Depts of 1 Pneumology and Critical Care Medicine, and 4 Thoracic Surgery, Thoraxklinik at the University of Heidelberg, Heidelberg, Germany. 2 Interventional Pulmonology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, USA. Depts of 3 Surgical Gastroenterology, and 5 Cardiothoracic Surgery, Gentofte University Hospital, Copenhagen, Denmark.
CORRESPONDENCE: F. J. F. Herth, Dept of Internal Medicine, Pneumology and Critical Care Medicine, Thoraxklinik Heidelberg, Amalienstr. 5, D-69126 Heidelberg, Germany. Fax: 49 62213961202. E-mail: Felix.Herth{at}thoraxklinik-heidelberg.de
Keywords: Endobronchial ultrasound, lung cancer, mediastinal lymphadenopathy, transbronchial needle aspiration
Received: October 26, 2005
Accepted June 8, 2006
| ABSTRACT |
|---|
|
|
|---|
1 cm in diameter. NSCLC patients with CT scans showing no enlarged lymph nodes (no node >1 cm) in the mediastinum underwent EBUS-TBNA. Identifiable lymph nodes at locations 2r, 2l, 4r, 4l, 7, 10r, 10l, 11r and 11l were aspirated. All patients underwent subsequent surgical staging. Diagnoses based on aspiration results were compared with those based on surgical results.
In 100 patients (mean age 58.9 yrs; 68 males), 119 lymph nodes ranging 510 mm in size were detected and sampled. Malignancy was detected in 19 patients but missed in two; all diagnoses were confirmed by surgical findings. The mean diameter of the punctured lymph nodes was 8.1 mm. The sensitivity of EBUS-TBNA for detecting malignancy was 92.3%, specificity was 100%, and the negative predictive value was 96.3%. No complications occurred.
In conclusion, endobronchial ultrasound-guided transbronchial needle aspiration can accurately sample even small mediastinal nodes, therefore avoiding unnecessary surgical exploration in one out of six patients who have no computed tomography evidence of mediastinal disease. Potentially operable patients with no signs of mediastinal involvement on computed tomography may benefit from pre-surgical endobronchial ultrasound-guided transbronchial needle aspiration and staging.
The staging of lung cancer 1, 2 not only provides important information with regard to survival, but also guides treatment. For patients whose disease has metastasised to mediastinal lymph nodes (stage III) or with tumours that have invaded mediastinal structures, the benefit of surgery as primary therapy is questionable. Combined chemoradiotherapy is most appropriate, although chemoradiotherapy followed by surgery may be considered 2. Mediastinal lymph node involvement is found in 26% of newly diagnosed lung cancer patients, and extrathoracic metastases are found in 49% 2. Thus, reliable staging of mediastinal lymph nodes and the mediastinum is essential for choosing appropriate therapy.
In most centres, computed tomography (CT) is the initial method for staging mediastinal nodes. Nodes detectable by CT that are considered abnormal generally have a short-axis diameter >1 cm. Smaller lymph nodes can harbour metastatic foci, and enlarged nodes may be benign, especially when central tumours are accompanied by inflammation. Therefore, the accuracy of CT for diagnosing mediastinal disease is low. In a recent meta-analysis of 20 studies and 3,829 patients 3, the pooled sensitivity was 57%, the pooled specificity was 82% and the pooled negative predictive value was 82% (range 6385%). Therefore, surgical mediastinal staging is commonly performed in patients with a radiologically normal mediastinum before a planned cancer resection, to rule out unexpected N2 or N3 disease. Approximately 18% (range 1537%) of patients with a negative CT scan who undergo surgical mediastinal staging are found to have metastatic disease.
Mediastinoscopy remains the reference standard for evaluating nodal disease. It has a sensitivity of 9095% 4. However, only certain mediastinal lymph node stations are accessible (levels 2, 4, and anterior level 7). For sampling levels 5 and 6, thoracoscopy, anterior mediastinotomy (the Chamberlain procedure), or extended mediastinoscopy can be performed. The inferior mediastinum may be evaluated by thoracoscopy. However, all these more aggressive staging procedures require general anaesthesia, surgical incision, and, therefore, are associated with considerable costs 5, 6. Positron emission tomography (PET) was expected to increase the accuracy of mediastinal staging in nonsmall cell lung cancer (NSCLC), and indeed, a meta-analysis has confirmed its superiority 7. However, more recent reports have tempered enthusiasm for using PET as the sole tool for evaluating the mediastinum 8.
Endoscopic ultrasound (EUS)-guided fine-needle aspiration has been reported as an endoscopic option to evaluate the mediastinum 912. Unfortunately, it does not allow for airway inspection during the procedure. Several authors 1315 have reported that a real-time procedure, endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is a highly accurate and safe method for sampling enlarged mediastinal lymph nodes. However, EBUS-TBNA has not been applied to patients with CT-negative scans. The aim of the present study was to determine the accuracy of EBUS-TBNA for mediastinal lymph nodes from NSCLC patients without enlarged mediastinal lymph nodes on chest CT scans.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Standard, conventional flexible bronchoscopy (model BF-T160 bronchoscope; Olympus, Tokyo, Japan) was first performed to examine the tracheobronchial tree, followed by EBUS-TBNA using the ultrasound bronchoscope (model XBF-UC160F-OL8; Olympus). Both bronchoscopy procedures were performed with the patient under local anaesthesia and sedation (midazolam) or under general anaesthesia. All patients underwent surgical staging within 10 days after the EBUS-TBNA.
Noninvasive imaging
Multislice CT was performed in all patients, and tumour/node/metastasis (TNM) staging, including identification of distant metastases, was recorded by two thoracic CT radiologists. On-site CT examinations were performed with a Siemens helical scanner (Erlangen, Germany), using a single breath-hold technique. Off-site CT scans were evaluated by the same radiologists and were included if their quality was similar to the on-site CT studies. If the quality of an off-site CT was inadequate, an on-site CT was performed. Lymph nodes were considered enlarged if the short-axis diameter was >1 cm.
EBUS-TNBA
Endobronchial ultrasound (EBUS) was performed using a linear array ultrasonic bronchoscope (XBF-UC 160F-OL8; Olympus; fig. 1
). The instrument is similar to a standard bronchoscope with an outer diameter of 6.9 mm, a 2.0-mm instrument channel, and 30° oblique forward-viewing optics. An electronic convex array ultrasound transducer is mounted at the distal tip and is covered by a water-inflatable balloon sheath. Scanning is performed at a frequency of 7.5 MHz. The angle of view is 90°, and the direction of view is 30° forward oblique. Image processing was performed by an Olympus ultrasound processor (EU-C60).
|
The endoscope was passed through the mouth and vocal cords to the main carina, the balloon, if used, was partially inflated (0.30.5 mL water), and the regional lymph node stations of the mediastinum and hilar regions (stations 2, 4, 7, 10 and 11) 16 were systematically imaged and measured (short-axis diameter) during slow withdrawal and rotation of the transducer.
All visualised nodes with a size of 510 mm were punctured. According to the size of normal lymph nodes, nodes <5 mm were not punctured 1719. Transbronchial needle aspiration was performed under real-time ultrasound control. Needle punctures were performed using the "jabbing" method 20. Integrated colour power Doppler ultrasound was used to avoid intervening vessels immediately before needle puncture (fig. 2
). Every node was punctured twice.
|
Statistical methods
The Chi-squared test was used, where appropriate, to compare proportional data. The type I error was set at 0.05 for all analyses. Confidence intervals were calculated to 95% using standard formulae. The sensitivity, specificity and accuracy were calculated using the standard definitions.
| RESULTS |
|---|
|
|
|---|
A known diagnosis of NSCLC was identified in 87 patients and 13 patients included in the population were highly suspicious for the disease (positive sputum cytology, increased tumour markers, CT criteria and tobacco abuse). All patients had at least one node identified by EBUS and a total of 119 lymph nodes 510 mm in size were detected and punctured by EBUS-TBNA (table 1
). The mean±SD diameter of the punctured lymph nodes was 8.1±0.7 mm (range 410 mm). Additionally, 89 nodes <5 mm were detected, which were not punctured. Despite the negative CT scans, EBUS-TBNA of mediastinal lymph nodes was positive for metastases in 19 patients. The stage changed from N0 in three patients to stage N3 disease, to stage N2 disease in 13 patients and to stage N1 disease in three patients. All punctures were adequate and in every smear lymphocytes were visible. In total, 78 patients were examined under general anaesthesia and 22 patients with moderate sedation and topical anaesthesia.
|
| DISCUSSION |
|---|
|
|
|---|
Compared with mediastinoscopy, EBUS-TBNA has the advantage that it is also able to routinely access posterior mediastinal (level 7) and hilar lymph nodes (levels 10 and 11) and that it can be performed using moderate sedation. Most of the study patients were examined under general anaesthesia, as the examination was done immediately prior to the scheduled surgical procedure. As reported in previous studies, there is no difference in yield or patient tolerance if the procedure is performed under moderate sedation or general anaesthesia 24.
Similar approaches using oesophageal ultrasound (OUS)-guided fine-needle aspiration (OUS-FNA) have been described with comparable results, albeit in smaller patient populations. A study by Wallace et al. 25 demonstrated 69 patients without enlarged mediastinal lymph nodes. OUS detected malignant mediastinal lymph nodes in 14 out of the 69 patients. The sensitivity of EUS for advanced mediastinal disease was 61% and the specificity was 98%. Additionally, Wallace et al. 25 found advanced stage in three other patients (one left adrenal metastasis and two with mediastinal invasion of tumour). LeBlanc et al. 26 examined 76 patients with NSCLC without mediastinal lymphadenopathy on CT. EUS-FNA was performed on sites that were suspicious for metastases. Of the 62 patients who underwent surgery, 23 (37%) patients had positive lymph nodes, six had peribronchial lymph node (N1) involvement, whereas the remaining 17 patients had ipsilateral or subcarinal lymph node (N2) involvement. Aspirates of lymph nodes yielding lymphocytes only, either by OUS-FNA or EBUS-TBNA, do require follow-up with a more definitive procedure.
Depending on the lymph node localisation, EBUS-TBNA seems at least comparable to the established EUS-FNA findings in the literature 15, 27, 28. It is important to note that small lymph nodes are more difficult to identify with any imaging modality (including EBUS) and probably contain a small number of malignant cells, making a cytological diagnosis difficult. Nonetheless, in 100 patients, EBUS-TBNA identified almost all (19 out of 21) patients with advanced disease. The present study supports the theory that EUS-TBNA has excellent potential, even in the patient with a normal mediastinum by CT scanning.
Limitations of the study
A limitation of the current study was the lack of routine PET scanning. Although PET has become increasingly available and is approved for lung cancer staging, it is still not routinely performed in practice. Recent guidelines for staging of NSCLC recommend PET staging 29, but the evidence supporting this recommendation is only moderate 7, 8. Furthermore, the limitations in accuracy discussed above make it unlikely that PET, with its current resolution, even in combination with CT, will replace the need for tissue sampling.
Lymph nodes are often grouped in stations. A potential exists that the surgically sampled lymph nodes in the examined stations were not identical to the ones punctured during endoscopy. This appears unlikely, as in the present routine all lymph nodes are dissected and the results of the endoscopically and surgically sampled lymph nodes are highly congruent.
Conclusions
The current findings suggest that endobronchial ultrasound-guided transbronchial needle aspiration should be considered in the pre-operative staging of all patients both with and without mediastinal lymph node enlargement on computed tomography scan. Further studies are needed to compare the different invasive and noninvasive staging techniques (computed tomography, positron emission tomography, endobronchial ultrasound-guided transbronchial needle aspiration, endobronchial ultrasound-guided fine-needle aspiration, mediastinoscopy and thoracoscopy) in patients with nonsmall cell lung cancer. To establish a definitive way of staging the mediastinum in patients with nonsmall cell lung cancer, a comparison study of all imaging, minimally invasive and also invasive techniques should be considered.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. S. Lee, G. K. Lee, H.-S. Lee, M. S. Kim, J. M. Lee, H. Y. Kim, B.-H. Nam, J. I. Zo, and B. Hwangbo Real-time Endobronchial Ultrasound-Guided Transbronchial Needle Aspiration in Mediastinal Staging of Non-Small Cell Lung Cancer: How Many Aspirations Per Target Lymph Node Station? Chest, August 1, 2008; 134(2): 368 - 374. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. J.F. Herth, R. K. Morgan, R. Eberhardt, and A. Ernst Endobronchial ultrasound-guided miniforceps biopsy in the biopsy of subcarinal masses in patients with low likelihood of non-small cell lung cancer. Ann. Thorac. Surg., June 1, 2008; 85(6): 1874 - 1878. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. J. F. Herth, R. Eberhardt, M. Krasnik, and A. Ernst Endobronchial Ultrasound-Guided Transbronchial Needle Aspiration of Lymph Nodes in the Radiologically and Positron Emission Tomography-Normal Mediastinum in Patients With Lung Cancer Chest, April 1, 2008; 133(4): 887 - 891. [Abstract] [Full Text] [PDF] |
||||
![]() |
M P Kennedy, C A Jimenez, J F Bruzzi, A D Mhatre, X Lei, F J Giles, T Fanning, R C Morice, and G A Eapen Endobronchial ultrasound-guided transbronchial needle aspiration in the diagnosis of lymphoma Thorax, April 1, 2008; 63(4): 360 - 365. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Wallace, J. M. S. Pascual, M. Raimondo, T. A. Woodward, B. L. McComb, J. E. Crook, M. M. Johnson, M. A. Al-Haddad, S. A. Gross, S. Pungpapong, et al. Minimally Invasive Endoscopic Staging of Suspected Lung Cancer JAMA, February 6, 2008; 299(5): 540 - 546. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. D. Vincent, E. El-Bayoumi, B. Hoffman, P. Doelken, J. DeRosimo, C. Reed, and G. A. Silvestri Real-Time Endobronchial Ultrasound-Guided Transbronchial Lymph Node Aspiration Ann. Thorac. Surg., January 1, 2008; 85(1): 224 - 230. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Nakajima, K. Yasufuku, A. Iyoda, S. Yoshida, M. Suzuki, Y. Sekine, K. Shibuya, K. Hiroshima, Y. Nakatani, and T. Fujisawa The evaluation of lymph node metastasis by endobronchial ultrasound-guided transbronchial needle aspiration: Crucial for selection of surgical candidates with metastatic lung tumors. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1485 - 1490. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Tino Clinical Year in Review I: Lung Cancer, Interventional Pulmonology, Pediatric Pulmonary Disease, and Pulmonary Vascular Disease Proceedings of the ATS, September 15, 2007; 4(6): 478 - 481. [Full Text] [PDF] |
||||
![]() |
F. C. Detterbeck, M. A. Jantz, M. Wallace, J. Vansteenkiste, and G. A. Silvestri Invasive Mediastinal Staging of Lung Cancer: ACCP Evidence-Based Clinical Practice Guidelines (2nd Edition) Chest, September 1, 2007; 132(3_suppl): 202S - 220S. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Nakajima, K. Yasufuku, M. Suzuki, K. Hiroshima, R. Kubo, S. Mohammed, Y. Miyagi, S. Matsukuma, Y. Sekine, and T. Fujisawa Assessment of Epidermal Growth Factor Receptor Mutation by Endobronchial Ultrasound-Guided Transbronchial Needle Aspiration Chest, August 1, 2007; 132(2): 597 - 602. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Silvestri Boys (and Girls) and Their Toys: A Look at New Technologies in the Bronchoscopy Suite Am. J. Respir. Crit. Care Med., July 1, 2007; 176(1): 1 - 2. [Full Text] [PDF] |
||||
![]() |
A. H.M. Piet, F. J. Lagerwaard, P. W.A. Kunst, J. R. van Sornsen de Koste, B. J. Slotman, and S. Senan Can Mediastinal Nodal Mobility Explain the Low Yield Rates for Transbronchial Needle Aspiration Without Real-Time Imaging? Chest, June 1, 2007; 131(6): 1783 - 1787. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Singh and A. Khan It's All About US: (E)US, (EB)US, and Their US(age) Am. J. Respir. Crit. Care Med., June 1, 2007; 175(11): 1208a - 1209. [Full Text] [PDF] |
||||
![]() |
P. J. Sterk and K. F. Rabe What else can you expect at www.erj.ersjournals.com? Eur. Respir. J., January 1, 2007; 29(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
A. H. Diacon, M. M. Schuurmans, J. Theron, K. Brundyn, M. Louw, C. A. Wright, and C. T. Bolliger Transbronchial needle aspirates: how many passes per target site? Eur. Respir. J., January 1, 2007; 29(1): 112 - 116. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. J. F. Herth, K. F. Rabe, S. Gasparini, and J. T. Annema Transbronchial and transoesophageal (ultrasound-guided) needle aspirations for the analysis of mediastinal lesions Eur. Respir. J., December 1, 2006; 28(6): 1264 - 1275. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |