1 Service de Physiologie, Explorations Fonctionnelles et Centre d'Investigations Technologiques, Hôpital Raymond Poincaré Garches, 2 Institut National de la Santé et de la Recherche Médicale, Hôpital Henri Mondor, Créteil, France and 3 Association d'Entraide des Polios et Handicapés, Puteaux, France
CORRESPONDENCE: F. Lofaso, Service de Physiologie-Explorations Fonctionnelles, Hôpital Raymond Poincaré, 92380, Garches, France. Fax: 33 147107943
Keywords: Continuous positive airway pressure, obstructive sleep apnoea
Received: February 8, 2001
Accepted September 14, 2001
This study was supported by the Institut national de la Santé et de la Recherche Médicale.
| Abstract |
|---|
|
|
|---|
Snoring was simulated by a loudspeaker connected to the lung model and abruptly activated during 1 s of each inspiratory period to induce pressure oscillation. The oscillation frequencies chosen were 30, 60, 90, and 120 Hz. For each frequency, the amplitude of the pressure oscillation produced by the loudspeaker was adjusted to find the threshold at which the auto-nCPAP devices detected snoring.
Differences in pressure-amplitude thresholds of up to three-fold were found across auto-nCPAP devices. A randomized clinical study to compare the effects of the least sensitive (Virtuoso LX; Respironics, Nantes, France) and one of the most sensitive, (Goodknight 418A; Malinckrodt, Nancy, France) devices, in two groups of six patients with obstructive sleep apnoea syndrome was then conducted. Goodknight 418A was more sensitive than Virtuoso LX for detecting snoring (mean±sd 92±11% versus 50±39% respectively, p=0.03).
To conclude, striking differences exist between auto-nasal continuous positive airway pressure devices in sensitivity for detecting snoring.
Nasal continuous positive airway pressure (nCPAP) was introduced in 1981 by Sullivan et al. 1 and has improved the treatment and prognosis of patients with obstructive sleep apnoea syndrome considerably. In practice, the optimal nCPAP level is a "trade-off" between pressure-related side-effects and efficacy in preventing upper airway obstruction during sleep 2. This optimal level is generally determined during a total or split-night study. Follow-up is needed to verify that the selected level remains appropriate for the patient's needs, since the minimal effective pressure can vary over time depending on weight variations, sleep deprivation, nasal obstruction, and alcohol or hypnotic agent ingestion 3. In addition, this minimal pressure can change during a given night, according to body position and/or sleep stage 3.
To try to improve the efficacy in relieving upper airway obstruction and the acceptability of nCPAP, new nCPAP devices have been developed that use noninvasive indirect assessment of upper airway obstruction to continuously adjust the pressure around the minimal level that prevents abnormal breathing and arousal. These devices use noninvasive indirect methods to detect respiratory events such as obstructive apnoeas, hypopnoea, snoring and flow limitation 4. Detection of flow limitation and differentiation between central and obstructive apnoeas are available only with the most sophisticated devices (AutoSet; ResCare, Sydney, Australia and Goodknight 418P, Malinckrodt, Nancy, France). Because auto-nCPAP devices cannot reliably differentiate central hypopnoeas from obstructive hypopnoeas, the latter being easily detected as flow limitation, hypopnoea detection is not used at all by some auto-nCPAP devices (Autoset) to adjust pressure and is used optionally by other devices (Goodknight 418P). In contrast, nearly all auto-nCPAP devices detect snoring. No studies have evaluated the snoring detection performance of these devices. The authors compared snoring detection by several auto-nCPAP devices.
A bench study was first conducted to examine the performance of six commercially available auto-nCPAP devices (table 1
). The second part of the study evaluated the clinical relevance of differences found between two auto-nCPAP devices during the first part of the study.
|
| Methods |
|---|
|
|
|---|
|
The auto-nCPAP devices were tested after pressure adjustment to 4, 8 and 12 cmH2O. To simulate snoring, a loudspeaker connected to the circuit between the auto-nCPAP and the CPAP pressurized chamber (fig. 1a
) was activated abruptly to produce a pressure oscillation during 1 s of each inspiratory period (fig. 1b
). The frequencies chosen were 30, 60, 90, and 120 Hz. For each frequency, the amplitude of the pressure oscillation produced by the loudspeaker was adjusted to identify the smallest value at which the auto-CPAP detected snoring. This threshold was validated by checking that no snoring was detected when an amplitude of
0.02 cmH2O below the threshold was maintained for 5 min. The effect of increasing and decreasing the duration of pressure oscillation from the total inspiratory time to 0 s or of substituting an expiratory pressure oscillation for the inspiratory pressure oscillation, on the snoring detection threshold was also evaluated.
Clinical study
Apparatus tested
Two auto-nCPAP devices were tested, the device that was least sensitive (Virtuoso LX; Respironics, Nantes, France) and one of the devices that were most sensitive (Goodknight 418A) in detecting snoring in the bench study. The Goodknight 418A was chosen because as with the Virtuoso LX, it detects snoring and apnoeas only.
Patients and clinical trial
This study was approved by the Research Ethics Committee of the authors' institution. All patients gave written informed consent to participate in the study. Twelve male patients with sleep apnoea syndrome requiring nCPAP were randomly allocated to two groups of six. Both groups received auto-nCPAP treatment during polysomnography. Auto-nCPAP was delivered by the Virtuoso LX in one group and by the Goodknight 418A in the other. Polysomnography included electroencephalography (EEG), electro-oculography (EOG), chin electromyography (EMG), nasal flow measured by a Fleisch No. 2 pneumotachograph connected to a differential pressure transducer (Validyne MP45±5 cmH2O), thoracic and abdominal movements, arterial pulse oximetry (Nellcor BS, Nellcor Inc., Hayward, CA, USA), nasal mask pressure measured by a differential pressure transducer (Validyne MP45±35 cmH2O), and oesophageal pressure (Gaeltec, Dunvegan, Isle of Skye, UK).
The initial nCPAP level was always set at the smallest possible value, i.e. 4 cmH2O. The algorithm provided by each manufacturer allowed the nCPAP level to increase according to whether snoring and/or apnoeas were detected.
Data analysis
On each polysomnography recording, the number of snoring events during objectively measured sleep was counted by one of the authors, who did not know which auto-nCPAP device had been used. A snoring event was defined as detection of pressure and flow oscillations 5, 6 during three consecutive respiratory cycles with an increase in the oesophageal pressure swing. The percentage of snoring events followed by a pressure increase was calculated for each patient. The percentage of pressure increases without concomitant snoring or apnoea events ("inappropriate pressure increases") was also calculated.
Statistics
Data are presented as means±sd. Comparisons were made using unpaired t-tests. The level of significance was set at 5%.
| Results |
|---|
|
|
|---|
0.2 cmH2O with the Autoset T and the Horizon auto-adjust, 1 cmH2O with the Goodknight 418P, Goodknight 418A, and Eclipse auto; and 2 cmH2O with the Virtuosa LX.
|
0.5 s for the Goodknight 418A, the Goodknight 418P and the Autoset T and about 0.3 s for the Virtuoso LX and Horizon. Snoring detection with the Eclipse auto required at least 30 oscillations and therefore, the minimal duration of pressure oscillation varied with the oscillation frequency, from <0.3 s at 120 Hz to 1 s at 30 Hz. In addition, Autoset T and Horizon auto adjust were apparently not able to detect snoring when pressure oscillation occurred during expiration.
Clinical study
The main characteristics of the two patient groups and the performance of the two auto-nCPAP devices are reported in table 2
. The two patient groups were similar, whereas snoring detection sensitivity was significantly lower with the Virtuoso LX than with the Goodknight 418A. No unexplained increases in airway pressure were observed.
|
| Discussion |
|---|
|
|
|---|
The oscillation frequencies chosen to simulate snoring were within the range of those previously observed during snoring in humans: Liistro and co-workers 5, 6 demonstrated that snoring was associated with high-frequency oscillation (4090 Hz during snoring through the nose and around 30 Hz during snoring through the mouth) of the soft palate, pharyngeal walls, epiglottis, and tongue, resulting in airway pressure and flow vibrations.
For all devices, it was observed that the sensitivity of snoring detection decreased when the nCPAP level increased. This finding is probably ascribable to the increase in turbine noise associated with increasing turbine speed. It suggests that snoring sensitivity may be lower in those patients who respond best to high nCPAP levels.
The clinical relevance of the differences observed in the bench study were also demonstrated; one of the most sensitive devices in the bench study was more sensitive for detecting snoring in patients than the least sensitive device in the bench study. In addition, the clinical study established that the most sensitive device remained as specific as the least sensitive device. The authors recently conducted a clinical study of an auto-nCPAP device (rapid eye movement (REM)+auto) that was the precursor of the Goodknight 418A and had similar sensitivity for snoring detection (unpublished data). It was found that this device induced no detrimental pressure increases 7, 8. Moreover, in one of the 15 patients, who was described by their bed partner as an almost continuous snorer, this auto-nCPAP device failed to detect snoring because the sound was not loud enough 7. This result was confirmed by a study reported in abstract form 9, in which snoring was recorded by a stethoscope secured to the trachea in 13 patients who used the precursor of the Goodknight 418A. The investigators reviewed the polysomnographical traces while listening with earphones to the stethoscope recording in order to identify snoring. The device increased its pressure in response to only 84±6% of snoring events 9. Similarly, using a less sensitive (more demanding) standard of reference for diagnosing snoring, it was observed that the Goodknight 418A increased its pressure response to 92±11% of snoring events. This suggests that one of the most sensitive devices may fail to respond consistently to low-intensity snoring and that similar failure may be even more common with less sensitive devices.
To conclude, important differences in snoring detection exist between commercially available auto-nasal continuous positive airway pressure devices, under both bench study and clinical conditions. The more sensitive devices seem acceptable, whereas the less sensitive fail to ensure effective prevention of snoring. In addition, the major differences between auto-nasal continuous positive airway pressure devices suggests that further studies are needed to evaluate and compare such devices, particularly the new devices that use detection of flow limitation or those that continuously measure the impedance and/or resistance of the respiratory system using noninvasive means such as the forced oscillation method 10, 11.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. Lofaso, G. Desmarais, K. Leroux, V. Zalc, R. Fodil, D. Isabey, and B. Louis Bench evaluation of flow limitation detection by automated continuous positive airway pressure devices. Chest, August 1, 2006; 130(2): 343 - 349. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Abdenbi, B. Chambille, and P. Escourrou Bench testing of auto-adjusting positive airway pressure devices Eur. Respir. J., October 1, 2004; 24(4): 649 - 658. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Noseda, C. Kempenaers, M. Kerkhofs, S. Braun, P. Linkowski, and E. Jann Constant vs Auto-Continuous Positive Airway Pressure in Patients With Sleep Apnea Hypopnea Syndrome and a High Variability in Pressure Requirement Chest, July 1, 2004; 126(1): 31 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |