For affiliations, please see Acknowledgements section
CORRESPONDENCE: V. Brusasco, Internal Medicine, University of Genoa, V.le Benedetto XV, 6, I-16132 Genova, Italy. Fax: 10 3537690. E-mail: vito.brusasco@unige.it
Keywords: Diffusing capacity, infections, lung function measurements, lung volume measurements, reference values, spirometry standardisation
Received: March 23, 2005
Accepted April 5, 2005
| BACKGROUND |
|---|
|
|
|---|
|
| DEFINITIONS |
|---|
|
|
|---|
Accuracy is the closeness of agreement between the result of a measurement and the conventional true value.
Repeatability is the closeness of agreement between the results of successive measurements of the same item carried out, subject to all of the following conditions: same method, same observer, same instrument, same location, same condition of use, and repeated over a short space of time. In previous documents, the term reproducibility was used in this context, and this represents a change towards bringing this document in line with the ISO.
Reproducibility is the closeness of agreement of the results of successive measurements of the same item where the individual measurements are carried out with changed conditions, such as: method of measurement, observer, instrument, location, conditions of use, and time. Thus, if a technician tests a subject several times, this is looking at the repeatability of the test. If the subject is then given a bronchodilator drug and tested again after 30 min, one needs to know the reproducibility of the test in order to make a decision on this comparison.
The measurement range for a recording device is the range over which the manufacturer indicates the device complies with the recommendations below.
Equipment resolution is the smallest detectable change in measurement.
| PATIENT CONSIDERATIONS |
|---|
|
|
|---|
|
| PATIENT DETAILS |
|---|
|
|
|---|
Therapy
The operator should record the type and dosage of any (inhaled or oral) medication that may alter lung function and when the drugs were last administered.
Subject preparation
Subjects should avoid the activities listed in table 2
, and these requirements should be given to the patient at the time of making the appointment. On arrival, all of these points should be checked, and any deviations from them recorded.
|
Patients should be asked to loosen tight-fitting clothing. Dentures should normally be left in place; if they are loose, they may interfere with performance and are, therefore, best removed.
| LABORATORY DETAILS |
|---|
|
|
|---|
Ideally, when patients return for repeat testing (e.g. at a clinic), the equipment and the operator should be the same, and the time of day should be within 2 h of previous test times.
The order for performing lung function tests should take into account the optimum work flow in the laboratory, potential influences of one test on another and the ability of the subject to undertake the test. One possible order is shown in table 3
.
|
10% of patients to obtain a measurement of diffusing capacity that was not possible pre-bronchodilator 10. | HYGIENE AND INFECTION CONTROL |
|---|
|
|
|---|
Infection can be transmitted by direct contact or by indirect means, which is discussed as follows.
Transmission by direct contact
There is potential for transmission of upper respiratory diseases, enteric infections and blood-borne infections through direct contact. Although hepatitis and HIV contagion are unlikely via saliva, transmission becomes a possibility with open sores on the oral mucosa or bleeding gums. The most likely surfaces for contact are mouthpieces and the immediate proximal surfaces of valves or tubing.
Transmission by indirect contact
There is potential for transmission of tuberculosis (TB), various viral infections, opportunistic infections and nosocomial pneumonia through aerosol droplets. The most likely surfaces for possible contamination by this route are mouthpieces, proximal valves and tubing.
Prevention
Transmission to technicians
Prevention of infection transmission to technicians exposed to contaminated spirometer surfaces can be accomplished through proper hand washing and use of barrier devices, such as suitable gloves. To avoid technician exposure and cross-contamination, hands should be washed immediately after direct handling of mouthpieces, tubing, breathing valves or interior spirometer surfaces. Gloves should be worn when handling potentially contaminated equipment if the technician has any open cuts or sores on his/her hands. Hands should always be washed between patients. Indications and techniques for hand washing during pulmonary function testing have previously been reviewed 13.
Cross-contamination
To avoid cross-contamination, reusable mouthpieces, breathing tubes, valves and manifolds should be disinfected or sterilised regularly. Mouthpieces, nose clips and any other equipment that comes into direct contact with mucosal surfaces should be disinfected, sterilised or, if disposable, discarded after each use. The optimal frequency for disinfection or sterilisation of tubing, valves or manifolds has not been established. However, any equipment surface showing visible condensation from expired air should be disinfected or sterilised before reuse.
Since the use of cold sterilising agents is not without risk, laboratory staff should take care to follow the manufacturer's recommendations concerning proper handling of these products. Some respiratory equipment may be damaged by some methods of sterilisation. For example, heat sterilisation or cold sterilisation chemicals could damage some flow sensors, tubing or seals. Manufacturers should explicitly describe acceptable methods of cleaning and disinfecting their equipment, including recommended chemicals and concentrations, as well as safety precautions for the technicians. Manufacturers' recommendations should be followed; however, a hospital infection-control department's requirements will probably supersede both manufacturers' recommendations and those in this document. If hospital infection-control recommendations have the potential to harm instruments, compromises may have to be negotiated.
Volume-based spirometers
Volume-based spirometers used with a closed-circuit technique should be flushed between subjects with room air at least five times over the entire volume range of the spirometer to enhance clearance of droplet nuclei. The breathing tube and mouthpiece should be decontaminated or changed between patients.
When the open-circuit technique is used and the patient/subject only exhales into the spirometer, only the portion of the circuit through which rebreathing occurs must be decontaminated between patients. For example, when a pneumotachometer system is used, either avoid having the patient inspire through the device, or decontaminate or replace the resistive element and tubing between subjects. Alternatively, a disposable sensor may be used. Disposable sensors, when appropriately used, avoid the need for decontamination of sensors and mouthpieces (see Disposable in-line filters section).
When an open-circuit technique (either volume or flow spirometers) is used without inspiration from the measuring system, only the mouthpiece would need to be changed or decontaminated between subjects. However, it is difficult, if not impossible, to assure that patients do not inhale through the device. A low-resistance one-way valve may be used to prevent inhalation, and, if used, must be demonstrated not to alter the spirometric measurements. Not having patients inspire through the device may make it difficult to assess test quality because of the absence of an inspiratory tracing. Hence, this technique should be used with caution. Disassembling, cleaning and/or sensor replacement will usually require recalibration of the spirometer.
Tuberculosis
In settings where TB or other diseases that are spread by droplet nuclei are likely to be encountered, proper attention to environmental engineering controls, such as ventilation, air filtration or ultraviolet decontamination of air, should be used to prevent disease transmission.
Haemoptysis and oral lesions
Special precautions should be taken when testing patients with haemoptysis, open sores on the oral mucosa or bleeding gums. Tubing and breathing valves should be decontaminated before reuse, and internal spirometer surfaces should be decontaminated with accepted disinfectants for blood-transmissible agents.
Other known transmissible infectious diseases
Extra precautions should be taken for patients with known transmissible infectious diseases. Possible precautions include the following: 1) reserving equipment for the sole purpose of testing infected patients; 2) testing such patients at the end of the day to allow time for spirometer disassembly and disinfection; and 3) testing patients in their own rooms with adequate ventilation and appropriate protection for the technician.
Disposable in-line filters
These may be an effective and less expensive method of preventing equipment contamination. The influence of commercially available in-line filters on forced expiratory measures, such as forced vital capacity (FVC) and forced expiratory volume in one second (FEV1) has not been well characterised. A low-impedance barrier device was found not to have a significant effect on FVC and FEV1 14, whereas a barrier filter has been shown to cause small but significant reductions in FEV1 (-44 mL) and peak expiratory flow (PEF; -0.47 L·s1), but did not appear to affect DL,CO, alveolar volume or TLC 15. Although significant differences between measurements with and without filters have been demonstrated for FVC, FEV1, airway resistance and specific airway conductance (sGaw) 16, these differences were unrelated to the average values of the measurements (except for sGaw), and the limits of agreement were within the range of intra-individual short-term repeatability for almost all of the function indices. Thus, the effect of a filter with optimal characteristics is not considered to be clinically significant, and no appreciable classification error was found in diagnostic tests.
If in-line filters are used, the measuring system should meet the minimum recommendations for accuracy, precision (reproducibility), flow resistance and back pressure with the filter installed. Airflow resistance must be measured with in-line filters in place if that is how patients are tested. Manufacturers of in-line filters should provide evidence that their filter does not alter standard lung function measurements (vital capacity, FVC, FEV1, PEF, mean forced expiratory flow between 25% and 75% of FVC, TLC and DL,CO).
In the absence of evidence for infection transmission during pulmonary function testing, and the absence of a clear-cut benefit, the regular use of in-line filters is not mandated when the precautions described in the previous Prevention sections are followed.
Use of such filters is an area of controversy. On the one hand, some spirometric equipment, particularly those incorporated in multipurpose testing systems, employ valve manifolds, which are situated proximal to breathing tubes. These valve arrangements provide internal surfaces on which the deposition of expired aerosol nuclei is likely. Given their complexity, they may be difficult to disassemble and disinfect between subjects. To the extent that in-line filters have been shown to remove microorganisms from the expiratory air stream and, thus, prevent their deposition as aerosol nuclei on spirometer surfaces, their use may be indicated. On the other hand, in-line filters have been relatively inefficient in excluding microorganisms at the high flows often seen in pulmonary testing, and instrument contamination has been observed when filters have been used 1720. However, barrier filters with a high efficiency (>99%) for excluding bacteria have been reported 21, 22, but their performance in excluding smaller microorganisms such as viruses is unknown. A reduction in overall costs with in-line filters, as compared with a disinfection approach to hygiene, in a pulmonary laboratory has been reported 17.
The use of in-line filters does not eliminate the need for regular cleaning and decontamination of lung function equipment.
Equipment design
Manufacturers of lung function equipment are encouraged to focus on designs that can be easily disassembled for cleaning and disinfection. Purchasers of pulmonary function equipment are encouraged to inquire about cleaning and disinfection issues prior to purchase of an instrument, which should involve an evaluation of the ease of cleaning and the clarity of written instructions, and an understanding of what equipment and chemicals will be required.
Level of infection risk
Lung function equipment has not been directly implicated in the transmission of infections, although there is indirect evidence of infection transmission during pulmonary function testing. Organisms from the respiratory tract of test subjects have been recovered from mouthpieces and the proximal surfaces of tubing through which subjects breathe 19, 23. The flows generated during spirometric manoeuvres may be high enough to aerosolise contaminant organisms, although such aerosolisation has not been demonstrated. There is one case report of a TB skin-test conversion following exposure to a spirometer previously used to test a patient with documented TB 24. Likewise, there is circumstantial evidence that contaminated lung function equipment may be implicated in increasing the prevalence of Burkholderia cepacia infections among cystic fibrosis patients at one centre 25. There is evidence that pneumotachometer-based systems are less susceptible to bacterial contamination than water-sealed spirometers 26. In addition, it is well documented that community water supplies can be contaminated with Mycobacteria spp. and Pseudomonas aeruginosa organisms 2729. Thus, there is a potential for both patients/subjects and healthcare workers to deposit microorganisms onto spirometer surfaces (including mouthpieces, nose clips, tubing and any internal or external machine surface), which could subsequently come into direct or indirect contact with other patients or healthcare workers.
This does not pose an appreciable threat to patients/subjects/workers with competent immune systems. It has been argued that immunocompromised patients may require only a relatively small infective dose of either opportunistic organisms or common pathogens for infection to occur. However, there is no direct evidence that routine pulmonary function testing poses an increased risk of infection to immunocompromised patients.
Concerns for the protection of immunocompromised patients, along with increased public and provider awareness of hospital infection-control issues since the 1990s, has led many laboratory directors to routinely use in-line filters to reassure patients and laboratory personnel that their protection has been considered.
| PERSONNEL QUALIFICATIONS AND TECHNICIAN'S ROLE IN QUALITY CONTROL |
|---|
|
|
|---|
The current guidelines suggest that completion of secondary education and at least 2 yrs of college education would be required to understand and fulfil the complete range of tasks undertaken by a pulmonary function technician.
For pulmonary function testing, an emphasis on health-related sciences (nursing, medical assistant, respiratory therapy, etc.) is desirable. Formal classroom-style training alone does not, however, establish competency in pulmonary function testing. Technicians who conduct pulmonary function testing need to be familiar with the theory and practical aspects of all commonly applied techniques, measurements, calibrations, hygiene, quality control and other aspects of testing, as well as having a basic background knowledge in lung physiology and pathology. In the USA, the National Institute for Occupational Safety and Health (NIOSH) has developed a model programme, and reviews and approves spirometry training courses. These 2- and 3-day courses include the fundamentals of spirometry standards and hands-on training. The workshop experience provides hands-on instruction in a small group setting with an experienced instructor. Students are expected to demonstrate their ability to properly prepare and administer a spirometric test, and demonstrate competency in other areas, such as calibration, recognition of unacceptable manoeuvres, etc.
This standard recommends training similar to the NIOSH-approved spirometry programme. Competency is demonstrated by passing a written and practical examination in the presence of an experienced instructor (i.e. hands-on testing and calibration). In Europe, training is being carried out differently in various countries. The ERS, through a specific Assembly (Assembly 9 for Allied Respiratory Professionals), regularly delivers relevant postgraduate course training at their annual Congress.
Spirometry refresher training is also recommended. Refresher training helps to ensure that testing technicians are informed of changes in spirometry standards and learn new skills. It also provides a mechanism for technicians to obtain answers to questions not foreseen during initial training. The need for refresher training has been recognised by several organisations, including the Lung Health Study 32, the National Health and Nutrition Examination Survey 33, 34 and the American College of Occupational and Environmental Medicine 35. The frequency of refresher training is dependent on many factors that differ among individuals. A recommended frequency of every 35 yrs is recommended, or shortly after changes to lung function standards are published. While in-house training may achieve the desired goals, laboratory directors should strongly consider the benefits of formal training programmes from outside providers.
Technician's role in quality control
Quality control is important to ensure that the laboratory is consistently meeting appropriate standards. In any quality-control programme, an important element is a manual of procedures that contains the following: calibration procedures, test-performance procedures, calculations, criteria, reference values source, and action to be taken when "panic" values are observed. A notebook or an equivalent method of recording and later producing these results should be maintained, which documents the daily instrument calibration, as well as any problems encountered with the system, corrective action required, and system hardware and software upgrades. Records of anomalous events involving either patients/subjects or the technician should be documented with the results of subsequent evaluation and responses to the event. The technician should also maintain records of continuing education and the results of evaluation and feedback provided by the medical director. The ATS has produced a complete procedure manual (Pulmonary Function Laboratory Management and Procedure Manual), which is available in paper and electronic format (www.thoracic.org/education/labmanual.asp), so as to be modifiable by laboratories to meet their individual needs.
In Europe, technical information on lung function tests is contained in a series of publications in the European Respiratory Journal 3642.
Perhaps the most important component in successful pulmonary function testing is a well-motivated, enthusiastic technician. The importance of a quality-control programme with feedback to technicians in obtaining adequate spirometry results has been documented 32. A quality-control programme that continuously monitors technician performance is critical to the collection of high-quality data. Feedback to the technicians concerning their performance should be provided on a routine basis, which should include, at a minimum: 1) information concerning the nature and extent of unacceptable manoeuvres and nonreproducible tests; 2) corrective action that the technician can take to improve the quality and number of acceptable manoeuvres; 3) positive feedback to technicians for good performance; and 4) comments regarding system set-up and reporting results.
Manufacturers are encouraged to include quality-control aids in their software packages. However, technicians should be trained not to rely exclusively on these quality-control prompts, since technical errors may occur that are not among those recognised by the software. An example of a quality-control aid is a calibration-logging program, which stores the date, time, technician name and the results of routine daily calibration checks. Additionally, the program could issue a warning if an acceptable daily calibration check has not been performed.
| REFERENCE VALUES |
|---|
|
|
|---|
| INTERPRETATION STRATEGIES |
|---|
|
|
|---|
The interpretation of lung function tests involves two tasks: 1) the classification of the derived values with respect to a reference population and assessment of the reliability of the data; and 2) the integration of the obtained values into the diagnosis, therapy and prognosis for an individual patient.
The first task is ordinarily the responsibility of the laboratory director or his/her designee, and not only serves to communicate information to referring healthcare providers, but is also an important aspect of laboratory quality control. The second task is usually the responsibility of the physician requesting the studies and is performed within the context of patient care.
It is the responsibility of the laboratory director to develop explicit procedures for the interpretation of lung function tests and to select appropriate reference values. The procedures for interpretation and choosing reference values may legitimately vary from laboratory to laboratory, depending upon geographical location and the characteristics of the population being tested. The interpretative strategy should be consistent and take into consideration the consequences of false-positive and false-negative errors. In this way, referring physicians will not infer a change in the condition of the patient from a change in interpretation when it is, in fact, the result of a change in the approach of the interpreting physician.
| ABBREVIATIONS |
|---|
|
|
|---|
|
| ACKNOWLEDGEMENTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. Lucas, F. Smeenk, I. Smeele, and C. van Schayck Overtreatment with inhaled corticosteroids and diagnostic problems in primary care patients, an exploratory study Fam. Pract., April 1, 2008; 25(2): 86 - 91. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Ofir, P. Laveneziana, K. A. Webb, Y.-M. Lam, and D. E. O'Donnell Mechanisms of Dyspnea during Cycle Exercise in Symptomatic Patients with GOLD Stage I Chronic Obstructive Pulmonary Disease Am. J. Respir. Crit. Care Med., March 15, 2008; 177(6): 622 - 629. [Abstract] [Full Text] [PDF] |
||||
![]() |
K.-L. Huang, C.-W. Chen, S.-J. Chu, W.-C. Perng, and C.-P. Wu Systemic Inflammation Caused by White Smoke Inhalation in a Combat Exercise Chest, March 1, 2008; 133(3): 722 - 728. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Cazzola, W. MacNee, F. J. Martinez, K. F. Rabe, L. G. Franciosi, P. J. Barnes, V. Brusasco, P. S. Burge, P. M. A. Calverley, B. R. Celli, et al. Outcomes for COPD pharmacological trials: from lung function to biomarkers Eur. Respir. J., February 1, 2008; 31(2): 416 - 469. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Kotz, P. Nelemans, C. P. van Schayck, and G. J. Wesseling External validation of a COPD diagnostic questionnaire Eur. Respir. J., February 1, 2008; 31(2): 298 - 303. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. G. Barr, S. Mesia-Vela, J. H. M. Austin, R. C. Basner, B. M. Keller, A. P. Reeves, D. Shimbo, and L. Stevenson Impaired Flow-mediated Dilation Is Associated with Low Pulmonary Function and Emphysema in Ex-smokers: The Emphysema and Cancer Action Project (EMCAP) Study Am. J. Respir. Crit. Care Med., December 15, 2007; 176(12): 1200 - 1207. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Viegi, F. Pistelli, D. L. Sherrill, S. Maio, S. Baldacci, and L. Carrozzi Definition, epidemiology and natural history of COPD Eur. Respir. J., November 1, 2007; 30(5): 993 - 1013. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. von Leupoldt, K. Taube, S. Schubert-Heukeshoven, H. Magnussen, and B. Dahme Distractive Auditory Stimuli Reduce the Unpleasantness of Dyspnea During Exercise in Patients With COPD Chest, November 1, 2007; 132(5): 1506 - 1512. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Ramsey, R. Mehra, and K. P. Strohl Variations in Physician Interpretation of Overnight Pulse Oximetry Monitoring Chest, September 1, 2007; 132(3): 852 - 859. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Normand, F. Normand, X. Le Coutour, M-A. Metges, and A. Mouadil Clinical evaluation of a screen pneumotachograph as an in-line filter Eur. Respir. J., August 1, 2007; 30(2): 358 - 363. [Abstract] [Full Text] [PDF] |
||||