ERJ
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Permissions
Right arrowRequest Permissions
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (7)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Broeders, M.E.A.C.
Right arrow Articles by Folgering, H.Th.M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Broeders, M.E.A.C.
Right arrow Articles by Folgering, H.Th.M.
Eur Respir J 2001; 18:780-783
Copyright ©ERS Journals Ltd 2001


Peak inspiratory flow rate and slope of the inhalation profiles in dry powder inhalers

M.E.A.C. Broeders1, J. Molema1, N.A. Vermue2 and H.Th.M. Folgering1

1 Dept of Pulmonary Diseases Dekkerswald, University of Nijmegen, Groesbeek, the Netherlands. 2 GlaxoSmithKline, Zeist, the Netherlands

CORRESPONDENCE: M.E.A.C. Broeders, Dept of Pulmonary Diseases Dekkerswald, University of Nijmegen, P.O. Box 9001, 6560 GB Groesbeek, The Netherlands. Fax: 31 246859290

Keywords: dry powder inhaler, inhalation therapy, peak inspiratory flow

Received: April 26, 2001
Accepted June 30, 2001

This study was supported by GlaxoSmithKline Zeist, the Netherlands.


    Abstract
 TOP
 Abstract
 Methods and measurements
 Results
 Discussion
 References
 
Aerosol delivery depends on device design and inhalation technique. In vitro device evaluations have shown that the emitted dose and fine particle mass of dry powder inhalers (DPIs) increase at high peak inspiratory flow rates (PIFR). Since the PIFR is mostly achieved after the release of the powder, slope of the pressure profile is also described as an important determinant. The aim of the present study was to assess whether the PIFR while using Diskus® and Turbuhaler® inhalers could be used to predict the slope of the inhalation pressure profile.

In a group of 10 stable asthma patients and three groups (mild, moderate and severe) of 16 chronic obstructive pulmonary disease (COPD) patients, lung function was measured, and for each device, 18 inhalation profiles were recorded with the inhalation profile recorder during six sessions over 10 weeks.

The values for the pressure slope and PIFR of both Diskus® and Turbuhaler® were significantly correlated. The r-;values were 0.865 and 0.882, respectively (p<0.01). Percentage explained variance was 74.8% for Diskus® and 77.8% for Turbuhaler®.

Significant correlations were found between peak inspiratory flow rates and slopes. It has been shown for two different dry powder inhalers that peak inspiratory flow rate and slope correlate well in a wide range of patient groups.

Drug delivery by dry powder inhalers (DPIs) depends on the inbuilt resistance of the inhaler and the inspiratory flow generated by the patient. This inspiratory flow depends on the respiratory muscle force and the airway resistance of the patient 1, 2. Patients must be able to generate a sufficient inspiratory flow in order to release the powder and deaggregate the drug to generate respirable particles. Consequently, lung deposition will generally increase at higher inhalation flows 3. Inhalation through a DPI is often described by a pressure or flow versus time curve (inhalation profile). The pressure profile can be integrated into a flow profile if the device's resistivity is known 4.

In the present study, two multidose dry powder inhalers were compared: Turbuhaler inhaler® (AstraZeneca, Sweden) and the Diskus inhaler® (GlaxoSmithKline, UK). In vitro device evaluations showed that the emitted dose and fine particle mass of the Turbuhaler® (higher resistivity) increase at higher peak inspiratory flow rates (PIFRs) 5, 6, whereas the Diskus® (lower resistivity) provided a more consistent dose delivery, relatively independent of different airflows 6, 7. However, there was therapeutic equivalence of salbutamol given via these devices 8.

The PIFR is the most referred parameter of the inhalation profiles. However, it is not the only characteristic parameter. Since the PIFR is mostly achieved after the release of the powder, the slope of the pressure profile (in kPa·s–1) has also been described as an important determinant 6, 9, 10. The present investigation was undertaken to assess whether the PIFR through a Diskus® and a Turbuhaler® inhaler could be used to predict the slope of the inhalation pressure profile.


    Methods and measurements
 TOP
 Abstract
 Methods and measurements
 Results
 Discussion
 References
 
Ten asthmatics and 48 patients with chronic obstructive pulmonary disease (COPD), according to European Respiratory Society (ERS) criteria 11, participated in the study. The patient characteristics are shown in table 1Go.


View this table:
[in this window]
[in a new window]
 
Table 1— Patient characteristics

 
Six identical measurements were performed within 10 weeks in order to measure reproducibility. First, a flow/volume curve was taken with an integrating pneumotachograph (Spiro Analyzer ST-;250, Fukuda Sangyo, Japan). Next, patients inhaled 400 µg salbutamol via a pressurized metered dose inhaler (pMDI) and a spacer to measure reversibility. After 20 min, routine lung function was performed and maximal inspiratory and expiratory mouth pressures (MIP and MEP, respectively) were measured as described by Wilson et al. 12. The inhalation pressure profiles were recorded by a pressure transducer (GlaxoSmithKline Research and Development Dept, Ware, UK), measuring pressures in the mouthpiece during inhalation through Diskus® and Turbuhaler®. The inhalation profiles were stored on the inhalation profile recorder (IPR). Patients inhaled three times through each device, hence, 18 profiles through each device were collected from each patient during the study. Both devices were blinded in a specially constructed box.

The parameters of the dry powder inhaler-inhalation profile recorder (DPI-;IPR) are: 1) pressure slope (Sl) (kPa·s–1); 2) peak pressure drop (PPD) (kPa); 3) time to peak pressure drop: the time between the onset of the inhalation and the moment of reaching the PPD (Tp) (s) (the start of the measuring time is the moment that the pressure passes the 0.2 kPa threshold); and 4) PIFR (L·min–1). If the PPD was reached within the first 0.5 s, the slope was calculated by PPD/Tp. If the PPD was achieved after 0.5 s, the slope was calculated by pressure at 0.5 s/0.5 s to approach the actual slope. The relationship between PIFR and PPD was calculated by PPD=(PIFRxR)2 4, where R is the resistivity of the devices used, measured in a separate experiment (RDiskus®=0.02133 kPa0.5·L–1·min and RTurbuhaler®=0.03223 kPa0.5·L–1·min) 13.

The study was approved by the local ethics committee and conducted in accordance with the Declaration of Helsinki. All patients gave written informed consent.

Statistical analysis
The means of slope and PPD of the 18 inhalation profiles were calculated for each patient. Also the mean lung function parameters were calculated. Relationships between slopes and PIFRs of all profiles of the Diskus® (SlD and PIFRD) and slopes and PIFRs of all profiles of the Turbuhaler® (SlT and PIFRT) were calculated using Spearman's correlation coefficient. The correlations between the maximal inspiratory pressure slope and PIFR for Diskus® and Turbuhaler® were again calculated using Spearman's correlation analysis. Data were expressed as mean±sd. A p-;value <0.05 was considered statistically significant.


    Results
 TOP
 Abstract
 Methods and measurements
 Results
 Discussion
 References
 
No significant difference was found between SlD and SlT (13.5±5.0 and 13.8±4.7 kPa·s–1, respectively). However, PIFR values of both devices were significantly different (PIFRD 108.3±20.4 and PIFRT 76.1±13.8 L·min–1; p<0.001), due to the different resistivities of the devices.

The values of the slope and PIFR of both Diskus® and Turbuhaler® devices were significantly correlated. The r-;values were 0.865 and 0.882, respectively (p<0.01). Figures 1Go and 2Go show the correlation of slope and PIFR of Diskus® and Turbuhaler®. The percentages of explained variance (r2x100) were 74.8% and 77.8%. Furthermore, significant correlations were found between the maximal inspiratory mouth pressure, slope and PIFR. The r-;values were 0.449 for MIP versus SlD and 0.571 for MIP versus PIFRD. The r-;value for correlation between MIP versus SlT was 0.484, and for MIP versus PIFRT was 0.632 (p<0.01).



View larger version (20K):
[in this window]
[in a new window]
 
Fig. 1.— Correlation of slope (SlD) and peak inspiratory flow rate (PIFRD) of Diskus®. Data were extrapolated. Arrows refer to the relationship between the minimal PIFRD (30 L·min–1) and the accompanying SlD. {square}: severe; {triangleup}: moderate; {diamond}: mild; {circ}: asthma. r=0.865.

 


View larger version (18K):
[in this window]
[in a new window]
 
Fig. 2.— Correlation of slope (SlT) and peak inspiratory flow rate (PIFRT) of Turbuhaler®. Data were extrapolated. Arrows refer to the relationship between the minimal PIFRT (60 L·min–1) and the accompanying SlT. {square}: severe; {triangleup}: moderate; {diamond}: mild; {circ}: asthma. r=0.882.

 

    Discussion
 TOP
 Abstract
 Methods and measurements
 Results
 Discussion
 References
 
The aim of present study was to assess whether the PIFR through Diskus® and Turbuhaler® inhalers could be used to predict the slope of the inhalation pressure profile. This study showed significant correlations between the PIFR and pressure slope of the inhalation irofiles of the two inhalers.

Drug particle release from DPIs is determined by a patient's peak inspiratory flow rate and pressure slope. Both depend on the patient's effort, related to the peak flow and the pressure build-up during the first part of the inhalation profile. The significant correlations between the maximal inspiratory mouth pressure and the inhalation profile parameters were in agreement with this principle. Therefore, it is important that patients should be instructed to inhale in a fast and powerful fashion.

Respiratory muscle function is often decreased in COPD patients due to hyperinflation, hypoxaemia and the use of corticosteriods. It must be noted that even the severe COPD patients in this study generated almost normal MIP values (96.3% of predicted). Nevertheless, their MIP values were considerably lower than the values of the asthmatics and the other COPD patients. Oral corticosteroids were not used, so hyperinflation must play a role. Although it is apparent that the PIFR and slope parameters are important determinants for an efficient use of a DPI (e.g. drug deposition), the results of this study indicate that they are also related to each other (r=0.865 (p<0.05) for the Diskus® and r=0.882 (p<0.05) for the Turbuhaler®). The significant correlation between PIFR and the pressure slope in this study may indicate that the measured PIFR will be a reasonable representation of slope. However, for a given PIFR, the slope can have a wide range (e.g. where PIFRD=120 L·min–1, SlD varied between 13 and 24 kPa·s–1; fig. 1Go).

It is remarkable that such good linear correlations were found in these systems, which are by definition alinear, due to the turbulent flows (Rohrer equation: P=k1xQ2+k2xQ). The devices are designed to generate turbulence at the mouthpiece in order to achieve optimal aerosolization of the drug 10, 14. However, the function is not a linear one over the entire range: the line was fitted through zero flow at zero slope. Comparing both parameters in different studiesGo should be done with some caution. No significant difference between slopes of Diskus® and Turbuhaler® were found. However, a decreasing slope by increasing resistivity could be expected. This may possibly be due to the small difference in the manufacturers' instructions. The Turbuhaler® instructions emphasize to inhale deeply and forcefully; whereas the instructions of the Diskus® emphasizes inhaling deeply and forcefully, but not too fast. It is possible that patients inhaled with less effort through the Diskus®, resulting in a lower slope value. Eventually a high effort through a high resistivity device (Turbuhaler®) resulted the same slope as a "not too fast effort" in a lower resistivity device (Diskus®). In the literature, only PIFR-;values have been shown, and minimal attention has been paid to other inhalation profile parameters. However, since the slope is such an important parameter, there must be a minimal slope for "optimal" delivery. Extrapolation of the present data gave a slope value of 3 kPa·s–1 for Diskus® at a PIFR of 30 L·min–1, (although these extrapolated values were outside the measured range) and 9 kPa·s–1 for Turbuhaler® at PIFR of 60 L·min–1.

In conclusion, reasonable correlations were found between peak inspiratory flow rates and slopes. It has been shown, for two different dry powder inhalers, that peak inspiratory flow rate and slope correlate well in a wide range of patient groups.


    References
 TOP
 Abstract
 Methods and measurements
 Results
 Discussion
 References
 

  1. Selroos O, Pietinhalho A, Riska H. Delivery devices for inhaled asthma medication: clinical implications of differences in effectiveness. Clin Immunother 1996;6:273–299.
  2. Ganderton D. General factors inflencing drug delivery to the lung. Respir Med 1997;91:13–16.[CrossRef]
  3. Pauwels R, Newman S, Borgstrom L. Airway deposition and airway effects of antiasthma drugs delivered from metered-dose inhalers. Eur Respir J 1997;10:2127–2138.[Abstract]
  4. Clark AR, Hollingworth AM. The relationship between powder inhaler resistance and peak inspiratory conditions in healthy volunteers-implications for in vitro testing. J Aerosol Med 1993;6:99–110.[Web of Science][Medline] [Order article via Infotrieve]
  5. Ganderton D. General factors influencing drug delivery to the lung. Respir Med 1997;91:Suppl. A, 13–16.
  6. Bisgaard H, Klug B, Sumby BS, Burnell PK. Fine particle mass from the Diskus inhaler and Turbuhaler inhaler in children with asthma. Eur Respir J 1998;11:1111–1115.[Abstract]
  7. Brindley A, Sumby BS, Smith IJ, Prime D, Haywood PA, Grant AC. Design, manufacture and dose consistency of the serevent Diskus inhaler. Pharm Tech Eur 1995;7:14–22.
  8. Arvidsson P, Mellen A, Palmqvist M, Lotvall J. Equivalent therapeutic ratio of salbutamol given by Turbuhaler and Diskus. Respir Med 2000;94:574–577.[Medline] [Order article via Infotrieve]
  9. Everard ML, Devadason SG, Le Souef PN. Flow early in the inspiratory manoeuvre affects the aerosol particle size distribution from a Turbuhaler. Respir Med 1997;91:624–628.[CrossRef][Medline] [Order article via Infotrieve]
  10. de Boer AH, Winter HMI, Lerk CF. Inhalation characteristics and their effects on in vitro drug delivery from dry powder inhalers. Part 1. Inhalation characteristics, work of breathing and volunteers' preference in dependence of the inhaler resistance. Int J Pharm 1996;130:231–244.[CrossRef]
  11. Siafakas NM, Vermeire P, Pride NB, et al. Optimal assessment and management of chronic obstructive pulmonary disease (COPD). The European Respiratory Society Task Force. Eur Respir J 1995;8:1398–1420.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  12. Wilson SH, Cooke NT, Edwards RI, Spiro SG. Predicted normal values for maximal respiratory pressures in Caucasian adults and children. Thorax 1984;39:535–538.[Abstract/Free Full Text]
  13. Broeders MEAC, Molema J, Folgering HThM. Resistivities of placebo and active Diskus® inhalers compared: statistically significant, clinically irrelevant. Am J Respir Crit Care Med 2001;163:A444.
  14. Wetterlin K. Turbuhaler: a new powder inhaler for administration of drugs to the airways. Pharm Res 1988;5:506–508.[Medline] [Order article via Infotrieve]



This article has been cited by other articles:


Home page
Eur Respir JHome page
W. Janssens, P. VandenBrande, E. Hardeman, E. De Langhe, T. Philps, T. Troosters, and M. Decramer
Inspiratory flow rates at different levels of resistance in elderly COPD patients
Eur. Respir. J., January 1, 2008; 31(1): 78 - 83.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Permissions
Right arrowRequest Permissions
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (7)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Broeders, M.E.A.C.
Right arrow Articles by Folgering, H.Th.M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Broeders, M.E.A.C.
Right arrow Articles by Folgering, H.Th.M.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS