|
|
||||||||
1 Dept of Medicine, University of Birmingham, Selly Oak Hospital, Birmingham, 2 Dept of Medicine, Good Hope NHS Trust, Sutton Coldfield, West Midlands and 3 Dept of Medicine, Solihull NHS Trust, Solihull, West Midlands
CORRESPONDENCE: M.R. Miller, Dept of Medicine, University of Birmingham, Selly Oak Hospital, Birmingham, B29 6JD, UK. Fax: 44 1216278292. E-mail: m.r.miller@bham.ac.uk
Keywords: fast Fourier analysis, lung function, peak expiratory flow, pneumotachography, spirometry
Received: March 18, 2001
Accepted August 22, 2001
| Abstract |
|---|
|
|
|---|
The frequency response of a Fleisch pneumotachograph (PT) was measured and used to record MFEM from 24 patients attending a lung-function laboratory and 26 normal volunteers. The first 1.024 s of the signal recorded at 1,000 Hz for that blow with maximum PEF, underwent fast Fourier transformation using a triangular window function, applied after 0.75 s to reduce flow linearly to zero. All the frequencies above a set limit were removed, followed by inverse transformation to reconstitute the blow. The limits for this frequency cut-off were progressively varied from 100 Hz down to 15 Hz, with the resulting PEF being compared with the PEF from the reconstituted blow with no frequency reduction.
The average±sd age for the group was 47±18 yrs and the average PEF was 450±187 L·min1, which, when expressed as a standardized residual, was 0.1±2.1, with a range from 4.53.9 indicating a good spread around normal values. Average rise time to PEF was 83±38 ms and dwell time >90% PEF was 45±25 ms. Cut-off >20 Hz reduced the mean PEF of the group by 8.5 L·min1 (95% confidence limit 5.511.4 L·min1), whereas cut-off >30 Hz reduced mean PEF by 4.4 L·min1 (2.66.2). In the present study subjects, 30 Hz was on the 95th percentile of frequencies for defining the upper limit for 98% of the power spectrum for the first second of the blow.
It has been shown that there are frequencies >20 Hz that contribute to peak expiratory flow enough to influence readings made using conventional hand-held peak expiratory flow meters, such as the mini-Wright. Devices used for recording flow from a maximum forced expiratory manoeuvre should therefore have an adequate frequency response of up to 30 Hz.
The frequency content of the maximal forced expiratory manoeuvre (MFEM) has been previously defined, with the finding of a significant amplitude content in frequencies up to 20 Hz for maximal breathing-capacity manoeuvres 1 and up to 4 Hz for tidal breathing and vital capacity manoeuvres recorded with a bell spirometer. Other studies found that <5% of the total frequency content was >10 Hz in males performing forced expiratory flow manoeuvres 2. Therefore, 20 Hz has been accepted as the limit for adequate recording devices when measuring the whole MFEM 3. One of the lung-function indices that might be most affected by alterations in frequency content is peak expiratory flow (PEF). It is possible that the frequency content at the beginning of an MFEM that contributes to PEF, may be higher than that suggested from studying the MFEM overall. This aspect of lung function may not be recorded with adequate fidelity if this fact is ignored.
Therefore, this study was undertaken to determine the frequency content of the first second of the MFEM, so that endeavours to test the frequency-response characteristics of PEF recording devices can be properly matched to the signal requirements.
| Method |
|---|
|
|
|---|
|
| (001) |
|
| (002) |
|
| (003) |
This PT system was used to record MFEM from 50 volunteer subjects. Twenty-four of these subjects were patients (11 males) attending a lung-function laboratory for routine testing and 26 (13 males) were volunteers from hospital staff. Three MFEM that met the acceptance criteria of the American Thoracic Society (ATS) 10 were recorded for each subject. Flow was sampled at 1-ms intervals and stored on computer disk. From the three blows, maximal PEF, forced vital capacity (FVC) and forced expiratory volume in one second (FEV1) were calculated and related to predicted values using the European Community for Steel and Coal (ECSC) equations and the method of standardized residuals 3. The blow with the largest PEF had a 1090% rise time to PEF and a dwell time for flow >90% of peak calculated. The data from the first 1.024 s of this blow underwent fast Fourier transform (FFT) analysis using Origin version 6.1 (OriginLab. Corp., MA, USA) with a triangular window function applied, so that from 0.75 s flow was linearly reduced to zero at 1.024 s. This was necessary in order to avoid spurious high-frequency content, due to the sudden transition to zero flow by truncation at 1.024 s. The derived FFT coefficients then underwent inverse transformation to regenerate the blow, with no alteration to the coefficients. PEF was derived from this reconstituted blow and called PEFR. This process of inverse transformation to reconstitute the blow was sequentially repeated with prior removal of all the frequency components at >100, 90, 80, 70 , 60, 50, 40, 30, 20, and 15 Hz. The PEF for these regenerated blows was calculated and stored as PEF100PEF15, respectively. Comparison between PEFR and each of the frequency-reduced PEF was performed by paired t-tests.
Any possible effect upon the recorded PEF from the frequency response of the PT system was checked by dividing the amplitude coefficients of the FFT from the flow/time data by the transfer function of the PT gain for each frequency and then performing an inverse FFT. This was only an approximation for the correction, since no account of phase changes could be taken.
| Results |
|---|
|
|
|---|
|
|
|
|
30 Hz for 98% of the power spectrum.
|
| Discussion |
|---|
|
|
|---|
To be certain that these conclusions are correct, the recording device for the original signals must be more than adequate for the purpose. The linearity of the PT has been previously verified using an adequate pump system 9, and its calibration has been performed by an accepted method that minimizes any bias across its range. The frequency response has been tested with a step test, which is one of the best ways to determine its characteristics. The method described is suitable for any device that has a continuous analogue output. An explosive decompression device fitted with a fast-response solenoid was used, but other flow-generating devices could be used. An "off" step test was used, as suddenly ceasing a flow has fewer potential artefacts when compared with suddenly generating a constant flow for an "on" step response.
The authors used a system that was under-damped, but with a natural resonant frequency well in excess of the frequency content of interest. An alternative would have been to have a system that had a lower resonant frequency but was critically damped. The authors believe that the PT system was more than adequate for the purpose of recording MFEM from humans. This has been verified by demonstrating the insignificant change in PEF when making an approximate correction for the PT's frequency-response characteristics. When using this PT system for subjects in a clinical setting, a low-pass analogue filter set at 50 Hz was employed. For this study, a much higher limit was used with a 200 Hz low-pass filter to ensure that none of the frequencies of interest could be affected.
The second requirement for this study to be representative was that the subject population adequately reflected the relevant client group. A sample of 50 subjects was examined. One-half of the sample were without symptoms or known disease and thus regarded as normal, and the other half were symptomatic subjects with airflow limitation. The results in table 1
indicate that these subjects ranged from 4.53.9 sd from predicted in PEF, with an average that was essentially as predicted. Whilst a larger sample might have shown tighter confidence limits in these findings, it is believed that the subjects were sufficiently representative for the findings to be accepted. The statistical significance of a change in PEF was at a cut-off frequency of 60 Hz, but it was at 30 Hz that the average change exceeded 5 L·min1, which would be the amount necessary to influence a reading from an analogue meter, such as the mini-Wright.
The present finding is important for verification of whether flow meters have an adequate frequency response to record peak expiratory flow and other flow phenomena. For meters such as the mini-Wright, which only have a single peak-flow output, test signals that span the range of input signals of interest are needed. The characteristics of these signals are probably best described by their rise and dwell time 11. Such signals, with different frequency content but identical flow, can be delivered to the meters, which should give the same reading irrespective of the frequency content of the input signals. Such tests can be verified by using a pneumotachograph to record the flow. This study presents evidence that the requirements for frequency response testing with regard to measuring peak expiratory flow should include frequencies up to 30 Hz.
| Acknowledgements |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. R. Miller, J. Hankinson, V. Brusasco, F. Burgos, R. Casaburi, A. Coates, R. Crapo, P. Enright, C. P. M. van der Grinten, P. Gustafsson, et al. Standardisation of spirometry Eur. Respir. J., August 1, 2005; 26(2): 319 - 338. [Abstract] [Full Text] [PDF] |
||||
![]() |
M R Miller, P R Atkins, and O F Pedersen Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device Thorax, May 1, 2003; 58(5): 411 - 416. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |