1 Pulmonary Unit, San Pedro de Alcántara Hospital, Cáceres, Spain. 2 Sleep Centre, University of Edinburgh, Royal Infirmary National Health Service Trust, Scotland, UK
CORRESPONDENCE: J.F. Masa, C/ Rafael Alberti 12, 10005, Cáceres, Spain. Fax: 34 927256202. E-mail: fmasa@separ.es
Keywords: respiratory effort-related arousal, sleep apnoea syndrome, thoracoabdominal bands, upper airway increased resistance syndrome
Received: January 31, 2003
Accepted June 1, 2003
This study was supported by Sociedad Extremeña de Patología Respiratoria and Asociación de Neumólogos de Cáceres.
| Abstract |
|---|
|
|
|---|
Altogether, 94 subjects suspected of having sleep apnoea but with an apnoea/hypopnoea index
10 in full polysomnography with oesophageal pressure were studied. A routine polysomnographical analysis was carried out. The polysomnographies were then reanalysed at random to determine which of the identified arousals were due to RERA, as determined either by oesophageal pressure or by induction bands without an oesophageal pressure signal. Altogether, 14,617 arousals were analysed.
The sensitivity and specificity to find RERA (arousal by arousal) from bands versus oesophageal pressure were both 94%. The average difference of RERA index between oesophageal pressure and bands was 0.6. The correlation between RERA index determined by oesophageal pressure and bands was 0.98. To evaluate the intra and interobserver agreement, 1183 arousals were additionally analysed. The intraobserver agreement was 91% for RERAs by oesophageal pressure and 80% by bands. The interobserver agreement was 89% by oesophageal pressure and 85% by bands.
The thoracoabdominal bands can be used to identify respiratory effort-related arousal (obstructive events not detected by thermistor) with similar efficacy to oesophageal pressure measurement. Since bands are routinely used in most polysomnographies, they can be used as the usual method to detect respiratory effort-related arousal, using a thermistor to evaluate apnoeas and hypopnoeas or as a complement to other methods, such as nasal cannula, which can detect apnoeas, hypopnoeas and respiratory effort-related arousal.
Disorders of sleep-related breathing are characterised by atotal or partial reduction of breathing. These events can be partially obstructive (hypopnoea), totally obstructive (apnoea) or very subtle upper airway obstructions, and require measurement of oesophageal pressure to detect respiratory effort-related arousal (RERA) 1. Sometimes, RERAs can appear predominantly (without a significant number of apnoeas andhypopnoeas) producing fatigue and daytime sleepiness, in what is called upper airway increased resistance syndrome 1. This excessive number of RERAs has been associated with raised blood hypertension 3 and car accidents 4.
In clinical practice, a thermistor is the most widely used method to detect oro-nasal flow in conventional polysomnography. However, thermistors have long time-constant responses 5, overestimate airflow on polysomnography 6 and the temperature of exhaled gas is relatively unaltered by changes in exhaled volume 7. Consequently, thermistors do not detect subtle airflow decreases, as observed in RERAs. Different methods have tried to identify RERAs without oesophageal pressure measurement: alteration in the flow curve obtained by nasal cannula 8 or continuous positive airway pressure (CPAP) 10, pulse transit time 12 and sum from inductance plethysmography 13.
Thoracoabdominal movements are an indirect flow measurement 7 in which reduction of amplitude and inspiratory curve alteration can be detected 11. RERA detection with thoracoabdominal bands has not been extensively investigated. The potential advantage of thoracoabdominal bands over nasal cannula is that the signal registered by the bands is not dependent on the patient having to breathe solely through the nose. Moreover, thoracoabdominal bands are already widely used in polysomnography to evaluate whether apnoeic events are obstructive or central. Therefore, if thoracoabdominal bands are useful to detect RERAs, theycould be used in combination with a thermistor or as acomplement to other methods that can detect apnoeas, hypopnoeas and RERAs, for example nasal cannula 1517.
The aim of the current study was to evaluate the analysis of the shape and amplitude of the signals from thoracoabdominal bands versus oesophageal pressure measurement to detect RERAs (obstructive events not detected by thermistor) in a population of subjects referred because sleep apnoea was suspected but whose apnoea/hypopnoea index was found to be<10.
| Methods |
|---|
|
|
|---|
10 on full polysomnography/oesophageal pressure was recorded for at least 2 h during another full polysomnography.
To perform the analysis the population was retrospectively divided into two groups, those with and without an Epworth Sleep Scale
9, in order to obtain two populations likely to have different prevalences of RERAs (table 1
).
|
10. In the second polysomnography an oesophageal catheter was introduced transnasally and advanced until it had obtained a positive pressure during the inspiration and then was pulled back until it had obtained a clear negative pressure during the inspiration. It was calibrated following a previously described technique 19. The signals from oesophageal pressure and bands were DC, the lower filter was 0.1 Hz, the higher filter was 5 Hz and the sampling rate was 100 Hz. The analysis of sleep stages, arousals, awakenings, apnoeas and hypopnoeas followed standard procedures 2022. Every arousal was numbered consecutively. The time of study with oesophageal pressure measurement and normal function of bands were recorded. When the routine analysis of sleep stages, arousals, awakenings, apnoeas and hypopnoeas was carried out in all the second polysomnographies, a technician reanalysed at random the recorded polysomnographies to determine if each arousal previously detected with standard procedures 20 (not caused by apnoeas and hypopnoeas) was due to RERAs or not, using the definitions based on oesophageal pressure or bands (see below). The signals were compressed into 120 s per epoch toallow the RERA episodes to be more easily seen. The technician scored blindly the same sleep studies twice: with and without oesophageal pressure in the scoring montage (the first score being made at random). Each time, the technician noted down if the previously numbered arousals were RERAs or not. Afterwards, an investigator examined the notes to determine the coincidence and noncoincidence between both methods in the total arousals studied (see statistical analysis). Only the times when both the oesophageal pressure and bands functioned normally were analysed.
An Institutional Committee on Investigation approved the protocol study, and written consent was obtained.
Definitions
Apnoea was defined as an absence of airflow of
10 s and hypopnoea when airflow (both in the thermistor signal) decreased
30% for
10 s with an oxygen saturation drop of
4% or final arousal 22.
Apnoea/hypopnoea index was defined as the total number of episodes of apnoea and hypopnoea divided into the number of hours slept.
When the scoring montage included oesophageal pressure measurement, a RERA was scored, if all the following were present: 1) there were increasing negative oesophageal pressures in the two or more breaths prior to an arousal; 2) oesophageal pressure became less negative in the arousal; and 3) increasing negative oesophageal pressure prior to the arousal did not coincide with increased oral-nasal flow.
When the scoring montage did not include oesophageal pressure measurement, an arousal was considered to be due to RERA if all the following were present: 1) alteration of the inspiratory contour in the waves from the thoracoabdominal bands (figs. 1
and 2
) in two or more breaths before the arousal, with or without discernible reduction of the thoracoabdominal bands amplitude; and 2) normalisation of the previous alteration coinciding with the arousal.
|
|
The number of RERAs (according to the above definition) was also counted with great reduction in bands amplitude, which had not been classified as hypopnoea according to the authors definitions using thermistor. These "true hypopnoeas" were defined as
50% of reduction in the sum of thoracoabdominal bands amplitude, lasting
10 s, with final arousal. The signal of the sum was calibrated automatically (not by pneumotacography) by the inductance plethysmography device (SomnoStar; SensorMedics, CA, USA).
Intra- and interobserver agreement
To evaluate the intraobserver agreement, the same technician who did the scoring randomly analysed eight subjects twice (1183 arousals each time) including (or not) the oesophageal pressure measurement in the scoring montage, using the same criteria as the first analysis. Additionally, to evaluate the interobserver agreement, another technician from the same sleep laboratory and with similar experience, analysed the same eight subjects twice with the same protocol. Likewise, when the analysis was done using the bands and when the arousals were recorded as RERAs, both technicians registered if RERA criteria were evident or not.
Statistical analysis
The comparison of proportions was made with the Fisher exact test and that of mean values with an unpaired t-test if data presented a normal distribution; otherwise, the nonparametric Mann-Whitney U-test was used.
Sensitivity and specificity tests were carried out to determine the efficacy to identify RERAs by thoracoabdominal bands compared with RERAs identified by oesophageal pressure. This analysis was carried out taking into account the number of arousals identified and not identified as RERAs by both methods from the total arousals evaluated (14,617). The sensitivity and specificity were performed in all the population and in each of the groups with and without sleepiness.
A correlation test was performed between the RERA index from the oesophageal pressure measurement and thoracoabdominal movements of each patient. The correlation significance was made using the Pearson test.
Bland and Altman 23 analysis was carried out with the differences in the RERA index between both methods to identify RERAs and the mean of RERA index also measured using both methods. This analysis was carried out to evaluate the agreement between the two methods and to know if the differences between them were independent of the measurement size. To determine the intraobserver and interobserver agreement excluding the random effects a Kappa test was carried out 24. A p-value <0.05 in two-sided test was considered statistically significant.
| Results |
|---|
|
|
|---|
Altogether, 14,617 arousals were analysed. Oesophageal pressure classified 4,082 as RERAs and bands classified 4,276. Most arousals (24±18 arousal·h1 of sleep per patient, 58% of total arousals) were not due to RERAs, apnoeas or hypopnoeas.
In the total sample of patients, 62% had >5 RERAs·h1 slept, with 30% having >15 RERAs·h1 slept (table 1
). In the group with sleepiness 88% had >5 and 52% >15 RERAs·h1 slept, compared with 29% and 2.4% in the group without sleepiness. The group with sleepiness had less sleep time, less deep and rapid eye movement sleep and a higher percentage of total sleep time with <90% of oxygen saturation.
From 4,276 RERAs detected by bands, 3,572 occurred in the sleepy group and 702 in the nonsleepy group. Alteration of the inspiratory contour and reduction in band amplitude were present together in most of RERAs, but only 684 (16%) of the total of RERAs detected by bands were true hypopneas, 607 RERAs (17%) in the sleepy group and 84 RERAs (12%) in the nonsleepy group.
Association between respiratory effort-related arousals from oesophageal pressure measurement and thoracoabdominal movements
The correlation between RERA index per patient determined by oesophageal pressure measurement and thoracoabdominal movements was close to one (r=0.98) (fig. 3
). The correlation index was 0.97 in the group with sleepiness and 0.96 in the group without sleepiness (fig. 3
). Data are close to the equality line in the three plots.
|
|
|
|
| Discussion |
|---|
|
|
|---|
The thoracoabdominal movement measurement with bands is an indirect tidal volume measurement and its usefulness for detecting apnoeas and hypopnoeas has been previously established 7. Loube et al. 13 tried to identify episodes of upper airway increased resistance in 14 patients with symptoms of sleep apnoea but with apnoea/hypopnea index <10, by means of the sum of thoracoabdominal bands measured by inductance plethysmography (Respitrace®). Measuring the relation inspiratory peak flow and midinspiratory flow immediately before the arousal, the sensitivity and specificity to detect patients with upper airway increased resistance versus oesophageal pressure measurement were 100%. However, when these authors tried to identify patients with increased upper airway resistance without taking into account the arousal, the sensitivity was 67% and the specificity was 80%. In another study with seven habitual snorers, the sensitivity and specificity to identify limitations of inspiratory flow measured in the sum (without taking into account arousals) secondary to upper airway resistance were between 7173% (depending on the limitation level to the airflow) compared with the sensitivity and specificity between 8284% with limitation of inspiratory flow measured by pneumotachograph 27.
Montserrat et al. 11 evaluated the usefulness of the thoracoabdominal bands output morphology and the inspiratory flow measured by pneumotachography to identify upper airway resistance (not necessarily associated with arousal) in nine patients with sleep apnoea during the CPAP titration. Both methods identified well episodes of increased upper airway resistance, although the inspiratory flow behaviour from pneumotachography was better than that of the bands.
The inspiratory flow limitation measured by nasal cannula in the identification of hypopneas or upper airway resistance has been evaluated in different studies 8, but few studies have evaluated this method in patients without the predominance of apnoeas or hypopnoeas versus the oesophageal pressure measurement (or pharyngeal pressure). In two studies with these characteristics, the sensitivity was estimated between 7688% and the specificity between 7781%, considering upper airway resistance not associated to arousal 8.
The present study has not compared the morphology of the thoracoabdominal bands with other methods for evaluating upper airway resistance noninvasively, but the sensitivity and specificity found are apparently higher than those found with other methods. This may be methodological and might perhaps be because the authors evaluated episodes of increased upper airway resistance always associated with arousal. Loube et al. 13 had similar findings. The morphology of band output in the arousal is the reference, therefore these waves can be compared with those that occur before arousal (fig. 2
). In 10 patients chosen at random from the 94, the authors reanalysed a period of the polysomnography starting from an epoch number chosen at random. All the signals except flow by thermistor and thoracoabdominal bands were removed from the scoring montage. The authors tried to identify upper airway resistance events according to the bands output contour (not associated to arousal) until 20 events in each patient was reached. Later, the analysis of the same periods was repeated, showing in the scoring montage only flow by thermistor and oesophageal pressure measurement, trying toidentify if the 20 episodes of upper airway resistance according to the bands, were associated with increasing negative oesophageal pressure. The sensitivity and specificity of thoracoabdominal bands versus oesophageal pressure to detect upper airway resistance were 74 and 81%, respectively. These data may explain why the results from the current study identifying RERAs (upper airway resistance associated with arousal) are better than those obtained in other studies that analysed upper airway resistance not associated with arousal.
The advantage of determining RERAs by means of the morphology of bands over the sum of bands is that calibration is not necessary 32 and according to the authors experience the alteration in the curve contour is better observed in the output of the individual bands than in their sum. This may be because alteration in the morphology of bands outputs does not happen in both bands in the same way (figs. 1
and 2
).
It is important to assess the time when the outputs were so poor they could not be evaluated. In the current study there was a mean loss of 20 min per patient (7.5% of total sleep time). Altogether, 59% of this lost time was due to a poor oesophageal pressure signal and 41% to a poor signal from the thoracoabdominal bands. This represents
8 min per patient due to thoracoabdominal bands signal failure. This small lost time of study may be due in part to the fact that the population of the current study is not very obese, since patients with sleep apnoea were excluded for definition. In the obese patients the signal from the bands tends to be poorer than in nonobese patients.
In the current study, the authors used a thermistor as a flow measurement, since this is the most widely used method to detect oro-nasal flow in the conventional polysomnography in clinical practice. However, a thermistor overestimates flow with respect to the pneumotachography 17 and the 16% of RERAs identified by bands were considered true hypopnoeas by the sum of thoracoabdominal bands. Nevertheless, the most important thing, in a practice point of view, is that thebands were able to recover obstructive events lost by the thermistor.
In the present study there are a large number of arousals, which were not associated with RERA, apnoeas or hypopnoeas. The oesophageal pressure measurement can disturb sleep 33 and this can explain the increase in the percentage of wakefulness found in the current study. Since studies with normal subjects (without oesophageal pressure measurement) 34 have found a similar number of arousals, it is possible that the cause is the polysomnography itself or that this is the normal arousal frequency.
The authors used inductive plethysmography, since this system has been validated to identify hypopneas 7, so future studies will have to determine if other types of bands are useful to identify RERAs. Since Montserrat et al. 11 using piezoelectric bands, observed similar findings to those in the current study, it seems likely that these type of bands are also useful in identifying RERAs.
In summary, the thoracic and abdominal bands used in this study can identify respiratory effort-related arousal (obstructive events that are not detected by thermistor) with efficacy similar to the oesophageal pressure measurement. Unlike nasal cannula it is independent of whether or not the patient breathes through the mouth. Since bands are widely used in polysomnography to determine if the apnoeas and hypopnoeas are obstructive or central, they can be used as the sole method to detect respiratory effort-related arousal, in conjunction with a thermistor to evaluate apnoeas and hypopnoeas or as a complement to other methods that can detect apnoeas, hypopnoeas and respiratory effort-related arousal, such as nasal cannula.
| Acknowledgements |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
L. H. Ibrahim, S. R. Patel, M. Modarres, N. L. Johnson, R. Mehra, H. L. Kirchner, and S. Redline A Measure of Ventilatory Variability at Wake-Sleep Transition Predicts Sleep Apnea Severity Chest, July 1, 2008; 134(1): 73 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Muzumdar and R. Arens Diagnostic Issues in Pediatric Obstructive Sleep Apnea Proceedings of the ATS, February 15, 2008; 5(2): 263 - 273. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Waag Carlson, V. J. Neelon, J. R. Carlson, M. Hartman, and S. Dogra Respiratory Periodicity and Electroencephalogram Arousals During Sleep in Older Adults Biol Res Nurs, April 1, 2007; 8(4): 249 - 260. [Abstract] [PDF] |
||||
![]() |
R. Farre, J.M. Montserrat, and D. Navajas Noninvasive monitoring of respiratory mechanics during sleep Eur. Respir. J., December 1, 2004; 24(6): 1052 - 1060. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |