1 Dept of Paediatrics, University of Western Australia, and Depts of 2 Respiratory Medicine and 3 Nuclear Medicine, Princess Margaret Hospital for Children
CORRESPONDENCE: S.G. Devadason, University Dept of Paediatrics, Princess Margaret Hospital for Children, Roberts Road, Subiaco WA 6008, Australia. Fax: 61 893882097. E-mail: sunalene@paed.uwa.edu.au
Keywords: aerosol delivery to children, lung deposition, paediatric asthma therapy
Received: September 24, 2002
Accepted December 6, 2002
This study was funded, in part, by 3M Pharmaceuticals Pty Ltd, Sydney, Australia. S.G. Devadason was supported by a grant from the National Health and Medical Research Council of Australia.
| Abstract |
|---|
|
|
|---|
Sixteen male asthmatic children (514 yrs) inhaled labelled drug (<4 MBq 99mTc; 100 µg beclomethasone dipropionate) via an Autohaler within 30 min after salbutamol (200 µg) administration. Simultaneous anterior and posterior planar scintigraphic scans(120 s acquisition time) were collected after inhalation of labelled drug.
Mean±sd lung deposition of labelled drug (attenuation-corrected; percentage of ex-actuator dose) was 36.9±9.2, 46.5±11.6 and 54.1±10.7% in children aged 57, 810 and 1114 yrs, respectively. Combined oropharyngeal and gastrointestinal deposition was 59.7±8.2, 48.9±12.3 and 40.3±11.8%. Lung deposition positively correlated with the forced expiratory volume in one second (FEV1) and forced vital capacity (FVC). Gastrointestinal dose negatively correlated with the FEV1, FVC, height and age.
In older children (1114 yrs), lung deposition was almost identical to that reported inadults using QVARTM. In children aged 510 yrs, lung deposition using QVARTM was greater than the levels measured using other commercial aerosol delivery systems. Oropharygeal and gastrointestinal deposition was inversely related toage.
The original devices used for inhalation therapy were designed for adult patients. Although problems with compliance and dose variability with inhaler therapy occur in adults, they are magnified when these devices are used by children. Added to this is a very unique set of considerations associated with choosing an appropriate aerosol device for use by young children. Hence choice of both device and formulation is critical when considering asthma therapy in children.
Children aged
5 yrs can, in theory, be taught specific inhalation techniques, which means that they should be able to use devices such as the TurbuhalerTM (AstraZeneca, Sydney, Australia) and the AutohalerTM (3M Pharmaceuticals, Sydney, Australia), a breath-actuated pressurised metered-dose inhaler (pMDI) requiring a relatively constant inspiratory flow over 24 s for optimal use. The AutohalerTM has been shown to improve lung deposition in adults with poor coordination compared to conventional "press-and-breathe" pMDIs 1. When using the same drug formulation, equivalent clinical efficacy has been shown using the AutohalerTM compared to efficient use of pMDIs in both adults 2, 3 and children 4. However, the efficacy of these devices and ability to consistently perform the appropriate inhalation technique still needs to be more formally assessed in children aged 57 yrs compared to older children.
The reformulation of pMDIs with hydrofluoroalkane (HFA) propellants has resulted in the development of an extrafine form of beclomethasone dipropionate (BDP), known as QVARTM (3M Pharmaceuticals). The mass median aerodynamic diameter (MMAD) of the QVARTM formulation is1.1 µm, which is much smaller than that of the chlorofluorocarbon (CFC)-based formulations (MMAD 3.3 µm) 5. Use of the QVARTM formulation has been shown to significantly increase lung deposition in adults compared to other pMDI formulations, when delivered by either pMDI 6 or AutohalerTM 7, 8. Clinical efficacy studies have also shown that 2.5-times more CFC/BDP is required to achieve the samebeneficial improvement in asthma control as obtained using QVARTM 9. Since children have a smaller airway diameter than adults, it was postulated that the extrafine formulation would be ideal for use in children with asthma.
Drug delivery to the lungs is likely to be affected by factors such as inspiratory flow, inspiratory volumes and structural aspects of the respiratory system, all of which arelikely tochange with increasing age. In addition, the pattern of deposition within the body is likely to differ in children using theextrafine QVARTM formulation compared with adults. In order to determine the efficacy of drug delivery from thisdevice and formulation, a gamma scintigraphic study wasundertaken in 16 children aged 514 yrs with mild, stable asthma toassess total and regional deposition of technetium-99m-labelled (99mTc) QVARTM, delivered via an AutohalerTM.
| Methods |
|---|
|
|
|---|
Approval for this study was granted by the Ethics Committee of Princess Margaret Hospital for Children, Perth, Australia. Informed consent was obtained from parents, and the children who completed the study were willing participants. The maximum level of radiation dispensed to each patient (4 MBq) was approved by the Radiation Safety Officer at Royal Perth Hospital (Perth, Australia), and was equivalent to the additional radiation exposure from a 12-h plane flight, or <3 weeks' natural background radiation.
Radiolabelling of QVARTM formulation
Radiolabelling of the HFA/BDP formulation was carried out using a previously described method 2. Briefly, a canister containing the commercial QVARTM (HFA/BDP) formulation was supercooled in dry ice. 99mTc in normal saline was eluted from a molybdenum/technetium generator (Australian Radioisotopes, Sydney, Australia) and extracted from the aqueous phase into chloroform using a separating funnel. Thechloroform was evaporated to dryness in an empty AutohalerTM canister under a slow stream of nitrogen gas. The AutohalerTM canister, now coated with 99mTc, was covered and cooled in dry ice. The supercooled commercial canister containing QVARTM was rapidly decrimped, the contents poured into the AutohalerTM canister containing 99mTc and anew valve crimped on top. This canister was then allowed towarm up in a water bath to ensure that no leakage of propellant occurred through the crimped valve.
In-vitro assessment of radiolabelling
The gamma-scintigraphic technique used in the present study measures the distribution of radiolabel within the body and not drug distribution. Hence, stringent in-vitro assessment of the output of radiolabelled BDP from the AutohalerTM device was employed to ensure that the radiolabelling method did not significantly alter the particle size distribution of the aerosol generated by the AutohalerTM, and to confirm that the distribution of 99mTc reflected that of the drug, thus acting as a suitable marker for BDP. Particle size distribution and total drug delivery was measured using an eight-stage nonviable cascade impactor (Thermo-Andersen, Smyrna, GA, USA) with an inhalation flow of 28 L·min1. The AutohalerTM was primed and then inserted into the "throat" of the cascade impactor. Twenty doses of radiolabelled BDP were drawn into the cascade impactor with the entraining air flow. Particles were deposited either in the mouthpiece of the AutohalerTM casing or on the "throat", jet stage, one of the eight stages or a filter. The location of particle deposition was determined by the aerodynamic size of the particle.
The mouthpiece and stages of the cascade impactor werewashed with 1025 mL ethanol. Activity was quantified using an ionisation chamber (Atomlab 200 dose calibrator; Gammasonics, Sydney, Australia). The absorbance (at 238 nm) of each sample was measured in duplicate using an ultraviolet spectrophotometric method. The standard curve for BDP waslinear (r2=1.00) for concentrations up to 20 µg·mL1. Thefine particle fraction of both radiolabel and drug was calculated as the proportion of the ex-valve dose of particles <4.7 µm in diameter (sum of Andersen Plate 3-filter). The extrafine particle fraction was calculated as the proportion of the ex-valve dose of particles <3.3 µm in diameter (sum of Andersen Plate 4-filter).
The particle size distribution of drug from each canister used for patient inhalation studies (n=9) was determined bothbefore and after labelling (fig. 1
). The distribution of radioactivity from these canisters after labelling was also measured. This was performed to confirm that the labelling procedure had been successfully carried out and to ensure that the ex-actuator dose of activity delivered per actuation did not exceed 4 MBq.
|
10 s after inhalation through the AutohalerTM. Patients then inhaled a single dose of 99mTc -labelled HFA/BDP (100 µg nominal dose) from the AutohalerTM. No more than 4 MBq of 99mTc (exactuator dose) was administered to each patient during the study. During all experiments, the AutohalerTM containing the radiolabelled drug was enclosed in lead sheeting, which allowed air to be entrained through the device as patients inhaled and also permitted measurement of inspiratory flow parameters using the Respiratory Training Device.
Quantification of distribution of activity in patients
An anterior transmission scan of each patient was initially obtained using a flood source containing 37 MBq 99mTc. Individual values for attenuation of activity due to absorption bybody tissues were calculated as described by Macey and Marshall 12.
Immediately after inhaling the 99mTc-labelled BDP, simultaneous anterior and posterior planar scintigraphic images (120 s acquisition time) of the chest and abdomen together with lateral images of the head and neck were obtained using a double-headed gamma camera (GCA 7200DI; Toshiba Australia, Perth, Australia). Areas of interest were defined for each of the images and separate count rates were determined for the right and left lungs, stomach, oesophagus, mouth and oropharynx. Each count rate was corrected for background counts and tissue attenuation, and the geometric means of the corresponding anterior and posterior count rates were then calculated.
The dose deposited in the lungs was then expressed as a percentage of the total dose deposited in the body or exhaled (i.e. the exactuator or emitted dose).
Statistical analysis
Statistical analysis was carried out using analysis of variance for unmatched data. Post-hoc analysis was performed using the Fisher protected least significant difference with a significance level of 95% (p<0.05), unless otherwise stated. Linear regression analysis was performed to determine significant relationships between variables. All data are presented as mean±sd.
| Results |
|---|
|
|
|---|
Patients were divided into three age groups: 57 (n=5), 810 (n=7), and 1114 yrs (n=4). The mean height, weight, FEV1 and forced vital capacity (FVC) of patients within each age group are shown in table 1
.
|
|
Anterior gamma scintigraphic scans illustrate the changes in distribution of radiolabelled drug within the body in children of different ages (fig. 2
). The distribution of label within the body (corrected for tissue attenuation), calculated as a proportion of the metered (exactuator) dose, is shown intable 3
. All further analyses were performed using the attenuation-corrected data.
|
|
|
|
| Discussion |
|---|
|
|
|---|
Although younger children (57 yrs) received significantly less drug than older children using this device and formulation, the lung dose was still higher than that found with other devices and formulations in this age group 1417. However, the combined oropharyngeal and gastrointestinal dose was higher than that found in older children, and the use of a spacer may still be indicated in this age group. Use of the extrafine BDP formulation delivered via pMDI and used with a low-volume spacer has been shown, in adults, to deliver the same amount of drug to the lungs with less drug deposition inthe mouth 18. This is still an important issue to consider when delivering corticosteroid medication to children of all ages. In addition, use of the AutohalerTM requires the ability to perform a slow maximal inhalation, rather than tidal breathing, which makes it unsuitable for infants and very young children. Hence, the efficacy of the extrafine formulation delivered via pMDI and low-volume spacer should be tested in this age group.
The issue of greater systemic availability through increased pulmonary absorption is of potential concern, particularly due to increased peripheral lung deposition with the extrafine aerosol. However, inflammation is detected throughout the airways in asthmatic individuals, and a more diffuse application of topical prophylactic therapy within the airways may be advantageous.
At the same dose of CFC/BDP and extrafine HFA/BDP (800 µg), an increase in systemic availability of drug in adults was noted with the extrafine formulation 19. However, there was no difference in adrenal effects found with the increased systemic levels of BDP after 14 days of treatment.
Owing to the increased drug delivery to the lungs, when switching over to the extrafine BDP formulation, a concomitant reduction in nominal dose (generally halving the dose) is required to maintain equivalent delivery. In children aged 511 yrs, no significant difference was noted in height velocity over a period of 1 yr, or in bone density or markers ofbone metabolism at half the dose of extrafine HFA/BDP (100200 µg) compared with CFC/BDP (200400 µg) 20. Atthe same doses, in children of the same age, no difference in24-h urinary free cortisol or response to low-dose adrenocorticotrophic hormone test was found between the two formulations after 6 months 21.
Hence, as long as the nominal dose is halved when switching over to the extrafine hydrofluoroalkane/beclomethasone dipropionate formulation, the increased efficiency of drug delivery to the lungs, using the extrafine aerosol, makes it ideal for use in children. Even when inhaled with a poor technique, a relatively high dose is still deposited in the lungs,in both adults 22 and children, as shown in the present study. The AutohalerTM combines the convenience andportability of the pressurised metered-dose inhaler (used without a spacer) and also eliminates problems with coordination of actuation and inhalation. This is of particular advantage in children, and should increase compliance in this age group.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. M. Roller, G. Zhang, R. G. Troedson, C. L. Leach, P. N. Le Souef, and S. G. Devadason Spacer inhalation technique and deposition of extrafine aerosol in asthmatic children Eur. Respir. J., February 1, 2007; 29(2): 299 - 306. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. L. Leach Inhalation Aspects of Therapeutic Aerosols Toxicol Pathol, January 1, 2007; 35(1): 23 - 26. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Kohler, P. Haidl, D. Dellweg, L. Bont, J. L.L. Kimpen, M. J.J. Ermers, E. Baraldi, M. Filippone, T. Guilbert, F. D. Martinez, et al. Inhaled corticosteroids and children. N. Engl. J. Med., August 10, 2006; 355(6): 624 - 624. [Full Text] [PDF] |
||||
![]() |
C. Leach, P. LeSouef, G. P. Chrousos, L. Ghaly, D. Iezzoni, A. Shedden, and A. G. Harris Mometasone and Beclomethasone Comparison Article Observations Chest, May 1, 2006; 129(5): 1389 - 1391. [Full Text] [PDF] |
||||
![]() |
R. Petersen, L. Agertoft, and S. Pedersen Treatment of exercise-induced asthma with beclomethasone dipropionate in children with asthma Eur. Respir. J., December 1, 2004; 24(6): 932 - 937. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |