1 Servei de Pneumologia i Allèrgia Respiratòria, Hospital Clínic, Barcelona, Spain, 2 Institut Clínic de Pneumologia i Cirurgia Toràcica (ICPCT), Hospital Clínic de Barcelona and Laboratori de Biofísica i Bioenginyeria, Facultat de Medicina, Universitat de Barcelona, Barcelona, Spain
CORRESPONDENCE: J.M. Montserrat, Servei de Pneumologia i Allèrgia Respiratòria, Hospital Clínic, Villarroel 170, 08036, Barcelona, Spain. Fax: 93 2275454
Keywords: Electromyography, forced oscillation technique, respiratory impedance, respiratory resistance, upper airway function, upper airway reflexes.
Received: September 28, 2000
Accepted March 20, 2001
This study was supported by grants CICYT (SAF99-0001), DGESIC (PM98-0027) and SEPAR 1999.
| Abstract |
|---|
|
|
|---|
The application of FOT did not increase EMGgg activity in any of the situations studied. In addition, no evidence of the effects on EEG was found: alpha/delta relationship: awake:0.70, baseline sleep:0.13, FOT(5 Hz):0.18, FOT(30 Hz):0.11.
The presented results suggest that the use of forced oscillation technique over the ranges of frequency and amplitude proposed for clinical sleep studies does not induce changes in upper airway muscle activity and neurological variables in patients with sleep apnoea/hypopnoea syndrome.
The forced oscillation technique (FOT) is a simple and noninvasive method that allows continuous on-line monitoring of airflow obstruction during sleep, even in the absence of flow, as occurs in apnoeas. FOT has been applied in clinical diagnostic studies in patients with sleep apnoea/hypopnoea syndrome (SAHS) [14] and in continuous positive airway pressure (CPAP) titration studies [5, 6] with highly encouraging results.
The upper airway is a complex structure with functional properties that are still poorly understood. Reflexes in response to pressure changes may play an important role in maintaining upper airway patency within the breathing cycle during sleep [710]. The effect of high-frequency oscillating pressures has been previously studied mainly in animals, but also in the sleeping human [1114]. In experimental conditions, high-frequency oscillation appeared to elicit an increase in upper airway muscle activity. The mechanism could be a direct activation of local receptors with an input on muscle activity. However, methodological artefacts must also be considered as they are not easy to avoid and may lead to nonvisible sleep fragmentation and neurological arousal [14]. Upon applying FOT extensively in sleep studies, an investigation was designed to verify that the low-frequency and low-amplitude oscillations applied in routine measurements of airway obstruction do not have any effect on upper airway patency, upper airway muscle electromyography and electroencephalogram (EEG) in a clinical setting in patients with moderate-to-severe sleep apnoea syndrome.
| Methods |
|---|
|
|
|---|
1.5 s and associated with EMGgg activity [17, 18]. In addition, an EEG signal analysis was performed by means of a fast Fourier frequency analysis of the signal [19, 20]. EMGgg was recorded using surface electrodes placed medially above and below the chin. The raw electromyogram (EMG) signals were acquired with a P55 A.C. Pre-Amplifier (Grass Instrument Co., West Warwick, RI, USA), band pass filtered at 301,000 Hz, and amplified and digitized at a sampling rate of 2,500 Hz. The EEG signal, EMGgg, airflow and CPAP pressure were also recorded and stored for further analysis in a CODAS system (Dataq Instruments Inc. Akron, OH, USA).
Stimulus
The oscillating pressure stimulus was a pressure wave generated with the same basic setting that the present group has described in FOT guided CPAP titration studies [13]. This setting allowed the application of a CPAP device to normalize sleep variables. The high-frequency oscillating pressure of 1 cmH2O of amplitude peak-to-peak was generated by a loudspeaker. In clinical studies, the amplitude is usually set at 1 cmH2O and oscillation frequency at 5 Hz, since in a model study [1] this frequency optimized the sensitivity of the system to obstruction as well as time resolution. However, in this investigation 5 Hz as well as 30 Hz were tested at the same pressure amplitude of 1 cmH2O. With this setting, both oscillation and CPAP pressure could be turned on or off automatically by the investigator from outside the patient's room. Therefore, the setting could be considered noiseless, thus minimizing possible environment-related arousal and methodological artefacts.
Protocol
All subjects arrived at the sleep unit
8 h before the beginning of the study. They were trained with CPAP and the best fitting mask was chosen. Once the setting was implemented and tested, awake maximal values of EMGgg were recorded and activity with voluntary contraction manoeuvres (tongue protrusion against resistance) were verified. All subjects were then allowed to sleep in their preferred position. No stimuli were applied until a stable sleep state was achieved (phase II or delta sleep for 10 min despite existence of transient arousal secondary to obstructive events). All patients were studied in nonrapid eye movement sleep. Oscillation stimuli were applied in the following two-step procedure. Step 1: during obstructive events and/or flow limitation periods; and step 2: continuously during a period of 20 s under optimal CPAP with absence of flow limitation.
Analysis
EMGgg activity was calculated as the root mean square (RMS) of the raw signal and expressed in arbitrary units (mean±sd). During the obstructive events or flow limited periods, the RMS of the breath with oscillation was compared with the mean of the RMS of the breath before oscillation application and the breath following the stimulus. During the periods of CPAP, the EMGgg RMS with pressure oscillation was compared with the EMGgg RMS of a period of the same length prior to the oscillation application. At least four applications were performed in each situation studied (16 applications for each subject: 5 Hz and 30 Hz, obstructive events and optimal CPAP). The results for each baseline or stimulation value are the mean of the four trials. The individual percentage of change between baseline and pressure oscillation stimuli was calculated for each of the four trials and, finally, the mean of these percentages of change was calculated.
A visual review of the EEG signal was performed. To complete the evaluation of the EEG, a fast Fourier spectral analysis of the signal was performed, comparing the amplitude of the alpha (812 Hz) and delta (0.24 Hz) bands of the signal with and without FOT applied at optimal CPAP with absence of flow limitation (step 2). Paired t-tests were used to compare baseline and stimuli results at both frequencies. Tests were assumed to be significant at p<0.05.
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
This lack of effect of forced oscillation on upper airway muscle activity can be considered, to some extent, as contradictory to results from other groups. Most of these previous studies were carried out in chronically tracheostomized dogs [1113]. These experimental conditions are very different from clinical sleep studies in the human. However, Henke and Sullivan [14] have shown evidence of EMG activation in the sleeping human, both in healthy subjects and in patients with obstructive sleep apnoea [14]. The presence of arousal or microarousals leading directly to upper airway muscle activation could be a possible explanation for these divergent results. In this study, trials that promoted immediate arousal were discarded, but trials in which arousal occurred shortly after stimulation were included in the analysis. In fact, the possibility that the responses observed are due to brief arousal is clearly pointed out by the authors. The present setting is believed to have improved the previous methodology, as it was completely automated. Forced oscillation was applied smoothly with a noiseless loudspeaker, and all the equipment was handled remotely from another room, thus minimizing possible methodological artefacts and the possibility of arousal. As a result, during the present investigations visible arousal or microarousal after the application of the forced oscillations were absolutely uncommon. Frequency analysis with the fast Fourier transformation confirmed that neurological activation was not present. Thus, in the absence of arousal and microarousals no evidence of direct effects of FOT on the upper airway muscles was found. In this investigation, surface EMG electrodes, which can be considered to be less sensitive than invasive wire electrodes, were used. However, it is felt that the quality of the signals obtained in the present study allows for a reliable interpretation. As can be seen in the figures provided, the electrodes accurately tracked EMG activation, not only during maximum voluntary contraction manoeuvres, but also at suboptimal CPAP and during snoring.
For the purposes of this study the effect of forced oscillations was examined at a frequency of 5 Hz and an amplitude of 1 cmH2O. These values have been found to provide the optimal balance between sensitivity to airflow obstruction and time resolution, and they are proposed for clinical studies [1]. In addition, 30 Hz was also tested in order to reproduce the methodology of the study of Henke and Sullivan. [14]. It is clear that larger pressure swings applied to the upper airway will cause increased activation of local mechanoreceptors and afferent pathways. In fact, all published animal model studies that demonstrate muscle activation applied a larger pressure amplitude than in the present study, ranging 24 cmH2O. It would be interesting to test a broader range of frequencies and amplitudes in order to identify thresholds of responses. However, this was beyond the scope of the present study as the aim was to test the clinical validity of FOT for sleep studies.
Another potential limitation of the present study may relate to the included population. All patients had moderate-to-severe sleep apnoea and had not been treated previously with CPAP. In addition, all were overweight or presented with obesity. These two conditions have been associated with impaired responses of the upper airway to diverse stimuli [21, 22]. Thus, it is possible that these patients could have a higher response threshold to the high-frequency pressure oscillation stimulus. Answering this question will require further investigation of a significant sample from the normal population.
When forced oscillation pressure is applied during apnoea the stimulus reaches only the upper airway above the point of obstruction. The present study was designed to include obstructed and nonobstructed breathing under optimal CPAP. Trials at optimal CPAP were included for two reasons. Firstly, in a previous study it was found that the measure of airflow obstruction obtained by means of FOT could be used for CPAP titration. Secondly, under optimal CPAP, the pressure oscillation is applied to the entire airway and thus could stimulate mechanoreceptors located below the site of obstruction. The presented findings were consistent and no evidence of upper airway muscle activation was found in any of the situations studied.
The results of this study show that the forced oscillation technique applied to assess airflow obstruction, as proposed for clinical studies, does not increase upper airway muscle tone or result in neurological activation in patients with sleep apnoea/hypopnoea syndrome.
| Acknowledgements |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. L. Dellaca, M. Rotger, A. Aliverti, D. Navajas, A. Pedotti, and R. Farre Noninvasive detection of expiratory flow limitation in COPD patients during nasal CPAP Eur. Respir. J., May 1, 2006; 27(5): 983 - 991. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Farre, J.M. Montserrat, and D. Navajas Noninvasive monitoring of respiratory mechanics during sleep Eur. Respir. J., December 1, 2004; 24(6): 1052 - 1060. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Steltner, R. Staats, J. Timmer, M. Vogel, J. Guttmann, H. Matthys, and J. Christian Virchow Diagnosis of Sleep Apnea by Automatic Analysis of Nasal Pressure and Forced Oscillation Impedance Am. J. Respir. Crit. Care Med., April 1, 2002; 165(7): 940 - 944. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||