Keywords: Autofluorescence thoracoscopy, lung cancer, malignant pleural disease, mesothelioma
Received: March 9, 2005
Accepted July 20, 2005
| ABSTRACT |
|---|
|
|
|---|
The present study displays early results of thoracoscopy performed consecutively with a normal light source and with autofluorescence light in 24 patients with exudative pleural effusion during 20032004.
In all cases of malignant pleuritis (carcinoma or mesothelioma), the colour of the affected area of the pleura changed from white/pink to red (sensitivity 100%). However, in two cases of chronic pleuritis, a colour change from white/pink to orange/red was recorded (specificity 75%).
In conclusion, the calculated positive predictive value of colour change for malignant pleuritis during autofluorescence thoracoscopy in this study was 92%. However, the clinical value of autofluorescence thoracoscopy in daily practice remains to be proven.
In 1981, Boutin et al. 1 reported that in a series of 1,000 consecutive patients with pleural effusions, thoracoscopy was indicated in 215 cases where it established the diagnosis in 97%. The diagnostic yield of medical thoracoscopy in malignant effusions is high: according to Loddenkemper and Boutin 2 it is 95%, whereas according to Blanc et al. 3 it is 93.3% and 90100% according to Colt 4. In any case, the sensitivity of thoracoscopy in detecting pleural malignancies is much higher than the sensitivity of percutaneous closed needle pleural biopsy (up to 50% 5), pleural fluid cytology (from 6280% when three separate specimens are collected 6, 7) or even when the two methods are combined (74% 8). Similar encouraging results have been obtained in studies on the role of thoracoscopy in the staging of patients with bronchial carcinoma or diffuse mesothelioma 2, 9.
The endoscopic appearance of pleural lesions during medical thoracoscopy is suggestive of malignancy in 86% of cases 10. Suggestive appearances include: nodules, polypoid lesions, masses, malignant thickening of the pleura, and localised "candle wax drops". However, there are malignancies which may resemble nonspecific inflammation and, conversely, certain inflammatory lesions can mimic tumours. Even mesotheliomas, which usually have a characteristic grape-like nodular appearance, can appear as ordinary inflammation.
To improve the diagnostic accuracy of thoracoscopy, autofluorescence was assessed in combination with the conventional method, as the effectiveness of autofluorescence techniques has been proven in both accurate detection of early malignant or invasive lesions and appropriate sampling of tissue during bronchoscopy. In addition, autofluorescence thoracoscopy (AFT) would be helpful for delineating the tumour margins and, therefore, for precise staging of intrathoracic lesions.
| PATIENTS AND METHODS |
|---|
|
|
|---|
AFT was performed in the endoscopy suite under local anaesthesia. The patients breathed spontaneously and 0.250.5 mg atropine and 10 mg morphine was used as the pre-medication. First, the entire pleural cavity in the examined hemithorax was thoroughly evaluated using conventional thoracoscopy (white light thoracoscopy; WLT), after removing all pleural fluid. The visceral and parietal pleura were inspected with WLT, and areas of abnormal pleura were mapped. The pleural cavity was then inspected with AFT and the change in colour of the areas of abnormal pleura was assessed. Furthermore, the pleural cavity was inspected with AFT for areas of abnormal pleura which were not noticed during WLT inspection. Biopsies of areas of abnormal pleura were subsequently taken for histology analysis. All thoracoscopic procedures under white or blue light were recorded on videotape and reviewed.
| THE VIDEOTHORACOSCOPIC AUTOFLUORESCENCE IMAGING DEVICE |
|---|
|
|
|---|
In the process of further development and technical evolution of the DAFE® system, it was recently discovered that the excitation wavelength yielding the highest sensitivity and specificity was >400 nm with a peak value near 405 nm, and that the positive predictive value (PPV) for the tumour detection endoscopies could be raised significantly from 38 to 100% by combining this method with the classical white light endoscopy 12.
With the exception of the imaging device as described before, the AFT system uses the same thoracoscopy equipment as WLT (Wolf thoracoscopy set; 7 mm optical telescope (as developed by C. Boutin); fluid light cables; and Wolf 5 mm biopsy forceps).
| RESULTS |
|---|
|
|
|---|
As shown in table 1
, in all cases of histologically proven malignant pleural disease (including nine patients with metastatic pleurae carcinomas and seven cases with mesotheliomas), the colour of the affected area of the pleura changed from white/pink to red in a darker or slighter attenuation (fig. 1
). Therefore, the sensitivity of AFT for detecting malignant lesions on the pleural surface was 100%.
|
|
| DISCUSSION |
|---|
|
|
|---|
100% sensitivity by precise assessment of disseminated intrathoracic tumour spread, was considered important for determining the resectability of the malignant lesions. Prosst et al. 17 induced diffuse pleural carcinosis affecting the entire pleural cavity of nude rats. After 57 weeks of tumour growth, a pleural lavage was performed with 1.5 or 3% ALA solution, followed by photosensitisation for 2, 4 or 6 h, after which a conventional white light VATS and a fluorescence examination using the D-Light (Karl Storz, Tuttlingen, Germany)/autofluorescence system were consecutively performed. The same researchers reported that the photosensitiser accumulation in the tumour was up to 11 times higher than in normal tissue and detected up to 30% additional pleural malignant lesions by thoracoscopic fluorescence diagnosis compared with conventional white light VATS alone. More recently, Noppen et al. 18 published a case report concerning fluorescein-inhanced autofluorescence thoracoscopy (FEAT) in primary spontaneous pneumothorax. The findings suggest substantial areas of parenchymal abnormality could be found using FEAT, which were undetected by WLT inspection of the parenchymal surface. Furthermore, satellite areas of parenchymal abnormalities could only be identified with FEAT.
Another recent report by Burgers et al. 19, studied 15 patients with suspected pleural malignancies using 5-aminolevulinic acid induced protoporphyrin IX fluorescence. No false positive fluorescence was observed, but 50% of cases had a false negative fluorescence signal.
A limitation of the current study was that biopsies were obtained from a rather restricted number of sites on the pleural space. Therefore, the estimations of the false negatives would not be accurate enough to be taken under consideration in calculating the sensitivity of the method. The ability to find a lesion if there was one (sensitivity) is the most important parameter to characterise the performance of a detection method, but requires the total number of false negatives to be known. It is impossible to have specimens from the entire pleural surface with the object of obtaining an approximately reliable value of false negatives.
The PPV, defined as the probability that a positive result represents a lesion, is thought to be a more realistic parameter for evaluation of AFT as a diagnostic method. In the group of patients who underwent AFT during the present study, the PPV, calculated as the ratio of the true positive results on the overall (true and false) positive results, was 92%.
In the authors' initial experience, AFT is easy to perform, without significant prolongation of the thoracoscopy procedure. In all patients with biopsy proven malignant pleural disease, either malignant mesothelioma or metastatic pleural carcinoma, a colour change from normal white or white/pink to red (in a deeper or slight attenuation) was found. In contrast to a sensitivity of 100% in patients with malignant effusions, two false positive results were found in patients with chronic pleuritis, which resulted in a specificity of 75% in this group.
In conclusion, an important limitation of the current study was that the patients had extended pleural disease at the moment of thoracoscopy, which was easy to diagnose with white light-guided biopsies. The same malignant lesions were easily found and more precisely mapped during autofluorescence thoracoscopy. The findings with blue light were more obvious than with white light, so it is expected that this new technique will be of help in cases without clear malignant appearance during white light thoracoscopy. To prove the additional value of autofluorescence thoracoscopy in clinical practice, a study including cases of malignant pleural disease in an early stage needs to be performed.