Copyright ©ERS Journals Ltd 2004 Sympathetic activity is reduced by nCPAP in hypertensive obstructive sleep apnoea patients1 Sleep Laboratory, Division of Pulmonary Diseases, Dept of Internal Medicine, and 2 Dept of Physiology, Marburg University, Germany. 3 Dept of Pulmonary Medicine, Sahlgrenska University Hospital, Gothenburg, Sweden CORRESPONDENCE: J. Heitmann, Klinik für Innere Medizin, SP Pneumologie, Schlafmedizinisches Labor, Klinikum der Philipps-Universität Marburg, Baldingerstrasse D-35033, Marburg, Germany. Fax: 49 6421284958. E-mail: heitmanj@med.uni-marburg.de Keywords: continuous positive airways pressure treatment, hypertension, sleep apnoea, sympathetic activity
Received: February 12, 2003
This study was supported by a scientific grant from the Deutsche Forschungs gemeinschaft (Eh 149/1-1).
There is increasing evidence that nasal continuous positive airway pressure (nCPAP) lowers blood pressure in obstructive sleep apnoea (OSA) patients, not only during sleep but also in the daytime. However, both the mechanisms of blood pressure reduction and the considerable differences in the magnitude of the effect in the studies presented to date are not fully understood. Therefore, the authors prospectively studied the effect of nCPAP on noradrenaline plasma levels (NApl), blood pressure and heart rate (HR) in 10 normotensive and eight hypertensive OSA patients before and after 41.6±16.9 days of nCPAP therapy. Polysomnography and invasive blood pressure were continuously monitored over 24 h in the supine position before and with nCPAP. NApl were analysed every 15 min. In hypertensives, nCPAP reduced NApl by 36±25%, lowered mean arterial blood pressure substantially (night-time: 8.89±14.09 mmHg; daytime: 7.94±10.47 mmHg) and decreased HR by 6.6±5.4 beats·min1, whereas in normotensives there were only minor changes. The decrease in heart rate was associated with a decrease in mean arterial blood pressure and noradrenaline plasma levels, suggesting a causal effect of nasal continuous positive airway pressure therapy. This nasal continuous positive airway pressure effect occurs mainly in hypertensive obstructive sleep apnoea patients, whereas the effect is small in normotensives. This may explain, at least in part, some of the discrepant results in previous treatment studies. Obstructive sleep apnoea (OSA) is characterised by repetitive reduction or cessation of airflow due to partial or complete upper airway obstruction 1. It results in repeated night-time hypoxaemia, sleep fragmentation and excessive daytime sleepiness. This disorder, which is prevalent in the adult population 2, has been associated with an increased incidence in cardiovascular morbidity, particularly in arterial hypertension 3. OSA is over-represented in the male population 2. Sleep fragmentation due to repeated arousal 4 and apnoea-related episodic hypoxaemia 5 have been suggested as possible mechanisms of the increased sympathetic activity that is present in OSA. There is growing evidence that this increased sympathetic activity may play a major pathogenic role in the development of cardiovascular complications 69. In OSA, plasma levels of noradrenaline (NApl), which correlate with sympathetic activity 10, have been found to be elevated both during sleep and wakefulness 11, 12.
Nasal continuous positive airway pressure (nCPAP), currently the treatment of choice in moderate-to-severe OSA 13, prevents upper airway obstruction and restores daytime symptoms 14, 15. A reduction in sympathetic activity in OSA patients on nCPAP has been shown in several studies 1618. Data concerning the effect of nCPAP on blood pressure are conflicting. Recently published controlled studies on the effects of nCPAP on arterial blood pressure demonstrated a small reduction (1.4 and 2.5 mmHg, respectively) in blood pressure with nCPAP therapy 19, 20, whereas others didn't find a blood pressure drop with nCPAP 21 or only during the night 22. However, post-hoc analysis in one study 19 revealed that the subgroups with severe hypertension and severe OSA may have more profound reductions in blood pressure. Conversely, a recently published controlled study from the current authors' group demonstrated There are a number of methodological problems in previous studies, which may explain the conflicting results from these trials: 1) exclusively or mainly normotensive patients were included in the studies that did not show an effect of nCPAP on blood pressure; 2) discontinuous blood pressure measurement may underestimate blood pressure changes; and 3) activity levels were not controlled. Therefore, the present authors conducted a study evaluating the effect of nCPAP on arterial blood pressure, heart rate (HR) and sympathetic activity. Normo- and hypertensive OSA patients were investigated using invasive continuous blood pressure monitoring during a 24-h constant routine with defined activity and resting levels.
Subjects Consecutive patients were recruited from the out-patient department of the Marburg Sleep Laboratory (Marburg, Germany). The indication for cardiorespiratory polysomnography due to suspected sleep apnoea was based on clinical symptoms, ambulatory nocturnal recording of HR, snoring, oxygen saturation and body position using a validated four-channel ambulatory device (MESAM IV®; Madaus Medizintechnik, Munich, Germany). Blood pressure and HR profiles were obtained over 24 h using the SpaceLabs 90207® (SpaceLabs Medical, Inc., Redmond, WA, USA), which provides oscillometric blood pressure measurements. Measurement intervals were 20 min during the day and 30 min during the night.
Inclusion criteria included: age 2065 yrs, apnoea-hypopnoea index (AHI) >20·h1. Exclusion criteria included: systolic blood pressure >220 mmHg or diastolic blood pressure >110 mmHg; myocardial infarction; stroke; heart failure; neurological disease; or a history of alcohol or drug abuse. Concomitant antihypertensive medication was discontinued at least 3 weeks before the study began. A patient was assigned to the hypertensive group when daytime average blood pressure in the ambulatory blood pressure measurement was
Demographic characteristics, AHI (at baseline and with nCPAP therapy), data of the ambulatory blood pressure measurements, and the duration of nCPAP use as well as nCPAP pressure of the normotensives and the hypertensives are shown in table 1
Study protocol Patients eligible for the study were hospitalised for a period of 5 days. Following baseline recordings (see below), all patients were put on nCPAP treatment. The follow-up measurements were carried out after a minimum of 4 weeks of home use of nCPAP therapy. To encourage the patients to use the nCPAP therapy regularly and to increase compliance, weekly telephone contacts and a follow-up in the out-patient department after 2 weeks were performed. Compliance with nCPAP treatment was measured with a run-time counter, which ran when the machine was switched on. A minimum of 5 h·night1 was set as the minimum threshold for compliance. This study was approved by the Ethics Committee of Marburg University. Patients gave written informed consent before the start of the study.
Data collection
Data analysis
The AHI (events per hour) was calculated from abnormal breathing events in relation to the total sleep time and classified according to standardised criteria. Oronasal airflow was recorded by the use of a thermistor and respiratory effort by inductive plethysmography (Respitrace®; Studley Data Systems, Oxford, UK). An apnoea was defined as the cessation of airflow for at least 10 s. Hypopnoea was defined as a reduction of airflow by Blood pressure and electrocardiogram (ECG) were digitised at 100 Hz. HR was derived from the ECG signal using an R-wave detection. Mean values of blood pressure and HR were calculated for 15-min periods corresponding to the blood sampling. Scoring of polysomnography, and analysis of blood pressure and HR were done by a technician otherwise not involved in the study. Adrenaline plasma levels (Apl) and NApl were measured by high-performance liquid chromatography with electrochemical detection using the method application of Chromosystems Instruments & Chemicals GmbH (Munich, Germany). The intra-assay coefficient of variation was 5.6 and 6.1% respectively, the sensitivity was 54.7 and 59.1 pmoL·L1 respectively.
Statistical analysis Hypothesis-generating post-hoc analysis was performed by means of correlation analysis and partial correlation analysis. Multiple testing was performed without adjustment for the significance level. Multivariate regression analysis was used to assess independent associations.
Missing data
Blood pressure, heart rate and catecholamines There was a substantial decrease in NApl in the hypertensives (1,679±539 to 1,081±489 pmol·L1; p=0.001) but not in the normotensives (1,449±447 to 1,385±524 pmol·L1). No significant changes in Apl levels, either in the normotensives nor in the hypertensives, occurred (normotensives: 143.1±28.9 to 126.7±54.7 pmol·L1; hypertensives: 137.4±36.8 to 137.8±36.21 pmol·L1). MAP decreased significantly with nCPAP in the hypertensive group from 103.2±13.7 to 95.0±12.3 mmHg (p=0.04), whereas in the normotensive group there was only a minor decrease in MAP (89.0±15.3 to 87.3±12.6 mmHg; p=0.7). In the hypertensive group, average HR decreased significantly from 76.3±5.16 to 69.8±6.0 beats·min1 (p=0.004), while no such effect was observed in the normotensive group (68.1±6.1 to 67.4±4.8 beats·min1) (fig. 1
Effects of nasal continuous positive airway pressure on sleep-disordered breathing and sleep In all 18 patients, good compliance with nCPAP therapy was achieved, the mean duration of nCPAP use being 5.7±0.68 h·night1. nCPAP compliance did not differ between the groups. The data on AHI and the sleep stages before and during treatment are shown in table 3
Interaction of factors associated with the observed change in 24-h blood pressure and heart rate Whole group analysis (n=18) There was a highly significant correlation of the individual baseline-therapy difference in NApl with the changes in HR (r=0.611, p=0.007, fig. 4
A multiple regression model was built taking baseline-therapy differences of systolic blood pressure, diastolic blood pressure, mean blood pressure, HR and NApl as dependent, and age, body mass index (BMI) and respiratory disturbance index (RDI) as independent variables. A statistical trend was seen in the correlation between the RDI at baseline and the decrease in NApl (r=0.579), the significance of the coefficient in the model was p=0.056. Patients with higher RDI showed a greater decrease in 24-h NApl with nCPAP therapy.
Subgroup analysis (8 normotensives and 10 hypertensives)
This study demonstrated that: 1) nCPAP significantly reduced NApl, blood pressure and HR in hypertensive but not in normotensive OSA patients; 2) the reduction in blood pressure during daytime and during sleep was in a similar range; and 3) the reduction of blood pressure was strongly associated with the change in HR. In the trial presented here there is a clear association between blood pressure reduction and changes in sympathetic activity. The reduction in NApl with nCPAP in the hypertensive patients was observed during the night and also during the day, suggesting that the elimination of OSA with nCPAP has an impact on sympathetic activity beyond the hours of sleep. This effect was not found in the normotensive subjects, which confirms the data of Marrone et al. 7. Hedner et al. 18 also demonstrated a marked reduction of catecholamine excretion in OSA patients with nCPAP both during wakefulness and sleep. For the first time, this study strongly suggests that the reduction of HR and cardiac output may be an important mechanism for the blood pressure-reducing effect of nCPAP therapy. This is suggested by the strong correlation between the change in HR and the change in blood pressure in the whole study group and in the hypertensive subgroup. Furthermore, the current data suggest that this effect may be directly mediated by the reduction of overall sympathetic activity, as indicated by the strong correlation between the changes in circulating NApl and the changes in HR induced by nCPAP. Alternatively, cardiac vagal activity might have been increased as a response to nCPAP therapy (not assessed in this study). Indeed, previous studies reported increased cardiac vagal tone after nCPAP therapy, as assessed by HR variability analysis 26, 27. There are further mechanisms potentially involved in the observed blood pressure reduction by nCPAP treatment which have not been assessed in this study, e.g. increase in vasodilatory function and nitric oxide content of the endothelium, reduced vasoconstriction due to normalised circulating NA, or change of reactivity of the renin angiotensin system. This lack of information limits the possibility to explain the mechanisms that caused the blood pressure reduction in hypertensives and not in normotensive OSA patients. However, the present study clearly indicates that blood pressure level at baseline and hypertensive state, the severity of OSA, and probably the change in HR with treatment are potential predictors for the blood pressure response to nCPAP therapy. Most uncontrolled studies have shown a blood pressure decrease with nCPAP 17, 28. In a few controlled studies there was either no effect 21, 22 or a small decrease in daytime MAP by 1.42.5 mmHg 19, 20. In the normotensive patients in the current study, the decrease in blood pressure had a similar magnitude (1.7 mmHg). However, the decrease in MAP in the hypertensive patients was much more pronounced (8.3 mmHg), similar to the data of Becker et al. 23. The discrepant results may be caused by the splitting between hypertensive and normotensive patients, the controlled conditions (activity level, food intake, alcohol consumption, etc.), the lack of confounding antihypertensive medication, and the invasive blood pressure measurement. These factors have not been considered in previous studies. An unexpected finding in the present study was the extent of the decrease in NApl in the hypertensives with nCPAP therapy; it reached levels below that of the normotensives. This finding remains unexplained. The studies of Jennum etal. 16 and Minemura et al. 17, which showed similar results, are not comparable because they did not differentiate between hypertensive and normotensive OSA patients. In the current study, there were no significant changes in adrenaline in the normotensive or in the hypertensive patients. This result confirms previous studies 17, 29. The strengths of this study were continuous intra-arterial blood pressure recording and the experimental protocol controlling for external conditions and vigilance. Continuous invasive blood pressure measurement was used because of its accuracy and absence of additional sleep disturbances, which is a limiting factor in ambulatory blood pressure monitoring 30. MAP is the true mean of the blood pressure curve and reflects the mechanical load to the vessels. Therefore, this parameter was chosen as primary end-point. Frequent antecubital plasma NA measurements were not done in OSA patients so far. This method can be used as a surrogate for systemic sympathetic activity 10, while other methods such as microneurography or catecholamine spill over are very cumbersome and cannot be used during a 24-h period.
The major limitation of the current study is the lack of a control group. At the time the study was designed (mid 90th) there was no accepted placebo treatment for OSA. However, this restriction clearly does not influence the main findings ofthe current study: the group difference between normo- and hypertensives, the association between blood pressure changes and NA changes, and the profound changes in HR by nCPAP. Furthermore, it has to be mentioned that the blood pressure reductions in the hypertensive OSA group by far exceeded previously reported placebo effects in other studies. In the screening procedure, the assignment of the OSA patients to the normotensive or hypertensive group had been done using ambulatory blood pressure monitoring (ABPM). Mean daytime values of 135 mmHg (systolic) and of 85 mmHg (diastolic) correspond to the casual blood pressure limits of 140 and 90 mmHg accepted as a cut-off for hypertension 24, 31. Beyond this, studies have shown that end-organ damage correlates better with ABPM measurement than with office blood pressure 32. Although this procedure was expected to better discriminate between the normo- and hypertensive state in the patients in the current study when compared with the conventional procedure of office blood pressure measurement, there was a considerable overlap in invasively assessed blood pressure during the baseline measurement between the two patient groups (fig. 3 The patients in the hypertensive group showed more severe OSA at baseline and were more obese when compared to normotensive subjects. Further analysis of the data using multiple regression analysis and partial correlation analysis revealed a statistical trend that the higher the RDI at baseline the greater the decrease in NApl with nCPAP therapy. However, such effects could not be demonstrated for blood pressure or HR. In this analysis, cardiovascular changes were not correlated to the BMI. This finding is in concordance with recent data suggesting that obesity in the absence of OSA is not accompanied by increased sympathetic activity 33. Therefore, NApl are not likely to be affected by the extent of obesity. In conclusion, the present data show that nasal continuous airway pressure substantially lowers sympathetic activity in hypertensive obstructive sleep apnoea patients, which occurred during both night- and daytime. This may be an important cause of the blood pressure reduction in these patients, although other well-known factors of blood pressure regulation have not been examined in this study. Interestingly, the change in heart rate with nasal continuous airway pressure may have a strong predictive value for the amount of blood pressure reduction. In normotensive obstructive sleep apnoea patients there is only a small decrease in both sympathetic activity and blood pressure. The fact that, in some previously published studies, blood pressure remained unchanged or showed only a small decrease with nasal continuous airway pressure may be due to the fact that most of the patients studied were normotensive. Therefore, studies evaluating the effect of nasal continuous airway pressure on blood pressure should focus on hypertensive obstructive sleep apnoea patients. The marked blood pressure reduction with nasal continuous airway pressure therapy seen in the obstructive sleep apnoea patients of this study may lead to a substantial reduction in cardiovascular risk in these patients.