|
|
||||||||
Respiratory Sleep Disorders Unit, St. Vincents University Hospital, Dublin, Ireland
CORRESPONDENCE: W. T. McNicholas, Dept of Respiratory Medicine, St. Vincents University Hospital, Elm Park, Dublin 4, Ireland, Fax: 353 12697949. E-mail: walter.mcnicholas{at}ucd.ie
Keywords: Auto-adjustable positive airway pressure therapy, compliance, continuous positive airway pressure therapy, obstructive sleep apnoea syndrome, quality of life
Received: November 1, 2005
Accepted March 13, 2006
| ABSTRACT |
|---|
|
|
|---|
Twenty-seven middle-aged patients (25 male) previously diagnosed with severe obstructive sleep apnoea syndrome (median (interquartile range) apnoea/hypopnoea index 48 (2976)), established on CPAP therapy for >3 yrs, were randomised to each APAP device for 4 weeks.
Mean pressure and patient compliance were significantly lower on the Breas PV 10i than on the other APAP devices. The devices were similar in terms of quality of life, daytime sleepiness and upper airway side-effects, but patients evaluated them significantly differently in terms of device features, sleep quality and pressure comfort, with the Breas PV 10i being the least popular.
Auto-adjustable positive airway pressure devices differ in pressure delivery and patient compliance in obstructive sleep apnoea syndrome patients.
Obstructive sleep apnoea syndrome (OSAS) is a common and potentially serious disorder, affecting
4% of adults 1. It is characterised by repeated episodes of apnoea during sleep, up to several hundred events per night, which lead to hypoxia, sleep fragmentation and excessive daytime sleepiness, with the associated risk of serious sequelae, such as road traffic accidents and significant cardiovascular morbidity and mortality 24. Nasal continuous positive airway pressure (CPAP) is the current therapy of choice, particularly in moderate and severe cases. It has been shown to improve quality of life, reduce the risk of road traffic and occupational accidents, and decrease cardiovascular morbidity and mortality 58. Since the mid-1990s, auto-adjustable positive airway pressure (APAP) devices have been developed to improve patient comfort and compliance. APAP devices are designed to continuously adjust the applied pressure to the optimal level throughout the night, according to an algorithm, and therefore minimise the mean overnight pressure requirements of patients with OSAS 9, 10. The effectiveness of APAP in reducing the apnoea/hypopnoea index (AHI), as well as the improvement in symptoms, is similar to that of fixed-pressure CPAP therapy 1113. At present, there is insufficient evidence of benefits of APAP devices over traditional CPAP treatment to justify their routine use in clinical practice 14. The role of APAP in increasing patient compliance, in particular, remains unsolved. Moreover, it is not known whether the different device algorithms have a comparable effect on patient comfort and compliance.
The present study was designed to compare the effect of three APAP devices on compliance, quality of life and side-effects of treatment in patients with OSAS, who were already established on fixed-pressure CPAP therapy.
| METHODS |
|---|
|
|
|---|
The baseline characteristics of the subjects are described in table 1
. The patients had been on fixed-pressure therapy for a median (interquartile range) of 53 (3785) months prior to the start of the study. The following CPAP devices were in use at that time: Sullivan Elite V (ResMed, Abingdon, UK; nine patients), Respironics Aria Lx (Respironics, Inc., Murrysville, PA, USA; 13 patients), RemStar Pro (Respironics, Inc.; four patients), and Fisher & Paykel HC221 (Fisher & Paykel Healthcare, Auckland, New Zealand; one patient). Each patient was given a 4-week home trial of three APAP devices: Autoset Spirit (ResMed), Breas PV 10i (Breas Medical, Mölnlycke, Sweden), and RemStar Auto (Respironics, Inc.), set at 416 cmH20, in a randomised crossover trial. The trial was fully blinded to the investigator performing the analysis as this person was not involved in immediate patient contact and was also blinded to the randomisation. It was not possible to fully blind the patients because of the different appearance of the devices. However, patients were not informed about the different technologies used in the APAP devices; they were simply told that these were newer treatment machines. Each patient used the same nasal mask throughout the trial. The devices were assigned to patients in random order by an independent person not involved in the study design, protocol or analysis. At the end of each APAP trial period, patients were evaluated by an experienced sleep technician and physician and asked to complete the Epworth Sleepiness Scale (ESS) 15, a questionnaire about physical and mental health (36-item short-form health survey (SF-36)) 16, and an in-house evaluation questionnaire specifically addressing device features, sleep quality and side-effects (see Appendix). Time-coded compliance data from the fixed-pressure CPAP device at the start of the trial and from the APAP devices at the end of each 4-week trial were downloaded. At the end of the study, patients were asked to place the four devices (three APAP plus CPAP) in order of preference and to give reasons for their first preference.
|
The Breas PV 10i creates a model of the patients breathing signal, which is then compared several times per second to a grid containing 64 pre-stored templates of normal and abnormal respiratory events. The template closest to the patient model determines the response of this unit.
The RemStar Auto compares the inspiratory flow shape of each breath to a rolling patient database, derived from the previous 23 min of breath shapes. Two pressure tests are then performed, one which reduces the pressure until the airway begins to collapse and another which increases the pressure until there is no further improvement in the flow shape. A therapy mode is then entered, providing a constant ideal pressure until the next change in respiration occurs.
All three devices also have a safety feature, which increases pressure rapidly in the presence of an unexpected abnormal event.
Statistical analysis
Subject baseline characteristics and downloaded device parameters are presented as median (interquartile range) or mean±SD, depending on distribution, and comparisons were made using a paired t-test or the Wilcoxon signed-rank test for paired samples. A significance level of p<0.05 was considered significant.
| RESULTS |
|---|
|
|
|---|
Compliance and treatment pressure
After each 4-week treatment period, the devices were downloaded for evaluation of compliance and applied pressure. The mean applied pressure was significantly lower on all three APAP devices compared to CPAP, with the lowest mean pressure of 5.8 cmH20 recorded on the Breas PV 10i. Overall, compliance was acceptable (used for >70% nights and >4 h·night-1) on CPAP and on the RemStar Auto and the Autoset Spirit, but was significantly lower on the Breas PV 10i (table 2
).
|
Patient evaluation of the three APAP devices
All subjects evaluated the devices after each treatment period in terms of device features, sleep quality and side-effects (figs. 1
3
). Overall, all three APAP devices were equally easy to use. The Autoset Spirit was the least preferable machine in terms of size and noise, whereas the RemStar Auto represented the best device features in the evaluation (fig. 1
). The Breas PV 10i provided significantly poorer sleep quality in comparison to the other two APAP devices, which were similar in this respect (fig. 2
). All three devices were similar in causing nasal and upper airways side-effects. Pressure discomfort was reported most frequently using the Breas PV 10i (fig. 3
).
|
|
|
| DISCUSSION |
|---|
|
|
|---|
The present study is the first report comparing three different APAP devices in terms of delivered pressure profile, patient compliance, quality of life and side-effects in patients already established on fixed-pressure CPAP therapy. Although all three of the APAP devices studied use the flow signal to determine the required pressure, the different algorithms account for a significant difference in delivered pressure, and, consequently, for differences in subjective outcome parameters and compliance. No previous study has compared these same three devices. However, in comparing different APAP devices, similar observations of efficacy and pressure profile have been obtained. Farre et al. 20 assessed five different APAP devices in a bench study utilising a breathing waveform generator and concluded that the responses of the devices to apnoeas, hypopnoeas, flow limitation and snoring were considerably different. Hertegonne et al. 21 came to a similar conclusion in a patient population. They compared the comfort parameters and pressure profiles of the AutoSet and SOMNOsmart in a split-night crossover protocol and detected significant differences.
Only one other trial to date has compared different APAP technologies in a patient group over a longer treatment period. Senn et al. 22 evaluated 29 CPAP-naive patients with OSAS and compared fixed-pressure CPAP and two APAP technologies, each for 1 month, in a crossover design. This study did not find any significant differences between the two APAP devices in terms of symptoms, AHI, mean applied pressure or compliance. At the end of the trial, 72% of patients had no preference for either APAP or CPAP, with four (14%) choosing APAP and four CPAP. These results differed from the present study, in which 52% preferred APAP and 48% CPAP. However, similar to the present results, they too found that details such as size, weight and noise level during use were the most frequently cited reasons for final preference. Furthermore, the present study evaluated a group of patients already established on fixed CPAP. Such patients were selected because they are familiar and compliant with this type of treatment process, and local airway pathology, such as oedema and inflammation, is likely to have already subsided with CPAP therapy. Also, these patients are established on their mask system, and could, therefore, more accurately evaluate the different APAP technologies without the inevitable limitations associated with CPAP-naive patients.
However, it is recognised that, since the patients were already established on CPAP, there may have been a bias towards CPAP over APAP in the present study. Some previous reports have indicated a preference towards APAP in treatment-naive patients 23, 24, whereas others have reported no such preference 25, 26. However, the present study design was specifically chosen to directly compare these three APAP devices, an objective that would have been more difficult in treatment-naive patients. There are only limited data on APAP therapy in patients with mild OSAS, although a preliminary report from the current authors' Respiratory Sleep Disorders Unit indicates no consistent preference for APAP over CPAP therapy in patients with mild disease 27.
Another potential limitation of the present study is the inclusion of only two female subjects. This distribution does not allow a conclusion regarding sex differences. However, these sex frequencies represent the distribution of severe OSAS in the current authors' Respiratory Sleep Disorders Unit, and separate studies investigating APAP evaluation in females should be performed specifically targeting this question.
It was found that the Breas PV 10i resulted in a significant lower pressure profile, but poorer sleep quality and more complaints of pressure discomfort; the lowest compliance rates were also observed with this device. Furthermore, it was the least popular choice among patients preferring APAP devices at the end of the study. Unfortunately, an evaluation of the AHI data is not available on this machine, making it impossible to propose ineffective treatment response as a cause of its negative impact. Both the RemStar Auto and the Autoset Spirit reduced the AHI to <5 events·h-1 (data not shown), but, given the significantly lower pressure with the Breas PV 10i, the same efficacy may not have been achieved. This could have led to impairment of sleep quality, with subsequent lower compliance.
Interestingly, none of the APAP devices led to a change in quality of life, as assessed by the SF-36, or daytime sleepiness, indicating that APAP treatment does not provide additional benefits in these domains. This is of clinical importance as it indicates that patients already established on fixed-pressure CPAP therapy will not obtain further benefit by switching to an APAP device.
It is concluded from the present findings that the different APAP algorithms create pressure profiles that are not equally effective in terms of compliance, sleep quality and side-effects in patients established on CPAP therapy. A routine switching from fixed-pressure CPAP to APAP treatment cannot be recommended, but, if a patient wishes to try APAP treatment, the specific device needs to be carefully assessed in order to ensure treatment effectiveness.
Furthermore, the data do not support the routine prescription of auto-adjustable positive airway pressure as initial therapy for severe obstructive sleep apnoea syndrome, and the present authors recommend that, at present, auto-adjustable positive airway pressure therapy be reserved for patients having difficulty with standard continuous positive airway pressure, or selected patients for whom there is objective evidence of auto-adjustable positive airway pressure superiority, such as those requiring high fixed-pressure levels with continuous positive airway pressure 28.
| APPENDIX: IN-HOUSE EVALUATION QUESTIONNAIRE |
|---|
|
|
|---|
NAME: DATE:
HOSP. NO: DEVICE:
The answers to these questions are graded on a scale of 110 where 1 indicates a strongly negative response and 10 a strongly positive response.
Please complete this questionnaire by circling the number between 1 and 10 that best reflects your opinion.
A. WHAT DO YOU THINK ABOUT THE SIZE OF THE DEVICE?
1 2 3 4 5 6 7 8 9 10
(1 = too bulky) (10 = perfect size)
B. DO YOU FIND THE DEVICE EASY TO USE?
1 2 3 4 5 6 7 8 9 10
(1 = too difficult) (10 = very easy)
C. IS THE DEVICE VERY NOISY?
1 2 3 4 5 6 7 8 9 10
(1 = too loud) (10 = very quiet)
D. WHAT ABOUT THE COST OF THE DEVICE?
1 2 3 4 5 6 7 8 9 10
(1 = too expensive) (10 = excellent value)
E. WHAT IS YOUR OVERALL IMPRESSION OF THE DEVICE?
1 2 3 4 5 6 7 8 9 10
(1 = not impressed at all) (10 = very impressed)
F. HOW RESTFUL WAS YOUR SLEEP?
1 2 3 4 5 6 7 8 9 10
(1 = not restful at all) (10 = very restful)
G. HOW MUCH TROUBLE DID YOU HAVE GETTING TO SLEEP?
1 2 3 4 5 6 7 8 9 10
(1 = a lot of trouble) (10 = no trouble at all)
H. DID YOU HAVE TROUBLE STAYING ASLEEP DURING THE NIGHT?
1 2 3 4 5 6 7 8 9 10
(1 = a lot of trouble) (10 = no trouble at all)
I. ARE YOU HAVING PROBLEMS WITH A BLOCKED, RUNNY OR STINGING NOSE?
1 2 3 4 5 6 7 8 9 10
(1 = a lot of problems) (10 = no problems at all)
J. DID YOU HAVE A SORE OR DRY MOUTH OR THROAT?
1 2 3 4 5 6 7 8 9 10
(1 = very much so) (10 = not at all)
K. HOW MUCH DISCOMFORT DID YOU GET FROM THE PRESSURE?
1 2 3 4 5 6 7 8 9 10
(1 = a lot of discomfort) (10 = none at all)
The authors would like to thank Resmed (Abingdon, UK), Breas Medical (Mölnlycke, Sweden) and Respironics, Inc. (Murrysville, PA, USA) for kindly supplying the devices and all of the patients for participating in this study.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. H. Sanders, J. M. Montserrat, R. Farre, and R. J. Givelber Positive Pressure Therapy: A Perspective on Evidence-based Outcomes and Methods of Application Proceedings of the ATS, February 15, 2008; 5(2): 161 - 172. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Pevernagie, J. F. Masa, J. C. Meurice, R. Farre, O. Marrone, and D. Rodenstein Treatment of obstructive sleep-disordered breathing with positive airway pressure systems Eur. Respir. Rev., December 1, 2007; 16(106): 125 - 131. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |