1 First Dept of Medicine, Hokkaido University School of Medicine, Sapporo, 2 Dept of Medicine, Iwamizawa Municipal General Hospital, Iwamizawa, and 3 Dept of Radiology, Hokkaido University School of Medicine, and 4 Dept of Surgery, National Hiroshima Hospital, Higashi-Hiroshima, Japan.
CORRESPONDENCE: K. Yamazaki, First Dept of Medicine, Hokkaido University School of Medicine, North 15, West 7, Kitaku, Sapporo, 060-8638, Japan. Fax: 81 117067899. E-mail: kyamazak@med.hokudai.ac.jp
Keywords: Bronchial brushing cytology, endobronchial ultrasonography, guide-sheath, peripheral pulmonary lesion, primary lung cancer, transbronchial biopsy
Received: December 15, 2003
Accepted June 23, 2004
| Abstract |
|---|
|
|
|---|
EBUS-GS-guided TBB was performed in 24 patients with 24 PPLs of
30 mm in diameter (average diameter=18.4 mm). A 20-MHz radial-type ultrasound probe, covered with GS was inserted via a working bronchoscope channel and advanced to the PPL in order to produce an EBUS image. The probe with the GS was confirmed to reach the lesion by EBUS imaging and X-ray fluoroscopy. When the lesion was not identified on the EBUS image, the probe was removed and a curette was used to lead the GS to the lesion. After localising the lesion, the probe was removed, and TBB and bronchial brushing were performed via the GS.
Nineteen peripheral lesions (79.2%) were visualised by EBUS. All patients whose PPLs were visible on EBUS images subsequently underwent an EBUS-GS-guided diagnostic procedure. A total of 14 lesions (58.3%) were diagnosed. Even when restricted to PPLs <20 mm in diameter, the diagnostic sensitivity was 53%.
In conclusion, endobronchial ultrasonography with guide sheath-guided transbronchial biopsy was feasible and effective for diagnosing peripheral pulmonary lesions.
The investigation of a peripheral pulmonary lesion (PPL) is a challenging diagnostic problem. A transbronchial approach using a fiberoptic bronchoscope (FB) under X-ray fluoroscopic guidance has been the most generally accepted method for diagnosing PPLs since the 1970s 13. However, the accuracy of diagnosing PPLs using FB under X-ray fluoroscopic guidance is reportedly 1471% 18. One factor that potentially limits the diagnostic accuracy of the standard bronchoscope is lesion size, as lesions <2 cm have very low yields ranging 1142% 610. Some respiratory physicians prefer to diagnose small PPLs from tissue samples obtained by percutaneous needle aspiration cytology or biopsy 1115. Although success rates of these techniques might be very high, with 7697% diagnostic accuracy 6, 15, these techniques have several problems. First, they have the potential to spread malignant cells from the tumour into the pleural cavity 11, 12. For patients with poor pulmonary function, these techniques result in an increased risk of pneumothorax. Moreover, systemic arterial air embolism is a rare but severe complication 13, 14.
Therefore, in many institutions, patients with small peripheral pulmonary lesions undergo surgical biopsy procedures, such as video-assisted thoracic surgery, although such invasive procedures are undesirable for elderly patients and patients with poor respiratory function. In order to spare patients from these operative procedures, new imaging and guidance technology is necessary.
Recently developed small-calibre ultrasound probes have enabled wider clinical application of ultrasonography (US) to tracheal and bronchial lesions. Guided by FB, a small-calibre US probe can be successfully introduced into the trachea and bronchus for assessment and diagnosis of endobronchial lesions or tumour metastasis to the mediastinal or hilar lymph nodes 1618. Transbronchial needle aspiration under endobronchial ultrasonography (EBUS) guidance has improved the N-staging yields for lung cancer without clear endoscopic landmarks 1720. EBUS has also been applied to the assessment of PPLs 16, 21. Kurimoto et al. 22 analysed EBUS images with the aim of distinguishing between benign and malignant disease, identifying the type of lung cancer, and determining the degree of differentiation. Moreover, Herth et al. 23 applied EBUS as imaging guidance for transbronchial biopsy (TBB) of PPLs. In the present study, the EBUS was further combined with the guide-sheath (GS) technique. Biopsy forceps covered with a GS can be moved to the lesions under EBUS guidance, after which biopsy and brushing specimens can be sequentially obtained by keeping the GS in the lesion. The present study was undertaken in order to examine the usefulness of a new technique, EBUS with GS (EBUS-GS), as a guide for TBB and bronchial brushing cytology for diagnosis of PPLs.
| Methods |
|---|
|
|
|---|
30 mm in mean diameter referred for diagnostic bronchoscopy were enrolled. PPLs were defined as those that were surrounded by pulmonary parenchyma and not visible by bronchoscopy (no evidence of endobronchial lesion, extrinsic compression, submucosal tumour, or narrowing, inflammation or bleeding of the bronchus). All chest computed tomographs (CT) were reviewed and the mean diameters of the PPLs were recorded. After written informed consent, patients underwent the bronchoscopic procedure.
EBUS-GS-guided TBB and bronchial brushing
Each patient was pre-medicated using 15 mg of pentazocine hydrochloride and 0.5 mg of atropine sulphates as a routine procedure for TBB in the two hospitals described above. Local anaesthesia of the upper respiratory tract was achieved using 4% lidocaine. EBUS was performed using an endoscopic ultrasound system (EU-M30S; Olympus, Tokyo, Japan), equipped with a 20-MHz mechanical radial-type probe (XUM-S20-17R; Olympus), having an external diameterof 1.4 mm (fig. 1
). FBs with a working channel of 2.0 mm in diameter were used (BF-P-260F, BF-P-240, BF-P-200; Olympus). The probe was inserted into a GS, and the GS-covered probe was inserted through the working channel of the FB and advanced to the peripheral pulmonary lesion in order to obtain an EBUS image. The probe and GS were confirmed to have reached the lesion by EBUS imaging and X-ray fluoroscopy. After localising the lesion using EBUS imaging, the probe was removed and the GS remained in the peripheral lesion. Biopsy forceps and a bronchial brush were introduced via the GS to perform pathological and cytological examination (fig. 2
ac). When the lesion was not identified on the EBUS image, a double-hinged curette was inserted intothe GS, and the appropriate bronchus was selected by manipulating the curette. The curette could be flexed and rotated by the proximal handle of the curette. When the lesion was not identified on the EBUS image or next to the EBUS image, using standard procedure, the GS was pulled, the next bronchus in the proximal position was selected by curette-loaded GS and the curette was manipulated. Once the bronchus was determined, the curette was removed. The probe was then inserted through the GS and the acquired EBUS images were again examined. These procedures were repeated by trial-and-error method until the probe and GS were found to have reached the correct position (fig. 2
dg).
|
|
| Results |
|---|
|
|
|---|
EBUS-GS-guided transbronchial biopsy and bronchial brushing
In 19 patients (79.2%) EBUS detected the PPL. In 10 patients the EBUS probe reached the peripheral lesions with curette navigation. All patients whose PPLs were visible by EBUS imaging subsequently underwent EBUS-GS-guided TBB and bronchial brushing. The mean number of specimens obtained using EBUS-GS-guided TBB was 3.5±1.3. Diagnosis was made in 14 cases (58.3%); 10 cases of primary lung cancer, one case of metastatic lung cancer of the thyroid, one case of metastatic lung cancer of the uterus and two cases of benign disease (pneumonia, one case; pneumoconiosis, one case). Pathological diagnosis was performed in 11 lesions (45.8%) and cytological diagnosis was performed in 11 lesions (45.8%) (table 1
). A total of five lesions that were visible by EBUS and for which diagnoses could not be obtained had insufficient biopsy and brush samples taken. When evaluating the position of EBUS against the 19 lesions visible by EBUS, the US probe was located in the bronchus inside the lesions in 14 cases (12 malignant diseases, two benign diseases). In 13 of these cases, cytopathological diagnosis was obtained. The US probe was located in the bronchus adjacent to the lesion in five cases (two malignant diseases, three benign diseases), and in one of these malignant cases cytopathological diagnosis was obtained.
|
|
When evaluated according to size (table 1
), 15 lesions were <20 mm in mean diameter. Ten lesions (66.7%) were visible on EBUS images and eight lesions (53.3%) were diagnosed using this procedure (six lesions by pathological diagnosis and five lesions by cytological diagnosis). In nine lesions with mean diameters of 2030 mm, all lesions were visible on EBUS images and six lesions (66.7%) were diagnosed (fivelesions by pathological diagnosis and six lesions by cytological diagnosis). The diagnostic accuracy of the EBUS-GS-guided bronchoscopic procedure in malignant and benign lesions was 66.7% and 33.3%, respectively.
When restricted to malignant lesions, the diagnostic sensitivity for a lesion <20 mm in mean diameter was 60%. There were no significant differences in diagnostic sensitivity of EBUS-GS with regard to bronchial segment.
Complications
One patient developed pneumothorax (4.2%) which was treated by tube thoracostomy. No major bleeding occurred during the diagnostic procedure.
| Discussion |
|---|
|
|
|---|
The present study has demonstrated that EBUS-GS-guided TBB and bronchial brushing were useful for evaluating PPLs. EBUS imaging shows the precise location of the pulmonary lesion and EBUS-GS enables diagnostic specimens to be obtained from the most appropriate position. The combination of EBUS and the curette-loaded GS technique has enabled visualisation of PPLs that are not accessible with a probe and allowed biopsy specimens to be obtained from PPLs that are not accessible by regular techniques. In the current study, 10 lesions were accessible with the use of the curette. Moreover, three lesions could not be detected by X-ray fluoroscopy in the present report, of which one lesion was diagnosed as adenocarcinoma by EBUS-GS. The diagnostic sensitivity of EBUS-GS-guided TBB and bronchial brushing for diagnosing PPLs
30 mm in mean diameter was 58.3% in the present study. When restricted to lesions having a mean diameter of <20 mm, the diagnostic sensitivity of this procedure was as high as 53.3%, and 40.0% of the lesions were diagnosed pathologically. The accuracy of diagnosing PPLs <20 mm in diameter using FB under X-ray fluoroscopy has been reported to be a very low, ranging 1142% 610. At the participating institute in this study, the diagnostic sensitivity of FB for small PPLs <20 mm in diameter under X-ray fluoroscopic guidance over the past year was 35% and the rate for obtaining diagnostic biopsy samples was as low as 13% (data not shown). Two cases of benign disease (one pneumoconiosis and one pneumonia) and two case of metastatic lung cancer were diagnosed. In these four cases, unnecessary surgical procedures could have been circumvented as a result of the diagnosis obtained by EBUS-GS.
Several improvements may be necessary in order to increase the yield of EBUS-GS-guided TBB and bronchial brushing. In the present study, 79.2% of PPLs were detected by EBUS imaging. However, five lesions (20.8%) that were visible by EBUS could not be diagnosed using the EBUS-GS-guided bronchoscopic procedure. This is largely because the biopsy forceps and bronchial brushes are small (diameter 1.2 mm) and sufficient specimens could not be obtained. In addition, the position of EBUS may affect the result. In the present study, out of 14 cases in which the US probe was located in the bronchus inside the lesion, 13 cases were diagnosed cytopathologically. In contrast, out of five cases in which the US probe was located in the bronchus adjacent to the lesion, only one case was diagnosed cytopathologically.
The authors of the present study normally experienced <1 mL of bleeding and coagulated blood was occasionally trapped in the GS, but bleeding outside of the GS was hardly seen even when the GS was removed. It is assumed that the wedged GS in the bronchus leading to the biopsied lesion is useful for stopping bleeding.
GS dislocation is occasionally encountered. The GS was occasionally dislocated from the target bronchus when the US probe or biopsy forceps was pulled out strongly from the GS or when a patient drew a deep breath. GS dislocation could be minimised when the procedure was performed with the bronchoscope wedging into the subsegment bronchus. Even in these undiagnosed lesions, the EBUS images were extremely useful to identify the appropriate bronchi during subsequent conventional bronchoscopic procedures using thicker forceps and bronchial brushes.
The benefits of EBUS-GS for TBB and bronchial brushing are as follows: 1) to confirm the precise location of PPLs by EBUS imaging even when such lesions are not visible on X-ray fluoroscopy; 2) to facilitate obtaining biopsy and brushing specimens repeatedly by leaving the GS in the PPLs; 3) to obtain biopsy specimens from PPLs that are accessible only through the use of a curette via the GS; 4) to decrease bleeding resulting from trapping the GS in the bronchus; and 5) to assess the internal structure of PPLs. Points 2), 3) and 4) are additional values of the GS technique above EBUS alone.
In summary, endobronchial ultrasonography with guide-sheath-guided transbronchial biopsy and bronchial brushing was performed safely and was found to be effective for diagnosing peripheral pulmonary lesions. The endobronchial ultrasonography and guide-sheath technique is a very promising new modality for diagnosing small peripheral pulmonary lesions. A randomised, prospective study of diagnosing peripheral pulmonary lesions and examining cost effectiveness by bronchoscopy, with or without endobronchial ultrasonography with guide-sheath, is currently being planned.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. Ost, R. Shah, E. Anasco, L. Lusardi, J. Doyle, C. Austin, and A. Fein A Randomized Trial of CT Fluoroscopic-Guided Bronchoscopy vs Conventional Bronchoscopy in Patients With Suspected Lung Cancer Chest, September 1, 2008; 134(3): 507 - 513. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Oki, H. Saka, C. Kitagawa, S. Tanaka, T. Shimokata, K. Mori, and S. Kajikawa Novel thin bronchoscope with a 1.7-mm working channel for peripheral pulmonary lesions Eur. Respir. J., August 1, 2008; 32(2): 465 - 471. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Yamada, K. Yamazaki, N. Kurimoto, H. Asahina, E. Kikuchi, N. Shinagawa, S. Oizumi, and M. Nishimura Factors Related to Diagnostic Yield of Transbronchial Biopsy Using Endobronchial Ultrasonography With a Guide Sheath in Small Peripheral Pulmonary Lesions Chest, August 1, 2007; 132(2): 603 - 608. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Eberhardt, D. Anantham, A. Ernst, D. Feller-Kopman, and F. Herth Multimodality Bronchoscopic Diagnosis of Peripheral Lung Lesions: A Randomized Controlled Trial Am. J. Respir. Crit. Care Med., July 1, 2007; 176(1): 36 - 41. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yoshikawa, N. Sukoh, K. Yamazaki, K. Kanazawa, S.-i. Fukumoto, M. Harada, E. Kikuchi, M. Munakata, M. Nishimura, and H. Isobe Diagnostic Value of Endobronchial Ultrasonography With a Guide Sheath for Peripheral Pulmonary Lesions Without X-Ray Fluoroscopy Chest, June 1, 2007; 131(6): 1788 - 1793. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Eberhardt, D. Anantham, F. Herth, D. Feller-Kopman, and A. Ernst Electromagnetic Navigation Diagnostic Bronchoscopy in Peripheral Lung Lesions Chest, June 1, 2007; 131(6): 1800 - 1805. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. R. Gildea, P. J. Mazzone, D. Karnak, M. Meziane, and A. C. Mehta Electromagnetic Navigation Diagnostic Bronchoscopy: A Prospective Study Am. J. Respir. Crit. Care Med., November 1, 2006; 174(9): 982 - 989. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Asahina, K. Yamazaki, Y. Onodera, E. Kikuchi, N. Shinagawa, F. Asano, and M. Nishimura Transbronchial Biopsy Using Endobronchial Ultrasonography With a Guide Sheath and Virtual Bronchoscopic Navigation Chest, September 1, 2005; 128(3): 1761 - 1765. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||