1 Dept of Pulmonary and Intensive Care Medicine, Otto-von-Guericke-Universität Magdeburg, Magdeburg, Germany. 2 Institute for Cardiovascular Research and 3 Dept of Respiratory Medicine, St James's University Hospital Leeds, Leeds, UK
CORRESPONDENCE: T. Köhnlein, Otto-von-Guericke-Universität Magdeburg, Zentrum Innere Medizin, Pneumologie und Internistische Intensivmedizin, Leipziger Str. 44, D-39120, Magdeburg, Germany. Fax: 49 3916715420. E-mail: thomas.koehnlein@web.de
Keywords: bilevel ventilation, central sleep apnoea syndrome, Cheyne-Stokes respiration, congestive heart failure, continuous positive airway pressure
Received: December 18, 2001
Accepted May 13, 2002
T. Köhnlein was funded with a long-term research fellowship from the European Respiratory Society.
| Abstract |
|---|
|
|
|---|
After excluding those with a history suggestive of obstructive sleep apnoea, 35 patients with CCF (left ventricular ejection fraction <35%) were screened with overnight oximetry and the diagnosis of CSAS was established with polysomnography in 18. Two 14-day cycles of CPAP (0.85 kPa (8.5 mbar)) or bilevel ventilation (0.85/0.3 kPa (8.5/3 mbar)) in random order, were compared in a crossover study.
Sixteen patients (13 males), mean age 62.0±7.4 yrs completed the study. The pretreatment apnoea/hypopnoea index of 26.7±10.7 was significantly reduced by CPAP and bilevel ventilation to 7.7±5.6 and 6.5±6.6, respectively. The arousal index fell from 31.1±10.0 per hour of sleep to 15.7±5.4 and 16.4±6.9, respectively. Significant and equal improvements with CPAP and bilevel ventilation were found for sleep quality, daytime fatigue, circulation time and New York Heart Association class.
The authors conclude that continuous positive airway pressure and bilevel ventilation equally and effectively improve Cheyne-Stokes respiration in patients with congestive cardiac failure.
Central sleep apnoea syndrome (CSAS) with Cheyne-Stokes respiration (CSR) is common in patients with severe cardiac failure. Up to 45% of patients with a left ventricular ejection fraction of
40% may suffer from this breathing disorder during sleep 1, 2. Current concepts of the pathophysiology of CSR suggest it is caused by enhanced carbon dioxide sensitivity, delayed transfer of blood gas tension changes to the chemoreceptors and the influence of adjacent brain centres (circadian rhythms, sleep wake centre) upon the central control of breathing 3. CSR is characterised by periodic changes of tidal volume, resulting in 3060-s cycles of hyperventilation, alternating with hypoventilation or apnoea in a crescendo decrescendo pattern. The resumption of respiration from apnoea is associated with arousals and episodes of sympathetic activation. These episodes may occur several hundred times during the night in patients with severe CSAS 4. The repeated arousals prevent the patient from entering the deeper recuperative stages of sleep, which leads to daytime sleepiness and fatigue. Activation of the sympathetic nervous system coincides with low levels of blood and tissue oxygenation. Heart rate, blood pressure, and the calculated work load for the heart are periodically increased during sleep. As a consequence, the stress upon the failing heart is multiplied 5. Sympathetic activation is a well-documented, independent risk factor for progression of cardiac dysfunction and sudden cardiac death 6.
Recent epidemiological studies have shown a significantly worse survival for patients with left heart insufficiency and CSAS 7, 8. Andreas et al. 9 concluded that CSR during sleep or wakefulness is associated with mortality or the need for transplantation within a few months. Therefore, treatment of CSR in combination with "optimised" medical therapy may influence the course of chronic heart failure, quality of life and the survival of patients 10.
Continuous positive airway pressure (CPAP) and time-controlled bilevel ventilation (bilevel spontaneous/timed) are both used as symptomatic approaches to treat sleep disordered breathing, correct hypoxaemia, reduce arousals, decrease sympathetic tone, and improve sleep quality 11, 12. However, the application of CPAP to patients with CHF may be problematic 13, 14. CPAP requires a spontaneously breathing patient and will be ineffective for central apnoeas, whereas timed bilevel ventilation is able to ventilate a patient even in the absence of spontaneous respiratory effort. Furthermore, some patients find breathing out against positive pressure uncomfortable. Bilevel ventilation has been suggested as an alternative, but little is known about the efficacy or tolerance of bilevel ventilation in these patients and, in particular, it has not been compared to CPAP. The authors hypothesised that bilevel ventilation might be more effective than CPAP in terms of reducing pathological respiratory events and be better tolerated by patients.
The aim of this prospective randomised, nonblinded, crossover, clinical trial was to compare the effect of CPAP and bilevel ventilation on the ventilatory pattern during sleep in a cohort of patients with CSAS during sleep. Secondary outcome parameters were the effect on blood oxygen saturation, arousals from sleep, sleep quality, circulation time, quality of life, daytime fatigue, and New York Heart Association (NYHA) functional class.
| Patients and methods |
|---|
|
|
|---|
Patients with a >1 yr history of chronic left heart failure, in NYHA functional class II and III due to ischaemic heart disease or dilated cardiomyopathy, qualified for the study. Furthermore, inclusion criteria were a left ventricular ejection fraction of
35%. The subjects had to be outpatients in stable clinical condition over the last 3 months before study entry and their medical treatment was optimised, based on the guidelines of the European Society of Cardiology 15. Only minor changes in the dosage of diuretics, but no introduction or withdrawal of any other medication with an effect on the circulation or control of breathing, were allowed in the 3 weeks prior to study entry and during participation in the study. The cardiologic assessment was provided by an independent cardiologist (L.B. Tan) who was blinded for the course of the study.
Exclusion criteria included any pulmonary or neurological disease that might affect ventilation or the control of breathing (such as central alveolar hypoventilation syndrome, chronic obstructive pulmonary disease, pulmonary fibrosis, chronic infections in the respiratory tract, previous stroke, transient ischaemic attack, neuropathy and tumour). Further exclusion criteria were myocardial infarction within 6 months of study entry, severe ventricular arrhythmia, atrial fibrillation 16, cardiac valve replacement, current smoking, known or a strong clinical suspicion of obstructive sleep apnoea (>10 obstructive apnoeas per hour of sleep or >10% of all detected apnoeas being obstructive in nature), diabetes mellitus, and (suspicion of) alcohol or drug abuse. Patients with any noncardiologic medication that would have a possible influence on the control of breathing (e.g. sedatives, tranquillisers, theophylline) were also excluded.
Any deterioration in the subject's condition as a result of participation in this study, judged by clinical assessment, polysomnography or from the patient's own impression, warranted immediate withdrawal from the study.
The study was approved by the Local Research Ethics Committee of the Leeds Teaching Hospitals NHS Trust. Informed written consent was obtained from every patient prior to study entry.
Experimental protocol
The procedure for identifying patients with CSAS consisted of two steps. First, potential patients were provided with a miniature oxygen saturation recording device (Minolta® Pulsox-3i; Minolta (UK) Ltd, Milton Keynes, UK) for an overnight measurement at home. This recording was interpreted according to the study of Staniforth et al. 17. If the oxygen saturation recording showed prolonged sections with regular fluctuations in oxygen saturation of >4% and if the percentage of the night (236 h) spent with an oxygen saturation of
90% was >5%, the recording was suggestive of nocturnal breathing abnormalities. Subjects with these findings underwent screening polysomnography in the sleep laboratory to establish or exclude the diagnosis of central sleep apnoea and CSR (see below for criteria).
CSR patients were randomised and entered a crossover study design (fig. 1
). The randomisation procedure was performed with closed envelopes containing a label for the first treatment mode. The interventions were treatment cycles with CPAP or bilevel ventilation, with each cycle consisting of 12 treatment nights at home. Polysomnography was performed at the beginning and at the end of each treatment cycle. Both cycles were separated by a 14-day washout period.
|
Polysomnography
Full polysomnography with simultaneous recording of electroencephalography (EEG) (C3A2; C4A1), eye movements, submental electromyography (EMG), airflow with a nose/mouth thermistor, respiratory efforts of the chest and abdomen (inductance plethysmography), oxygen saturation, body position, electrocardiography (ECG), heart beat-to-beat interval, respiratory sound, mask air flow and mask pressure, was performed with an Alice 4 system (software version 1.7.16; Respironics Inc., Pittsburgh, PA, USA). Polysomnography was performed according to the standards of the American Thoracic Society 18 and scored according to the criteria of Rechtschaffen and Kales 19. The typical ventilatory pattern of CSR, with regular fluctuation of tidal volume, was identified visually. All scoring was performed by the same investigator (T. Köhnlein). It was not possible to blind either the patients or the investigator to the type of device used.
Circulation time was calculated as described in detail elsewhere 20. In brief, 15 randomly selected sequences with periodic breathing in sleep stage II were analysed for the time interval between the resumption of respiratory efforts and the nadir of the oxygen saturation at the finger tip of the index finger of the left hand (fig. 2
). Assessment of the circulation time was performed in high-resolution printouts with oxygen saturation recording in 1-s intervals.
|
Continuous positive airway pressure and noninvasive ventilation
Nasal CPAP (nCPAP) was delivered with a Sullivan V flow generator (Resmed Inc., Poway, CA, USA). All patients, independent of age, body mass index (BMI) or sex, were treated with 0.85 kPa (8.5 mbar) mask pressure. Bilevel ventilation was performed with a VPAP II flow generator (Resmed Inc.) in the spontaneous-timed mode: during spontaneous breathing, subjects were ventilated in assisted mode with fixed pressure levels of 0.85 kPa (8.5 mbar) during inspiration and 0.3 kPa (3 mbar) during expiration. In case of an apnoea, the ventilator switched to controlled (timed) ventilation after a latency of 5 s. During controlled ventilation, the breathing frequency was fixed at 12 breaths·min1, and the applied pressures were the same as during assisted ventilation. Controlled ventilation lasted until sufficient spontaneous breathing to trigger the assisted ventilation mode reappeared 24.
The run time of all ventilators was recorded with an hour meter in the device recording the actual time the patient received assisted ventilation within the last fourteen 24-h cycles (not visible to the patient). Ultra Mirage masks (Resmed Inc.) or Respironics Special nasal masks (Respironics Inc.) were used according to the subject's preference.
Data analysis
All variables are given as mean±SEM. All data were analysed for a normal distribution with the Kolmogorov-Smirnov test. One-way repeated measures analysis of variance (ANOVA) was applied for all parametric data. A p-value of
0.05 was regarded as statistically significant.
| Results |
|---|
|
|
|---|
|
After 2 weeks of treatment, both CPAP and bilevel ventilation reduced CSR significantly from 37.1±16.6% to 14.1±17.5% (p<0.001) and 13.8±17.6% (p<0.001) of TST, respectively (fig. 3a
). The high amount of CSR was reflected in a pretreatment apnoea/hypopnoea index (AHI) of 26.7±10.7 (range 1546), which was significantly reduced by CPAP and bilevel ventilation to 7.7±5.6 (p<0.00) and 6.5±6.6 (p<0.00, respectively (fig. 3b
)). There were no significant differences in any of the measured variables between CPAP and bilevel ventilation, as they both markedly improved sleep disordered breathing.
|
|
|
|
| Discussion |
|---|
|
|
|---|
The finding that CPAP was well tolerated was surprising, since it has not been experienced by the authors in routine clinical practice. Only two subjects were unable to tolerate noninvasive positive pressure in this study, both withdrawing after 1 night of treatment with CPAP. The remaining 16 patients managed to fulfil all requirements of the protocol.
There was not a high proportion of nonresponders to CPAP, as described by Javaheri 28, or bilevel ventilation. However, the baseline AHI in the patients in the current study (26.7±10.7, range 1546) was considerably lower than in the total cohort (51±23) in the study by Javaheri 28, but was close to the mean AHI of responders to CPAP (36±15) in the latter study 28.
Similar effects of bilevel ventilation on CSA were demonstrated by Teschler et al. 29, who investigated four different treatment modes (oxygen, CPAP, bilevel ventilation and adaptive pressure support servo-ventilation) during 1 treatment night each. In contrast to the present study, Teschler et al. 29 did not find that CPAP was effective. The current study evaluated each treatment after a 12-day treatment period at home to rule out the influence of adaptation problems and first night effects on the assessment of the treatment mode.
The daily ventilator use was satisfactory for both interventions. The run time for CPAP and bilevel ventilation exceeded the average daily run time from the authors' laboratory for patients with obstructive sleep apnoea syndrome (4.3 h·day1). The involvement of a full time doctor in the study might have resulted in greater involvement with the patient than would be seen in normal every day clinical practice and the relatively short period of 14 days might also have been a factor.
The choice of pressure settings was guided by the possible impact of increased intrathoracic pressure on the cardiovascular system and this may have been a further reason for the good compliance with CPAP in this study. Buckle et al. 14 found that their congestive heart failure patients did not tolerate CPAP pressures of >0.50.75 kPa (57.5 mbar). In this study, relatively low pressures of 0.85 kPa (8.5 mbar) and 0.850.3 kPa (8.5/3 mbar) were chosen for CPAP or bilevel ventilation, respectively, similar to those of Naughton et al. 11, who used pressures from 0.51.25 kPa (5 to 12.5 mbar), because of the observation of Becker et al. 30 that there was a decrease in cardiac output with CPAP of >1 kPa (10 mbar) or with bilevel ventilation of
1/0.5 kPa (10/5 mbar) in patients with sleep apnoea.
The applied pressures in this study were effective in all subjects, regardless of the severity of CSR, their BMI or body constitution. Treatment of CSAS does not seem to be critically dependent on the applied pressure, in contrast to patients with obstructive sleep apnoea syndrome 31. The fact that ventilatory parameters during sleep were improved in all patients with the same settings suggests that the beneficial effects of positive intrathoracic pressure upon the failing heart are the predominant mechanism of benefit. In various studies, Bradley and co-workers 11, 26 demonstrated that CPAP is effective not only in improving respiratory events but also in reducing cardiac pre and after load, transmural pressure of intrathoracic blood vessels 32 and cardiac dimensions 33. The current finding of a reduction in circulation time is also consistent with improved cardiac performance.
There were no adverse effects of assisted ventilation in the patients in this study. Hyperventilation is a major aspect of the pathophysiology of CSAS in chronic heart failure patients. CPAP and bilevel ventilation might further increase hyperventilation by reducing the work of breathing. Although ventilation was assessed by visual analysis of the ventilation signals in the polysomnography recordings only, hyperventilation was not detected with any intervention in any patient. Ventilation was regularised with CPAP and bilevel ventilation, i.e. the pre-existing pathological breathing pattern was improved in all subjects. This was true for the relatively low pressure settings in this study. In view of the satisfactory results, higher ventilatory pressures were not studied, and therefore no information on the effect of ventilatory pressure was obtained.
Carbon dioxide tensions were not measured. Direct measurement is not feasible during sleep unless an arterial line is available, but this was not performed with respect to patient safety and comfort. Transcutaneous measurement techniques are not sufficiently responsive for patients with rapid alterations of their breathing pattern and the measurement of gas samples in the exhaled air stream in the nose is quick, but during CPAP or bilevel ventilation there is no undisturbed exhaled air stream for analysis 34. Assisted ventilation could theoretically worsen CSR by reducing the carbon dioxide tension in arterial blood below the apnoeic threshold. The fact that an improvement in CSR was found suggests that this did not occur even though carbon dioxide tensions were not measured. This was probably because of the low pressures that were chosen and higher pressures may have had a deleterious effect on CSR through this mechanism.
The lack of a placebo limb is a potential limitation of this study. However, in other studies, CPAP has been shown to have a beneficial effect on CSR, but because implementation may be difficult, the primary interest was to compare CPAP and bilevel ventilation in CHF patients, and the addition of a placebo limb would have considerably added to the complexity of the protocol for the patients. This study does support the observation in other studies of a beneficial effect of assisted ventilation, even though it was not the primary aim of this study. One advantage of this study compared to previous work is that the selection of subjects was guided by restrictive inclusion and exclusion criteria to avoid the confounding effects of comorbidity. Previous studies with similar design 11, 28, 35 might have been biased, because they were not controlled for cofounders influencing the control of breathing, such as medication and stroke, or the underlying cardiac rhythm.
Health status showed a trend towards improvement and no patient became worse with positive pressure ventilation. The generic questionnaire, SF-36, is probably not sensitive enough to detect changes that might be picked up with disease-specific questionnaires. The improvement in the NYHA functional class of many patients is a further indicator of improved physical health under either treatment. However, the estimation of the NYHA class by the investigators is subjective and therefore open to bias.
In conclusion, this randomised, crossover, open clinical trial confirmed the finding of other authors of significant benefits of continuous positive airway pressure and bilevel ventilation in patients with stable chronic heart failure due to ischaemic heart disease or idiopathic cardiomyopathy and Cheyne-Stokes respiration. However, in the absence of a control group, the fact that this might have been because of other factors, such as closer observation, cannot be excluded. Continuous positive airway pressure and bilevel ventilation with low pressures were equally effective in improving sleep quality, blood oxygenation, arousals from sleep and daytime symptoms. Both interventions were well accepted over the 14-day treatment periods and these data do not suggest a role for bilevel ventilation, the equipment for which is more expensive than that for continuous positive airway pressure. Low pressures are effective and well tolerated. The impact of treatment on the pathophysiology of chronic heart disease and the long-term effects are still unknown, and the results of a multicentre Canadian study of continuous positive airway pressure in chronic left ventricular failure are therefore awaited 36. The data presented in this study support the use of continuous positive airway pressure as an effective treatment for reducing sleep disordered breathing in patients with cardiac failure.
| Acknowledgements |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Arzt, R. Wensel, S. Montalvan, T. Schichtl, S. Schroll, S. Budweiser, F. C. Blumberg, G. A. J. Riegger, and M. Pfeifer Effects of Dynamic Bilevel Positive Airway Pressure Support on Central Sleep Apnea in Men With Heart Failure Chest, July 1, 2008; 134(1): 61 - 66. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ozsancak, C. D'Ambrosio, and N. S. Hill Nocturnal Noninvasive Ventilation Chest, May 1, 2008; 133(5): 1275 - 1286. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Pevernagie, J. P. Janssens, W. De Backer, M. Elliott, J. Pepperell, and S. Andreas Ventilatory support and pharmacological treatment of patients with central apnoea or hypoventilation during sleep Eur. Respir. Rev., December 1, 2007; 16(106): 115 - 124. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Arzt and T. D. Bradley Treatment of Sleep Apnea in Heart Failure Am. J. Respir. Crit. Care Med., June 15, 2006; 173(12): 1300 - 1308. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Philippe, M Stoica-Herman, X Drouot, B Raffestin, P Escourrou, L Hittinger, P-L Michel, S Rouault, and M-P d'Ortho Compliance with and effectiveness of adaptive servoventilation versus continuous positive airway pressure in the treatment of Cheyne-Stokes respiration in heart failure over a six month period Heart, March 1, 2006; 92(3): 337 - 342. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. G. Johnson and D. C. Johnson Bilevel Positive Airway Pressure Worsens Central Apneas During Sleep Chest, October 1, 2005; 128(4): 2141 - 2150. [Abstract] [Full Text] [PDF] |
||||
![]() |
L J Cormican and A Williams Sleep disordered breathing and its treatment in congestive heart failure Heart, October 1, 2005; 91(10): 1265 - 1270. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. K. Gehlbach and E. Geppert The Pulmonary Manifestations of Left Heart Failure Chest, February 1, 2004; 125(2): 669 - 682. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. T. Pepperell, N. A. Maskell, D. R. Jones, B. A. Langford-Wiley, N. Crosthwaite, J. R. Stradling, and R. J. O. Davies A Randomized Controlled Trial of Adaptive Ventilation for Cheyne-Stokes Breathing in Heart Failure Am. J. Respir. Crit. Care Med., November 1, 2003; 168(9): 1109 - 1114. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||