|
|
||||||||
1 Dept of Paediatrics/Respiratory Medicine, 3 Dept of Epidemiology and Biostatistics, 4 Central Instrumentation Dept, Erasmus MC University Medical Center - Sophia Children's Hospital, Rotterdam, The Netherlands. 2 Dept of Paediatrics/Respiratory Medicine, University Hospital Padova, Padova, Italy
CORRESPONDENCE: J.C. de Jongste, Erasmus Medical Center - Sophia Children's Hospital, P.O. Box 2060, 3000 CB, Rotterdam, The Netherlands. Fax: 31 104636801. E-mail: dejongste@lons.azr.nl
Keywords: children, dynamic flow restriction, exhaled nitric oxide, normal values, off-line, on-line
Received: February 15, 2002
Accepted May 31, 2002
M.W.H. Pijnenburg was supported by a grant from the Dutch Asthma Foundation and L. Ghiro was supported by a grant of the Italian Nitric Oxide Club (INOC).
| Abstract |
|---|
|
|
|---|
The authors compared off-line with on-line FEno, assessed feasibility, and obtained normal values for FEno in children aged 48 yrs. Subjects inhaled nitric oxide (NO)-free air and exhaled into the device, where DFR kept exhalation flow constant at 50 mL·s1. Dead space air was discarded. Exhaled air was collected in a 150 mL mylar balloon. On-line measurements were performed and values compared with off-line FEno in 19 adult volunteers. Seventy-nine children performed off-line sampling. All samples were analysed with a chemiluminescence NO-analyser. Normal values were obtained in 34 healthy children.
There was an excellent correlation between on- and off-line values. Bland and Altman plots showed good agreement between on- and off-line FEno. Seventy-four out of 79 children were able to perform a correct off-line procedure. Geometric mean±sem FEno in healthy children was 4.9±1.2 parts per billion (ppb) for male children and 7.6±1.1 ppb for female children.
It can be concluded that off-line fraction of exhaled nitric oxide measurements with dynamic flow restriction are feasible in young children and correspond to on-line values.
Fractional exhaled nitric oxide concentration (FEno) is a noninvasive marker of eosinophilic airway inflammation in asthma 1, 2. The American Thoracic Society (ATS) published guidelines for the sampling of FEno in adults and older children 3. In addition, a joint European Respiratory Society (ERS)/ATS task force on exhaled nitric oxide (NO) measurements in children has summarised its recommendations 4.
FEno values are highly dependent on exhalation flow 5. Hence, ATS guidelines recommend measurements during a single, slow exhalation with constant, low flow. However, 48-yr-old children are often unable to perform the required controlled flow procedures needed for standardised sampling 6, 7. In children unable to cooperate, tidal breathing methods or exhalations with uncontrolled flow have been used, but these are ill standardised and allow for contamination with ambient and nasal NO 3. A dynamic flow restrictor (DFR) that automatically varies resistance depending on blowing pressure overcomes these problems. The ERS/ATS Task Force 2001 is the first to give recommendations on measuring FEno in children, and it mentions DFR as a method of choice 4. However, published data on DFR concern techniques where manual adjustments are made and this limits their applicability. Especially in young children, FEno as a noninvasive, simple, reproducible inflammatory marker would be very useful as a diagnostic and monitoring tool in respiratory disease. Therefore, the authors developed a new, flow-constant device for off-line measurement of FEno in children aged
4 yrs, who can cooperate but have difficulty in performing controlled expiratory manoeuvres. In the present study, the authors first validated the device against on-line FEno in young adult volunteers. Next, FEno was measured in schoolchildren aged 48 yrs in order to assess feasibility and obtain normal values.
| Subjects and methods |
|---|
|
|
|---|
Off-line exhaled nitric oxide measurements
Exhaled air was collected with a new, custom-built device, which contained a mechanical variable flow restrictor that kept exhalation flow constant at 50 mL·s1 over a pressure range of 520 cmH2O (fig. 1
). Preliminary experiments with calibration gas of 115 parts per billion (ppb) indicated that the device was NO inert. Subjects were asked to take a deep breath through a mouthpiece attached to a bag with NO-free medical air via a nonrebreathing valve. They immediately exhaled into the DFR. A manometer checked mouth pressure and manoeuvres were accepted when pressures were 510 cmH2O. A whistling feedback signal indicated a pressure of >5 cmH2O. The dead space air of the subject and device was discarded by a side tube. Air exhaled during the last 34 s was collected in a 150 mL NO-impermeable mylar balloon (Jurjen de Vries BV, Leeuwarden, the Netherlands). Total duration of the measurement was
10 s. All subjects performed five manoeuvres with a varying exhalation time, discarding exhaled air during 3, 4, 6, 8 and 10 s corresponding to 150500 mL, in random order. Between measurements, there was a minimum of 30 s rest. All balloons were sealed, and analysed within 2 h. Samples of ambient air were collected and analysed. A chemiluminescence analyser (Sievers 280 NOA, Boulder, CO, USA) with a sensitivity of <0.1 ppb and a detection range of <0.1500,000 ppb was used for measuring NO. The analyser was checked once daily before the measurements, using certified NO-free gas and certified 115 ppb gas (BOC, Herenthout, Belgium). The sampling flow was 200 mL·min1 and the response time was 200 ms.
|
3 s was required to approve the measurement. Exhalation flow was kept constant through visual feedback and a DFR at 50 mL·s1. Mouth pressure was checked.
Off-line exhaled nitric oxide measurements in children
Exhaled air was collected while children attended school, during the morning hours of 1 week. The first 4 s of exhaled air (200 mL) were discarded, thus the total duration of the measurement was
7 s. All balloons were analysed within 6 h, a period in which NO stays stable in balloons 9. Samples of ambient air were collected every hour.
Data analysis
The method of Bland and Altman 10 was used to assess reproducibility of duplicate measurements and agreement between off-line and on-line values. For correlations between FEno and diary card data, age, height, sex and ambient NO-values, bivariate correlations analysis was used (Pearson's correlation coefficient). Data are presented as geometric means (log transformed, processed and backtransformed) and standard error of the mean±sem. The level of significance was set at p=0.05.
| Results |
|---|
|
|
|---|
|
|
|
Duplicate measurements
Bland and Altman 10 analysis showed no difference and a high reproducibility of duplicate measurements (fig. 3
). The intraclass correlation coefficient was 0.84.
|
| Discussion |
|---|
|
|
|---|
Correlations of different on- and off-line sampling techniques have been reported in the literature 7, 1114. Several authors found an excellent correlation between on- and off-line values; however, agreement between both methods differed in different studies. While Silkoff et al. 12 found lower off-line values compared to on-line values, Paredi et al. 11 found that agreement was flow dependent and best with expiratory flows of
150 mL·s1. At the recommended expiratory flow of 50 mL·s1, these authors found low agreement. In contrast, the present findings show that with a single-breath, low-flow exhalation according to ATS guidelines, controlled by dynamic flow restriction and discarding dead space air, both good correlation and agreement between on- and off-line values can be obtained. As the NIOX NO-analyser (Aerocrine) works with DFR as well, this finding is not surprising.
Several groups described different off-line techniques for sampling exhaled NO 7, 1119. These include tidal breathing techniques and uncontrolled or controlled single exhalations into a reservoir. Tidal breathing techniques require no active cooperation and are feasible down to the newborn age. On the other hand, exhalation flow cannot easily be controlled and it is difficult to prevent nasal contamination adequately, leading to increased variation in FEno values. Uncontrolled single exhalations against a resistance are easy to perform and reproducible even in young children, and there is no need for expensive or complicated equipment 7, 14, 16. However, in the absence of flow control, FEno values may vary and be unsuitable to monitor individual patients. Controlled single exhalation manoeuvres improve reproducibility and agreement with on-line procedures. Biofeedback signals can be used to facilitate the procedure. Nevertheless, many young children will still have difficulty in performing these manoeuvres. Dynamic flow restriction overcomes this problem. The majority of all children in the present study successfully completed the sampling procedure with the off-line device with dynamic flow restriction. Hence, it is now feasible to obtain reliable FEno values at a constant, low flow in children aged
4 yrs, using a simple mechanical sampling device.
The present data are the first on 48-yr-old children, measured with a constant flow of 50 mL·s1. Although the number of children was relatively small, the values found are in agreement with values from previous studies in older children 9, 15, 20. Moreover, symptom scores in the whole study population were low, and FEno values in children with and without symptoms were not significantly different. This suggests that, in this young age group, airway symptoms are less likely to be related to eosinophilic airway inflammation than in older subjects, thus confirming earlier observations 21. Future studies using larger groups should establish reference values.
The authors found significantly higher FEno values in female compared to male children. Only in one previous study was a difference in FEno between males and females found. However, in this study higher FEno levels as well as higher plasma nitrate levels were found in males 22. Male and female children in the present study did not differ in weight, height or age. Female children were not expected to have a higher prevalence of atopy or mild asthma and symptom scores did not differ between males and females (p=0.48) 23, 24. As atopy was not measured specifically, a difference in atopic status between males and females may still be possible in this relatively small sample. The authors could not confirm the correlation between age and FEno that was reported for older children, nor was a correlation with height found 9.
No correlation between diary card symptom score or questionnaire answers and FEno was shown. However, symptom scores in the authors' group (measured in summer) were very low, and only four children were treated for respiratory symptoms with inhaled steroids. Therefore, lack of correlation may be due to the relatively healthy population that was included.
The authors found a significant correlation between FEno and ambient NO levels >7 ppb. Several studies investigated the correlation between FEno and ambient NO using different sampling techniques, resulting in conflicting data 2527. Ambient air can have NO values as high as several hundred ppb and may contaminate dead space of the airways, thus giving false high FEno values after expiration in a reservoir. Inspiration of NO-free air or discarding dead space air have been proposed as methods to eliminate this disturbing factor 3. However, despite these measures, the authors still found a correlation between FEno and ambient NO. Several explanations for the present findings are possible. According to ATS recommendations, children took only one deep breath of NO free air, which may be not enough to effectively wash out dead space NO. Although special attention was paid to closure of the mouth around the mouthpiece during inspiration, some children may have inhaled ambient air along the mouthpiece. Based on the authors' findings in adult volunteers where agreement did not improve with higher discarded volumes, the first 150200 mL of each exhalation was discarded before exhaled air was sampled. Because the anatomical dead space of children is smaller than in adults, discarding this amount of exhaled air should be sufficient to avoid contamination with dead space air. The low FEno in these healthy children may reveal any con<1?show=[fo]>tamination more readily than higher levels in asthmatics. The authors recommend to record ambient NO levels when measuring FEno.
In conclusion, the authors found an excellent agreement and correlation between on- and off-line fractional exhaled nitric oxide values, with a new constant low-flow off-line device for single-breath measurement with dynamic flow restriction. With this device, fractional exhaled nitric oxide can be reproducibly measured in children aged 48 yrs, with a high success rate. Normal fractional exhaled nitric oxide values for healthy male and female children are reported. Furthermore, there was a weak correlation between ambient nitric oxide >7 parts per billion and fractional exhaled nitric oxide levels, despite the inhalation of nitric oxide-free air and dead space discarding. Thus, the authors propose this off-line technique and dynamic flow restriction as a suitable method to collect samples for fractional exhaled nitric oxide measurements in children too young to cooperate with recommended on-line procedures.
| Acknowledgements |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Pijnenburg Exhaled nitric oxide in childhood asthma: methodological issues and clinical applications Eur. Respir. Rev., December 1, 2006; 15(101): 215 - 217. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. W. Turner Exhaled nitric oxide in 4-year-old children: relationship with asthma and atopy Eur. Respir. J., August 1, 2005; 26(2): 359 - 359. [Full Text] [PDF] |
||||
![]() |
ATS/ERS Recommendations for Standardized Procedures for the Online and Offline Measurement of Exhaled Lower Respiratory Nitric Oxide and Nasal Nitric Oxide, 2005 Am. J. Respir. Crit. Care Med., April 15, 2005; 171(8): 912 - 930. [Full Text] [PDF] |
||||
![]() |
J. E. Brussee, H. A. Smit, M. Kerkhof, L. P. Koopman, A. H. Wijga, D. S. Postma, J. Gerritsen, D. E. Grobbee, B. Brunekreef, and J. C. de Jongste Exhaled nitric oxide in 4-year-old children: relationship with asthma and atopy Eur. Respir. J., March 1, 2005; 25(3): 455 - 461. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Franklin, S. M. Stick, P. N. Le Souef, J. G. Ayres, and S. W. Turner Measuring Exhaled Nitric Oxide Levels in Adults: The Importance of Atopy and Airway Responsiveness Chest, November 1, 2004; 126(5): 1540 - 1545. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.J. Franklin, S.W. Turner, R.C. Mutch, and S.M. Stick Comparison of single-breath and tidal breathing exhaled nitric oxide levels in infants Eur. Respir. J., March 1, 2004; 23(3): 369 - 372. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |