Copyright ©ERS Journals Ltd 2007 Hot biopsy forceps in the diagnosis of endobronchial lesions1 Divisions of Respiratory Medicine, Thoracic Surgery and Medical Oncology, and 2 Dept of Pathology and Laboratory Medicine, University of Calgary, and 3 Southern Alberta Cancer Research Institute, Calgary, AB, Canada. CORRESPONDENCE: A. Tremblay, Division of Respiratory Medicine, Health Sciences Center, 3330 Hospital Drive NW, Calgary, AB, T2N 4N1, Canada. Fax: 1 4039441577. E-mail: alain.tremblay{at}ucalgary.ca Keywords: Biopsy, bronchoscopy, lung cancer
Received: May 30, 2006
Electrocoagulation bronchoscopy biopsy forceps may prevent bleeding, but could also impair the quality of the specimens obtained. Patients with endobronchial lesions during bronchoscopy underwent six endobronchial biopsies each with a hot biopsy forceps, alternating between with electrocoagulation ("hot") and without ("cold"). Bleeding was quantified on a scale of 14, with 1 being no bleeding. The generator was set on "soft coagulation" mode, with power settings of 40, 60, 80 and 100 W for each group of 10 patients in a sequential fashion. Clinical pathology results were recorded before samples were reviewed by a second, blinded, pulmonary pathologist. A total of 39 patients with 40 endobronchial lesions had six biopsies performed (one patient had only four samples taken), giving a total of 238 biopsy samples. Concordance between hot and cold samples was 92.5% for the clinical pathologist and 87% for the blinded pathologist. Paired analysis suggested lower average bleeding score with the use of hot forceps. Overall bleeding rates for cold and hot biopsies, respectively, were as follows: grade 1: 30.3 and 41.2%; grade 2: 62.2 and 49.6%; grade 3: 7.6 and 9.2%; and grade 4: 0 and 0%. In conclusion, the use of hot biopsy forceps for endobronchial biopsy does not appear to have a negative impact on the pathological samples. Hot biopsy forceps showed a statistically significant reduction in bleeding score, which is unlikely to be of clinical significance. The traditional approach to diagnosis of endobronchial lesions relies on forceps biopsy during flexible bronchoscopic examination. A new electrocautery "hot" bronchoscopy biopsy forceps is now commercially available and may prevent bleeding following biopsy. Unfortunately, no published data exist regarding the use of this device. The primary objective of the present study was to determine the impact of electrocautery forceps on pathological diagnosis following biopsy of endobronchial lesions. The secondary objective was to compare bleeding rates between the two techniques.
The present study was conducted as a prospective controlled study, with each patient serving as his or her own control. The study protocol was approved by the Conjoint Health Research Ethics Board of the University of Calgary (Calgary, AB, Canada). Written informed consent was obtained for all patients included in the study.
Inclusion/exclusion criteria
Procedure Once a suitable lesion was identified, six biopsies were taken, alternating between "hot" and "cold" without specifying the method for the initial biopsy. The samples were placed in two separate formalin containers. Quantification of bleeding was carried out and recorded by the bronchoscopist (unblinded to biopsy method) after each biopsy according to a four-point scale (1: no bleeding; 2: minimal bleeding not requiring intervention; 3: mild-to-moderate bleeding requiring intervention, such as ice-cold saline, topical adrenaline or electrocoagulation; 4: moderate-to-severe bleeding requiring termination of procedure, endotracheal intubation or other invasive treatment). Additional biopsies or other sampling were performed at the discretion of the bronchoscopist, but were placed in separate containers. All biopsies were taken with an electrocoagulation-capable biopsy forceps (FD-6C-1A; Olympus America, Melville, NY, USA) with ("hot") or without ("cold") the application of an electrocoagulation current (ICC 350 electrosurgical generator; ERBE, Tübingen, Germany) set on "soft coagulation" mode. The power setting on the generator was set at 40, 60, 80 and 100 W for each group of 10 patients, in a sequential fashion.
Pathology If the endobronchial biopsies were nondiagnostic, other samples taken at the time of bronchoscopy or additional samples taken at a later date were reviewed and recorded for study purposes.
Analysis The primary outcome measure for the study was the discordance of pathological results obtained with the hot and cold methods. A McNemar's test for changes to a 4x4 table was applied comparing results from both diagnostic techniques. Fisher's exact test was also used to compare the diagnostic yield between the two approaches and a 95% confidence interval of the frequency of positive concordance between the two approaches was calculated. The study sample size of 40 patients was calculated using Fisher's exact test (two-sided) as if the samples were unmatched, with a power of 0.80 to detect a decrease in diagnostic yield from 95% to 70%, with an alpha of 0.05. This approach was used as there was no way to determine the expected frequency of discordant results between each diagnostic test required for the McNemar's test to be used in sample size calculations. Ten discordant results would be required to show the above difference with the McNemar's test. Comparison of bleeding rates was performed by comparing the average bleeding score for hot versus cold biopsies in a paired nonparametric analysis (Wilcoxon signed rank tests).
A total of 39 patients with 40 endobronchial lesions were entered into the study and had the six biopsies performed as per protocol (one patient had only four samples taken), giving a total of 238 biopsy samples. Of the subjects, 56% were male (22 out of 39). The mean (range) age was 60.8 yrs (2085).
Locations of biopsies and final diagnoses are listed in table 1
Concordance between hot and cold samples was 92.5% for clinical pathologist and 87% for research pathologist (p>0.05, McNemar's test), suggesting lack of significant impact of hot biopsy on diagnostic yield (table 2
Paired analysis of the average bleeding score between hot and cold biopsies showed a statistically significant lower bleeding score for hot samples (p = 0.03). Overall bleeding rates for cold and hot biopsies were, respectively, as follows: grade 1: 30.3 and 41.2%; grade 2: 62.2 and 49.6%; grade 3: 7.6 and 9.2%; and grade 4: 0 and 0% (p>0.05; KruskalWallis test; fig. 1a
The diagnostic yield following endobronchial biopsy is excellent, with the largest case series demonstrating sensitivities ranging 7697% 1, 2. It has also been demonstrated that three samples are sufficient to maximise diagnostic yield 3, although the collection of additional samples has been justified by others 4, 5.
The only significant complication following endobronchial biopsy is bleeding. Mild, self-limited bleeding is common 6 and may lead to difficult visualisation, potentially impeding the remainder of the examination. Severe life-threatening bleeding is fortunately a rare event. One report described two deaths attributed to bleeding in a questionnaire study of Endobronchial electrocoagulation is an effective tool in the management of endobronchial lesions and can be very effective at controlling bleeding from an identified bleeding site 810. This makes new biopsy forceps capable of applying an electrocoagulation current at the time of biopsy promising in terms of reducing bleeding risk following this common procedure. Despite the commercial availability of such a forceps, no published investigations address the potential impact of this approach on the quality of the biopsy specimen obtained, despite the fact that electrocautery applied to the airway can lead to significant pathological changes in the tissues 11, 12. Nor has the impact of hot forceps on bleeding rate been reported. The present study was performed to better understand the performance characteristics of a new commercially available device designed to obtain endobronchial biopsies with simultaneous application of an electrocautery current aimed at preventing or reducing bleeding. The study was powered to detect a significant impact on the diagnostic yield of the samples, as it was felt that a greater harm could come to patients if this device lowered the diagnostic yield from bronchoscopy biopsy than if the device simply failed to prevent bleeding. Fortunately, the hot biopsy forceps did not appear to impair the quality of the specimens obtained or hinder the pathologist's ability to make a diagnosis. The next issue was to determine whether the device actually does what it is designed to do: reduce bleeding rates. Although a statistically significant decrease in mean bleeding score was noted with the use of the electrocoagulation current, this seemed to be limited to a reduction in the number of minor bleeding episodes (grade 2), which are unlikely to be of clinical significance. No decrease in moderate bleeding was noted, although the number of these events was low and a larger study would be needed to have sufficient power to confirm this finding. Similarly, severe bleeding episodes were rare enough that it would be difficult to show a significant difference in this outcome. Given that the population studied did not have any particular increased risk for bleeding, the present authors cannot exclude the possibility that this technique may reduce bleeding in high-risk patients or lesions. During the study period, only one patient had severe bleeding resulting in termination of the procedure. Interestingly, this occurred during a hot biopsy, but the patient was not included in the study as per protocol, given that no matching cold biopsy was obtained. The lack of blinding at the time of quantification of bleeding could potentially bias the results of the study, but this is difficult to remedy. The use of a simple scoring system and the initial equipoise of the current investigators regarding the effect of this device should have minimised any bias. Another potential bias relates to the lack of randomisation of the sequence of the biopsies. Given an alternating biopsy protocol (i.e. coldhotcoldhotcoldhot or vice versa), it seems unlikely that starting with one approach rather than the other would impact bleeding rates or the quality of the samples. The cost of the biopsy forceps is relatively minor, as it is a re-usable device; however, the infection-control implications of reusable biopsy forceps need to be considered. The cost of the electrosurgical generator is much more important, but such devices may be easily accessible in most hospitals. Given these findings, it would not seem that routine use of the electrocoagulation bronchoscopy forceps is warranted. The present authors have occasionally found it useful to apply electrocautery when moderate bleeding occurs following endobronchial biopsy. This can be done using the biopsy forceps as the electrocautery probe and is of course facilitated if all the equipment is already set up. Familiarity with this equipment is also important and may lead to improved patient and bronchoscope safety. Other simpler methods, such as topical application of ice-cold saline or adrenaline, may be easier to apply and more easily accessible to bronchoscopists.