ERJ
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Permissions
Right arrowRequest Permissions
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chavasse, R.J.
Right arrow Articles by Seddon, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chavasse, R.J.
Right arrow Articles by Seddon, P.
Eur Respir J 2001; 18:330-334
Copyright ©ERS Journals Ltd 2001


Comparison of resistance measured by the interrupter technique and by passive mechanics in sedated infants

R.J. Chavasse, Y. Bastian-Lee and P. Seddon

The Royal Alexandra Hospital for Sick Children, Dyke Road, Brighton, UK

CORRESPONDENCE: P. Seddon, The Research Centre, The Royal Alexandra Hospital for Sick Children, Dyke Road, Brighton, BN1 3JN, UK. Fax: 44 1273321441

Keywords: Chest wall mechanics, infant, interrupter technique, interrupter resistance in infants, lung mechanics, passive mechanics

Received: February 28, 2000
Accepted April 25, 2001


    Abstract
 TOP
 Abstract
 Methods
 Results
 Discussion
 References
 
Airways resistance measured by the interrupter technique (Rint) requires little patient cooperation and has been successfully used in young children, but little studied in infants. The authors aimed to evaluate the measurement of Rint in infants, using a commercially available device (the MicroRint), by comparing it with an established technique to measure respiratory resistance: the single breath occlusion technique (SBT); and a measure of airflow obstruction during forced expiration.

Infants <18 months old with a history of wheeze, sedated with triclofos for pulmonary function testing, had measurements taken and compared to Rint (using the MicroRint), respiratory system resistance (Rrs) by SBT, and to maximal flow at functional residual capacity (V'maxFRC).

Paired data from 25 of 37 infants studied was obtained. There was a significant difference between Rint (mean 2.94±0.68) and Rrs (4.02±0.87), but the two measures were strongly correlated (r=0.7). Rint was negatively correlated with V'maxFRC (r=–0.63). Smaller infants failed to trigger the MicroRint.

Interrupter resistance values in infants are significantly lower than values of respiratory system resistance obtained by passive mechanics. However, there is a strong correlation between the two measurements, as well as between resistance measured using the interrupter technique and maximal flow at functional residual capacity, which indicates that resistance measured using the interrupter technique may be a useful marker of airway obstruction in infants. There remain a number of theoretical and technical problems which require further exploration.

Respiratory disorders form a major part of infant morbidity. Monitoring the progression of a condition, the response to treatment or the requirement to change therapy is usually based on subjective observation or indirect measurement. Measuring pulmonary function may give further objective evidence on which to base clinical decisions. Conventional measures of pulmonary function in infants remain complex procedures. The requirement for sedation and the duration of the tests have limited the use of most to research purposes only.

The interrupter technique is not new: it was first described in 1927 by von Neergaard and Wirz 1, and further evaluated in 1954 by Mead and Whittenberger 2. The principle is simple: airflow is briefly occluded during inspiration or expiration, and measured pressure at the mouth (Pao) equilibrates with alveolar pressure (PA). This initial rapid change in Pao ({Delta}Pinit), an estimate of driving pressure across the airways, is divided by flow immediately before (or after) the occlusion to give a measure of airflow resistance termed interrupter resistance (Rint). Unfortunately, pressure oscillations immediately after airway occlusion make it impossible to measure {Delta}Pinit directly, and it is also necessary for valve closure to occur very rapidly. Because of such theoretical and technical concerns, the method was mostly neglected from the 1950s until the 1980s, when these concerns were addressed in a series of papers by Bates and coworkers 37; and studies exploring clinical potential have followed 812.

The technique requires no active subject cooperation as the measurements are made during quiet tidal breathing: it is, therefore, particularly suitable for young children, and the recent published clinical studies reflect this interest. Perhaps surprisingly, there has been little previous work in infants.

The MicroRint (Micro Medical, Rochester, Kent, UK) is a commercially available device that has been developed to measure Rint, based on a prototype reported by Chowienczyk et al. 8, 13. It is marketed for use in all age ranges from neonates to adults. Its use has been well documented in preschool children 12, 14. The authors investigated whether this device could also be used, and produce meaningful results, in young infants. They aimed to do this by comparing measurements of Rint with measurements of respiratory system resistance (Rrs) obtained by the widely-used single breath occlusion technique (SBT) 15.


    Methods
 TOP
 Abstract
 Methods
 Results
 Discussion
 References
 
Infants (n=34) with a history of recurrent or persistent wheeze had pulmonary function tests performed as part of a larger study into bronchodilator response. All infants had been reported by parents to have either persistent wheeze for ≥6 weeks, or at least three episodes of wheeze over 3 months, and all had a family or personal history of atopy. Their characteristics are summarized in table 1Go. The pulmonary function tests were performed when the infants were asymptomatic or had only mild symptoms.


View this table:
[in this window]
[in a new window]
 
Table 1— Infant characteristics

 
The infants were sedated with 100–120 mg·kg–1 triclofos sodium syrup. Once asleep they were laid supine on a cot with a neck roll. A semirigid, transparent silicon facemask (Size 2; Pari, Starnberg, Germany) was placed over their mouth and nose with therapeutic putty (Carters, Wiltshire, UK) used to form an airtight seal. Measurements were made when the infants were in quiet sleep, as judged by a regular respiratory rate, stable end-expiratory volume, and no obvious eye movements. Oxygenation was monitored continuously by pulse oximetry.

The order of testing was: SBT; rapid thoracoabdominal compression technique (RTC); interrupter resistance (IR). For the SBT and RTC measurements, a pneumotachograph (series 3500; Hans Rudolph Inc., Kansas City, Mo, USA) was attached to the mask. Airflow and mouth pressure were measured by Validyne transducers (Validyne MP45, Validyne Corp., Miltipas, CA, USA) and the data was then digitized and recorded by RASP (PhysioLogic Ltd, Newbury, Berks, UK). The IR measurements were carried out after the other procedures to avoid unnecessary disturbance to the infant, due to the need to change between pneumotachograph and the MicroRint apparatus.

Rrs and respiratory system compliance (Crs) were measured using the SBT. This technique and analysis of data is well described elsewhere 15. Briefly, airway occlusions were made at end-inspiration to induce a Hering-Breuer reflex, then released: the resulting Pao plateau and relaxed expiration were analysed to derive Crs, the respiratory system time constant, and hence Rrs. The mean Rrs from the five most technically satisfactory occlusions were used, with a minimum correlation coefficient of ≥0.995 over at least 40% of the expiratory flow/volume trace. Up to 20 manual occlusions were performed to obtain technically satisfactory data suitable for analysis.

Partial forced expiratory flow measurements were performed by RTC, with the mean maximal flow at functional residual capacity (V'maxFRC) taken from three technically acceptable manoeuvres as described previously 16.

Once this data had been collected, the pneumotachograph was disconnected and the MicroRint connected to the face mask. The cheeks were supported and the airway was manually supported by a chin lift technique to optimize airway patency. The MicroRint apparatus incorporates a screen pneumotachograph, a piezo-resistive transducer, and an oval shutter pivoted about its short axis, placed distal to the pneumotachograph and transducer ports. The shutter is actively rotated about its short axis to close the airway for a duration of 100 ms then actively rotated to reopen the airway. The time for shutter closure (and opening) to take place is 5–6 ms 17, which satisfies the criteria established by Bates et al. 7. The device also incorporates a microcomputer, which controls the shutter, samples the flow and pressure data at 100 Hz, and performs the analysis. {Delta}Pinit is estimated by two-point linear regression from mean pressure readings over 10-ms intervals, centred at 30 ms and 70 ms into the occlusion. This regression line is back extrapolated to the point of complete occlusion. Complete occlusion is defined as the time at which 25% of the peak of the first oscillation pressure upstroke is reached.

The device can be set to occlude in inspiration or expiration, at a preset flow or at peak flow. The authors opted to occlude during expiration at a flow of 100 mL·s–1. This was the minimum fixed flow available for triggering from the MicroRint software, and this option was elected rather than triggering at peak flow to avoid variability due to flow dependence of resistance. Expiratory occlusion was chosen to be more comparable with the SBT, which measures resistance during expiration. The median Rint was calculated automatically by the device software from five consecutive satisfactory occlusions.

Analysis
Rint was compared with Rrs and with V'maxFRC by correlation. Paired t-tests and Bland-Altman analysis 18 were also performed to compare Rint with Rrs. Characteristics of the infants in whom Rint was successfully measured were compared with the measurement failure group by unpaired t-tests and confidence interval analysis, except for V'maxFRC, which was non-normally distributed and compared by the Mann-Whitney U-test.


    Results
 TOP
 Abstract
 Methods
 Results
 Discussion
 References
 
The infants (n=37) were studied over a 1-yr period. The authors obtained 25 sets of paired data for Rint and Rrs. In 11 patients the MicroRint failed to occlude and in one patient Rrs could not be calculated due to the alinearity of the flow/volume trace. Paired data for Rint and V'maxFRC were obtained in 21 infants.

The infants in whom there was a failure to obtain values for Rint were younger, shorter and had lower peak tidal expiratory flow values compared to those in whom successful values were obtained (table 1Go). They did not have more airway obstruction as gauged by Rrs or V'maxFRC. Although their peak tidal flows were significantly lower, the majority still had flows exceeding the target value of 100 mL·s–1 yet failed to trigger the MicroRint.

For the 25 infants with paired data, mean±sd Rint was 2.94±0.68 kPa·L·s–1, and was substantially lower than Rrs at 4.02±0.87 kPa·L·s–1. However, there was a strong correlation between the two measurements (fig. 1Go) with a correlation coefficient of 0.7 (p<0.01). The Bland-Altman plot (fig. 2Go) demonstrates the bias, with all the Rint values being lower than the corresponding Rrs. There was also a strong (negative) correlation between Rint and V'maxFRC (fig. 3Go) with a coefficient of 0.63 (p<0.01) in the 21 infants with paired data.



View larger version (12K):
[in this window]
[in a new window]
 
Fig. 1.— Correlation of airways resistance measured by the single breath occlusion technique and interrupter technique (Rint). Rrs: respiratory system resistance. r=0.70.

 


View larger version (16K):
[in this window]
[in a new window]
 
Fig. 2.— Bland-Altman plot. Rint: interrupter technique; Rrs: respiratory system resistance. -----: mean difference –1.08 kPa·L–1·s; : limits of agreement i.e. mean±sd, –2.31±0.16.

 


View larger version (12K):
[in this window]
[in a new window]
 
Fig. 3.— Interrupter technique (Rint) correlation with maximal flow at functional residual capacity (V'maxFRC). R=0.63.

 

    Discussion
 TOP
 Abstract
 Methods
 Results
 Discussion
 References
 
It has been demonstrated that, in sedated infants with a history of recurrent wheeze, respiratory resistance measured by Rint with a commercially available device shows a strong correlation with two other widely-used measures of airway narrowing: Rrs and V'maxFRC. This suggests that the technique may be capable of giving a meaningful indication of airflow resistance in infants. However, the values obtained for Rint were markedly lower than for Rrs. This is not entirely surprising in view of what is known firstly about the physiology of the two techniques, and secondly about their respective assumptions and sources of error.

Rrs, as measured by single breath passive mechanics, gives a measure of resistance to airflow of the entire respiratory system during expiration: i.e. airways, lung tissue and chest wall. It is obtained indirectly from calculating compliance and the time constant of the respiratory system, and carries the assumptions of absent respiratory muscle activity and a single compartment model, with a single resistance value applicable over the range of the expiration 15.

Conversely, Rint gives a measure of resistance to airflow at one point during (in this study) expiration, and hence at the particular volume and flow at which occlusion occurred. It does not assume muscle relaxation nor a single value for resistance independent of flow and volume. The precise components of the respiratory system which contribute to Rint remain controversial, have been extensively reviewed by Bates and Milic-Emili 19, and will depend on exactly how the pressure change used to calculate Rint is estimated.

The pressure changes occurring after airflow interruption can be divided into three phases: a rapid rise ({Delta}Pinit), a series of damped oscillations masking the transition, and a slower rise ({Delta}Pdif). {Delta}Pinit is caused by equilibration between Pao and PA, and if it could be measured directly (impossible due to the oscillations), Rint calculated from this would correspond very closely to pure airway resistance 3, possibly with a small component from chest wall resistance 4. Hence, even without sources of error; Rint would be expected to be lower than Rrs, which includes airway, tissue and chest wall resistance. {Delta}Pdif, on the other hand, is a result of the visco-elastic behaviour of lung tissues and, in disease, "pendelluft" from equilibration between disparate lung compartments 20. The greater the component of {Delta}Pdif which is incorporated into Rint measurement, the more Rint might be expected to approach Rrs. When Rint has been compared to airway resistance (Raw, measured by body plethysmography) in adults and older children, it has again correlated strongly, but been higher than Raw 8, 10. This may be because of an added chest wall element in {Delta}Pinit and/or because of incorporation of an element of {Delta}Pdif, as well as {Delta}Pinit, in the calculation of Rint.

The two main sources of inaccuracy in Rint measurements are: insufficiently rapid airway occlusion, and inaccurate estimation of {Delta}Pinit. The first is a problem because airflow is measured before valve closure, but pressure afterwards. If during the time the valve takes to close fully, significant airflow in relation to lung volume occurs, the pressure change will not accurately reflect the driving pressure which existed at the time flow was measured. The MicroRint shutter meets the criteria suggested by Bates et al. 7 in this regard, and therefore, would not be expected to be a serious source of error in these measurements. Accurate estimation of {Delta}Pinit remains a major problem. The linear back-extrapolation used by the MicroRint is acceptable provided that {Delta}Pdif follows a linear trajectory: the more curvilinear or irregular it becomes, the more inaccurate the estimation. The direction and magnitude of the error will depend on the nature of the curve and the points chosen to back-extrapolate. The MicroRint (unlike an earlier prototype) does not give the option to visually inspect the pressure waveform before accepting a result. Other approaches suggested have included: curvilinear back-extrapolation 6, or taking the pressure at a set time after occlusion and accepting that this includes part of {Delta}Pdif as well as {Delta}Pinit 17. This latter approach is the one taken by the other current commercially available Rint system (Masterscreen Rocc, Jaeger, Hoechberg, Germany) which calculates Rint from the pressure just before release of occlusion (i.e. 100 ms after flow interruption).

The other important issue in discussing the results is that of upper airway compliance. Infants have relatively large and compliant upper airways, and with the present experimental setup, "upper airway" includes the face mask. The effect of airway compliance is both to reduce {Delta}Pinit (and hence reduce calculated Rint) and to blur the distinction between {Delta}Pinit and {Delta}Pdif (and hence add a further inaccuracy to estimating {Delta}Pinit) 19. These effects are more marked in the presence of high airway resistance and low lung volumes, as found in infants particularly with airway disease 21.

In summary, Rint as measured by the MicroRint device in infants is closely related to other measures of airway obstruction commonly used in infants, but is consistently and substantially lower than Rrs. Likely reasons for this difference are: the different components of the respiratory system which contribute, the effect of upper airway compliance, errors in the estimation of {Delta}Pinit, and errors and unjustified assumptions in the SBT for measuring Rrs. These sources of error in both techniques probably explain why the correlation demonstrated between Rint and Rrs was somewhat less good than that previously reported between Rint and Raw in adults and older children 8, 10. The results also demonstrate the difficulty of using a preset flow value for triggering in small infants whose maximal flows may barely exceed this value. The Rint appears promising, but requires further assessment in infants, before it can be recommended for widespread use, either as a clinical or research tool.

The device studied is not suitable in its present form to study very small infants or those with very low peak tidal flows due to lung disease. Apart from the concerns discussed earlier, this study did not address the issue of reproducibility, though this has been looked at in preschool children 12. It is also unclear whether measurements are best made in inspiration or expiration 14, whether occlusion should be triggered at a preset flow, peak flow or some other standard, and what type (if any) of back-extrapolation of mouth pressure is optimal 17. Finally, and perhaps most critically, the measurements were made under optimal conditions in sedated infants, yet the most likely niche for the interrupter technique would seem to be as a "quick and easy" measure of airway obstruction at the cotside in clinical and epidemiological work. Use in unsedated infants is likely to pose a whole new crop of problems which will need further study.


    References
 TOP
 Abstract
 Methods
 Results
 Discussion
 References
 

  1. von Neergaard K, Wirz K. Die Messung der Ströungswiderstände in den Atemwegen des Menschen, insbesondere bei Asthma und Emphysem. Z Klin Med 1927;105:51–82.
  2. Mead J, Whittenberger JL. Evaluation of airway interruption technique as a method for measuring pulmonary air-flow resistance. J Appl Physiol 1954;6:408–416.[Free Full Text]
  3. Bates JHT, Ludwig MS, Sly PD, Brown K, Martin JG, Fredberg JJ. Interrupter resistance elucidated by alveolar pressure measurement in open-chest normal dogs. J Appl Physiol 1988;65:408–414.[Abstract/Free Full Text]
  4. Bates JHT, Abe T, Romero PV, Sato J. Measurement of alveolar pressure in closed-chest dogs during flow interruption. J Appl Physiol 1989;67:488–492.[Abstract/Free Full Text]
  5. Bates JHT, Brown KA, Kochi T. Respiratory mechanics in the normal dog determined by expiratory flow interruption. J Appl Physiol 1989;67:2276–2285.[Abstract/Free Full Text]
  6. Bates JHT, Sly PD, Okubo S. General method for describing and extrapolating monotonic transients and its application to respiratory mechanics. Med Biol Eng Comput 1987;25:131–135.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  7. Bates JHT, Hunter IW, Sly PD, Okubo S, Filiatrault S, Milic-Emili J. Effect of valve closure time on the determination of respiratory resistance by flow interruption. Med Biol Eng Comput 1987;25:136–140.[CrossRef][Web of Science][Medline] [Order article via Infotrieve]
  8. Chowienczyk PJ, Lawson CP, Lane S, et al. A flow interruption device for the measurement of airway resistance. Eur Respir J 2000;4:623–628.[CrossRef]
  9. Phagoo SB, Watson RA, Pride NB, Silverman M. Accuracy and sensitivity of the interrupter technique for measuring the response to bronchial challenge in normal subjects. Eur Respir J 1993;6:996–1003.[Abstract]
  10. Carter ER, Stecenko M, Pollock BH, Jaeger MJ. Evaluation of the interrupter technique for the use of assessing airway obstruction in children. Pediatr Pulmonol 1994;17:211–217.[Web of Science][Medline] [Order article via Infotrieve]
  11. Bisgaard H, Klug B. Lung function measurements in awake young children. Eur Respir J 1995;8:2067–2075.[Abstract]
  12. Bridge PD, Ranganathan S, McKenzie SA. Measurement of airway resistance using the interrupter technique in preschool children in the ambulatory setting. Eur Respir J 1999;13:792–796.[Abstract]
  13. Bridge PD, Lee H, Silverman M. A portable device based on the interrupter technique to measure bronchodilator response in schoolchildren. Eur Respir J 1996;9:1368–1373.[Abstract]
  14. Bridge PD, McKenzie SA. Airway resistance measured by the interrupter technique (Rint) in preschool children in inspiration and expiration. Eur Respir J 1999;14:Suppl. 30, 44s.
  15. Fletcher ME, Baraldi E, Steinbrugger B. Passive Respiratory Mechanics. In: Stocks J, Sly PD, Tepper RS, Morgan WJ, editors. Infant Respiratory Function TestingNew York, Wiley-Liss, 1996.
  16. Morgan WJ, Castile R, Turner DJ, Motoyama E, Morgan WJ. Forced Expiratory Maneuvers. In: Stocks J, Sly PD, Tepper RS, Morgan WJ, editors. Infant Respiratory Function TestingNew York, Wiley-Liss, 1996.
  17. Phagoo SB, Watson RA, Silverman M, Pride NB. Comparison of four methods of assessing airflow resistance before and after induced airway narrowing in normal subjects. J Appl Physiol 1995;79:518–525.[Abstract/Free Full Text]
  18. Bland JM, Altman DG. Statistical method for assessing agreement between two methods of clinical measurement. Lancet 1986;i:307–310.
  19. Bates JHT, Milic-Emili J. The flow interruption technique for measuring respiratory resistance. J Crit Care 1991;6:227–238.
  20. Ludwig MS, Romero PV, Sly PD, Fredberg JJ, Bates JHT. Interpretation of interrupter resistance after histamine-induced constriction in the dog. J Appl Physiol 1990;68:1656.
  21. Sly PD, Bates JHT. Computer analysis of the physical factors affecting the use of the interrupter technique in infants. Pediatr Pulmonol 1988;4:219–224.[Web of Science][Medline] [Order article via Infotrieve]



This article has been cited by other articles:


Home page
Arch. Dis. Child. Fetal Neonatal Ed.Home page
M R Thomas, G F Rafferty, R Blowes, J L Peacock, N Marlow, S Calvert, A Milner, and A Greenough
Plethysmograph and interrupter resistance measurements in prematurely born young children
Arch. Dis. Child. Fetal Neonatal Ed., May 1, 2006; 91(3): F193 - F196.
[Abstract] [Full Text] [PDF]


Home page
CMAJHome page
Diagnosis of asthma
Can. Med. Assoc. J., September 13, 2005; 173(6_suppl): S15 - S19.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Permissions
Right arrowRequest Permissions
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chavasse, R.J.
Right arrow Articles by Seddon, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chavasse, R.J.
Right arrow Articles by Seddon, P.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS