Copyright ©ERS Journals Ltd 2001 Off-line sampling of exhaled air for nitric oxide measurement in children: methodological aspects1 Dept of Paediatrics, Division Paediatric Respiratory Medicine, and 2 Dept of Epidemiology and Biostatistics, Erasmus University and University Hospital/Sophia Children's Hospital, Rotterdam and 3 Dept of Paediatrics, University Hospital Maastricht, the Netherlands CORRESPONDENCE: J.C. de Jongste, Sophia Children's Hospital, Room Sp-2465, P.O. Box 2060, 3000 CB, Rotterdam, the Netherlands Keywords: asthma, breath-hold, children, exhaled air, nitric oxide, off-line sampling
Received: April 26, 1999
This study was supported by a research grant of the Netherlands Asthma Foundation (NAF 94.14).
Measurement of nitric oxide in exhaled air is a noninvasive method to assess airway inflammation in asthma. This study was undertaken to establish the reference range of exhaled NO in healthy school-aged children and to determine the influence of ambient NO, noseclip and breath-holding on exhaled NO, using an off-line balloon sampling method. All children attending a primary school (age range 813 yrs) underwent NO measurements on two occasions with high and low ambient NO. Each time, the children performed four expiratory manoeuvres into NO-impermeable balloons, with and without 10 s of breath-holding and with and without wearing a noseclip. Exhalation flow and pressure were not controlled. NO was measured within 4 h after collection, by means of chemiluminescence. All children completed a questionnaire on respiratory and allergic disorders, and performed flow/volume spirometry. With low ambient NO, the mean exhaled NO value of 72 healthy children with negative questionnaires and normal lung function was 5.1±0.2 parts per billion (ppb) versus a mean of 6.8±0.3 ppb in the remaining 49 children with positive questionnaires for asthma and allergy, and/or recent symptoms of cold (p=0.001). Exhaled and ambient NO were significantly related, especially with ambient NO >10 ppb (r=0.86, p=0.0001 versus r=0.34, p=0.004 for ambient values <10 ppb). The use of a noseclip, with low ambient NO and without breath-holding, caused a small decrease in exhaled NO values (p=0.001). The effect of breath-holding on exhaled NO depended on ambient NO. With ambient NO >10 ppb, exhaled NO decreased, whereas with ambient NO <10 ppb, exhaled NO increased after 10 s breath-hold. It is concluded that off-line sampling in balloons is a simple and, hence, attractive method for exhaled nitric oxide measurements in children which differentiates between groups with and without self-reported asthma, allergy and colds, when ambient nitric oxide is <10 parts per billion. Wearing a noseclip and breath-holding affected measured values and should, therefore be standardized or, preferably, avoided. Nitric oxide (NO) is a reactive free radical gas produced by various cells within the respiratory tract. It plays an important role in the pathophysiology of inflammatory airway disease and can be detected in exhaled air 14. There is now convincing evidence that NO in exhaled air (eNO) is a noninvasive marker of airway inflammation in various airway disorders 1, 5, 6. Recently, slow exhalation from total lung capacity with a constant flow against a resistance was recommended for measurement of eNO 6. However, this manoeuvre is difficult to perform for many children 7, and requires the presence of an NO-analyser and equipment to measure flow with a visual feedback control. The present authors have developed a simple off-line, single breath, balloon sampling method to measure eNO. Blowing air in a balloon is feasible even for young children, and produces values similar to those of end-expiratory plateaus 7, 8. Off-line sampling has the advantage that children may be studied at home or school, and that it can be used for large epidemiological studies, independent from the direct presence of an NO-analyser. The aim of this study was to establish a reference range of eNO sampled in balloons, of a large group of healthy school-aged children. Furthermore, the influence of ambient NO, the effect of 10 s of breath-holding and the use of a noseclip was examined.
All pupils of a primary school (n=129) participated. They were interviewed with validated questionnaires on asthma, allergy, eczema and rhinitis, translated from the core questionnaires on wheezing, rhinitis amd eczema of the International Study of Asthma and Allergies in Childhood (ISAAC) 9. In addition they were asked about recent symptoms of cold (stuffy or running nose, sneezing, coughing, sore throat, with or without fever). Measurement of eNO was performed on two occasions in the same children. On the first occasion, the ambient NO was high (median 35 parts per billion (ppb), range 15188 ppb). At that time, 91 children (mean age 10 yrs) had negative questionnaires on asthma, allergy and eczema (all questions answered negatively) and did not report symptoms of cold within the 3 weeks before the measurements. On the second occasion 3.5 months later, ambient NO was low (median 7.0 ppb, range 1.59.0 ppb). At that time, 8 children were absent, 72 of the 91 who initially had negative ISAAC questionnaires reported no symptoms of cold, and 49 had either positive ISAAC questionnaires or reported a recent cold. Measurements took place during several sessions at each occasion, always between 14.00 and 16.00 h. Ambient NO was drawn and stored for later analysis before and after each session, and the mean of these two samples was taken as the ambient NO concentration at that session. All children were self-reported never-smokers, never had had any other chronic illness, and used no medication. The study was approved by the hospital ethical committee.
Exhaled air sampling and nitric oxide measurements All subjects performed expiratory manoeuvres into four balloons: 1) immediate exhalation with a noseclip; 2) immediate exhalation without a noseclip; 3) exhalation after breath-holding for 10 s with a noseclip; 4) exhalation after breath-holding for 10 s without a noseclip. All measurements were performed with the children seated. The NO concentration in balloons was measured within 4 h after collection. Air was drawn out of the balloons at 200 mL·min1 by a chemiluminescence analyser (Sievers 280, Boulder, CO, USA) with a response time of 200 ms. The NO profile showed a rapid rise to a steady NO plateau, at which eNO was calculated. The analyser was regulary calibrated according to the manufacturer's guidelines, with two different certified NO gases (100 ppb and 9 parts per million (ppm)) and a certified NO free air mixture (HoekLoos, Barendrecht, the Netherlands).
Lung function
Stability of exhaled nitric oxide in balloons
Data analysis
Characteristics of the study population are given in table 1
Reference values All children performed the sampling procedures without difficulty. Values obtained in 72 healthy children, not wearing noseclips, without breath-hold, and while ambient NO was below 10 ppb, are considered as reference values and show a normal, symmetrical distribution with a mean level of 5.1±0.2 ppb (95% reference interval 1.78.5 ppb). Exhaled NO was similar in male and female children (means 5.1 ppb and 4.9 ppb, respectively, p=0.15). There was no significant correlation between eNO and FVC or FEV1 (p=0.56). A significant positive correlation was found between age and eNO; for every 10 months of increasing age eNO increased 0.3±0.1 ppb (p=0.013). Individual data and 95% reference interval are given in figure 1
Effect of ambient nitric oxide High concentrations of ambient NO led to significantly higher eNO values in healthy children: 19.2±1.6 ppb with ambient NO >10 ppb versus 5.1±0.2 ppb with ambient NO <10 ppb (p<0.0001). Exhaled NO values and ambient NO concentrations were significantly related (fig. 2
Effect of noseclip The noseclip significantly reduced eNO values when ambient NO was <10 ppb. In 72 healthy children, the mean eNO value was 4.5±0.2 ppb with, and 5.1±0.2 ppb without noseclip (p=0.001). There was no significant effect of a noseclip on eNO values under all other measurement conditions: high ambient NO levels with (p=0.2) or without breath-holding (p=0.9) and low ambient NO levels with breath-holding (p=0.4).
Effect of breath-holding
Stability of exhaled nitric oxide in balloons
In this study, a reference range of eNO obtained off-line with balloon sampling in a large group of healthy school-aged children was defined. The mean eNO concentration of healthy children in the study is in agreement with eNO concentrations of healthy children from a number of previous studies 1216. Furthermore, it has been shown that a number of methodological factors influence eNO collected off-line in balloons, including ambient NO, the use of a noseclip and breath-holding. Significant differences between children with and without self-reported asthma, allergy and colds were deteted, suggesting that this method is sufficiently sensitive to detect minor degrees of airway inflammation in groups of otherwise healthy school-aged children. There are a number of factors that should be considered when comparing the present simple off-line sampling technique with more complicated on-line sampling techniques where flow is controlled and the NO signal displayed during exhalation.
It has previously been shown that despite lack of flow standardization, off-line sampling in balloons produces eNO values that are reproducible and similar to those of end-expiratory plateaus measured on-line with constant flow 7. It has also been shown that eNO is flow-dependent, but mainly at flows of In a previous study, an excellent within-subject short-term repeatability of eNO was shown, with an intraclass correlation coefficient (ri) of 0.91, using the same technique without wearing a noseclip 7. Small alterations of this original sampling technique, like the use of a noseclip or 10 s breath-holding, could potentially influence the repeatability. It has previously been shown that short-term repeatibility in clinicallly stable allergic asthmatic children was alsoexcellent when a noseclip was used (ri=0.96) 7, 8, 14, 21. This study did not specifically investigate the influence of 10 s breath-holding on the repeatability of eNO obtained with the off-line balloon method, and no such data are available in the literature. Contamination with nasal NO is another potential problem with off-line sampling. An expiratory resistance is recommended to isolate the lower airway gas from gas with much higher NO concentrations produced in the nose and paranasal sinuses 6. In the present study, balloons were filled via a plastic tube during a single oral exhalation. The tube worked as a fixed flow restrictor which causes a positive mouth pressure during the exhalation manoeuvre. It has been shown in adults that a mouth pressure of 4 cmH2O is sufficient to keep the soft palate closed, thus preventing contamination with NO from upper airways 22. Studies on mouth pressure and velum closure were performed in adults; it may well be that lower pressures close the velum in children. The lowest estimated mean exhalation flow of 250 mL·s1 in this study produced a mouth pressure of 6 cmH2O. This should be sufficient to close the velum, and the authors are therefore confident that contamination with NO from the upper airways, despite the simple set-up without monitoring of flow or pressure during exhalation, did not occur to an important degree. The potential effect of NO in the dead space air volume on eNO in the balloon is another point of concern. Children have a smaller dead space volume than adults which, in a similar balloon volume, will cause a smaller effect on NO values. The authors speculate that, in children, the effect of the dead space air on the eNO value will be minimal when ambient NO is <10 ppb, and when subjects inhale orally before exhaling into the balloon, as was done in the present study.
Previous studies on the effect of ambient NO on eNO vary remarkably. Some authors found no influence of ambient NO on eNO values 2327, others found important effects 2832. This study found a strong positive relationship between ambient NO concentrations and eNO in balloons. The variation of sampling techniques used in the different studies probably explains the differences. As an example, Corradi et al. 32 found no relationship between ambient NO and eNO in balloons, with ambient NO <35 ppb, but the expiration in their study in adults took
The present study found a significant correlation between an increase in eNO concentrations and increasing age in healthy children. A possible explanation is the influence of the different dead space air volumes on eNO values in balloons. It could also be speculated that this correlation is caused by increasing nasal contamination of the eNO with age, due to development and pneumatization of the paranasal sinuses in childhood. Indeed, nasal NO increases with age, reaching levels similar to those in adults at the age of It has been stated that using a noseclip enhances nasal contamination of eNO, although there are no data to support this 2, 10, 17. Several investigators routinely require subjects to wear noseclips during NO measurements 10, 26, 31, 35, while others do not 17, 21, 36. Without a noseclip, diffusion of NO through the nares as well as into the posterior nasopharynx is possible; with a noseclip only diffusion towards the pharynx can occur, and this would increase contamination with high nasal NO levels. Using the present off-line, single breath, balloon sampling method with an expiratory resistance, it was found that a noseclip in combination with low ambient NO levels causes a small but significant decrease in exhaled NO values. A possible explanation is that a noseclip prevents a minimal nasal inspiration, thus preventing contamination of eNO with nasal NO. Although a noseclip only minimally affected eNO, the present data do not argue against the European Respiratory Society Task Force recommendation not to use noseclip during NO measurement 6. Prior studies have shown that breath-holding increases eNO in asthmatics and in healthy subjects in a time dependent fashion 10, 17, 37, 38. However, others have shown that when 40 ppm is inhaled, a breath-hold of several seconds reduced NO in exhaled air to 13 ppm 25. This decrease may be caused by rapid uptake of NO by haemoglobin in the lung capillaries 17, 25, 39. This present study found that the effect of breath-holding on eNO depends on ambient NO; with high levels, breath-holding reduces NO, whereas with low levels it increases eNO. It is not likely that these results are influenced by nasal contamination during breath-holding or in the oral expiration phase, as shown earlier with an inert tracer gas 22. It is concluded that off-line measurement of nitric oxide sampled in balloons, is a simple and feasible method of measuring exhaled nitric oxide in children. Off-line sampling has the important advantage that it offers independence from the presence of a nitric oxide analyser. Off-line balloon sampling without use of a noseclip and without breath-hold, and with ambient nitric oxide levels <10 parts per billion, shows differences between healthy children and those with self-reported asthma, allergy or cold, confirming that exhaled nitric oxide may be used as a surrogate marker of airway inflammation. The simplicity of this method makes it attractive for studies in larger populations, where differences between groups can be detected.
The authors are indebted to the children and teachers of the De Wilgenstam school in Rotterdam, who participated in this study.