
Received: August 26, 2006
The aim of the current study was to assess the safety of bronchoscopy-guided radiofrequency ablation (RFA) and compare the effectiveness between new internal cooled-RFA and standard noncooled-RFA. Normal lungs from sheep were used (n = 6). Internal cooled-RFA and standard noncooled-RFA were set to assess the most suitable RFA conditions, such as power output, flow rate and ablation time. Internal cooled-RFA was then applied under the most optimal conditions of power output and flow rate for 15, 30, 60 and 120 s, and two water temperatures either room temperature (RT) water or cold water. Criteria for the most appropriate conditions were set over 15 s of ablation time and 50°C of the tip's temperature. Internal cooled-RFA had no complications. Standard noncooled-RFA was complicated with bronchial bleeding after RFA. On the basis of the histological findings, average temperature and average output, the most appropriate conditions of the cooled-RFA were a power output of 30 W and flow rate of 30 or 40 mL·min1. The cooled-RFA using cold water caused a smaller, more discrete, lesion compared with that using RT water. Bronchoscopy-guided internal cooled-radiofrequency ablation was an effective, safe and feasible procedure that could become a potential therapeutic tool in managing lung pathology. Radiofrequency ablation (RFA) uses an electromagnetic wave with a frequency band the same as an electric scalpel used in surgery and a radiofrequency interchange electric current. RFA is a new treatment offered as a minimally invasive treatment and is most commonly used to treat patients with liver tumours 1, kidney cancer 2, breast cancer 3 or lung cancer 4. Tumour cells are generally more susceptible to heat than normal cells. This procedure is safe, technically feasible and suitable for use with imaging guidance technologies, such as ultrasonography, radiograph and computed tomography (CT). Mortality from lung cancer remains high and establishment of new therapeutic modalities are needed. RFA has been described as a new technique for the treatment of lung cancer, and Miao et al. 5 suggested that cooled-RFA is an effective alternative to lobectomy, in certain patients, for the minimally invasive treatment of lung cancer. In the USA, standard RFA has been more widely used than percutaneous imaging-guided therapy 6. Complications from this method include pain, pneumothorax, haemothorax and pleural effusion 7. Tsushima et al. 8 reported fibreoptic bronchoscopy-guidance under real-time CT fluoroscopy for the diagnosis of peripheral lung lesions. A critical aspect is that this procedure may reduce the above complications if RFA can be applied using real-time chest CT images under fibreoptic bronchoscopy. The current authors developed a new internally cooled electrode (Japan Application No. 2006-88228) suitable for forceps channel bronchoscopy. The initial aim of the present study was to compare the effective differences between the new internal-cooled electrode and the standard noncooled electrode and to assess the feasibility and safety of bronchoscopy-guided RFA using normal sheep lungs.
Ethical considerations The study protocol was approved by the Institutional Review Board for the care of animals at Shinshu University (Asahi, Matsumoto, Japan). The care and handling of animals was in accordance with the guidelines of the National Institute of Health. The animals had free access to commercial food and drinking water.
Preparation
Fibreoptic bronchoscopy
RFA preparation
RFA procedure Various conditions were set to assess the most suitable RFA condition according to the methods of cardiac conduction abnormality for procedure 1 911. An X-ray fluoroscopic image was obtained to correlate the location of the electrode tip in peripheral lung lesions with 1-cm margin from pleura to prevent pneumothorax complications. Once the appropriate peripheral location of the electrode tip was confirmed and impedance was <300 , RFA was used. The settings regarding the areas which were ablated were as follows.
Procedure 1: To compare between standard noncooled-RFA and internal cooled-RFA, standard noncooled-RFA (power output; 30 W) ablated each right lobe only once and internal cooled-RFA (power output; 30W, flow rate 30mL·min1) ablated each left lobe only once. RFA was initially applied for 120 s under several conditions of power output and flow rate (table 1
Procedure 21: To obtain an optimal output of 30 W, several sets of flow rates were performed. Internal cooled-RFA with flow rates of 20, 30 and 40 mL·min1 were ablated once in the upper, middle and lower lobes, respectively. One sheep was harvested immediately and another harvested 7 days after ablation. Procedure 22: Different output (20 W and 40 W) and flow rate (20, 40 and 50 mL·min1) parameters were used to compare the data at 30 W. Internal cooled-RFA was ablated once in each right lobe and left lobe using a power output of 20 and 40 W, respectively. Lungs obtained from one sheep were harvested at 7 days after ablation. Procedure 3: Internal cooled-RFA was applied under the most optimal conditions of power output and flow rate for 15, 30 and 60 s, using two different temperatures of flow water (room temperature water or cold (4°C) water). Internal cooled-RFA was used once for 15, 30 and 60 s in the upper, middle and lower lobes, respectively. Two sheep were harvested 7 days after ablation. Internal cooled-RFA was performed on the bronchial mucosa using the same methods as described previously.
Histopathological analysis
Data analysis
As shown in figure 2 . Macroscopic findings showed bronchial haemorrhage and necrosis of 25x15 mm. Standard noncooled-RFA showed diffuse bronchial bleeding from all ablated lobes. All histological findings showed diffuse alveolar haemorrhage, lung oedema and destruction of the alveolar space. It was concluded that, despite achieving a sufficient area of coagulation necrosis, this electrode was not suitable for lung ablation because of the increase of impedance and bronchial haemorrhage.
Internal cooled-RFA had no immediate complications. As shown in figure 3 . Histological findings showed diffuse alveolar haemorrhage and preserved structure of the alveolar space. With a flow rate of 40 mL·min1, histological findings did not differ from those with a flow rate of 20 mL·min1 (table 2
As shown in figure 4 . Macroscopic findings showed necrosis of 15x15 mm, and no bronchial bleeding. Histological findings showed diffuse alveolar haemorrhage and necrosis. With 30 W power output and a flow rate of 20 mL·min1, sufficient power output could not be achieved because of the rapid increase in the tip's temperature; however, histopathological findings showed the same lesion and size as with a flow rate of 40 mL·min1. The findings between a flow rate of 30 and 40 mL·min1 were almost identical (table 2 . Histological findings showed diffuse alveolar haemorrhage, lung oedema and destruction of the alveolar space. A power output of 40 W achieved almost the same size of coagulation necrosis as 30 W, for clinical safety; therefore, a 40 W power output is suggested to be unnecessary for ablation of the lung tissue. On the basis of these histological findings and the average temperature and output, the most appropriate conditions for the cooled electrode were a 30 W power output and a flow rate of 30 or 40 mL·min1. In addition, the temperature of the electrode tip needed to ablate the lung tissue was 50°C.
As shown in figure 5 7 mm in diameter after 15 s of ablation. There was a major difference histologically between cold water and room temperature water. The macroscopic findings with room temperature water showed larger burn lesions than with cold water (40x45 and 20x15 mm, respectively) after 60 s of ablation; however, after 30 s of ablation, the burn size was almost identical between room temperature and cold water (11x7 mm and 7x5 mm, respectively). From these histological results, a 30-s exposure of RFA affected deeper lung tissues similarly, independent of the temperature of water used.
On ablation of the bronchial membrane, a sufficient necrotic effect was achieved with white denaturation of the mucous membrane within 10 s using a cooled-RFA with cold water. No bronchial perforation was observed within 30 s of ablation; however, since impedance increased faster compared with the peripheral lung tissue, RFA could not continue for >30 s. This was due to contact resistance around the electrode tip.
Percutaneous guided-RFA has found clinical applications for lung cancer with good results reported 47, 12. Since the electrode is placed percutaneously directly into the tumour, under cross-sectional imaging guidance such as chest CT, complications such as pneumothorax occur with frequency. The frequency of pneumothorax or haemothorax using a percutaneous method in a rabbit model was 47% 13 and clinical complications occur in 1635% of percutaneously directly RFA 6, 7. However, it is possible to avoid these complications using the present fibreoptic bronchoscopy guidance; this is the greatest advantage of the authors internal cooled electrode. Fiberoptic bronchoscopy-guided cooled-RFA is both safe and technically feasible as the tip of the electrode is confirmed with CT or radiographic imaging guidance. As the present study has shown, cooled-RFA had no complications, such as bronchial bleeding or pneumothorax, and can be used with CT or radiographic imaging guidance technologies. Therefore, to the current authors knowledge, the present study is the first to demonstrate fibreoptic bronchoscopy-guided cooled-RFA as a potential therapeutic tool. The peripheral lung is referred to as the safety zone; however, the anatomically central and middle bronchi are accompanied by the pulmonary artery branches. One complication anticipated with bronchoscopy is bronchial bleeding. The authors tried to ablate several areas of lesions from peripheral lung tissues to central lung tissues using power output of 30 W and flow rate of 40 mL·min1 (data not shown). On day 7 after ablation there were no complications of macro bronchial haemorrhage from the middle or central artery or atelectasis. Therefore, the authors confirm that cooled-RFA achieved coagulation and haemostasis in the normal lung tissues although the safety in other lung tissues that may bleed more than normal lungs needs to be assessed. This internal cooled-RFA is constructed as a thin catheter (diameter 1.67 mm) to pass through the bronchoscopy channel. The necrotic size obtained by this electrode is a critical consideration for this method. To achieve higher power output from the electrode and sufficient coagulation necrosis, it was necessary to increase the power output and to reduce impedance around the electrode tip. Using the standard noncooled electrode, the temperature around the tip rose rapidly and the pop phenomenon occurred in the lung tissues. This phenomenon means that coagulated necrotic tissue is formed around the electrode tip and tissue impedance increases rapidly. To avoid this phenomenon, the electrode tip should be cooled using water. On the basis of these results using a cooled-RFA for cardiac conduction disease 14, 15, it is predicted that the lesion occurs in lung tissues away from the cooled-RFA. The cooled-RFA enables greater power output for a longer time compared with the standard noncooled-RFA, resulting in larger coagulation necrosis area. The electrode tip is cooled by circulating water in the electrode catheter. As a result, the tissue temperature just around the electrode tip does not reach an excessive temperature. Based on this theory, the cooled-RFA can reach deeper and wider areas of ablation using the same power output. Therefore, as shown in the current results, the cooled-RFA could obtain coagulation necrosis compared with the standard noncooled-RFA, and the cooled-RFA using circulating room temperature water could obtain larger burn lesions than that using circulating cold water. RFA is regarded as suitable for use on lung cancer as the radiofrequency energy to a tumour relies on the insulation effect by the existence of normal surrounding lungs 5. However, surgery is recognised as the standard treatment for patients with localised primary lung cancer and metastatic resectable lesions. A minimally invasive procedure such as RFA is usually considered in patients with a generally poor outcome following treatment with systemic chemotherapy and radiotherapy, patients with a poor cardiopulmonary status or coexistent medical problems for surgery. There have been some reports of RFA applied clinically for lung tumours 16, 17. The role of RFA in lung cancer is not sufficiently clear. However, bronchoscopy-guided RFA achieves only local coagulation necrosis and is likely to minimise injury to lung tissues and perhaps have a beneficial effect on a patient's condition. When institutions use RFA percutaneously, an epidural anaesthetic consisting of lidocaine, fentanyl or both as needed, needs to be administered percutaneously with an epidural tube 7. If this procedure could be applied to advanced human lung cancer, cooled-RFA could be used with the same method as usual biopsy forceps under topical lidocaine anaesthesia only. The advantages of bronchoscopy-guided cooled-RFA could include the ability to administer treatment nonsurgically as well as a shorter hospital stay for the patient. Based on the current results, it is suggested that RFA for advanced lung cancer might have numerous benefits over systemic chemotherapy or radiation therapy, although trials would be needed to confirm the potential risks/benefits.
One limitation of the present study is that potential tumours requiring ablation would most likely have a different behaviour, and data obtained on RFA in normal lungs cannot be extrapolated to tumour-bearing lungs. As VX2 tumour strains have strong malignant potential, some studies using RFA have been reported 13. As the current authors needed to perform bronchoscopy-guided RFA, larger animals such as sheep, dogs or pigs were required. The authors tried to inject VX2 tumour cells into the right or left lower lung of sheep, but the tumours cells did not grow in the lung. Secondly, as sheep skin is covered with wool, initial impedance was high compared with human lungs (initial impedance of humans: 130200 In conclusion, internal bronchoscopy-guided cooled-radiofrequency ablation was a safe, effective and feasible procedure without major complications. The ideal power output of 30 W and a flow rate of 40 mL·min1 were assessed as safe and technically feasible conditions for bronchoscopy-guided cooled-radiofrequency ablation for this normal sheep lung.
The authors would like to thank F. D'Alessio from Johns Hopkins University School of Medicine (Baltimore, MD, USA) for assistance with the language of the manuscript.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||