Copyright ©ERS Journals Ltd 2001 Evaluation of impulse oscillation system: comparison with forced oscillation technique and body plethysmography1 Dept of Paediatrics and 2 Laboratory of Pneumology, Katholieke Universiteit, Leuven, Belgium CORRESPONDENCE: M. Demedts, Laboratory for Pneumology, U.Z. Gasthuisberg, Herestraat 49, B-3000, Leuven, Belgium. Fax: 016346803 Keywords: airway resistance, body plethysmography, forced oscillation technique, impulse oscillation system, impedance, reactance
Received: May 23, 2000
This study was funded by the National Fund for Scientific Research action "Care for Life", project numbers 7.0033.94, 7.0047.94 and 7.0078.94.
The impulse oscillation system (IOS) has been developed recently to measure respiratory system resistance (Rrs) and reactance (Xrs) at different frequencies up to 25 Hz. IOS has, however, not been validated against established techniques. This study compared IOS with the classical pseudorandom noise forced oscillation technique (FOT) and body plethysmographic airway resistance (Raw) in 49 subjects with a variety of lung disorders and a wide range of Raw (0.101.28 kPa·L1·s). Rrs,IOS was slightly greater than Rrs,FOT, especially at lower frequencies, with a mean±sd difference at 56 Hz of 0.14±0.09 kPa·L1·s. Comparisons with the wave-tube technique applied on two analogues indicated an overestimation by IOS. Xrs,IOS and Xrs,FOT were very similar, with a slightly higher resonant frequency with IOS than with FOT (mean difference±sd 1.35±3.40 Hz). Raw was only moderately correlated with Rrs-FOT and Rrs-IOS; although the mean differences were small (0.04±0.14 kPa·L1·s for Rrs6,FOT and 0.10±0.14 kPa·L1·s for Rrs5,IOS), IOS and FOT markedly underestimated high resistance values. In conclusion, the impulse oscillation system yields respiratory system resistance and reactance values similar, but not identical to those provided by the forced oscillation technique. Recently, the Jaeger impulse oscillation system (IOS, Erich Jaeger, Hoechberg, Germany) has been introduced as a user-friendly commercial version of the forced oscillation technique (FOT). IOS offers data-analysis and an elaborate report, containing total respiratory system resistance (Rrs) and reactance (Xrs) at a wide range of frequencies. It also contains estimations of central and peripheral pulmonary mechanics based on a simple model. However, only limited data have been published on this technique and these reports were mainly related to results in asthmatic and healthy children 13. The FOT was introduced by Dubois et al. 4 in 1956 as a method to characterize respiratory impedance and its two components, Rrs and Xrs, over a wide range of frequencies. Briefly, flow oscillations generated by means of a loudspeaker are applied at the subject's mouth and superimposed on normal breathing. The resulting pressure signal, as well as the flow signal, are recorded and analysed. These signals are, in general, waveforms containing several frequencies. For each of these frequencies, the ratio of pressure to flow can be considered (i.e. the impedance), which is a complex number that contains information about both the ratio of the magnitude of pressure to flow and about the phase shift between these signals. Most often this complex number is represented by its real part, the respiratory resistance (Rrs), and its imaginary part, the respiratory reactance (Xrs). Many studies have been published on FOT, especially since microprocessor techniques became available in the 1970s, allowing the analysis of complex signals by Fourier transform 57. The clinical potential of the method became apparent because it is rapid, demands only passive cooperation (i.e. no forced manoeuvres), and needs neither introduction of annoying devices (e.g. oesophageal balloon) nor frightening measuring conditions (e.g. closed body plethysmograph). It is especially appealing to children as it can be used routinely from 3 yrs of age onwards 1. The FOT has also proved its usefulness in many pathological conditions 9. In addition, the characteristics of the FOT have been widely studied 11. The IOS is, however, different from the classical FOT because an impulse (a rectangular wave form) rather than a pseudorandom noise signal (a mixture of several sinusoidal wave forms) is applied by the loudspeaker, and because of differences in data processing. No published data are available on accuracy of equipment and data handling, e.g. criteria for acceptance of data based on the coherence function 13 and on the applicability of implemented simple models simulating mechanics of the central and peripheral parts of the respiratory system 10. The aim of the present study was, therefore, to compare the results obtained with IOS, FOT and body plethysmography over wide ranges of resistances in patients. Preliminary data have been published as an abstract 17. In addition, the accuracies of IOS and FOT were evaluated on two mechanical structures by comparing the results with those obtained with the wave-tube technique 18, which can be considered as a reference technique for the measurement of acoustic impedance.
Forty-nine subjects with widely different resistances were included in the study. Some were healthy, while others suffered from a variety of diseases including asthma, cystic fibrosis, chronic obstructive pulmonary disease and lung fibrosis. Ages ranged from 870 yrs (mean±sd: 24±19 yrs). At random, resistance was measured by the SensorMedics 6200 body plethysmograph (SensorMedics, Yorbe Linda, CA, USA; airway resistance (Raw)), the Jaeger impulse oscillation system (IOS, Erich Jaeger) 1 and the Landsèr forced oscillation technique 5 within a time period of 3060 min. Raw was measured during rhythmic breathing at 0.5 Hz whilst keeping the cheeks supported in a constant-volume body plethysmograph, according to the technique of Dubois et al. 19, following the guidelines of the European Respiratory Society 6. Raw was obtained as the pressure/flow slope between ±0.5 L·s1; the mean of three values was retained.
With the IOS, Raw and reactance (Xaw) were calculated from the pressure/flow relationship obtained from impulses applied at the mouth during
With the FOT, a pseudorandom noise signal was applied 5 containing all the harmonics of 226 Hz, and Rrs and Xrs were calculated as the mean value of three measurements of 16 s each. The signals were analysed up to Data analysis consisted of calculating mean±sd, linear regressions and dispersions from the line of identity, according to the method of Bland and Altman 20. In addition, resistance and reactance of two mechanical structures were measured: one consisted of three layers of meshed wire fitted inside a short tube, and the other had an additional layer of sintered copper resulting in a much higher resistance. The impedances obtained with the FOT and IOS were compared with those obtained with the wave-tube technique 18. The latter is similar to the FOT, but the pneumotachograph is replaced by a 2-m long cylindrical tube.
This heterogeneous group of subjects showed a wide range of resistances, which thus made a reliable comparison of the three techniques possible. The mean value of Rrs5,IOS was 0.57 kPa·L1·s (range 0.181.06), of Rrs6,FOT 0.43 kPa·L1·s (range 0.140.80) and of Raw 0.47 kPa·L1·s (range 0.101.28).
Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
These data show that, although there is a fairly good agreement between Rrs values measured with IOS and FOT at higher frequencies, the latter are smaller than those measured with IOS, especially at lower frequencies and for higher resistances. It is unlikely that a poor signal to noise ratio can account for this difference. Indeed, measuring high impedance values at lower frequencies is unfavourable for the signal-to-noise ratio (quantified by the coherence function). It has been verified for FOT, that a coherence function with a value of <0.95 indicates an unreliable result. Such verification, however, has not been performed for IOS. This means that the value of the coherence function that must be selected as a threshold for the reliability of the IOS results is not known, so no values were discarded.
Accordingly, all FOT data were also retained for further analysis, including those with a coherence value <0.95. Figures 1 and 3 It might be tempting to attribute this increasing difference at the lowest frequencies and in the patients with the highest impedances to the fact that FOT is estimating resistance at 6 Hz and IOS at 5 Hz where higher resistance values can be expected due to the negative frequency dependence of resistance observed in those patients. However, this is unlikely to explain all the differences because Rrs,FOT at 4 Hz was also smaller than Rrs5,IOS, although the former data were less reliable (18 out of 49 scored a coherence of <0.95).
It is more likely that this difference is due to an overestimation of the resistance by IOS. Indeed, the resistance and reactance of two mechanical structures (one with a low resistance and the other with a much higher resistance) were measured with FOT, IOS and the wave-tube technique 18. The latter technique does not estimate mechanical impedance from the ratio of pressure to flow, but from the ratio of inlet to outlet pressure across the tube, and from the physics of the gas inside the tube. The ratio of two pressures can be measured more easily and accurately than the ratio of pressure to flow, so this technique can be considered as a reference technique for the measurement of acoustic impedances. The data in figure 5
The reactance values of IOS and FOT, as well as f0, were very similar to each other, except for Xrs5,IOS and Xrs6,FOT. From figure 4 The agreement between Rrs,IOS and Rrs,FOT estimates of Raw is only moderately good. The correlation coefficients are rather poor (R2=0.590.27) and the values are not superimposable. For resistance values in the normal range, Rrs with FOT is comparable with Raw, although somewhat larger. This has been attributed to the fact that the former technique measures total respiratory resistance, while body plethysmography measures only airway resistance 10. Rrs with IOS is clearly larger than Raw, even for resistance values at 5 Hz exceeding the normal range. This might be another indication that IOS is overestimating respiratory resistance at lower frequencies. For higher resistance values, Rrs becomes progressively smaller than Raw and this decrease was more pronounced at higher frequencies. This may be explained by the upper airway shunt (i.e. the loss of oscillatory flow into the cheeks) 16 and results in the frequency dependence of Rrs. This, therefore, makes these higher frequencies less accurate for clinical purposes. These resistances at higher frequencies are theoretically, however, not without importance. Indeed, the clinician should never be confined to one isolated frequency, but rather should consider the resistance/frequency curves and reactance/frequency curves as a whole. For clinical applications, the value at 56 Hz and the slope of the resistance/frequency curve may be most relevant. The fact that the Rrs values obtained with IOS and FOT are related to each other, and behave similarly in comparison with body plethysmography, should not lead to the conclusion that they are interchangeable. Firstly, pseudorandom noise is applied in the FOT while an impulse is applied in the IOS. The former signal contains a limited number of frequencies while the latter does not have this limitation. This is in favour of the signal-to-noise ratio for the FOT. Indeed, keeping the magnitude of the overall signal within acceptable limits therefore reducing the number of frequencies, increases the power at each frequency. Secondly, the FOT recommendations have been formulated on the basis of apparatus characteristics, calibration, input signals and frequencies, data processing and criteria for data acceptance 14. No such evaluations of IOS have been published. Further investigations of the impulse oscillation system are warranted to confirm its reliability. In particular, measurements with standard calibrating systems should be considered 13. The present authors are aware that the different forced oscillation technique apparatus each have their own characteristics, and can yield some variation in results. However, it would be worthwhile to validate the impulse oscillation system apparatus against standard systems because this is built according to specific technical standards, which are different from those of the forced oscillation technique. Consequently, the impulse oscillation system may give different results for some pathophysiological events. Furthermore, normal values for the impulse oscillation system in different age categories 1 have to be established, and the degree and pattern of changes in different disease states (e.g. chronic obstructive pulmonary disease, upper airway obstruction, lung fibrosis etc.) have to be evaluated. Finally, although this issue was not addressed in the present study, the impulse oscillation system provides estimates of central and peripheral pulmonary mechanics based on a model of the respiratory system. These estimates have not been critically investigated and no evidence in the literature has been found to support their validity. Until this validity is established, these estimates should be viewed with suspicion.
This article has been cited by other articles:
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||