Copyright ©ERS Journals Ltd 2004 Scintigraphic measurement of tracheal mucus velocity in vivo1 Dept of Thoracic Medicine, and 2 Dept of Nuclear Medicine, Concord Repatriation General Hospital, Concord, and 3 Anatomy and Cell Biology, University of Melbourne, Parkville, Australia CORRESPONDENCE: L. Morgan, Dept of Thoracic Medicine (C31), Concord Hospital, Hospital Rd, Concord NSW 2139, Australia. Fax: 61 297677605. E-mail: morganl@email.cs.nsw.gov.au Keywords: chronic obstructive pulmonary disease, cilia, mucociliary clearance, tracheal mucus velocity
Received: July 10, 2002
Mucociliary clearance (MCC) is an important mechanism for removing inhaled particles, secretions and cellular debris from the respiratory tract. Here, a direct measurement of tracheal mucus velocity (TMV) for assessment of MCC, suitable for clinical and research use, is reported, and a comparison is made of TMV in normal subjects and patients with chronic obstructive pulmonary disease (COPD). A 0.1-mL bolus of radiolabelled (25 MBq), technetium-labelled macroaggregated human albumin (99mTc MAA) was injected through the cricothyroid membrane into the trachea of 20 young (<50 yrs) and 12 older (>50 yrs) normal subjects and 34 patients with COPD. Repeat studies were carried out in 13 normal subjects and 16 COPD patients. Movement of the bolus in the trachea was recorded (15 min) using a gamma camera interfaced to a computer. Data were analysed using specifically designed software. The test was well tolerated by subjects and patients, and no significant adverse events were reported. No significant differences were observed between data recorded from different regions of the bolus (leading edge, peak, trailing edge) by analysis of variance (ANOVA). Bland-Altman plots of the repeat studies indicated that data were more variable in normal subjects (coefficient of repeatability (COR) 10.3 mm·min1) than in COPD patients (COR 5.5 mm·min1). TMV (mean±sd) in young normal subjects (n=20) was 10.7±3.5 mm·min1. TMV was reduced in older normal subjects (n=12; 6.5±2.6 mm·min1) and further reduced in COPD (n=34; 2.1±2.7 mm·min1). In conclusion, this technique can be used to measure tracheal mucus velocity rapidly and safely in healthy subjects and patients with respiratory tract disease. This study has confirmed that tracheal mucus velocity declines with age and is further impaired in patients with chronic obstructive pulmonary disease. Mucociliary clearance (MCC) is an important defence mechanism in the respiratory tract, resulting in the removal of inhaled particles and respiratory secretions. Impaired MCC is known to predispose to bronchial infection and may lead to the development of chronic lung diseases such as bronchiectasis. Measurement of MCC allows a better understanding of such chronic lung diseases and may provide an objective measure of response to therapeutic strategies. MCC is most commonly measured using inhaled radiolabelled aerosols (IRLA) and scintigraphy. This is a valuable technique but has some limitations. Optimal large airway deposition requires specialised aerosol generation and is affected by airway disease. The acquisition period is relatively long, limiting the use of the test in breathless subjects or those with cough. Data analysis requires correction for isotope decay and signal attenuation. Furthermore, clearance of aerosol occurs in three dimensions and the analysis is based on a two-dimensional representation of the lungs. As a result, the equipment required and the technical features of IRLA studies restrict its use to specialised units. Some of these problems can be circumvented by direct measurement of particle clearance from the trachea. Measurement of tracheal mucus velocity (TMV) has been investigated using a variety of markers 13, deposition techniques 46 and detection systems 7. However, no technique has proved sufficiently practical or reliable for use in clinical practice. The aim of this project was to develop a simple, reliable and reproducible technique for direct measurement of TMV that could be used easily in a clinical setting. This technique was used to compare TMV in normal subjects with that in subjects with COPD.
Subjects Twenty young (<50 yrs) and 12 older (>50 yrs) normal subjects, and 34 patients with COPD were studied. All normal subjects were nonsmokers with no history of lung disease. COPD patients were current or ex-smokers of 20 pack-yrs, with a forced expiratory volume in one second (FEV1) <70% predicted and FEV1/forced vital capacity ratio <70% (table 1
Patients were instructed not to use short-acting bronchodilators for 4 h prior to study. All procedures in this study were performed after obtaining informed consent. The Institutional Ethics Committee of Central Sydney Area Health Service (CRGH Zone) approved the study.
Radiolabelling
Bolus droplet deposition
Measurement
Analysis Further analysis of the movement of the peak of activity was made using CiliaC© (RSI Inc., Boulder, CO, USA), a software program written in interactive data language specifically for this project. CiliaC follows the peak of activity by applying a two-dimensional Gaussian fit to each image to find the peak location. Peak positions were plotted as a function of time and then smoothed with a polynomial function. A reference point was selected arbitrarily by the operator on the smoothed curve midway along the steepest part of the slope (the region of highest velocity). CiliaC provides a semi-automated calculation of velocity at that point and then identifies the frame on either side of the reference point at which this slope changes by >20%. The midpoint between these limits is chosen automatically as a new reference point, and the process is repeated until the start and finish of the region of steepest slope has been delineated. Maximum velocity was calculated as the gradient of this fitted line. Measurements were accepted only from segments of the trace >12 frames (2 min) in duration. To determine whether different regions of the bolus moved at different rates over the course of the study, the TMV of the leading and trailing edges were measured, as well as the apparent peak of activity. As analysis of variance showed no significant regional variability, peak activity measurements were used subsequently. In order to determine the repeatability of this test, repeat TMV measurements of 13 young normal subjects and 16 patients with COPD were carried out at least 1 week after the first study.
Statistics
A total of 95 studies were completed (45 in normal subjects and 50 in patients with COPD including repeat studies). All were well tolerated, even in those subjects with severe COPD. One normal subject developed minor bruising at the injection site and two reported minor discomfort on swallowing for up to 12 h after injection. Cough at the time of injection occurred in five studies and was short lived. There were no other adverse events.
A representative TMV trace showing the movement of the peak of activity of the radiolabelled bolus along the tracheal mucosa of a normal subject is shown in figure 1
In some studies the shape of the bolus changed during the 15-min acquisition period, dividing into separate, smaller boluses that subsequently re-coalesced. To ensure that this did not influence the measurement, separate measurements were obtained from three regions of the bolus (leading edge, trailing edge and peak of activity) in 38 studies of normal subjects and 21 studies of COPD patients. No significant differences were observed between the mean velocities for each region in the young normal subjects ((mean±sd) 10.0±4.2, 10.7±4.6 and 10.4±4.1 mm·min1, respectively) or in the COPD patients (1.5±2.4, 1.3±2.5 and 1.4±2.4 mm·min1, respectively) by ANOVA or by t-test (p>0.05).
Analysis of repeated measurements of TMV, made on separate occasions in 13 young normal subjects and 16 COPD patients, by Bland-Altman plots (fig. 2
Figure 3
Cough following the intra-tracheal injection occurred in only a few subjects. When the bolus was at or just below the vocal cords, cough sometimes resulted in movement of the bolus through the cords. When the bolus was further distal to the cords, no appreciable change in bolus position, peak activity or shape was observed.
The mean TMV in the young normal subjects was 10.7±3.5 mm·min1. This was significantly greater than the mean TMV for older normal volunteers, 6.5±2.6 mm·min1 (p<0.001). In turn, TMV in COPD patients (2.1±2.7 mm·min1) was significantly less (p<0.001) than in older normal subjects. As is shown in figure 4
MCC is an important respiratory tract defence mechanism. The authors of this study have developed a rapid, safe and reproducible technique to measure TMV, which can be used to investigate abnormal clearance in disease and the effectiveness of therapy. This technique represents a practical evolution of a methodology explored by several investigators in the past. It has several advantages over the more commonly used aerosol techniques of measuring lung clearance. In particular, it is rapid, requires ten-fold lower doses of ionising radiation, and label deposition and data analysis have been simplified. This measurement is limited to providing information from the trachea only and thus may not accurately reflect clearance throughout the tracheobronchial tree. Various methods of droplet deposition were explored during the development of the technique. Initially, this was attempted through a fine catheter placed through the biopsy channel of a bronchoscope in a manner similar to that used by Santa Cruz et al. 10. The bronchoscope was placed either just above or just below the vocal cords. However, this procedure was slow and poorly tolerated without sedation or local anaesthetic, both of which have the potential to impair MCC. Deposition of the droplet without provoking paroxysms of cough was critical and proved to be difficult with this approach. A further problem was the difficulty in controlling a small (100 µL) droplet at the end of a long, narrow gauge catheter. Cricothyroid injection is a simple procedure that is used commonly with lignocaine to produce topical anaesthesia prior to bronchoscopy. It was found that injection with a 25-gauge needle was well tolerated and caused minimal bruising. Most subjects did not cough and haemoptysis did not occur. In addition, it allowed a small volume of radiolabel to be handled easily. With practice, injection and immediate detection of the droplet was straightforward. 99mTc MAA was selected as the tracer since it is stable, not absorbed and possesses high specific activity. Furthermore, it is relatively inexpensive, readily available in kit form and widely used for other scintigraphic studies. The macro-aggregates have a diameter of 1040 µm, a particle size known to be cleared by MCC 11. When deposited on the tracheal wall, the agent remains in droplet form on the mucus layer. Preparation, performance and analysis of the studies were not time consuming. Label preparation took <5 min, injection was completed in <15 s and the images were acquired in 15 min. Analysis is largely automated and a result can be obtained within 10 min after completion of the data acquisition. In the initial studies performed in normal subjects, the droplet was seen to move over a distance of 5080 mm, which was adequate to allow analysis. Both normal and COPD patients were comfortable lying still for the period of the test. Subject comfort is important in order to minimise body movement. By comparison, measuring MCC using inhaled radiolabelled aerosols and scintigraphy entails a prolonged acquisition time, during which the subject is required to lie supine for 4560 min. Such standard aerosol studies also require active patient cooperation and training to ensure appropriate inhalation patterns and deposition of the aerosol. On rare occasions, the deposition of the bolus elicited a cough response. However, in these subjects no movement of the bolus due to cough was observed. This lack of effective expulsion or increase in TMV following cough seems counterintuitive but has been reported previously 6. The increase in MCC seen with cough in radiolabelled aerosol studies may reflect an effect on lower airways 12. An absence of major effect of cough on TMV suggests that this technique may have increased utility in patient groups for whom the suppression of cough, even over 15 min, may be difficult (e.g. cystic fibrosis). In addition, this technique does not require any specific respiratory manoeuvres or breathing patterns.
This method of measuring TMV results in values for normal subjects in the same range (table 2
This technique is not directly comparable to IRLA studies. IRLA studies can provide information on regional lung clearance, small airway ventilation and airway obstruction that are not possible by measurements of TMV. However, for measurement of large airway clearance, this technique has advantages over IRLA and previous methods of measuring TMV. It is more rapid, more comfortable and exposes subjects to ten-fold lower levels of radiation than required for a typical aerosol study. Resolution and image quality is not compromised by this dose minimisation. The authors have shown previously that standard scintillation detection systems can detect this very low level of activity and accurately track the movement of such a small bolus over small distances 16. The information obtained by measuring TMV is complementary to that provided by IRLA studies. Further investigation is needed to determine the relationship between the different methods in normal subjects and patients with pulmonary disease. It has been shown here that, in normal subjects, TMV was slower in a group aged >50 yrs (mean±sd 6.5±2.6 mm·min1) than a group aged <50 yrs (10.7±3.5 mm·min1). Other groups have reported a reduction in MCC of similar magnitude with increasing age. Goodman et al. 8 showed that mean TMV in a group of young nonsmoking subjects (mean age 23 yrs) was 10.1 mm·min1 and in a group of seven elderly nonsmokers (mean age 63 yrs) 5.8 mm·min1. Puchelle et al. 17, using an inhaled radiolabelled aerosol technique, also demonstrated that clearance was reduced in older, as compared to younger, normal subjects 17. Consistent with previous reports, impaired MCC (2.1±2.7 mm·min1) in patients with COPD has been demonstrated here. Goodman et al. 8 showed that smokers and ex-smokers with simple and obstructive chronic bronchitis had markedly decreased TMV (0.7±1.9 and 0.8 mm·min1, respectively) compared to an age-matched nonsmoking group (5.8±2.6 mm·min1; p<0.05). Several studies of MCC measured by IRLA have also reported impaired clearance in patients with COPD 18. Putative disease-specific factors include the effects of cigarette smoke and recurrent infections. Possible mechanisms include progressive deciliation of airways 20 and altered rheology of respiratory secretions 22. Further studies are underway to determine whether TMV correlates with other clinical features of COPD (e.g. clinical progression, FEV1, smoking history). A technique for measuring tracheal mucus velocity is reported here, which uses cricothyroid injection for deposition of a radiolabel and automated analysis of dynamic scintigraphic data. The technique is rapid, simple, safe and reproducible. It uses materials and methods that are widely available and relatively inexpensive. The relative simplicity of the technique should make measurements of tracheal mucus velocity more accessible to nonspecialist centres. This technique was used to measure tracheal mucus velocity in two groups of normal subjects and patients with chronic obstructive pulmonary disease. The results confirm that mucociliary clearance decreases with age and is markedly impaired in chronic obstructive pulmonary disease. This technique should facilitate evaluation of the role of abnormal clearance in disease, and the effects of time, acute infection and pharmacological intervention on mucociliary clearance.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||