Keywords: Power spectral analysis, screening, signal processing, sleep apnoea
Received: July 11, 2006
Accepted December 28, 2006
| ABSTRACT |
|---|
|
|
|---|
A total of 399 polysomnography (PSG) records were employed, including a thermal sensor signal. The first 100 records were used in the development of the algorithm and the remainder for validation. In addition, 119 PSG records, including a thermocouple signal and a nasal pressure signal, were used for the validation. The algorithm was designed to obtain a time series (flow-power) using power spectral analysis, which expresses fluctuation in the airflow signal amplitude. From the time series the algorithm detects transient falls of the flow-power and calculates flow-respiratory disturbance index (RDI), defined as the number of falls per hour.
In the validation group, the areas under receiver operating characteristic curves for diagnosis of SDB (apnoea/hypopnoea index
5) were 0.96, 0.95 and 0.95, for the records of the thermal sensor, thermocouple and nasal pressure system, respectively. The diagnostic sensitivity/specificity ratios of the flow-RDI were 96/76, 88/80 and 97%/77%, respectively.
The present results suggest that a single-channel airflow monitor can be used to detect sleep-disordered breathing automatically if the analytic algorithm is optimised.
Sleep-disordered breathing (SDB) is known to be a relatively common condition which can affect 24 and 9% of middle-aged males and females, respectively, when it is defined by five or more episodes of apnoea or hypopnoea per hour of sleep regardless of symptoms 1. Many studies have indicated that SDB, even in a mild stage, is strongly associated with cardiovascular morbidity 24, as well as traffic accidents 56. Thus, undiagnosed SDB is considered an important public health problem 7. In a clinical setting, the exact diagnosis of SDB is made by polysomnography (PSG), which is a labour-intensive and time-consuming process, and is thus impractical to perform as part of a general population health check-up. In Japan, several surveys using pulse oximetry have been performed and a high prevalence of SDB was found in working populations 8. However, this method has a limitation since oximetry cannot always detect apnoea events, especially for nonobese subjects 9. Recently, Ayappa et al. 10 reported that manual analysis of the flow signal could provide a result similar to that obtained by full PSG analysis. Furthermore, several single-channel airflow monitors have been developed for automated detection of SDB 1113. Studies for the accuracy of these monitors have provided conflicting results and it is unclear whether performance depends on the signal acquisition system, including the sensor, or on the analytical algorithm. It was hypothesised that automated analysis of a single-channel airflow record may have a good diagnostic performance for SDB when the analytic algorithm is fully optimised. In order to test the hypothesis, a novel computer algorithm for flow signal analysis was developed to detect SDB and it was validated against a series of PSG records. The algorithm was developed for a thermal flow sensor, but it was also validated for another type of thermal sensor and a nasal pressure sensor.
| METHOD |
|---|
|
|
|---|
|
PSG
The sensors for PSG were fixed by technicians in the Fukuoka National Hospital (Fukuoka, Japan) but the sensors were not monitored after the start of the recording. PSG was recorded using a polygraph system (EEG7414; Nihon Kohden, Tokyo, Japan). Oronasal airflow was monitored with a thermal sensor as follows: 1) for group 1, a polyvinylidene fluoride (PVDF) film (PVDF sensor; Dymedix Corporation, Minneapolis, MN, USA); and 2) for group 2, a thermocouple sensor (Pro-Tec, Mukilteo, WA, USA). In group 2 a nasal prong-pressure transducer (PTAF; Pro-Tec) was also used to monitor airflow. Thoracic and abdominal respiratory movements were monitored by respiratory inductive plethysmography (RIP; Respitrace; Ambulatory Monitoring Inc., Ardsley, NY, USA), which was calibrated by an iso-volume manoeuvre before the test. Signals of airflow and respiratory movement were recorded at a sampling frequency of 10 Hz. Oxyhaemoglobin saturation was monitored using a pulse oximeter (OLV-3100; Nihon Kohden) at the fastest response mode and recorded at a sampling frequency of 1 Hz. Sleep stages were scored manually according to standard criteria 14. The oximetry data were analysed automatically with a personal computer to detect any arterial oxygen saturation (Sa,O2) dip
3% lasting for 10120 s. The oxygen desaturation index (ODI) was defined by the number of Sa,O2 dips per hour of examination. Hypopnoea (including apnoea) was defined as follows: 1) a
50% reduction of amplitude in the RIP sum signal lasting
10 s; or 2) a discernible reduction (
30<50%, duration
10 s) in the RIP sum signal associated with
3% oxygen desaturation or an arousal 15, 16. The scoring of apnoeahypopnoea events was performed by automated detection based on the RIP amplitude criteria, followed by a thorough manual review and editing. The manual review was performed on the computer screen and the amplitude of each signal was adjusted to be adequate for visual analysis. The RIP sum channel was used primarily for hypopnoea identification; however, in a few cases where airflow signal correlated with change in oxygen saturation better than RIP signal, the raw airflow channel was used. The scoring was carried out before the development of the algorithm for the current study. The apnoea/hypopnoea index (AHI) was calculated as the number of apnoea and hypopnoea events per hour of sleep. Respiratory effort-related arousals (RERAs) were not scored. Due to the scoring method outlined previously, in most cases the thermal sensor signal did not contribute to the AHI value.
Development of algorithm
For the automated analysis of the airflow signal, the following procedures (fig. 1
) were developed using the PSG data from the development group.
|
Power dB = 10xlog10 (power/maximal power) (1)
Thereafter, the flow-power time series was low-pass filtered to obtain a smooth contour.
Detection of the flow-power dip
The second step of the algorithm was detection of the SDB events. The typical change in flow-power at the SDB event was a transient decrease, which was detected when there was a decrease greater than the threshold value within 30 s followed by a normalisation part starting within 90 s (fig. 1b
). Such a decrease in the flow-power was designated as a "flow-power dip."
Tuning of the algorithm
To determine the aforementioned threshold value for the detection of flow-power dips, the present authors compared the concordance between the number of apnoeahypopnoea events detected by PSG and the number of flow-power dips at various threshold values. Thereafter, flow-power dips were detected at the determined threshold and a respiratory disturbance index (RDI) was obtained, defined as the number of flow-power dips per hour of examination. To determine an appropriate flow-RDI cut-off value, a receiver operating characteristic curve (ROC) was constructed to diagnose three levels of SDB (AHI
5,
10 and
15).
Validation of the algorithm
The developed algorithm was applied to the PVDF thermal sensor record of the validation group in group 1 and to the nasal pressure record and thermocouple record in group 2 in order to calculate flow-RDI. The square root of the pressure was used as the flow signal for the nasal prong-pressure transducer system 17.
The relationship between the flow-RDI and the AHI was evaluated using Pearson's correlation coefficient. The agreement between these data was analysed as described by a BlandAltman plot 18. The diagnostic ability of the flow-RDI for SDB was evaluated in terms of sensitivity and specificity.
To evaluate event-by-event agreement, the present authors compared the events detected using three methods of automated detection according to the algorithm, visual inspection of the flow sensor record and of full PSG records, during mid-1-h sleep in 100 consecutive PSG records. Visual detection of apnoeahypopnoea events (
50% reduction of amplitude, lasting
10 s) from the flow sensor record was performed on the computer screen, which visualised only the flow sensor record. The event-by-event agreement was evaluated using the proportion of specific agreement (PSA) 19, 20. In addition, in order to evaluate the effect on flow-RDI of artefacts and irregular breathing while awake, the awake flow-RDI for
10-min continuous awake segments was calculated in 50 patients in group 1 who had low sleep efficiency (<66%).
Comparison with other methods
Comparison with a conventional automatic analysis of flow signal
The diagnostic performance of the developed algorithm was compared with that of a conventional automated analysis using the flow-channel data of the validation group in group 1. A computer program was employed (Screening.exe 21 for flow signal alone; NGK SparkPlug Co. Ltd, Nagoya, Japan) as the conventional analysis of the airflow signal. This program is designed to detect a flow signal amplitude decrease (>50%; duration
10 s) from baseline. The number of events per hour of examination obtained by the program was designated as "conventional RDI." The diagnostic performance of conventional RDI was evaluated as described previously. To compare the diagnostic sensitivity and specificity between the flow-RDI and the conventional RDI, the ROCs were constructed and the areas under the curves (AUCs) calculated. The comparisons of the AUCs were performed by the Hanley and McNeil method 22.
Comparison with oximetry
The diagnostic ability of flow-RDI and ODI were compared in the validation group in group 1. A separate comparison was also conducted for normal-weight group (body mass index (BMI) <25) and overweight/obese group (BMI
25).
| RESULTS |
|---|
|
|
|---|
|
|
Determination of the cut-off value for flow-RDI
From the ROCs for the diagnosis of three levels of SDB in the development group, the appropriate flow-RDI cut-off values were determined to be 5, 7.5, and 10 for the SDBs of AHI
5,
10 and
15, respectively. The sensitivity and specificity at these cut-off values were 92 and 80, 93 and 84, and 90% and 83%, respectively.
Validation of the algorithm
Agreement between flow-RDI and AHI
The correlation coefficients between flow-RDI and AHI were high regardless of flow sensor type (table 2
). BlandAltman analysis showed that the difference between flow-RDI and AHI was narrower for the nasal pressure sensor compared with the other sensors. The bias in group 1 (PVDF sensor) is shown in figure 4
. The bias was highly correlated with AHI (fig. 4b
).
|
|
5,
10 and
15, respectively.
|
|
Analysis of waking time
The mean±SD flow-RDI while awake was 13.2±11.5. The awake flow-RDI had weak but significant correlation with patient's AHI (r = 0.28, p = 0.045). The mean awake RDI for patients with AHI <15 was significantly lower than that for patients with AHI
15 (6.9±6.1 versus 15.2±12.2; p = 0.0036 by Welch's t-test).
Comparison with other methods
Comparison with a conventional automated analysis programme
The correlation coefficient between the conventional RDI and the AHI was 0.888 (95%confidence interval 0.8610.910). BlandAltman analysis showed that the mean±1.96 SD difference between the flow-RDI and the AHI was -3.2±22.2. The AUCs were 0.903, 0.924 and 0.919 for AHI
5,
10 and
15, respectively, all of which were significantly lower than those of the flow-RDI (p = 0.000002, 0.000020 and 0.000035 for AHI
5,
10 and
15, respectively).
Comparison with oximetry analysis
The diagnostic sensitivity and specificity of the ODI at the same cut-off value as flow-RDI were 94 and 84, 93 and 87, and 92% and 82% for the AHI cut-off values of
5,
10 and
15, respectively. The AUCs were 0.966, 0.977 and 0.961 for AHI
5, 10 and 15, respectively. All these parameters for diagnostic ability were similar to those of the flow-RDI. However, the diagnostic sensitivity of ODI was influenced by body habitus. The diagnostic sensitivity of ODI in the normal-weight group was 86, 87 and 84% for AHI
5,
10 and
15, respectively, while in the overweight/obese group it was 99, 97 and 97%, respectively. By contrast, the diagnostic sensitivity of the flow-RDI in the normal-weight group was 90, 95 and 96% for AHI
5,
10 and
15, respectively, while in the overweight/obese group it was 99, 96 and 94%, respectively.
| DISCUSSION |
|---|
|
|
|---|
Technical aspect
The algorithm is unique in several points. First, it utilises power spectral analysis, by which it can lessen the effect of noise contamination since fluctuation of the airflow signal outside the possible frequency range of respiration can be eliminated. Second, it generates a time series of logarithmic power of airflow fluctuation (flow-power). The change in logarithmic power is proportional, not to the difference but to the ratio between two points. When a 6-dB decrease is observed in the time series, it means that the mean amplitude during
10 s decreased to one-half of that value during the preceding part. This property provides an advantage since the same scale can be used even when the level of flow signal changes due to change in the condition of the airflow sensor (e.g. a change in location) during a whole night's recording. Third, it processes the flow-power time series by a low-pass filter, which can eliminate changes irrelevant to apnoeahypopnoea cycles. Fourth, it detects a transient decrease in flow-power (flow-power dip). The algorithm can detect the flow-power dip in a simple automated way. It is believed that these characteristics of the algorithm make the single-channel monitoring relatively accurate in the diagnosis of SDB.
Most automated analyses of airflow to detect apnoeahypopnoea events are considered to employ waveform analysis, which detects the events based on decrease in amplitude of the airflow waveform. Although the conventional waveform analysis programme that was tested in the present study follows the apnoeahypopnoea definition literally, a considerable number of the events detected by the programme appeared not to be real events upon visual inspection. A pattern recognition method on waveform is thought to be susceptible to various kinds of artefact. To the present authors knowledge, only one prior study 23 has used power spectral analysis for airflow signal. Matrot et al. 23 used short-time Fourier transform to calculate the spectral power profile of the respiratory signal, which is the same method used by the present authors. However, the second part of the analysis is different from the method used in the present study. Matrot et al. 23 used normalisation of the signal before power spectral analysis and detected apnoeahypopnoea using a threshold for the spectral power. No normalisation of the signal was used in the present study, but logarithmic transformation was employed for the spectral power. Such different strategy was considered to be relevant to the different setting. Matrot et al. 23 developed the programme to monitor activity and apnoeas of mice for relatively short times, while the present authors developed the programme to monitor breathing events of humans throughout the night. In the present setting, the combination of normalisation and threshold is thought to work incompletely. In a different study by Nakano et al. 24, power spectral analysis was also used to detect SDB from tracheal sound record. However, the purpose of power spectral analysis is different, since in the tracheal sound analysis the power spectral analysis is used to calculate sound power, which is related to airflow velocity, but in the present algorithm the power spectral analysis is used to obtain the magnitude of airflow signal fluctuation corresponding to respiration.
Performance of the algorithm
The flow-RDI obtained by the algorithm has a relatively accurate diagnostic performance despite the fact that it uses only one channel and it uses completely automated analysis.
It is difficult to compare the performance of the algorithm with other algorithms since almost all of them can only be applicable for the monitors originally equipped with them. The performance of the developed algorithm could be compared with that of a conventional automatic analysis programme on the market, demonstrating the superiority of the algorithm. Several previous studies on the diagnostic accuracy of portable monitors have shown that automated scoring of respiratory signals is far less accurate compared with manual scoring 25, 26. By contrast, it was demonstrated that the performance of the algorithm is almost comparable with manual analysis of flow signal alone according to the event-by-event analysis. From these facts, it is believed that the programme developed is superior to many conventional algorithms.
The BlandAltman plot showed considerable systematic bias between the flow-RDI and the AHI. The bias, which was highly correlated with the AHI, means systematic underestimation of the flow-RDI. The underestimation is considered to be due to two reasons. One is the failure of the algorithm to detect a certain proportion of hypopnoea events, which was demonstrated by the event-by-event analysis and could cause underestimation of 23%. The other reason is that the flow-RDI uses examination time as the denominator while the AHI uses sleep time. Mean sleep efficiency in the validation group (78%) can cause underestimation of 22%. However, this underestimation is considered to be partially counterbalanced by detection of events during waking time. The counterbalance is believed not to be accidental but relevant to SDB, since the awake flow-RDI was correlated with the AHI. The resulting actual effect of underestimation was 26% of AHI. The event-by-event analysis also showed false detection of 200 events·100 h1. This false detection was not correlated to the AHI and could cause flow-RDI to be overestimated by 2.0 events·h1. This overestimation explains the positive bias for subjects with low AHI in the AHI versus bias plot.
The AHI-dependent systematic bias between the AHI and the flow-RDI is considered to be an important advantageous feature of the algorithm in two points. One is that such bias can be partially compensated by multiplying the flow-RDI by a specific coefficient. The other is that subjects with relatively low AHI (AHI <20) are less affected by the bias, resulting in less adverse effects for the diagnostic ability. Other reports 11, 13, 25, 26 on validation of automated analysis do not show such an AHI-dependent systematic bias but rather a random and larger bias.
Performance of sensors
The algorithm was originally developed for the analysis of thermal flow sensor signals. It is well known that the change in thermal flow signal is not linear to the change in actual airflow 27 and hypopnoea detection by thermal flow sensor may underestimate SDB.
The thermal sensor used in group 1 utilises a PVDF film, which is reported to have a very fast response to temperature change and an excellent ability to detect hypopnoea when compared with conventional thermal sensors 28. Although the PVDF and thermocouple sensors could not be compared directly, the event-by-event agreement was better for the PVDF sensor in group 1 than for the thermocouple sensor in group 2. Furthermore, it was shown that the nasal pressure sensor was far better than the thermocouple sensor in view of the event-by-event agreement in group 2. These findings agree with the known difference in sensitivity among these sensors.
However, diagnostic ability of the flow-RDI expressed as AUC of ROC was not so different among them. The high AUC means that the flow-RDI at an appropriate cut-off value can have high diagnostic ability. Actually, the appropriate cut-off values were lower for the thermocouple sensor and equivalent or higher for the nasal pressure sensor than those for the PVDF sensor, corresponding to the difference of the sensor characteristics. It is believed that the aforementioned characteristics of the algorithm may make such adjustment capable to improve the diagnostic ability of the low-performance sensor. Another possible way to adjust the sensor sensitivity is to choose an appropriate threshold value to detect the flow-power dip, which was not tested in the current study.
Study limitations
The present study had several limitations. First, the diagnostic ability of a monitor to identify SDB must be dependent on the population studied. In the current study, the subjects belonged to a population with relatively high probability of presenting SDB since they had symptoms suggestive of the condition. Therefore, the present findings may not be applicable to the general population. However, the characteristic of the flow-RDI by which the diagnostic sensitivity was less affected by subjects' body habitus when compared with oximetry analysis is one of its advantageous features, since mean BMI of the general population tends to be lower than that of clinical samples. Secondly, the present study used a flow signal channel from PSG records to validate the algorithm as a simulation of single-channel monitoring. A real validation of the algorithm needs to use the flow signal record in the home. Thirdly, RERAs were not included in SDB events. If these were included, the estimated diagnostic ability may have become worse. Fourthly, although the algorithm is relatively immune to noise contamination, a record which is occupied exclusively by noise may be interpreted as having no events. For practical use, it may not be so difficult to register noise segments as invalid segments based on the power spectral property. However, in the present study such a process was not implemented, since the present authors intended to show the usefulness of flow-power dip analysis. Finally, the single-channel flow analysis has inevitable disadvantages; one is the inability to calculate sleep time as the denominator of AHI and the other is the inability to distinguish obstructive and central events. However, for the purpose of mass screening for SDB, the advantage of simplicity and easy applicability might predominate over such disadvantages.
In conclusion, a new computer algorithm has been developed which can automatically detect sleep-disordered breathing from the airflow record alone. The system needs no manual edit and is relatively accurate. It is believed that not only the performance of sensors but also the analytical algorithm is crucial for the development of simple portable monitors applicable for mass screening for sleep-disordered breathing.